K233008 · Aap Implantate AG · HWC · Jun 5, 2024 · Orthopedic
Device Facts
Record ID
K233008
Device Name
Cannulated Headless Bone Screws
Applicant
Aap Implantate AG
Product Code
HWC · Orthopedic
Decision Date
Jun 5, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The aap Cannulated Headless Bone Screws are indicated for use in non-spinal fragment fixation, osteotomy and arthrodesis of human bones appropriate for the size of the screw. The devices are intended for single use only. The aap Cannulated Headless Bone Screws are indicated for adults, and children (2 years to less than 12 years) and adolescents (aged 12 through 21, up to but not including the 22nd birthday) in which growth plates have fused or in which growth plates will not be crossed by fixation. K-Wires are indicated for use as guide wire for osteosynthesis implants and for application as implant according to the AO/ASIF principles of fracture management. The K-Wires are indicated for adults, and children (2 years to less than 12 years) and adolescents (aged 12 through 21, up to but not including the 22nd birthday) in which growth plates have fused or in which growth plates will not be crossed by fixation.
Device Story
Cannulated headless bone screws and K-wires for orthopedic fixation; screws feature cannulated design at leading/trailing ends to facilitate fracture gap closure and compression. Screws provided in various diameters (2.5/3.4 mm to 7.5/7.9 mm) and lengths. K-wire (2.4 mm diameter, 250 mm length) serves as guide for implant placement. Used in clinical settings by surgeons for bone fragment fixation, osteotomy, and arthrodesis. Devices provided non-sterile; require moist heat sterilization by end-user. Mechanical performance verified via ASTM F543 testing.
Clinical Evidence
No clinical data included. Substantial equivalence supported by bench testing, including mechanical testing per ASTM F543, biocompatibility assessment, and sterilization validation (moist heat, SAL 10^-6).
Technological Characteristics
Screws: Ti-6Al-4V alloy (ASTM F136, ISO 5832-3). K-Wire: Stainless steel (ASTM F138, ISO 5832-1). Cannulated design; trocar point for K-wire. Sterilization: Moist heat (ISO 17665-1/2).
Indications for Use
Indicated for non-spinal fragment fixation, osteotomy, and arthrodesis in adults, children (2-12 years), and adolescents (12-21 years) where growth plates are fused or not crossed by fixation. K-Wires indicated as guide wires for osteosynthesis implants and fracture management.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
June 5, 2024
aap Implantate AG % Kevin Thomas Vice President & Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K233008
Trade/Device Name: Cannulated Headless Bone Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: May 14, 2024 Received: May 14, 2024
Dear Kevin Thomas:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Shumaya Ali
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices
{2}------------------------------------------------
OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K233008
#### Device Name
Cannulated Headless Bone Screws
#### Indications for Use (Describe)
#### Cannulated Headless Bone Screws
The aap Cannulated Headless Bone Screws are indicated for use in non-spinal fragment fixation, osteotomy and arthrodesis of human bones appropriate for the size of the screw. The devices are intended for single use only. The aap Cannulated Headless Bone Screws are indicated for adults, and children (2 years to less than 12 years) and adolescents (aged 12 through 21, up to but not including the 22nd birthday) in which growth plates have fused or in which growth plates will not be crossed by fixation.
#### K-Wire
K-Wires are indicated for use as guide wire for osteosynthesis implants and for application as implant according to the AO/ASIF principles of fracture management.
The K-Wires are indicated for adults, and children (2 years to less than 12 years) and adolescents (aged 12 through 21, up to but not including the 22nd birthday) in which growth plates have fused or in which growth plates will not be crossed by fixation.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------|------------------------------------------------------------------------------|
| <div> <span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K233008 Cannulated Headless Bone Screws aap Implantate AG
May 30, 2024
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | aap Implantate AG |
|---------------------------|-------------------------------------------------|
| | Lorenzweg 5 |
| | 12099 Berlin Germany |
| | Telephone +49 30 750 19 129 |
| | Fax +49 30 750 19 111 |
| Official Contact | Agnieszka Mierzejewska, Chief Operating Officer |
| Representative/Consultant | Kevin A. Thomas, PhD |
| | Floyd G. Larson, MS, MBA |
| | PaxMed International, LLC |
| | 12264 El Camino Real, Suite 400 |
| | San Diego, CA 92130 |
| | Telephone +1 858-792-1235 |
| | Fax +1 858-792-1236 |
| | Email kthomas@paxmed.com |
| | flarson@paxmed.com |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Device Name<br>Common Name | Cannulated Headless Bone Screws<br>Screw, fixation, bone |
|----------------------------------|----------------------------------------------------------------------------|
| Regulation Number | 21 CFR 888.3040 |
| Regulation Name | Smooth or threaded metallic bone fixation fastener |
| Regulatory Class | Class II |
| Primary Product Code | HWC |
| Additional Product Codes | HTY |
| Classification Panel | Orthopedic |
| Reviewing Office | Office of Health Technology 6 (OHT 6: Orthopedic Devices) |
| Reviewing Division | Division of Health Technology 6 C (Restorative, Repair and Trauma Devices) |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device K111316, aap Cannulated Screws, aap Implantate AG
Reference Devices K130590, aap Cannulated Screw 2.0, aap Implantate AG K131459, aap Wire Bone, aap Implantate AG K 161616, DePuy Synthes 4.0 mm and 4.5 mm Cortex Screws, DePuy Synthes 2.4 mm Cannulated Screws, DePuy Synthes 3.5 mm and 4.0 mm Cannulated Screws, DePuy Synthes 4.5 mm Cannulated Screws, DePuy Synthes 6.5 mm Cannulated Screws, DePuy Synthes 7.0 mm Cannulated Screws, DePuy Synthes 1.5 mm Headless, Synthes USA Products, LLC
{5}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
#### Cannulated Headless Bone Screws
The aap cannulated headless bone screws are indicated for use in non-spinal fracture and fragment fixation, osteotomy and arthrodesis of human bones appropriate for the screw. The devices are intended for single use only. The aap Cannulated Headless Bone Screws are indicated for adults, and children (2 years to less than 12 years) and adolescents (aged 12 through 21, up to but not including the 22nd birthday) in which growth plates have fused or in which growth plates will not be crossed by fixation.
#### K-Wire
K-Wires are indicated for use as guide wire for osteosynthesis implants and for application as implant according to the AO/ASIF principles of fracture management.
The K-Wires are indicated for adults, and children (2 years) and adolescents (aged 12 through 21, up to but not including the 22nd birthday) in which growth plates have fused or in which growth plates will not be crossed by fixation.
#### SUBJECT DEVICE DESCRIPTION
The subject device Cannulated Headless Bone Screws are cannulated at both the leading end (tip) and at the trailing end. The leading/trailing diameters are: 2.5/3.4 mm. 3.5/4.5 mm. 4.0/4.8 mm. 4.5/5.3 mm, 5.5/6.2 mm, 6.5/7.2 mm, and 7.5/7.9 mm., The leading thread have different pitches which facilitates closure of any fracture gap and generates compression across the bone fragments. The subject screws are provided in overall lengths ranging from 12 mm, and with threaded lengths (at the leading end or tip) ranging from 4 mm to 40 mm. The subject screws are manufactured from Ti-6Al-4V alloy conforming to ASTM F136 and ISO 5832-3.
The subject K-Wire is provided in one size with a diameter of 2.4 mm and a length of 250 mm. The subject K-Wire has a trocar point, is threaded, and is manufactured from stainless steel conforming to ASTM F138 and ISO 5832-1.
## PERFORMANCE DATA
Recognized standards used in the non-clinical performance testing included:
ASTM F136 (2013) Standard Specification for Wrought Titanium-4 Vanadium ELI (Extra Low Interstitial) Allov for Surgical Implant Applications (UNS R56401);
ISO 5832-3 (2021) Implants for surgery - Metallic materials - Part 3: Wrought titanium 6-aluminium 4vanadium alloy;
ASTM F138 (2019) Standard Specification for Wrought 18Chromium-14Nickel-2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants (UNS S31673);
ISO 5832-1 (2016) Implants for surgery - Metallic materials - Part 1: Wrought stainless steel;
ISO 17665-1 (2006) Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation, and routine control of a sterilization process for medical devices:
ISO 17665-2 (2009) Sterilization of health care products – Moist heat – Part 2: Guidance on the application of ISO 17665-1;
ISO 10993-1 (2018) Biological evaluation of medical devices – Part I : Evaluation and testing within a risk management process; and
ASTM F543 (2017) Standard Specification and Test Methods for Metallic Medical Bone Screws.
Non-clinical data submitted or referenced to demonstrate substantial equivalence included:
- referenced from for the primary predicate K111316 and the reference devices K130590 and K131459 . was moist heat sterilization validation to a sterility assurance level of 10° by the overkill method according to
{6}------------------------------------------------
ANSI/AAMI/ISO 17665-1, ANSI/AAMI/ISO TIR 17665-2, and ANSI/AAMI/ISO 14937; analysis showed that the subject devices do not create a new worst case for moist heat sterilization;
- referenced from K111316, K130590, and K131459 was packaging and shelf life information for the subject . devices:
- referenced from K111316, K130590, K131459 was biocompatibility for the subject device materials; ●
- provided in this submission was mechanical testing of the subject device screws according to ASTM F543. ●
No clinical data were included in this submission.
# EQUIVALENCE TO MARKETED DEVICES
The subject device, the primary predicate device K111316, and the reference devices K130590 and K131459 have the same technological characteristics and use the same operating principles for bone fracture fixation (bone screws and K-Wire) or for use as an instrument (K-Wire to facilitate placement of cannulated screws). The reference device K161616 is to support the use of the subject screws in the pediatric population.
The subject device bone screws are provided in ranges of thread diameters, threaded lengths that are similar to the primary predicate K111316 and the reference device K130590. The subject screw designs also are similar to the designs of screws cleared in K111316 and K130590.
The subject device K-Wire has a diameter of 2.4 mm, an overall length of 250 mm, and a trocar point. These dimensions are within the range of K-Wires (with trocar point) cleared in K131459.
The subject device screws are manufactured from titanium alloy conforming to ASTM F136 and ISO 5832-3 using materials and processes identical to that used for screws cleared in the primary predicate device K111316 and the reference device K130590.
The subject device K-Wire is manufactured from stainless steel confirming to ASTM F138 or ISO 5832-1 using processes identical to that used for K-Wires cleared in the reference device submission K131459.
All subject device final finished components are manufactured in the same facilities using identical and identical manufacturing processes as used for the previously cleared aap device components. Therefore, the subject devices are substantially equivalent to predicate devices regarding biocompatibility.
The subject device components are provided non-sterile in the same packaging as in the primary predicate K111316 and the reference devices K130590 and K131459, and are to be sterilized to a sterility assurance level (SAL) of 10 the end user using the same sterilization method (moist heat) and parameters as the previously cleared devices. The subject devices do not represent a new worst-case for the sterilization.
In support of substantial equivalence of the subject device screws mechanical testing was performed according to ASTM F543.
## CONCLUSION
The subject devices, the primary predicate device, and the reference devices have the same intended use and similar technological characteristics. They encompass a similar range of physical dimensions, are manufactured from the same materials, and are to be sterilized using identical methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.