K141237 · Alma, Inc. · ONF · Mar 13, 2015 · General, Plastic Surgery
Device Facts
Record ID
K141237
Device Name
HARMONY LITE
Applicant
Alma, Inc.
Product Code
ONF · General, Plastic Surgery
Decision Date
Mar 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Harmony Lite is intended for use in aesthetic, cosmetic, and surgical applications requiring the ablation, vaporization, excision, incision, and photothermolysis (photocoagulation) of soft tissue in the medical specialties of dermatology, general and plastic surgery, and aesthetic applications, as follows: 500-600 nm Dye VL Pro Module AFT Hand piece The Advanced Fluorescence Technology (AFT) Dye VL Pro Module handpiece (with and without contact-cooling) is indicated for: • The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and legtelangiectasias, rosacea, erythema of rosacea, angiomas, poikiloderma of Civatte, leg veins and venous malformations. • For use on Fitzpatrick skin types I-V. 540-950 nm VP PRO AFT Hand piece The Advanced Fluorescence Technology (AFT) VP ProModule handpieces (with and without contact-cooling) are indicated for: ·The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, and ephelides (freckles),lentigines, nevi, melasma, and cafe-au-lait macules. · The treatment of cutaneous lesions including warts, scars and striae. • The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and legtelangiectasias, rosacea, erythema of rosacea, angiomas, poikiloderma of Civatte, leg veins and venous malformations. · Use on all skin types (Fitzpatrick I-VI). 570-950 nm SSR Pro Module AFT Hand piece The Advanced Fluorescence Technology (AFT) 570-950 nm SSR Module handpiece (with and without contact-cooling) is indicated for: ·The treatment of benign pigmented epidermal lesions, including dyschromia , hyperpigmentation, melasma and ephelides (freckles). · The treatment of face and body vascular and pigmented lesions. · The treatment of cutaneous lesions, including scars and striae. · The treatment of benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias,rosacea, melasma, angiomas, poikiloderma of Civatte, leg veins and venous malformations. •Use on all skin types (Fitzpatrick I-VI). 590 nm LED Module Handpiece The 590 nm LED module handpiece is indicated to: · Provide topical heating to promote increased blood flow for temporary relaxation of muscle and relief of pain . · Provide topical heating for the purpose of elevating and/or maintaining tissue temperature. 780-950 nm SHR PRO AFT Module Hand piece The Advanced Fluorescence Technology (AFT) 780-950 nm SHR PRO (with and without contactcooling) is indicated for: •The treatment of pseudofolliculitis barbae (PFB). •The removal of unwanted hair and to effect stable long-term or permanent hair reduction in hairregrowth is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12months after the completion of the treatment regime. NIR Large and Small Modules : The NIR Modules are intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating the tissue temperature for the temporary relief of minor muscle pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and the relaxation of muscles: may also help muscle spasms, minor sprains and strains, and minor muscle back pain. Thermoelectric Cooler (TEC) The thermoelectric cooler, integrated into the light and laser handpieces, is indicated for use in cooling the epidermis at the treatment site prior to, during and after light or laser treatment in general surgery, plastic surgery and dermatology to: · Reduce pain during and/or associated with light or laser treatment; · Reduce discomfort during and/or associated with light or laser treatment; ·Minimize thermal injury, including thermal necrosis, to non-target skin and skin structures during, and/or associated with light orlaser treatment, thus reducing possible complications such as scabbing, scarring, hyper- and/or hypopigmentation: · Allow the use of higher light or laser fluences for light or laser treatments (such as for hair removal and the treatment of vascular orpigmented lesions); and · Reduce potential side effects of light or laser treatments (such as for hair removal and the treatment of vascular or pigmented lesions).
Device Story
Multi-application platform utilizing various handpieces (IPL, LED, NIR) for dermatologic and surgical procedures; console includes touch-screen, power supply, and cooling system. Handpieces deliver light/infrared energy to target tissue via photothermolysis or topical heating. Operated by clinicians in medical settings. Integrated thermoelectric cooling (TEC) protects epidermis, reduces pain, and allows higher fluences. Output affects tissue via thermal energy; clinical benefit includes lesion removal, hair reduction, and pain relief. System uses footswitch control.
Clinical Evidence
No human clinical trials provided. Evidence includes bench testing (IEC 60601-1, 60601-1-2, 60601-2-57) and collateral animal data (pig study) comparing Dye VL Pro IPL to pulsed dye laser (PDL) data, supported by Monte Carlo simulations of energy absorption and penetration depth.
Technological Characteristics
Platform uses IPL (AFT), LED, and NIR modules. Materials: Quartz tubes for NIR. Energy: Light (500-950nm) and IR (1300nm). Cooling: Integrated thermoelectric cooler (TEC). Connectivity: Standalone console with touch-screen. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-57.
Indications for Use
Indicated for aesthetic, cosmetic, and surgical soft tissue applications including treatment of benign vascular/pigmented lesions, hair removal, pseudofolliculitis barbae, and topical heating for pain relief/muscle relaxation. Applicable to Fitzpatrick skin types I-VI depending on module.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
The Modified Alma Lasers Soprano XL Family of Multi-Application and Multi-Technology Platforms (K140009)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized emblem. The emblem consists of three overlapping profiles facing right, resembling a family or group of people.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 13, 2015
Alma Lasers Incorporated % Ms. Kathy Maynor USA Regulatory 26 Rebecca Court Homosassa, Florida 34446
Re: K141237 Trade/Device Name: Alma Harmony Lite Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONF, ILY Dated: February 4, 2015 Received: February 6, 2015
Dear Ms. Maynor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K141237
Device Name Alma Harmony Lite
## Indications for Use (Describe)
Indications for Use :
The Harmony Lite is intended for use in aesthetic, cosmetic, and surgical applications requiring the ablation, vaporization, excision, incision, and photothermolysis (photocoagulation) of soft tissue in the medical specialties of dermatology, general and plastic surgery, and aesthetic applications, as follows:
500-600 nm Dye VL Pro Module AFT Hand piece
The Advanced Fluorescence Technology (AFT) Dye VL Pro Module handpiece (with and without contact-cooling) is indicated for:
• The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and legtelangiectasias, rosacea, erythema of rosacea, angiomas, poikiloderma of Civatte, leg veins and venous malformations.
• For use on Fitzpatrick skin types I-V.
## 540-950 nm VP PRO AFT Hand piece
The Advanced Fluorescence Technology (AFT) VP ProModule handpieces (with and without contact-cooling) are indicated for:
·The treatment of benign pigmented epidermal lesions including dyschromia, hyperpigmentation, melasma, and ephelides (freckles),lentigines, nevi, melasma, and cafe-au-lait macules.
· The treatment of cutaneous lesions including warts, scars and striae.
• The treatment of benign cutaneous vascular lesions including port wine stains, hemangiomas, facial, truncal and legtelangiectasias, rosacea, erythema of rosacea, angiomas, poikiloderma of Civatte, leg veins and venous malformations.
· Use on all skin types (Fitzpatrick I-VI).
570-950 nm SSR Pro Module AFT Hand piece
The Advanced Fluorescence Technology (AFT) 570-950 nm SSR Module handpiece (with and without contact-cooling) is indicated for:
·The treatment of benign pigmented epidermal lesions, including dyschromia , hyperpigmentation, melasma and ephelides (freckles).
· The treatment of face and body vascular and pigmented lesions.
· The treatment of cutaneous lesions, including scars and striae.
· The treatment of benign cutaneous vascular lesions, including port wine stains, hemangiomas, facial, truncal and leg telangiectasias,rosacea, melasma, angiomas, poikiloderma of Civatte, leg veins and venous malformations.
•Use on all skin types (Fitzpatrick I-VI).
590 nm LED Module Handpiece
The 590 nm LED module handpiece is indicated to:
· Provide topical heating to promote increased blood flow for temporary relaxation of muscle and relief of pain .
· Provide topical heating for the purpose of elevating and/or maintaining tissue temperature.
780-950 nm SHR PRO AFT Module Hand piece
The Advanced Fluorescence Technology (AFT) 780-950 nm SHR PRO (with and without contactcooling) is indicated for: •The treatment of pseudofolliculitis barbae (PFB).
{3}------------------------------------------------
•The removal of unwanted hair and to effect stable long-term or permanent hair reduction in hairregrowth is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12months after the completion of the treatment regime.
NIR Large and Small Modules :
The NIR Modules are intended to emit energy in the infrared spectrum to provide topical heating for the purpose of elevating the tissue temperature for the temporary relief of minor muscle pain and stiffness, the temporary relief of minor joint pain associated with arthritis, the temporary increase in local circulation where applied, and the relaxation of muscles: may also help muscle spasms, minor sprains and strains, and minor muscle back pain.
## Thermoelectric Cooler (TEC)
The thermoelectric cooler, integrated into the light and laser handpieces, is indicated for use in cooling the epidermis at the treatment site prior to, during and after light or laser treatment in general surgery, plastic surgery and dermatology to: · Reduce pain during and/or associated with light or laser treatment;
· Reduce discomfort during and/or associated with light or laser treatment;
·Minimize thermal injury, including thermal necrosis, to non-target skin and skin structures during, and/or associated with light orlaser treatment, thus reducing possible complications such as scabbing, scarring, hyper- and/or hypopigmentation:
• Allow the use of higher light or laser fluences for light or laser treatments (such as for hair removal and the treatment of vascular orpigmented lesions); and
· Reduce potential side effects of light or laser treatments (such as for hair removal and the treatment of vascular or pigmented lesions).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## Section 8 - 510(k) Summary or 510(k) Statement
#### I. General Information
| Submitter: | Alma Lasers Inc.<br>485 Half Day Rd. Suite 100<br>Buffalo Grove, IL 60089 |
|---------------------------|---------------------------------------------------------------------------|
| Contact Person: | Avi Farbstein<br>General Manager, EVP<br>Alma Lasers Inc. |
| | Kathy Maynor<br>USA Regulatory<br>Alma Lasers Inc. |
| Summary Preparation Date: | March 12, 2015 |
#### II. Names
| Device Names: | Harmony Lite Multi-Application Platform |
|-------------------------------|-----------------------------------------------------------------------------------------------------|
| Primary Classification Names: | Surgical Powered Light Instrument, Ultraviolet Dermatological<br>Light, and LED Phototherapy device |
#### III. Predicate Devices
- K072564 Alma Harmony XL Plarform
- K140009 The Modified Alma Lasers Soprano XL Family of Multi-Application and Multi-Technology Platforms
#### IV. Product Description
The Alma Lasers Harmony Lite Multi-Application Platform is comprised of the following main components:
- . The main console unit that incorporates the touch-screen control panel, power supply modules, cooling system, switching module, service panel and isolating transformer.
- Variety of handpieces, including IPL, LED and NIR modules ●
- Footswitch.
#### V. Indications for Use
The Alma Lasers Harmony Lite Multi-Application Platform is intended for use in dermatologic and general surgical procedures.
{5}------------------------------------------------
The Indications for Use of the Harmony Lite Multi-Application Platform are provided in the tables below.
| | Dye VL Pro IPL Handpiece – this submission | K072564 - Alma<br>VL515 IPL handpiece |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | |
| | | Primary Predicate |
| Characteristic | | |
| Wavelength<br>(nm) | 500-600nm | VL515 Handpiece<br>(515nm-950nm) |
| Intended Use | Intended for use in aesthetic and cosmetic<br>applications requiring selective photothermolysis<br>(photocoagulation or coagulation) and hemostasis<br>of soft tissue in the medical specialties of general<br>and plastic surgery, and dermatology | Intended for use in aesthetic and cosmetic<br>applications requiring selective photothermolysis<br>(photocoagulation or coagulation) and hemostasis<br>of soft tissue in the medical specialties of general<br>and plastic surgery, and dermatology |
| Indications for<br>Use | Indicated for the treatment of:<br>• Benign cutaneous vascular lesions,<br>including port wine stains, hemangiomas,<br>facial, truncal and leg telangiectasias,<br>rosacea, erythema of rosacea,, angiomas<br>and spider angiomas, poikiloderma of<br>Civatte, leg veins and venous<br>malformations<br>• For Use on skin types (I-V) | Indicated for the treatment of:<br>• Moderate inflammatory acne (acne<br>vulgaris)<br>• Tattoos and benign pigmented epidermal<br>and cutaneous lesions including warts,<br>scars ,striae; dyschromia,<br>hyperpigmentation, melasma, epithelides<br>(freckles), lentigines, nevi, and café-au-lait<br>macules<br>• Benign cutaneous vascular lesions,<br>including port wine stains, hemangiomas,<br>facial, truncal and leg telangiectasias,<br>rosacea, erythema of rosacea, angiomas<br>and spider angiomas, poikiloderma of<br>Civatte, leg veins and venous<br>malformations<br>• For use on skin types (I-V) |
| Pulse Width | 10,12,15 | 10, 12, 15 |
| (msec) | | |
| Pulse<br>Repetition<br>Rate (Hz) | 2/3 | 2/3 |
| Energy<br>Density<br>(Fluence)<br>(J/cm²) | 10-30 | 10-30 |
| Spot Size<br>(mm) | 3 cm² | 3 cm² |
| | VP PRO – this submission | Previously Cleared K072564 – Alma IPL<br>handpiece VL/PL, VP and SSR |
| | | Primary Predicate |
| Characteristic | | |
| Wavelength | 540-950nm | 540-950nm |
| Intended Use | Intended for use in aesthetic and cosmetic<br>applications requiring selective<br>photothermolysis (photocoagulation or<br>coagulation) and hemostasis of soft tissue<br>in the medical specialties of general and<br>plastic surgery, and dermatology | Intended for use in aesthetic and cosmetic<br>applications requiring selective<br>photothermolysis (photocoagulation or<br>coagulation) and hemostasis of soft tissue in<br>the medical specialties of general and plastic<br>surgery, and dermatology |
| Indications for Use | Indicated for the treatment of:<br>The treatment of benign pigmented<br>epidermal lesions including<br>dyschromia, hyperpigmentation,<br>melasma, and ephelides (freckles),<br>lentigines, nevi, melasma, and<br>cafe-au-lait macules. The treatment of benign cutaneous<br>vascular lesions including port<br>wine stains, hemangiomas, facial,<br>truncal and leg telangiectasias,<br>rosacea, erythema of rosacea,<br>angiomas and spider angiomas,<br>poikiloderma of Civatte, leg veins<br>and venous malformations. Use on all skin types (Fitzpatrick I-<br>VI). | The treatment of moderate<br>inflammatory acne vulgaris. The treatment of benign pigmented<br>epidermal lesions including dyschromia<br>hyperpigmentation, melasma, and<br>ephelides (freckles), lentigines, nevi,<br>melasma, and cafe-au-lait macules. The treatment of cutaneous lesions<br>including warts, scars and striae. The treatment of benign cutaneous<br>vascular lesions including port wine<br>stains, hemangiomas, facial, truncal and<br>leg telangiectasias, rosacea, erythema of<br>rosacea, angiomas and spider angiomas,<br>poikiloderma of Civatte, leg veins and<br>venous malformations. The removal of unwanted hair to effect<br>stable long-term or permanent hair<br>reduction. Use on all skin types (Fitzpatrick I-VI). |
| Energy Density<br>(Fluence) (J/cm2) | 10-30 | 10-30 |
| Pulse Repetition<br>Rate (Hz) | 2/3 | 2/3 |
| Pulse Width<br>(msec) | 10, 12, 15 | 10, 12, 15 |
| Spot size | 3 cm2 | 3 cm2 |
| | This submission – SSR IPL handpiece | K072564 - Alma |
| | | SR IPL handpiece |
| | | Primary Predicate |
| Characteristic | | |
| Wavelength<br>(nm) | 570-950 | 570-950 |
| Intended Use | Intended for use in aesthetic and cosmetic<br>applications requiring selective<br>photothermolysis (photocoagulation or<br>coagulation) and hemostasis of soft tissue in<br>the medical specialties of general and<br>plastic surgery, and dermatology | Intended for use in aesthetic and cosmetic<br>applications requiring selective<br>photothermolysis (photocoagulation or<br>coagulation) and hemostasis of soft tissue in<br>the medical specialties of general and plastic<br>surgery, and dermatology |
| Indications for<br>Use | The Advanced Fluorescence<br>Technology (AFT) 570-950 nm SSR<br>Module handpiece (with and without<br>contact-cooling) is indicated for:<br>● The treatment of benign pigmented<br>epidermal lesions, including<br>dyschromia, hyperpigmentation,<br>melasma and ephelides (freckles).<br>● The treatment of face and body<br>vascular and pigmented lesions.<br>● The treatment of cutaneous lesions,<br>including scars and striae.<br>● The treatment of benign cutaneous<br>vascular lesions, including port<br>wine stains, hemangiomas, facial,<br>truncal and leg telangiectasias,<br>rosacea, melasma, angiomas and<br>spider angiomas, poikiloderma of<br>Civatte, leg veins and venous<br>malformations.<br>● Use on all skin types (Fitzpatrick I-<br>VI). | The Advanced Fluorescence Technology<br>(AFT) 570-950 nm SR Module handpiece<br>(with and without contact-cooling) is<br>indicated for:<br>● The treatment of moderate<br>inflammatory acne vulgaris.<br>● The treatment of benign pigmented<br>epidermal lesions, including<br>dyschromia, hyperpigmentation,<br>melasma and ephelides (freckles).<br>● The treatment of face and body<br>vascular and pigmented lesions.<br>● The treatment of cutaneous lesions,<br>including scars and striae.<br>● The treatment of benign cutaneous<br>vascular lesions, including port wine<br>stains, hemangiomas, facial, truncal<br>and leg telangiectasias, rosacea,<br>melasma, angiomas and spider<br>angiomas, poikiloderma of Civatte,<br>leg veins and venous malformations.<br>● The removal of unwanted hair to<br>effect stable long-term or permanent<br>hair reduction.<br>● Use on all skin types (Fitzpatrick I-<br>VI). |
| Energy Density<br>(Fluence)<br>(J/cm²) | 10-25 J/cm² | 10-25 J/cm² |
| Pulse<br>Repetition Rate<br>(Hz) | 2/3 | 2/3 |
| Pulse Width<br>(msec) | 10,12,15 msec | 10, 12, 15 msec |
| Spot size (mm) | 3 cm² | 3, 6 cm² |
| | This submission<br>SHR IPL handpiece | Previously Cleared K072564 - Alma SHR IPL |
| | | Primary Predicate |
| Characteristic | | |
| Wavelength<br>(nm) | 780-950 | 780-950 |
| Intended Use | Intended for use in aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery, and dermatology | Intended for use in aesthetic and cosmetic applications requiring selective photothermolysis (photocoagulation or coagulation) and hemostasis of soft tissue in the medical specialties of general and plastic surgery, and dermatology |
| Indications for Use | The Advanced Fluorescence Technology (AFT) 780-950 rnm SHR Module handpiece (with and without contact-cooling) is indicated for:<br>• Removal of unwanted hair<br>Effect stable or permanent hair reduction.<br>Permanent reduction in hair regrowth is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after the completion of a treatment regime<br>• The treatment of pseudofolliculitis barbae (PFB).<br>• Use on all skin types (Fitzpatrick I-VI), including tanned skin | The Advanced Fluorescence Technology (AFT) 780-950 rnm SHR handpieces (with and without contact-cooling) is indicated for:<br>• The treatment of pseudofolliculitis barbae (PFB).<br>• The removal of unwanted hair and to effect stable long-term or permanent hair reduction.<br>• Use on all skin types (Fitzpatrick I-VI), including tanned skin. |
| Pulse Width<br>(msec) | 30,40,50 msec…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.