K122995 · Beijing Kes Biology Technology Co., Ltd. · ONF · Jan 22, 2013 · General, Plastic Surgery
Device Facts
Record ID
K122995
Device Name
INTENSE PULSED LIGHT (IPL) SYSTEMS
Applicant
Beijing Kes Biology Technology Co., Ltd.
Product Code
ONF · General, Plastic Surgery
Decision Date
Jan 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Intense Pulsed Light (IPL) Systems (inclusive of the handpiece used to deliver pulsed-light energy) are indicated for use in surgical, aesthetic and cosmetic applications in permanent hair reduction, treatment of pigmented lesions, moderate inflammatory acne vulgaris, ephelides (freckles), and vascular lesions.
Device Story
Intense Pulsed Light (IPL) system (models MED-210, MED-230) delivers pulsed light energy (430nm-1200nm) via a sapphire handpiece. Operates on selective thermolysis principle: light pulses exceed chromophore thermal relaxation time to cause thermal damage while limiting pulse width to spare surrounding skin. Unlike lasers, IPL delivers multiple wavelengths per pulse to enhance penetration and target specific chromophores. Used by clinicians in surgical, aesthetic, and cosmetic settings. Output parameters (wavelength, energy density, pulse sequence, pulse width) are adjusted based on skin type and condition. Clinical benefit includes hair reduction, lesion treatment, and acne therapy. Device includes integrated thermal cooling to protect epidermis, reduce pain/discomfort, and minimize thermal injury risks like scarring or pigmentation changes.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 10993-5 (in vitro cytotoxicity), and ISO 10993-10 (irritation and skin sensitization) was demonstrated to verify design specifications.
Technological Characteristics
IPL system; sapphire light delivery; energy density 10-60J/cm²; pulse sequence 1-15 pulses; pulse width 1-20ms; wavelength 430-1200nm. Power: 220V/50Hz or 110V/60Hz. Standards: IEC 60601-1, IEC 60601-1-2, ISO 10993-5, ISO 10993-10. Integrated thermal cooling.
Indications for Use
Indicated for permanent hair reduction, treatment of pigmented lesions, moderate inflammatory acne vulgaris, ephelides (freckles), and vascular lesions in patients across Fitzpatrick skin types I-V.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
# JAN 2 2 2013
# Attachment II 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number: K122995
- l. Date of Submission: September 17, 2012
- 2. Sponsor
Beijing KES Biology Technology Co., Ltd. Building E, No.28 Langfa Yifa Industrial Park, Daxing District, Beijing, 102613, China
Contact Person: Mr. Xun Zeng Position: Vice President Tel: +86-10-84405668 Fax: +86-10-84404920 Email: info@keslaser.com
- Submission Correspondent 3. Ms. Diana Hong & Mr. Tarzan Wang Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
- Proposed Device Identification 4.
Proposed Device Name: Intense Pulsed Light (IPL) Systems Proposed Device Model: MED-210, MED-230 -
Classification: Class II Product Code: ONF Regulation Number: 21 CFR 878.4810 Review Panel: General and Plastic Surgery
{1}------------------------------------------------
Intended Use Statement:
The Intense Pulsed Light (IPL) Systems (inclusive of the handpiece used to deliver pulsed-light energy) are indicated for use in surgical, aesthetic and cosmetic applications in permanent hair reduction, treatment of pigmented lesions, moderate inflammatory acne vulgaris, ephelides (freckles), and vascular lesions.
| | Skin Types | | | | | |
|-------------------|------------------------------------|----------|----------|----------|----------|----------|
| Conditions | | I | II | III | IV | V |
| Hair | Filter setting and wavelength rang | 640-1200 | 640-1200 | 640-1200 | 690-1200 | 750-1200 |
| Pigmented lesions | | 480-1200 | 530-1200 | 530-1200 | 560-1200 | 560-1200 |
| Acne vulgaris | | 430-1200 | 430-1200 | 430-1200 | 480-1200 | 530-1200 |
| Ephelides | | 480-1200 | 480-1200 | 530-1200 | 530-1200 | 530-1200 |
| Vascular lesions | | 530-1200 | 530-1200 | 560-1200 | 560-1200 | 590-1200 |
#### 5. Predicate Device Identification
510(k) Number: K093627
Product Name: IPULSELIGHT IPL SYSTEM
Manufacturer: Shanghai APOLO Medical Technology Co., Ltd
- Device Description 6.
The Intense Pulsed Light (IPL) Systems are intense pulsed light system which delivers intense pulsed light at a wavelength ranging from 430nm-1200nm. Intense Pulsed Light (IPL) systems work on the principles of selective thermolysis. That is, causing thermal damage to target chromophores by using light of appropriate wavelength in pulses that exceeds the chromophores' thermal relaxation time but sparing normal skin by limiting the pulse width below the thermal relaxation time for skin.
IPL Systems are different from lasers in that they deliver many wavelengths in each pulse of light instead of just one wavelength. Generally, IPL enhances penetration without using excessive energy levels and enables targeting of specific chromophores.
Based on this, the IPL Systems (inclusive of the handpiece used to deliver pulsed-light energy) are indicated for use in surgical, acsthetic and cosmetic applications in permanent hair reduction, treatment of pigmented lesions, moderate inflammatory acne vulgaris, ephelides (freckles), and vascular lesions.
- 7. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was
{2}------------------------------------------------
Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A 1EC 60601-1: 1988 +A1:1991+A2:1995, Medical Electrical Equipment - Part 1: General requirements for safety.
- A IEC 60601-1-2: 2007, Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic compatibility -Requirements and tests.
- A ISO 10993-5: 2009, Biological evaluation of medical devices - Part 5: Tests for In Vitro cytotoxicity.
- ン ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization.
| Table II-1 SE Comparison | | |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ITEM | Proposed Device | Predicate Device |
| | Intense Pulsed Light (IPL) Systems | IPULSELIGHT IPL SYSTEM (K093627) |
| Product Code | ONF | ONF |
| Regulation No. | 21 CFR 878.4810 | 21 CFR 878.4810 |
| Class | II | II |
| Intended Use | TheIntense Pulsed Light (IPL) Systems (inclusive of<br>the handpiece used to deliver pulsed-light energy) are<br>indicated for use in surgical, aesthetic and cosmetic<br>applications in permanent hair reduction, treatment of<br>pigmented lesions, moderate inflammatory acne<br>vulgaris, ephelides (freckles), and vascular lesions. | IPulseLight IPL System (HS 300C, HS 650)<br>are identical with regard to indications for<br>use including recommended filters to be<br>used with Fitzpatrick skin type. Both models<br>are intended for medical use in the treatment<br>of the following dermatologic conditions:<br>Permanent hair reduction- long-term stable<br>reduction in number of hairs re-growing<br>after a treatment regimen. Treatment of: -<br>Moderate inflammatory acne vulgaris -<br>Benign pigmented epidermal lesions<br>including dyschromia, hyperpigmentation,<br>melasma, ephelides (freckles). - Cutaneous<br>lesions including scars - Benign cutaneous<br>vascular lesions including port wine stains,<br>hemangiomas, facial, truncal and leg<br>telangiectasias, erythema of rosacea, leg<br>veins, spider anglomas and venous<br>malformations. The integrated thermal<br>cooling is indicated for use in cooling the<br>epidermis at the treatment site prior to,<br>during, and after light treatment in general<br>aesthetic dermatologic and plastic surgery<br>procedures. - Reduce pain during light<br>treatment (via partial anesthesia from<br>cooling) - Reduce discomfort during and/or<br>associated with light treatment - Minimize<br>thermal injury, including thermal necrosis, to<br>non-target skin and skin structures during<br>and/or associated with light treatment, thus<br>reducing possible complications such as<br>scabbing, scarring, hyper - and/or hypo<br>pigmentation - Allow the use of higher light<br>or laser fluencies for light treatments (such<br>as for hair removal and the treatment of<br>vascular or pigmented lesions) - Reduce<br>potential side effects of light treatments<br>(such as for hair removal and the treatment |
#### 8. Substantially Equivalent Comparison
Table II-1 SE Comparison
{3}------------------------------------------------
:
:
.
.
| | | of vascular or pigmented lesions) |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Light source | Intense pulsed light | Intense pulsed light |
| Wavelength | 430-1200nm, 530-1200nm, 640-1200nm,<br>Optional: 480-1200nm, 560-1200nm, 590-1200nm,<br>690-1200nm, 750 -1200nm | 420nm-1200, 510-1200nm, 560-1200nm,<br>610-1200nm, 640-1200nm (Standard);<br>480nm-1200, 585-1200nm, 690-1200nm,<br>755-1200nm (Options) |
| Deliver system | Sapphire | Sapphire |
| Energy density | 10-60J/cm² | 10-60J/cm² |
| Pulse sequence | 1~15 pulses | 1~15 pulses |
| Pulse delay | 5 - 50ms | 5 - 50ms |
| Pulse width | 1-20ms | 2-20ms |
| Spot size | MED-210: 15mmX50mm (optional: 12mmX33mm,<br>15mmX35mm )<br>MED-230: A: 12mm X33mm; B: 15mmX50mm<br>(optional: 15mmX35mm) | 12x35mm, 15x50mm (Standard); 12x12mm,<br>12x25mm, 12x50mm, 15x35mm, 15x45mm<br>(Options) |
| Power Supply | 220V±20V 50Hz or 110V±20V 60Hz | 100/110V, 50~60Hz or 230~260V, 50~60Hz |
| Max. power<br>consumption | MED-210: 1400 W MED-230: 2000 W | 1200 W |
| Electrical Safety | Comply with IEC 60601-1 | Comply with IEC 60601-1 |
| EMC | Comply with IEC 60601-1-2 | Comply with IEC 60601-1-2 |
Table II-2 Setting Comparison of Specified Indication for Use
| ITEM | Proposed Device | Predicate Device |
|-------------------------------------|------------------------------------|---------------------------------------------|
| | Intense Pulsed Light (IPL) Systems | IPULSELIGHT IPL SYSTEM (K093627) |
| | (MED-210, MED-230) | (HS 300C, HS 650) |
| permanent hair reduction | | |
| Wavelength Range (nm) | 640-1200/690-1200/750-1200 | 610-1200/640-1200/690-1200 |
| Energy Range (J/cm2) | 10-44 | 10-46 |
| Pulse Duration (ms) | 3-14 | 4-15 |
| Pulse Delay (ms) | 16-32 | 15-30 |
| Spot Size (mm) | 12mm X33mm; 15mmX50mm<br>15mmX35mm | 12x25mm,12x35mm,12x50mm,15x35mm,<br>15x45mm |
| pigmented lesions | | |
| Wavelength Range (nm) | 480-1200/530-1200/560-1200 | 510-1200/560-1200 |
| Energy Range (J/cm2) | 12-44 | 14-46 |
| Pulse Duration (ms) | 3-9 | 2-9 |
| Pulse Delay (ms) | 16-32 | 15-30 |
| Spot Size (mm) | 12mm X33mm; 15mmX50mm<br>15mmX35mm | 12x25mm,12x35mm,12x50mm,15x35mm,<br>15x45mm |
| moderate inflammatory acne vulgaris | | |
| Wavelength Range (nm) | 430-1200/480-1200/530-1200 | 420-1200/510-1200/560-1200 |
| Energy Range (J/cm2) | 10-40 | 10-42 |
| Pulse Duration (ms) | 3-8 | 3-9 |
:
{4}------------------------------------------------
| Pulse Delay (ms) | 16-32 | 15-30 |
|-----------------------|------------------------------------|------------------------------------------------|
| Spot Size (mm) | 12mm X33mm; 15mmX50mm<br>15mmX35mm | 12x25mm, 12x35mm, 12x50mm, 15x35mm,<br>15x45mm |
| ephelides (freckles) | | |
| Wavelength Range (nm) | 480-1200/530-1200 | 510-1200 |
| Energy Range (J/cm2) | 10-42 | 12-44 |
| Pulse Duration (ms) | 3-8 | 3-9 |
| Pulse Delay (ms) | 16-32 | 15-30 |
| Spot Size (mm) | 12mm X33mm; 15mmX50mm<br>15mmX35mm | 12x25mm, 12x35mm, 12x50mm, 15x35mm,<br>15x45mm |
| vascular lesions | | |
| Wavelength Range (nm) | 530-1200/560-1200/<br>590-1200 | 510-1200/560-1200 |
| Energy Range (J/cm2) | 10-42 | 10-44 |
| Pulse Duration (ms) | 3-8 | 3-9 |
| Pulse Delay (ms) | 16-32 | 15-30 |
| Spot Size (mm) | 12mm X33mm; 15mmX50mm<br>15mmX35mm | 12x25mm, 12x35mm, 12x50mm, 15x35mm,<br>15x45mm |
Difference in intended use, wavelength, spot size, power supply, Max. power consumption and setting of specified indication for use between the proposed and predicate device are discussed in the 510(k) submission documents, it is concluded that these differences will not affect the effectiveness and safety of the proposed device.
The proposed device, Intense Pulsed Light (IPL) Systems, are determined to be Substantially Equivalent (SE) to the predicate device, IPULSELIGHT IPL SYSTEM (K093627), in respect of safety and effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized eagle-like symbol with three curved lines representing its body and wings.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
January 22, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Beijing Kes Biology Technology Company, Limited % Mid-Link Consulting Company, Limited Ms. Diana Hong Correspondent, General Manager P.O. Box 237-023 Shanghai, China 200237
Re: K122995
Trade/Device Name: Intense Pulsed Light (IPL) Systems Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
- Regulatory Class: II Product Code: ONF Dated: December 26, 2012 Received: December 31, 2012
#### Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{6}------------------------------------------------
### Page 2 - Ms. Diana Hong
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
# Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Section II Indications for Use
510(k) Number:
Device Name: Intense Pulsed Light (IPL) Systems
Indications for Use:
The Intense Pulsed Light (IPL) Systems (inclusive of the handpiece used to deliver pulsed-light energy) are indicated for use in surgical, aesthetic and cosmetic applications in permanent hair removal, skin rejuvenation, reduction of pigmented lesions, acne therapy, freckle, vascular lesions and facial blemish removal.
ZPRESCRIPTION USE (Part 21 CFR 801 Subpart D)
## OVER-THE-COUNTER USE (21 CFR 801 Subpart C)
(PI.EASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Neil R Ogden 2013.01.18 08:52:24 -05'00'
(Division Sign-Off)
Division of Surgical Devices
510(k) Number K 122995
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.