K152790 · Venus Concept , Ltd. · ONF · Jan 20, 2016 · General, Plastic Surgery
Device Facts
Record ID
K152790
Device Name
Venus Versa System
Applicant
Venus Concept , Ltd.
Product Code
ONF · General, Plastic Surgery
Decision Date
Jan 20, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Venus Versa System is a multi-application device intended to be used in aesthetic and cosmetic procedures. The SR515 and SR580 IPL applicators are indicated for the following: Treatment of benign pigmented epidermal and cutaneous lesions including: hyperpigmentation, melasma, ephelides (freckles), lentigines, nevi, and cafe-au-lait macules. Treatment of benign cutaneous vascular lesions including: port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, angiomas, poikiloderma of civatte, leg veins and venous malformations. The HR650 and HR690 IPL applicators are indicated for the removal of unwanted hair and to effect stable long-term or permanent hair reduction for skin types I-IV. Permanent hair reduction is the long-term stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.
Device Story
Venus Versa System is a multi-application console-based device delivering Intense Pulsed Light (IPL) energy for aesthetic procedures. System includes a console with user interface and four interchangeable IPL applicators (SR515, SR580, HR650, HR690). Device operates via selective photothermolysis; light energy is absorbed by target chromophores in skin. Features include adjustable light fluence, pulse duration, and integrated sapphire light guide cooling. Used in clinical settings by trained professionals. Output is controlled light pulses; healthcare providers use device to treat vascular/pigmented lesions and remove unwanted hair. Benefits include non-invasive treatment of dermatological conditions and long-term hair reduction.
Clinical Evidence
Bench testing only. Performance demonstrated via electromagnetic compatibility and electrical safety testing (AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-57), biocompatibility testing (ISO 10993-5), cleaning/disinfection validation, and bench testing of energy output specifications. No clinical study data provided.
Technological Characteristics
IPL system; console with user interface and interchangeable applicators. Wavelengths: 515-950 nm (SR515/SR580), 650-950 nm (HR650/HR690). Fluence: up to 25 J/cm². Pulse duration: up to 50 ms. Frequency: up to 3 Hz. Spot size: 10 x 30 mm. Sapphire light guide. Integrated cooling system. Biocompatible patient-contacting materials (ISO 10993-5). Electrical safety per AAMI/ANSI ES60601-1.
Indications for Use
Indicated for aesthetic/cosmetic procedures including treatment of benign pigmented epidermal/cutaneous lesions (hyperpigmentation, melasma, freckles, lentigines, nevi, cafe-au-lait macules) and benign cutaneous vascular lesions (port wine stains, hemangiomas, telangiectasias, rosacea, angiomas, poikiloderma of civatte, leg veins, venous malformations). Also indicated for hair removal and permanent hair reduction in skin types I-IV.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 20, 2016
Venus Concept Ltd. % Ms. Janice Hogan Regulatory Counsel Hogan Lovells US LLP 1835 Market Street, 29th FL Philadelphia, Pennsylvania 19103
Re: K152790
Trade/Device Name: Venus Versa System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument for Use in General and Plastic Surgery and In Dermatology Regulatory Class: Class II Product Code: ONF Dated: December 21, 2015 Received: December 21, 2015
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152790
Device Name Venus Versa System
#### Indications for Use (Describe)
The Venus Versa System is a multi-application device intended to be used in aesthetic procedures.
The SR515 and SR580 IPL applicators are indicated for the following:
· Treatment of benign pigmented epidermal and cutaneous lesions including: hyperpigmentation, melasma, ephelides (freckles), lentigines, nevi, and cafe-au-lait macules.
· Treatment of benign cutaneous vascular lesions including: port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, angiomas, poikiloderma of civatte, leg veins and venous malformations. The HR650 and HR690 IPL applicators are indicated for the removal of unwanted hair and to effect stable long-term or permanent hair reduction for skin types I-IV. Permanent hair reduction is the long-term stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regimen.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="text-decoration: line-through;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
## Venus Concept's Venus Versa System
#### Sponsor/Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Venus Concept USA Ltd. 4556 N. Hiatus Road Sunrise, FL 33351 Telephone: 416.907.0115 Fax: 954.572.5680
Contact Person: Tal Bresler-Stramer, Ph.D., RAC VP QA/RA Venus Concept, Ltd.
Date Prepared: December 21, 2015
#### Trade Name of Device
Venus Versa System
#### Classification Regulation, Name, Product Code
Laser surgical instrument for use in general and plastic surgery and in dermatology (21 CFR 878.4810), Product Code: ONF
#### Predicate and Reference Devices
Predicate Devices: Danish Dermatologic Development A/S's Ellipse I PL System (K060516) (Primary Predicate), MATTIOLI Pulse TWO/THREE Plus Family (K100053)
Reference Devices: Alma Laser's Harmony Lite Multi-Application Platform (K141237); Emvera Diolux (K123257); Lumenis Inc.'s LightSheer Duet Laser System (K053628); Venus Legacy CX (K143554); Palomar Medical Products, Inc.'s StarLux Pulsed Light System (K041086); Beijing KES Biology's IPL System (K122995); BTL Industries, Inc.'s Exilite (K150051); Beijing Syntech Laser Co.'s APOLLO V+ (K113018); Syneron's ePlus Treatment System (K113868); Cynosure's Icon (K142376); Lynton Lumina (K063427)
#### Intended Use / Indications for Use
The Venus Versa System is a multi-application device intended to be used in aesthetic and cosmetic procedures.
{4}------------------------------------------------
The SR515 and SR580 IPL applicators are indicated for the following:
- Treatment of benign pigmented epidermal and cutaneous lesions, including: . hyperpigmentation, melasma, ephelides (freckles), lentigines, nevi, and cafe-au-lait macules.
- . Treatment of benign cutaneous vascular lesions, including: port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, angiomas and spider angiomas, poikiloderma of civatte, leg veins and venous malformations,
The HR650 and HR690 IPL applicators are indicated for the removal of unwanted hair and to effect stable long-term or permanent hair reduction for skin types I-IV. Permanent hair reduction is defined as the long-term stable reduction in the number of hairs regrowing when measured at 6. 9. and 12 months after the completion of a treatment regimen.
# Device Description
The Venus Versa System is comprised of a console (controller/CPU) and applicators that deliver Intense Pulsed Light (IPL) optical energy to the treatment area. Four IPL handpieces (SR515, SR580, HR650, HR690) are included with the device system. The Versa device provides individual adjustment of light fluency and pulse duration, and has integrated skin cooling features. Using the IPL applicators, the light energy delivered to the patient's skin is used to treat various conditions via the mechanism of selective photothermolysis.
# Technological Characteristics
The Venus Versa System has similar technological characteristics to its predicates. All three devices are IPL systems requlated under 21 CFR 878.4810 as class II devices. The Versa is a multi-application system that allows delivery of IPL enerqy, similar to its predicates. The Versa and the predicate devices are each designed as a system console with a user interface and applicator(s) indicated for the desired treatment effects. The technological characteristics of the Venus Versa as a system and when used for each of the three proposed indications (vascular lesions, pigmented lesions, hair removal/reduction) are similar to the corresponding parameters of the predicate devices. The technological differences between the Venus Versa and its predicates are minor, including some differences in terms of the spot size, frequency, and pulse duration. However, the kev parameters of the Versa affecting treatment outcomes (i.e., IPL wavelengths, fluence levels) are the same or encompassed within the range of the predicate devices. In addition, each of the systems utilizes a light quide and cooling system in delivering treatment. See Table 1 below.
{5}------------------------------------------------
| Medical Device | Venus Concept Ltd.'s Venus Versa<br>System (K152790) | Danish Dermatologic Development A/S<br>Ellipse I²PL System (K060516) | Mattioli Pulse TWO/THREE Plus Family<br>(K100053) |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Class, Product<br>Code, Regulation | Class II, ONF, 21 CFR 878.4810 | Class II, GEX, 21 CFR 878.4810 | Class II, ONF, 21 CFR 878.4810 |
| Indications for<br>Use | The Venus Versa System is a multi-<br>application device intended to be used in<br>aesthetic and cosmetic procedures.<br>The SR515 and SR580 IPL applicators<br>are indicated for the following:<br>· Treatment of benign pigmented<br>epidermal and cutaneous lesions<br>including: hyperpigmentation, melasma,<br>ephelides (freckles), lentigines, nevi, and<br>cafe-au-lait macules.<br>· Treatment of benign cutaneous vascular<br>lesions including port wine stains,<br>hemangiomas, facial, truncal and leg<br>telangiectasias, rosacea, angiomas and<br>spider angiomas, poikiloderma of civatte,<br>leg veins and venous malformations.<br>The HR650 and HR690 IPL applicators<br>are indicated for the removal of unwanted<br>hair and to effect stable long-term or<br>permanent hair reduction for skin types I-<br>IV. Permanent hair reduction is defined<br>as the long-term stable reduction in the<br>number of hairs regrowing when<br>measured at 6, 9, and 12 months after the<br>completion of a treatment regimen. | Ellipse I²PL System is intended for use in<br>dermatology:<br>* Hair removal (permanent hair reduction).<br>* Treatment of benign pigmented lesions<br>(including, but not limited to solar<br>lentigines, ephilides, mottled<br>pigmentation) and benign vascular lesions<br>(including but not limited to diffuse<br>redness, telangiectasias, port wine<br>stains).<br>* Treatment of inflammatory acne.<br>[Table included in indications for use form<br>not included here.] | The MATTIOLI PULSE TWO/THREE<br>PLUS is an Intense Pulse Light (IPL)<br>device family indicated for use in aesthetic<br>applications (based on selective<br>photothermolysis), in the treatment of<br>various benign pigmented lesions and hair<br>removal and that produce different effects<br>depending on the applicator that is used:<br>SA APPLICATOR:<br>Model SN: Wavelengths from 560 - 1200<br>nm are indicated for treatment of benign<br>pigmented (epidermal and cutaneous)<br>lesions, including hyperpigmentation,<br>warts, lentigines, nevi, melasma, and<br>cafe-au-lait.<br>VA APPLICATOR<br>Model VN: Wavelengths from 510 - 1200<br>nm are indicated for the treatment of<br>benign vascular lesions including port<br>wine stains, hemangiomas, facial, truncal<br>and leg telangiectasias, rosacea,<br>melasma, angiomas and spider<br>angiomas, poikiloderma of civatte; leg<br>veins, facial veins and venous<br>malformations.<br>HR APPLICATORS indicated for the<br>treatment of unwanted hair (i.e. hair<br>removal).<br>Model HN: Wavelengths from 650-1200<br>nm for skin types I-IV;<br>Model HE: Wavelengths from 710 - 1200 |
# Table 1: Venus Versa System Substantial Equivalence Comparison
{6}------------------------------------------------
| Medical Device | Venus Concept Ltd.'s Venus Versa<br>System (K152790) | Danish Dermatologic Development A/S<br>Ellipse I²PL System (K060516) | Mattioli Pulse TWO/THREE Plus Family<br>(K100053) |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| | | | nm for skin type V<br>The equipment should only be used under<br>medical supervision. |
| Energy Type | IPL | IPL | IPL |
| IPL Spectrum | 515-950 nm<br>o Lesions: 515-950 nm (SR515); 580-<br>950 nm (SR580)<br>▪ Pigmented Lesions: 515-950 nm<br>(SR515); 580-950 nm (SR580)<br>▪ Vascular Lesions: 515-950 nm<br>(SR515); 580-950 nm (SR580)<br>o Hair Removal: 650-950 nm (HR650);<br>690-950 nm (HR690) | 400-950<br>o Lesions:<br>▪ Pigmented Lesions: 400-720 nm<br>(PL)<br>▪ Vascular Lesions: 555-950 nm (VL);<br>530-750 nm (PR)<br>o Hair Removal: 600-950nm (HR, HR-S);<br>645-950 nm (HR-D) | 510-1200 nm<br>o Lesions: 560-1200 nm (SA); 510-1200<br>nm (VA)<br>o Hair Removal: 650-1200 nm (HN);<br>710-1200 nm (HF) |
| Frequency | Up to 3 Hz | 1.5-2.0 s charge time/repetition rate | Up to 1 Hz |
| Spot Size | 10 x 30 mm | 10 x 48 mm | 10 x 40 mm<br>15 x 50 mm |
| Pulse Duration | Up to 50 ms<br>o Lesions (SR515, SR580): 5-20 ms<br>▪ Pigmented Lesions: 10-20 ms<br>▪ Vascular Lesions: 10-20 ms<br>o Hair Removal (HR650, HR690): 20-50<br>ms | 1.5-100 ms | 14-35 ms |
| Energy Density<br>(Fluence) | Up to 25 J/cm²<br>o Lesions:<br>▪ Pigmented Lesions: 5-20 J/cm²<br>▪ Vascular Lesions: 6-22 J/cm²<br>o Hair Removal: 6-20 J/cm² | Up to 26 J/cm²<br>o Lesions:<br>▪ Pigmented Lesions: up to 18 J/cm²<br>▪ Vascular Lesions: up to 26 J/cm²<br>o Hair Removal: up to 21 J/cm² | Up to 30 J/cm²<br>o Skin Rejuvenation (SR): 6-30 J/cm²<br>o Hair Removal (HR): 6-25 J/cm² |
| Components | System console (with user interface)<br>Applicators<br>Ultrasonic gel | System console (with user interface)<br>Applicators<br>Optical coupling gel | System console (with user interface)<br>Applicators |
| Light guide | Sapphire light guide | Light guide | Crystal Light Guide |
| Cooling system | Cooling system | Cooling system (circulating water) | Cooling system (water cooling) |
{7}------------------------------------------------
Therefore, the dimensional differences and other minor technological differences between the Versa and its predicates do not present any new or different issues of safety or effectiveness. Therefore, the Venus Versa System presents similar technological characteristics as its predicates, in support of substantial equivalence.
#### Performance Data and Standards
The performance of the Venus Versa has been demonstrated in testing. Electromagnetic Compatibility and Electrical Safety testing was conducted per AAMI/ANSI ES60601-1, IEC 60601-1-2, and IEC 60601-2-57. All results were passing. In addition, the patient contacting materials are biocompatible per ISO 10993-5. Cleaning and disinfection validation testing has also been completed.
The Venus Versa System also underwent software validation and results demonstrate that the software was appropriate for release. Bench testing further confirmed that the energy outputs of the Versa met specifications.
# Substantial Equivalence
The Venus Versa has the same intended use and similar indications for use, technological characteristics and principles of operation as its predicate devices. The minor technological differences between the Venus Versa and its predicate devices do not raise any new or different issues of safety or effectiveness, given that the key energy parameters are the same or similar for the devices.
#### Conclusions
The tests outlined above demonstrate that the device performs as intended for the indicated uses of each applicator. Minor differences between the subject and predicate devices do not present any new or different types of safety or effectiveness questions, as confirmed by device performance evaluations. In sum, the Venus Versa System is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.