K211582 · Aap Implantate AG · HRS · Oct 18, 2021 · Orthopedic
Device Facts
Record ID
K211582
Device Name
LOQTEQ Proximal Humerus Plate 3.5 System
Applicant
Aap Implantate AG
Product Code
HRS · Orthopedic
Decision Date
Oct 18, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LOQTEQ® Proximal Humerus Plate 3.5 System is intended for fractures and fracture dislocations, osteotomies, and nonunions of the proximal humerus, particularly in osteopenic bone.
Device Story
The LOQTEQ® Proximal Humerus Plate 3.5 System consists of metallic bone plates and compatible screws used for internal fixation of bone fragments in the proximal humerus. The device is provided non-sterile and is intended for single-patient, single-use. It is designed for surgical implantation by orthopedic surgeons to stabilize fractures, osteotomies, or nonunions. The plates feature anatomic designs with various lengths and hole configurations to accommodate locking and non-locking cortical and periprosthetic bone screws. The system utilizes established mechanical principles for bone fixation, providing stability to the bone segments to facilitate healing. The device is manufactured from titanium alloy (ASTM F136). Clinical benefit is derived from the stabilization of bone injuries, particularly in patients with compromised bone quality (osteopenic bone).
Clinical Evidence
No clinical data were submitted. Substantial equivalence was demonstrated through non-clinical testing, including biocompatibility (referenced from K121495), validation of moist heat sterilization per ANSI AAMI ISO 17665-1, and biomechanical performance assessment via cross-sectional analysis compared to the predicate device.
Technological Characteristics
Metallic bone fixation plates and screws made of Ti-6Al-4V alloy (ASTM F136). Features gliding-holes (locking/compression) and round holes (locking only). Available in lengths 85-247 mm. Sterilization: moist heat (end-user).
Indications for Use
Indicated for patients with fractures, fracture dislocations, osteotomies, and nonunions of the proximal humerus, particularly those with osteopenic bone.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
aap LOQTEQ® Proximal Humerus Plate 3.5 System (K121495)
Synthes LCP Proximal Humerus Plate, Long (K041860)
{0}------------------------------------------------
October 18, 2021
aap Implantate AG % Melissa Burbage Sr. Regulatory Specialist PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K211582
Trade/Device Name: LOQTEQ® Proximal Humerus Plate 3.5 System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: May 19, 2021 Received: May 21, 2021
Dear Melissa Burbage:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K211582
Device Name
LOQTEQ® Proximal Humerus Plate 3.5 System
Indications for Use (Describe)
The LOQTEQ® Proximal Humerus Plate 3.5 System is intended for fracture dislocations, osteotomies, and nonunions of the proximal humerus, particularly in osteopenic bone.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
## K211582
## LOQTEQ® Proximal Humerus Plate 3.5 System
## aap Implantate AG
October 15, 2021
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | aap Implantate AG<br>Lorenzweg 5<br>12099 Berlin Germany<br>Telephone: +49 30 75019129<br>Fax: +49 30 75019111 | | |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Official Contact | Agnieszka Mierzejewska, Chief Operating Officer | | |
| Representative/Consultant | Melissa Burbage<br>Kevin A. Thomas, PhD; Floyd G. Larson, MS, MBA<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 858-792-1235<br>Fax: +1 858-792-1236<br>Email: mburbage@paxmed.com<br>kthomas@paxmed.com; flarson@paxmed.com | | |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Device Name | LOQTEQ® Proximal Humerus Plate 3.5 System |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name | Plate, Fixation, Bone |
| Regulation Number | 21 CFR 888.3030 |
| Regulation Name | Single/multiple component metallic bone fixation appliances and<br>accessories |
| Regulatory Class | Class II |
| Product Code | HRS |
| Classification Panel | Orthopedic |
| Reviewing Division | Office of Orthopedic Devices (OHT6)<br>Division of Restorative, Repair and Trauma Devices (DHT6C)<br>Stereotaxic, Bone Growth Stimulators and Fracture Fixation<br>Devices Team |
#### PREDICATE DEVICE INFORMATION
Primary Predicate K121495, aap LOQTEQ® Proximal Humerus Plate 3.5 System Additional Predicate K041860, Synthes LCP Proximal Humerus Plate, Long
{4}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
The LOQTEQ® Proximal Humerus Plate 3.5 System is intended for fractures and fracture dislocations, osteotomies, and nonunions of the proximal humerus, particularly in osteopenic bone.
### SUBJECT DEVICE DESCRIPTION
The subject device includes a total of nine (9) bones plates for internal fixation of bone fragments. i.e., for treatment of bone fracture and other bone injuries. All LOOTEO® Proximal Humerus Plate 3.5 are available non-sterile.
The subject device is provided in anatomic designs in overall lengths of 85 mm, 92 mm, 105 mm, 118 mm. 143 mm. 169 mm. 195 mm. 221 mm. and 247 mm. The plates are provided with 3, 4, 5, 6. 8, 10, 12, 14, and 16 holes, respectively, in plate. The lengths of 92 mm, 118 mm, 143 mm, 169 mm, and 195 mm are identical to the plates previously cleared in K121495. Subject device lengths of 85 mm, 221 mm, and 247 mm are identical to the plates previously cleared in K121495 except for the length and number of holes.
The subject device plates include screw holes designed to accommodate appropriately sized locking and non-locking cortical and periprosthetic bone screws, previously cleared in K121495, and K-wires cleared in K131459, presented marked as part of the LOQTEQ® Proximal Humerus Plate 3.5 System. The compatible screws are 3.5 mm in diameter. The subject device plates are also compatible with the 1.6 mm diameter app K-Wires.
## PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence include biocompatibility referenced from K121495; validation of the recommended end-user moist heat sterilization cycle according to ANSI AAMI ISO 17665-1 Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices. An assessment of the biomechanical performance of the subject device to determine substantial equivalence was performed based on cross sectional analysis of the subject device as compared to the predicate device plates in submission K121495. Clinical data were not submitted in this premarket notification.
## COMPARISION OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICES
The primary predicate device is the manufacture's own device K121495, aap LOQTEQ® Proximal Humerus Plate 3.5 System. The additional predicate device is K041860, Synthes LCP Proximal Humerus Plate, Long, Synthes (USA).
The subject device is substantially equivalent in indications and design principles to the predicate devices listed above. The subject device has similar Indications for Use Statements (IFUS) to those of devices previously cleared in K121495 and K041860. Differences between the subject device and primary predicate device K121495 include the statement "The aap LOQTEQ®
{5}------------------------------------------------
Proximal Humerus Plate 3.5 System includes LOQTEQ® Proximal Humerus Plate 3.5 short and LOQTEQ® Proximal Humerus Plate 3.5 long. The plates accept 3.5 mm cortical locking screws and 3.5 mm cortical screws as well as 3.8 mm cancellous locking screws." The statement was removed from the subject device IFUS because the statement describes the device rather than defining "... the disease or condition the device will diagnose, treat, prevent, cure or mitigate, including a description of the patient population for which the device is intended" per 21 CFR §814.20(b)(3)(i). The compatibility of the subject device plates and previously cleared screws is described in the Instructions for Use and Surgical Technique. Differences between the subject device and the additional predicate device K041860 are limited to the device trade or proprietary name. These differences do not impact the substantial equivalence because all IFUS express the same intended use for fractures and fracture dislocations, osteotomies, and nonunions of the proximal humerus, particularly in osteopenic bone.
The subject device, the primary predicate device K121495, and the additional predicate device K041860 have the same technological characteristics and use the same operating principles for bone fixation. Furthermore, the subject device, the primary predicate device, and the additional predicate device include similar anatomic designs for placement on the proximal humerus, with screw holes to accommodate locking and non-locking screws. The subject device designs in lengths of 92 mm, 105 mm, 118 mm, 169 mm, and 195 mm with 4, 5, 6, 8,10, and 12 holes, respectively, are identical to the plates cleared in K121495. Subject device lengths of 85 mm, 221 mm, and 247 mm, with 3, 14 and 16 holes, respectively, have identical design characteristics as plates cleared in K121495, and encompass a similar range of dimensions as K041860 (secondary predicate). These two new lengths do not create a new worst case for mechanical testing.
The subject device and the primary predicate device K121495 are manufactured from the identical titanium alloy material conforming to ASTM F136. All subject device final finished components are manufactured in the same facilities using identical materials and identical manufacturing processes as used for the previously cleared aap device components (K121494). Similarly, the Class II device-specific accessories and Class I instruments are manufactured in the same facilities using identical materials and identical manufacturing processes as the Class II accessories and Class I instruments previously cleared in K121495. Therefore, the subject devices are substantially equivalent to K121495 regarding biocompatibility.
The subject device includes components provided non-sterile in the same packaging and are to be sterilized to a sterility assurance level (SAL) of 10 by the end user using the same sterilization method (moist heat) and parameters as devices previously cleared in K121495. The subject devices do not represent a new worst case for the sterilization validation.
| Feature | Subject Device | Primary Predicate Device | Additional Predicate Device |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | K211582 | K121495 | K041860 |
| | LOQTEQ Proximal Humerus<br>Plates 3.5 | LOQTEQ Proximal Humerus<br>Plates 3.5 | Synthes LCP Proximal Humerus<br>Plates, Long |
| | aap Implantate AG | aap Implantate AG | Synthes USA |
| | Product Code | HRS | HRS, HWC |
| Indications for Use | The LOQTEQ Proximal Humerus<br>Plate 3.5 System is intended for<br>fractures and fracture dislocations,<br>osteotomies, and nonunions of the<br>proximal humerus, particularly in<br>osteopenic bone. | The aap LOQTEQ Proximal Humerus<br>Plate 3.5 System includes LOQTEQ<br>Proximal Humerus Plate 3.5 short and<br>LOQTEQ Proximal Humerus Plate 3.5<br>long. The plates accept 3.5 mm cortical<br>locking screws and 3.5 mm cortical | Synthes (USA) LCP Proximal<br>Humerus Plate, long is indicated for<br>fractures and fracture dislocations,<br>osteotomies, and nonunions of the<br>proximal humerus, particularly in<br>osteopenic bone. |
| Feature | Subject Device<br>K211582<br>LOQTEQ Proximal Humerus<br>Plates 3.5<br>aap Implantate AG | Primary Predicate Device<br>K121495<br>LOQTEQ Proximal Humerus<br>Plates 3.5<br>aap Implantate AG | Additional Predicate Device<br>K041860<br>Synthes LCP Proximal Humerus<br>Plates, Long<br>Synthes USA |
| | | screws as well as 3.8 mm cancellous<br>locking screws.<br>The LOQTEQ Proximal Humerus Plate<br>3.5 System is intended for fractures and<br>fracture dislocations, osteotomies, and<br>nonunions of the proximal humerus,<br>particularly in osteopenic bone. | |
| Reason for<br>Predicate/Reference<br>Device | Not applicable | Plate designs, including use with<br>locking and non-locking screws;<br>Compatible screws;<br>Materials ASTM F136 | Plate length, plate designs,<br>including use with locking and<br>non-locking screws;<br>Reference devices for<br>comparison mechanical testing: |
| Plate Materials | Ti-6Al-4V alloy, ASTM F136 | Ti-6Al-4V alloy, ASTM F136 | Ti-6Al-4VNb alloy or Stainless<br>Steel 316L |
| Type of Holes | Gliding-holes<br>(locking/compression)<br>Round holes (locking only) | Gliding-holes<br>(locking/compression)<br>Round holes (locking only) | Gliding-holes<br>(locking/compression)<br>Round holes (locking only) |
| Plate Length | Short Plates<br>3 holes, 85 mm<br>4 holes, 92 mm<br>5 holes, 105 mm<br>6 holes 118 mm<br>Long Plates<br>8 holes, 143 mm<br>10 holes, 169 mm<br>12 holes, 195 mm<br>14 holes, 221 mm<br>16 holes, 247 mm | Short Plates<br>4 holes, 92 mm<br>5 holes, 105 mm<br>6 holes 118 mm<br>Long Plates<br>8 holes, 143 mm<br>10 holes, 169 mm<br>12 holes, 195 mm | Short Plates<br>3 holes, 90 mm<br>5 holes, 114 mm<br>Long Plates<br>5 holes 142 mm<br>6 holes, 160 mm<br>8 holes, 196 mm<br>10 holes, 232 mm<br>12 holes, 268 mm |
| Head & Shaft<br>Dimensions | Short Plates<br>Head width: 23 mm<br>Head thickness: 3.5 mm<br>Shaft width: 12 mm<br>Shaft thickness: 3.5 mm<br>Long Plates<br>Head width: 23 mm<br>Head thickness: 3.5 mm<br>Shaft width: 13 mm<br>Shaft thickness: 4.5 mm | Short Plates<br>Head width: 23 mm<br>Head thickness: 3.5 mm<br>Shaft width: 12 mm<br>Shaft thickness: 3.5 mm<br>Long Plates<br>Head width: 23 mm<br>Head thickness: 3.5 mm<br>Shaft width: 13 mm<br>Shaft thickness: 4.5 mm | Short Plates<br>Head width: 20 mm<br>Head thickness: 2.5 mm<br>Shaft width: 12 mm<br>Shaft thickness: 2.5 mm<br>Long Plates<br>Head width: 20 mm<br>Head thickness: 3.0 mm<br>Shaft width: 12 mm<br>Shaft thickness: 3.7 mm |
| Surface Treatment | Polished | Polished | Type III anodization |
| Screw Diameters | <i>No new screws</i><br>Cancellous screws:<br>3.8 mm diameter, 28-60 mm<br>length<br>Cortical screws, self-tapping:<br>3.5 mm diameter, 18-38 mm<br>length | Cancellous screws:<br>3.8 mm diameter, 28-60 mm length<br>Cortical screws, self-tapping:<br>3.5 mm diameter, 18-38 mm length | Cancellous screws, fully<br>threaded:<br>4.0 mm diameter, 10-60 mm<br>length<br>Cancellous screws, partially<br>threaded:<br>4.0 mm diameter, 10-50 mm<br>length<br>Cortex screws:<br>3.5 mm diameter, 10-60 mm<br>length<br>4.0 mm diameter, 10-60 mm<br>length (SS material only)<br>Locking screws (SS material and |
| Feature | Subject Device<br>K211582<br>LOQTEQ Proximal Humerus<br>Plates 3.5<br>aap Implantate AG | Primary Predicate Device<br>K121495<br>LOQTEQ Proximal Humerus<br>Plates 3.5<br>aap Implantate AG | Additional Predicate Device<br>K041860<br>Synthes LCP Proximal Humerus<br>Plates, Long<br>Synthes USA |
| | | | Ti-6Al-4VNb alloy only):<br>3.5 mm diameter, 10-60 length |
| Screw Materials | Ti-6Al-4V alloy, ASTM F136 | Ti-6Al-4V alloy, ASTM F136 | TiCP, Ti6Al7Nb, Ti-6Al-4V<br>alloy, SS 316L |
| Surface Treatment | Anodized Type III (locking<br>cancellous screw green)<br>Partially anodized Type III<br>(locking compression screw)<br>Anodized Type III (non-<br>locking screws gold) | Anodized Type III (locking<br>cancellous screw green)<br>Partially anodized Type III (locking<br>compression screw)<br>Anodized Type III (non-locking<br>screws gold) | Anodized Type III (all screws) |
| How Provided | | | |
| Sterility | Provided non-sterile | Provided non-sterile | Provided non-sterile and<br>Provided sterile |
| Sterilization | Non-sterile: to be sterilized by<br>moist heat | Non-sterile: to be sterilized by<br>moist heat | Non-sterile: to be sterilized by<br>moist heat<br>Sterile: method not stated |
| Usage | Single-patient, single-use | Single-patient, single-use | Single-patient, single-use |
{6}------------------------------------------------
{7}------------------------------------------------
In summary, non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence include: biocompatibility referenced from K121495; validation of the recommended end-user moist heat sterilization cycle according to ANSI AAMI ISO 17665-1 Sterilization of health care products – Moist heat – Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices. An assessment of the biomechanical performance of the subject device to determine substantial equivalence was performed based on cross sectional analysis of the subject device as compared to the predicate device plates in submission K121495. Clinical data were not submitted in this premarket notification.
Any differences in the technological characteristics among the subject and predicate devices do not raise different questions of safety and effectiveness. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
## CONCLUSION
The subject devices, the primary predicate device, and the additional predicate device have the same intended use and similar technological characteristics. They encompass a similar range of physical dimensions, are manufactured from the same materials, and are to be sterilized using identical methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.