APOLLO V+ MEDICAL PLATFORM, APOLLO IV+ MEDICAL PLATFORM, NICE STATION LIGHT BASED PLATFORM
K113018 · BEIJING SYNTECH LASER CO., LTD. · GEX · Dec 14, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113018
Device Name
APOLLO V+ MEDICAL PLATFORM, APOLLO IV+ MEDICAL PLATFORM, NICE STATION LIGHT BASED PLATFORM
Applicant
BEIJING SYNTECH LASER CO., LTD.
Product Code
GEX · General, Plastic Surgery
Decision Date
Dec 14, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
APOLLO V+ Medical Platform / APOLLO IV+ Medical Platform / Nice Station Light Based Platform, the three models have the same intended use. They can be used in dermatology, cosmetic medicine, and other surgical applications according to the different handpieces. The specific indications should reference to the indications of each handpiece.
Device Story
Modular multifunction platform (APOLLO V+, IV+, Nice Station) for dermatology and cosmetic surgery. System utilizes interchangeable handpieces: Er:YAG laser (2940 nm), Long Pulse Nd:YAG laser (1064 nm), Q-Switch Nd:YAG laser (1064 nm), and IPL (515-1200 nm). Energy generated via Xenon lamps and laser crystals; delivered to tissue to induce thermal effects (selective photothermolysis for IPL). Operated by clinicians in clinical settings via footswitch; parameters set via user interface. Includes cooling system (semiconductor-based for IPL), power supply, and safety features (emergency stop, interlock). Output enables tissue ablation, hair reduction, and lesion treatment. Device provides thermal energy to target tissues; clinical benefit includes cosmetic improvement and lesion removal.
Clinical Evidence
Bench testing only. Compliance verified against IEC 60825-1 (laser safety), IEC 60601-2-22 (therapeutic laser equipment), IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), and ISO 10993-5/10 (biocompatibility). No clinical data presented.
Indicated for incision, excision, ablation, and vaporization of soft tissue; non-ablative treatment of facial wrinkles; removal of unwanted hair for stable long-term or permanent reduction (skin types I-V); treatment of PFB; tattoo removal (dark ink: black/blue); treatment of Nevus of Ota; skin resurfacing for acne scars and wrinkles; treatment of benign pigmented lesions (e.g., warts); treatment of benign vascular lesions (e.g., port wine stains, hemangiomas, telangiectasias, rosacea, melasma, angiomas, spider angiomas, poikiloderma of civatte, leg veins, venous malformations).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Section III 510(k) Summary
Project #: MIDLINK0382010Bd
### Section III 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number:
- 1. Date of Submission: Apr 15, 2011
- 2. Sponsor
· Beijing Syntech Laser Co., Ltd.
5F, No.1 Building, No.65 Jiancaicheng West Road, Xisanqi, Haidian District, Beijing, 100096, P.R.China Contact Person: Jin Jinghua
Position: Vice general manager
Tel: +86-10-82939866-234
Fax: +86-10-82935528
Email: sales@syntechlaser.com
- 3. Submission Correspondent
Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 237-023, Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: info@mid-link.net
- 4. Proposed Device Identification
Proposed Device Name: APOLLO V+ Medical Piatform Proposed Device Model: MLP-01 Product Code: GEX Regulation Number: 21 CFR 878.4810 Review Panel: General& Plastic Surgery
{1}------------------------------------------------
Project #: MIDLINK0382010Bd
Proposed Device Name: APOLLO IV+ Medical Platform Proposed Device Model: MLP-02 Product Code: GEX Regulation Number: 21 CFR 878.4810 Review Panel: General& Plastic Surgery
Proposed Device Name: Nice Station Light Based Platform Proposed Device Model: MLP-04 Product Code: GEX Regulation Number: 21CFR 878.4810 Review Panel: General& Plastic Surgery
Intended Use Statement:
APOLLO V+ Medical Platform / APOLLO IV+ Medical Platform / Nice Station Light Based Platform, the three models have the same intended use. They can be used in dermatology, cosmetic medicine, and other surgical applications according to the different handpieces.
The specific indications should reference to the indications of each handpiece.
- 1) Er:YAG laser handpiece:
- Incision, excision, ablation, vaporization of soft tissue .
- The non-ablative treatment of facial wrinkles -
2) Long Pulse Nd:YAG Laser Handpiece:
- Removal of unwanted hair, for stable long term or permanent hair reduction and for treatment . of PFB.
- 3) Q-Switch Nd:YAG Laser Handpiece: Tattoo Removal: Dark Ink: (Black & Blue) Nevus of Ota Skin resurfacing procedures for the treatment of acne scars and wrinkles
- 4) IPL Handpiece:
The Intense Pulsed Light wavelengths are 515-1200 nm
HR (Hair Removal): 650-1200 nm
The removal of unwanted hair from skin types I-V, and to effect stable long-term, or permanent, hair reduction in skin types I-V through selective targeting of melanin in hair follicles
{2}------------------------------------------------
..
SR (Skin Remove): 570-1200nm
The treatment of benign pigmente (epidermal and cutaneous) lesions, such as warts.
VR (Vascular and Pigmented Lesion Removal): 515-1200nm
- - - - --
The treatment of benign vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, melasma, angiomas and spider anglormas, polkiloderma of civate; leg veins, and venous malformations.
{3}------------------------------------------------
...... ・・・・
- Predicate Device Identification 5.
- a. ER:YAG Laser Hanpiece K080374/ Bios Italia s.r.l/ BIOSYAG MEDICAL SYSTEM
- b. Long Pulse Nd: Yag Laser Hanpiece K072564/ Alma Laser, Ltd / Harmony XL Multiple Application Plateform
- Q-Switch Nd: Yag Laser Hanpiece c. K.072564/ Alma Laser, Ltd / Harmony XL Multiple Application Plateform
- d. IPL Laser Hanpiece K024093/ Lumenis Family of Intense Pulsed-Light (IPL) and IPL/Nd:YAG Laser Systems/ Lumenis, Inc.
{4}------------------------------------------------
#### Device Description 6.
. APOLLO V+ Medical Platform / APOLLO IV+ Medical Platform / Nice Station Light. Based Platform, the three models have the same modules and handpieces. The equipment configuration is the only difference among the three models.
APOLLO V+ Medical Platform is a modular multifunction device. The equipment can be used in dermatology, cosmetic medicine and other surgical according to the different handpieces. The APOLLO V+ Medical Platform has three laser handpieces of Er: YAG laser, Long Pulse Nd: YAG Laser, and Q-Switch Nd: YAG Laser, and one IPL handpiece.
APOLLO IV+ Medical Platform is a modular multifunction device. The equipment can be used in dermatology, cosmetic medicine and other surgical according to the different handpieces. The APOLLO IV+ Medical Platform has three laser handpieces of Er: YAG laser, Long Pulse Nd: YAG Laser, and Q-Switch Nd:YAG Laser, and one IPL handpiece. There is one difference between APOLLO IV+ Medical Platform and APOLLO V+ Medical Platform, it is "desktop" style for APOLLO IV+ Medical Platform, and a standard case, with wheels to allow easy movement on the floor, for APOLLO V+ Medical Platform.
Nice Station Light Based Platform is a modular multifunction device. The equipment can be used in dermatology, cosmetic medicine and other surgical according to the different handpieces. The Nice Station Light Based Platform has three laser handpieces of Er:YAG laser, Long Pulse Nd:YAG Laser, and Q-Switch Nd: YAG Laser, and one IPL handpiece. The Nice Station Light Based Platform has a standard case, with wheels to allow easy movement on the floor. The difference between Nice Station Light Based Platform and APOLLO V+ Medical Platform is the appearance.
The user can activate laser emission, IPL emission by means of a footswitch.
APOLLO V+ Medical Platform / APOLLO IV+ Medical Platform / Nice Station Light Based Platform includes:
- a) Power supply
The system provides energy for the operation of equipment.
- b) Control system
The system controls the cooling system, Laser emission and delivery system and safety features according the setting parameter which is set by user via the user interface.
- c) User interface
.
User sets treatment parameter via the user interface.
- d) Cooling system,
The cooling system consists of whole device cooling and IPL handpiece skin cooling.
- Laser emission and delivery system e)
111-2
{5}------------------------------------------------
Project #: MIDLINK0382010Bd
- f) Safety features
The safety features are used to ensure the equipments is operated properly, such as key switch, -- emergency stop switch and interlock control unit.
- g) Many handpiece types
- Er: YAG handpiece, 2940 nm wavelength t
- . Long Pulse Nd: YAG laser handpiece, 1064 nm wavelength
- . Q-Switch Nd: YAG laser handpiece, 1064 nm wavelength
- IPL handpiece. 650-1200nm wavelengths in RH filter, 570-1200nm wavelengths in SR filter, 515-1200 nm wavelengths in VR filter.
All of the housing of handpieces are made of ABS (Acrylonitrile Butadiene Stryene), These handpieces can be removed by the user and interchanged.
### The principle of generation and treatment of energy sources
There are three kinds of energy sources, laser energy, IPL energy
- The principle of generation and treatment of laser energy .
In laser handpiece, there is one optical cavity containing the laser crystal (such as Er. Y AG crystal,
- Nd:YAG crystal). The laser beam is activated by means of the use of Xenon lamps.
This beam is directed to the treatment zone by means of a laser viewfinder. When the laser beam contacts human tissue, the energy in the beam is absorbed, resulting in a very rapidly, highly localized temperature increase to the target tissue. The instantaneous temperature increase (thermal effect) causes the cells change of target tissue to achieve laser treatment effect.
- The principle of generation and treatment of IPL energy
In IPL module, the IPL (Intense Pulsed Light) which is activated by means of the use of Xenon lamp is directed to the treatment zone via the sapphire in selectable wavelengths. The IPL works on the principle of selective photothermolysis. That is, causing thermal damage to target tissue by using light of appropriate wavelength. The IPL is different from lasers in that it has many wavelengths in each pulse of light instead of just one wavelength. There are three filtered light guide for the selection of the wavelengths range.
- 7. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- . IEC 60825-1: 2007, Safety of laser products - Part 1: Equipment classification, requirements.
- IEC 60601-2-22: 1995, Medical Electrical Equipment - Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipment.
- . IEC 60601-1:1988+A1:1991+A2:1995, Medical Electrical Equipment - Partl: General requirements for safety.
- IEC60601-1-2:2001+A1:2004, Medical Electrical Equipment - Part 1: General requirements for III-Q
{6}------------------------------------------------
safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
- . ISO 10993-5:2009 Biocompatibility Evaluation of Medical Device, Part 5- Tests for Vitro cytotoxicity.
- ISO 10993-10:2002 Biocompatibility Evaluation of Medical Device, Part 10- Test for irritation and delay-type hypersensitivity.
#### 8. Substantially Equivalent Conclusion
About Laser handpiece and IPL handpiece, the subject device and the predicate device have the same output energy, pulse width, repetition rate and spot size.
The subject device does not have a aiming beam compared to the proposed device has a focusing device (viewfinder) fitted on the handpiece head instead of aiming beam, the focusing device has the same intended use as the aiming beam, and the Biocompatibility test of the viewfinder demonstrates that the focusing device (viewfinder) does not raise the problems of safety and effectiveness.
Although the cooling method for treated skin area between the subject device with IPL handpiece and the predicate devices is different, the subject device with IPL handpiece complies with IEC 60601-1. The temperature-of-IPL-handpiece-meets-the-requirements-which-are-specified-in-1EC-60601-1-The semiconductor absorbs the heat of the surrounding material to achieve the equivalent cooling effect.
We can't identify the power calibration (method and frequency) and disinfection method for the predicated devices, but we can provide test report to prove that the power calibration (method and . frequency) and disinfection method for the subject device are safe and effective.
The proposed devices have the same energy source, similar main technological characteristics and intended use as the predicate device, based on the comparison and analysis, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, sans-serif font. The text is arranged on a single line and is centered horizontally.
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular fashion around the eagle. The logo is black and white and appears to be a simple, clean design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Underwriters Laboratories, Incorporated % Mr. Marc Mouser Engineering Leader, FDA Office Coordinator 2600 Northwest Lake Road Camas, Washington 98607-9526
December 12, 2012
Re: K113018
Trade/Device Name: APOLLO V+ Medical Platform/APOLLO IV+ Medical Platform/Nice -Station Light Based Platform
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology
Regulatory Class: Class II Product Code: GEX Dated: November 26, 2012 Received: November 29, 2012
Dear Mr. Mouser:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
Page 2 - Mr. Marc Mouser
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
## Peter D. Rumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Apollo V+ Medical platform / Apollo IV+ Medical Platform / Nice Station Light Based Platform, the three models have the same intended use. They can be used in dermatology, cosmetic surgery, and other surgical applications according to the different hand pieces. The specific indications should reference to the indications of each hand piece.
1. Er: YAG Laser handpiece:
Incision, excision, ablation, vaporization of soft tissue
The non-ablative treatment of facial wrinkles
2. Long pulse Nd:YAG Laser handpiece:
Removal of unwanted hair, for stable long term or permanent hair reduction, and for treatment of PFB.
3. Q-Switched Nd: YAG Laser Hand piece:
1064nm wavelength
Tattoo removal: Dark ink: (Black
Nevus of Ota
Skin resurfacing procedures for the treatment of acne scars and wrinkles
5. IPL Handpiec
The Intense Pulse Light wavelength are
HR (Hair removal): 650-1200
- V, and to effect stable long-term, or The removal of unwanted hair from skin type through selective targeting of melanin in hair permanent hair reduction in skin types I follicles.
# Peter D. Rumm -S
(Division Sign-Off)
Division of Surgical Devices
510(k) Number_K113018
{10}------------------------------------------------
### SR (skin remove): 570-1200nm
The treatment of benign pigmente (epidermal and coutaneous) lesions, such as warts.
VR (Vascular and Pigmented Lesions Removal
..............................................................................................................................................................................
The treatment of benign vascular lesions including port wine stains, hemangiomas, facial, truncal and leg telangiectasias, rosacea, melasma, angiomas and spider anglormas, polkiloderma of civatte; leg veins and venous malformations.
# Peter D. I
(Division Sign-Off)
Division of Surgical Devices
510(k) Number_K113018
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.