K254253 · Aap Implantate AG · HRS · Mar 24, 2026 · Orthopedic
Device Facts
Record ID
K254253
Device Name
LOQTEQ® VA Proximal Humerus Plate 3.5
Applicant
Aap Implantate AG
Product Code
HRS · Orthopedic
Decision Date
Mar 24, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LOQTEQ® VA Proximal Humerus Plate 3.5 is indicated for fractures and fracture dislocations, osteotomies, and nonunions of the proximal humerus, particularly in osteopenic bone.
Device Story
LOQTEQ® VA Proximal Humerus Plate 3.5 is a metallic bone fixation system designed for surgical stabilization of proximal humerus fractures. The system consists of titanium alloy plates (short and long variants) and locking screws (cortical and cancellous). Plates feature variable angle (VA) locking holes allowing for polyaxial screw placement, providing stable fixation in osteopenic bone. The device is used by orthopedic surgeons in an operating room setting. The surgeon selects the appropriate plate and screw configuration based on fracture morphology, secures the plate to the bone using the locking screws, and achieves rigid fixation. This stabilizes the fracture site, facilitating bone healing and restoring anatomical alignment. The device is provided non-sterile and requires moist heat sterilization by the end user prior to surgery.
Clinical Evidence
No clinical data were included in this submission. Substantial equivalence is supported by non-clinical bench testing, including mechanical testing of plate/screw constructs using a fracture gap model, ASTM F543 screw testing, and MR environment testing (ASTM F2182, F2052, F2213, F2119). Sterilization validation is referenced from K233008.
Technological Characteristics
Materials: Ti-6Al-4V alloy (ASTM F136, ISO 5832-3). Design: Metallic bone plates with variable angle (VA) locking screw holes and cortical/cancellous locking screws. Dimensions: Plates 3.2-3.9 mm thick, 12.4-21.5 mm wide. Sterilization: Moist heat (autoclave) to SAL 10^-6. Connectivity: None (mechanical implant).
Indications for Use
Indicated for patients with fractures, fracture dislocations, osteotomies, and nonunions of the proximal humerus, including those with osteopenic bone.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
LOQTEQ® Proximal Humerus Plate 3.5 System (K211582)
Reference Devices
LOQTEQ® Proximal Humerus Plate 3.5 System (K121495)
Synthes (USA) LCP® Proximal Humerus Plates, Long (K041860)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 24, 2026
Aap Implantate AG
% Kevin Thomas
Vice President & Director of Regulatory Affairs
PaxMed International, LLC
1925 Palomar Oaks Way, Suite 210
Carlsbad, California 92008
Re: K254253
Trade/Device Name: LOQTEQ® VA Proximal Humerus Plate 3.5
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: HRS, HWC
Dated: December 29, 2025
Received: December 29, 2025
Dear Kevin Thomas:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254253 - Kevin Thomas
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13484 clause 8.3 (Nonconforming product), and ISO 13485 clause 8.5 (Corrective and preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254253 - Kevin Thomas
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K254253
Device Name
LOQTEQ® VA Proximal Humerus Plate 3.5
Indications for Use (Describe)
LOQTEQ® VA Proximal Humerus Plate 3.5 is indicated for fractures and fracture dislocations, osteotomies, and nonunions of the proximal humerus, particularly in osteopenic bone.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
{4}
510(k) Summary
Page 1 of 4
LOQTEQ® VA Proximal Humerus Plate 3.5
K254253
510(k) Summary
K254253
LOQTEQ® VA Proximal Humerus Plate 3.5
*aap Implantate AG*
March 12, 2026
## ADMINISTRATIVE INFORMATION
Manufacturer Name
*aap Implantate AG*
Lorenzweg 5
12099 Berlin Germany
Telephone +49 30 750 19 0
Official Contact
Andreas Reuter
Director of Research and Development Trauma, and Regulatory Affairs
Representative/Consultant
Kevin A. Thomas, PhD
Rebecca E. Kattan, PhD
Floyd G. Larson, MS, MBA
PaxMed International, LLC
1925 Palomar Oaks Way, Suite 210
Carlsbad, CA 92008
Telephone +1 858-792-1235
Email kthomas@paxmed.com
rkattan@paxmed.com
flarson@paxmed.com
## DEVICE NAME AND CLASSIFICATION
Trade/Device Name
LOQTEQ® VA Proximal Humerus Plate 3.5
Common Name
Plate, Fixation, Bone
Primary Product Code
HRS – Plate, Fixation, Bone
Regulation Number
21 CFR 888.3030
Regulation Name
Single/multiple component metallic bone fixation appliances and accessories
Regulatory Class
Class II
Secondary Product Code
HWC – Screw, Fixation, Bone
Regulation Number
21 CFR 888.3040
Regulation Name
Smooth or threaded metallic bone fixation fastener
Regulatory Class
Class II
Classification Panel
Orthopedic
Reviewing Division
Office of Orthopedic Devices (OHT6)
Division of Restorative, Repair and Trauma Devices (DHT6C)
Restorative, Repair and Trauma Devices
## PREDICATE DEVICE INFORMATION
Primary Predicate Device
K211582, LOQTEQ® Proximal Humerus Plate 3.5 System, *aap Implantate AG*
Reference Devices
K121495, LOQTEQ® Proximal Humerus Plate 3.5 System, *aap Implantate AG*
K041860, Synthes (USA) LCP® Proximal Humerus Plates, Long, Synthes (USA)
{5}
510(k) Summary
Page 2 of 4
LOQTEQ® VA Proximal Humerus Plate 3.5
K254253
K230141, LOQTEQ® VA Elbow Plates 2.7/3.5 System, aap Implantate AG
K113652, aap LOQTEQ® Small Fragment Set, aap Implantate AG
K233008, Cannulated Headless Bone Screws, aap Implantate AG
## INDICATIONS FOR USE STATEMENT
LOQTEQ® VA Proximal Humerus Plate 3.5 is indicated for fractures and fracture dislocations, osteotomies, and nonunions of the proximal humerus, particularly in osteopenic bone.
## SUBJECT DEVICE DESCRIPTION
The subject device LOQTEQ® VA Proximal Humerus Short Plates are 21.5 mm wide at the proximal end, 12.4 mm wide along the shaft, and 3.2 mm thick. The shaft of each of the subject device short plates has one (1) oblong, non-locking screw hole, and 2, 3, 4, 5, or 6 VA (variable angle) locking 3.5 mm screw holes.
The subject device LOQTEQ® VA Proximal Humerus Long Plates are 21.5 mm wide at the proximal end, 12.8 mm wide along the shaft, and 3.9 mm thick. The shaft of each of the subject device long plates has one (1) oblong, non-locking screw hole, one (1) VA 3.5 mm screw hole, and 6, 8, 10, 12, or 14 LOQTEQ® 3.5 mm gliding locking-compression screw holes.
The subject device LOQTEQ® Cortical Screws have a standard cortical thread, 3.5 mm in diameter that tapers slightly at the tip. The heads of the screws are threaded to lock into the VA variable angle holes in the subject device plates. The screws are provided in overall lengths of 20 mm to 60 mm, in 2 mm increments.
The subject device LOQTEQ® Cancellous Screws have a standard cancellous thread, 4.0 mm thread diameter that tapers slightly at the tip. The heads of the screws are threaded to lock into the VA variable angle holes in the subject device plates. The screws are provided in overall lengths of 20 mm to 60 mm, in 2 mm increments.
The subject device plates and screws are manufactured from Ti-6Al-4V alloy conforming to ASTM F136 and ISO 5832-3.
## PERFORMANCE DATA
Recognized standards used in the non-clinical performance testing included:
ASTM F136 Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401);
ISO 5832-3 Implants for surgery – Metallic materials – Part 3: Wrought titanium 6-aluminium 4-vanadium alloy;
ASTM F2182 Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance;
ASTM F2052 Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the Magnetic Resonance Environment;
ASTM F2213 Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the Magnetic Resonance Environment;
ASTM F2119 Standard Test Method for Evaluation of MR Image Artifacts from Passive Implants;
ASTM F2182 Standard Test Method for Measurement of Radio Frequency Induced Heating On or Near Passive Implants During Magnetic Resonance;
ANSI/AAMI/ISO 17665-1 Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices;
ANSI/AAMI/ISO TIR 17665-2 Sterilization of health care products – Moist heat – Part 2: Guidance on the application of ISO 17665-1; and
{6}
510(k) Summary
Page 3 of 4
LOQTEQ® VA Proximal Humerus Plate 3.5
K254253
ASTM F543 Standard Specification and Test Methods for Metallic Medical Bone Screws.
Non-clinical data submitted or referenced to demonstrate substantial equivalence included:
- referenced from the reference device K233008 was moist heat sterilization validation to a sterility assurance level of 10⁻⁶ by the overkill method according to ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO TIR 17665-2; analysis showed that the subject devices do not create a new worst case for moist heat sterilization;
- referenced from the primary predicate device K211582 and the reference device K121495 was biocompatibility for the subject device materials;
- provided in this submission was testing of the subject device within the magnetic resonance environment according to ASTM F2182, ASTM F2052, ASTM F2213, ASTM F2119, and ASTM F2182;
- provided in this submission was mechanical testing of the subject device plate and screw constructs and mechanical testing of predicate device plate and screw constructs using a fracture gap model; and
- provided in this submission was mechanical testing according to ASTM F543.
No clinical data were included in this submission.
## EQUIVALENCE TO MARKETED DEVICES
The subject device and the primary predicate device K211582 have the same technological characteristics and use the same operating principles for bone fixation. The subject device and the primary predicate device include similar anatomic designs for placement on the proximal humerus, with screw holes to accommodate locking and non-locking screws. The subject device designs in lengths ranging from 84 mm to 255 mm with 3-16 holes, are similar in lengths (85 mm to 247 mm) to the plates cleared in K211582 and K041860.
The subject device plates incorporate multidirectional polyaxial or VA (variable angle) locking screw hole technology that is identical to that cleared in the reference device K230141.
The subject device LOQTEQ® cortical locking screws (3.5 mm diameter, 20-60 mm length) are substantially equivalent in design to screws cleared in the reference device K113652 (LOQTEQ® cortical locking screws 3.5 mm diameter, 10-90 mm length). The subject device LOQTEQ® cancellous locking screws are substantially equivalent to LOQTEQ® cancellous locking screws cleared in the reference device K121495.
The subject devices are compatible with aap 3.5 mm cortical screws previously cleared in K141949, and with 3.5 mm LOQTEQ® 3.5 mm cortical locking screws cleared in K113652. The subject device plates also are compatible with K Wires previously cleared in K131459.
The subject device plates, the plates cleared in the primary predicate device K211582, and the plates cleared in the reference devices K121495 and K230141 are manufactured from the identical titanium alloy material conforming to ASTM F136 and ISO 5832-3. Similarly, the subject device screws, and the screws cleared in the reference devices K121495 and K233008 are manufactured from the identical titanium alloy material conforming to ASTM F136 and ISO 5832-3, and are anodized using a process identical to that used for screws cleared in K211582.
All subject device final finished components are manufactured in the same facilities using identical materials and identical manufacturing processes as used for previously cleared aap components (K211582, K121495, K230141, and K233008). Similarly, the Class II device-specific accessories and Class I instruments are manufactured in the same facilities using identical materials and identical manufacturing processes as the Class I instruments and cleared Class II accessories in K211582 and K233008. Therefore, the subject devices are substantially equivalent to the primary predicate and reference devices (K211582, K121495, K230141, and K233008) with regard to biocompatibility.
{7}
510(k) Summary
Page 4 of 4
LOQTEQ® VA Proximal Humerus Plate 3.5
K254253
The subject device includes components provided non-sterile in the same packaging as devices previously cleared in K230141, to be sterilized to a sterility assurance level (SAL) of 10-6 by the end user using the same sterilization method (moist heat) and parameters as devices previously cleared in K233008.
Sterilization validation for the subject device components assembled in trays is referenced from K233008. The moist heat sterilization validation referenced from K233008 was performed to a sterility assurance level of $10^{-6}$ by the overkill method according to ISO 17665-1 and ISO TS 17665-2. The referenced sterilization validation included a worst case kit and configuration based on design features that are the most challenging to the moist heat sterilization process, such as presence of lumen, geometry, and surface area. The subject device plates and screws do not create a new worst-case and do not introduce new design features that create a greater challenge to the sterilization process.
In support of substantial equivalence of the strength of the subject plates dynamic mechanical testing was performed using a fracture gap construct model. In support of substantial equivalence of the strength of the subject device screws mechanical testing was performed according to ASTM F543.
## CONCLUSION
The subject devices, the primary predicate device, and the reference devices have the same intended use and similar technological characteristics. They encompass a similar range of physical dimensions, are manufactured from the same materials, and are to be sterilized using identical methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.