Active-V Total Knee System; World Total Knee System
Applicant
Signature Orthopaedics Pty, Ltd.
Product Code
MBH · Orthopedic
Decision Date
Dec 23, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3565
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The patient should be skeletally mature to receive a knee replacement. Patients should have adequate bone stock and size to support and accept the prosthesis. The patient’s needs for knee replacement should be due to one or more of the following conditions: • Non-inflammatory degenerative joint disease including osteoarthritis, traumatic arthritis or avascular necrosis • Inflammatory degenerative joint disease including rheumatoid arthritis • Functional deformity such as varus, valgus or flexion deformities • Revision procedures where other treatments or device have failed • Fractures that are unmanageable using other techniques Signature Orthopaedics’ World Knee/Active-V Knee replacement components may be intended for cemented or cementless use.
Device Story
Modular total knee replacement system; includes femoral components, meniscal inserts, patella, tibial baseplates, and external tibial augments. Designed for cemented or cementless fixation. Femoral components (CoCrMo alloy) and tibial components (Ti6Al4V alloy) provide structural support; meniscal inserts (Vitamin-E stabilized UHMWPE) provide bearing surfaces. Used by orthopedic surgeons in clinical settings to replace damaged knee joints. Device restores joint function, corrects deformities, and alleviates pain associated with degenerative or inflammatory joint disease. Surgeon selects components based on patient anatomy and pathology; implants are secured to bone to facilitate joint articulation. Benefits include improved mobility and pain relief for patients with severe knee pathology.
Clinical Evidence
Bench testing only. Engineering evaluations performed on worst-case sizes included Finite Element Analysis (FEA) of the tibial baseplate and Range of Motion (ROM) analysis for constrained condylar knee (CCK) and medial-constrained (MC) variants. No clinical data provided.
Indicated for skeletally mature patients requiring total knee replacement due to non-inflammatory degenerative joint disease (osteoarthritis, traumatic arthritis, avascular necrosis), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity (varus, valgus, flexion), revision procedures, or unmanageable fractures. Requires adequate bone stock.
Regulatory Classification
Identification
A knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device is designed to achieve biological fixation to bone without the use of bone cement. This identification includes fixed-bearing knee prostheses where the ultra high molecular weight polyethylene tibial bearing is rigidly secured to the metal tibial base plate.
Special Controls
*Classification.* Class II (special controls). The special control is FDA's guidance: “Class II Special Controls Guidance Document: Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-Coated Uncemented Prostheses; Guidance for Industry and FDA.” See § 888.1 for the availability of this guidance.
Predicate Devices
Signature Orthopaedics, World Total Knee System (K181530)
Signature Orthopaedics, World Total Knee System (K223062)
Signature Orthopaedics, Triverse Total Knee System (K222380)
Signature Orthopaedics, World Total Knee System (K220737)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 23, 2025
Signature Orthopaedics Pty, Ltd.
Declan Brazil, Managing Director
7 Sirius Rd.
Lane Cove West, NSW 2066
Australia
Re: K253239
Trade/Device Name: Active-V Total Knee System; World Total Knee System
Regulation Number: 21 CFR 888.3565
Regulation Name: Knee Joint Patellofemorotibial Metal/Polymer Porous-Coated Uncemented Prosthesis
Regulatory Class: Class II
Product Code: MBH, JWH
Dated: September 29, 2025
Received: September 29, 2025
Dear Declan Brazil:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253239 - Declan Brazil
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253239 - Declan Brazil
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Lixin Liu, Ph.D.
Assistant Director
DHT6A: Division of Joint Arthroplasty Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K253239 | ? |
| Please provide the device trade name(s). | | ? |
| Active-V Total Knee System;
World Total Knee System | | |
| Please provide your Indications for Use below. | | ? |
| The patient should be skeletally mature to receive a knee replacement. Patients should have adequate
bone stock and size to support and accept the prosthesis.
The patient’s needs for knee replacement should be due to one or more of the following conditions:
• Non-inflammatory degenerative joint disease including osteoarthritis, traumatic arthritis or avascular
necrosis
• Inflammatory degenerative joint disease including rheumatoid arthritis
• Functional deformity such as varus, valgus or flexion deformities
• Revision procedures where other treatments or device have failed
• Fractures that are unmanageable using other techniques
Signature Orthopaedics’ World Knee/Active-V Knee replacement components may be intended for
cemented or cementless use. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K253239
Page 1 of 4
SignatureOrtho
# 510(k) SUMMARY
**Manufacturer:** Signature Orthopaedics Pty Ltd
7 Sirius Road
Lane Cove, NSW 2066
Australia
**Device Trade Name:**
1. Active-V Total Knee System
2. World Total Knee System
**Common Name:**
1. Total Knee Replacement System
2. Total Knee Replacement System
**Contact:**
Dr. Declan Brazil
Managing Director of Signature Orthopaedics
**Prepared By:**
Signature Orthopaedics Pty Ltd
7 Sirius Road
Lane Cove, NSW 2066
Australia
Phone: +61 (2) 9428 5181
Fax: +61 (2) 8456 6065
**Date Prepared:** 22nd December 2025
**Classification:**
888.3565 - Knee joint patellofemorotibial metal/polymer porous-coated uncemented prosthesis
MBH
888.3560 - Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis
JWH
**Predicate Devices:**
Substantial equivalence to the following devices is claimed:
**Active-V Total Knee System**
**Primary Predicate**
- Signature Orthopaedics, World Total Knee System (K181530)
**Additional Predicates**
- Signature Orthopaedics, World Total Knee System (K223062)
- Signature Orthopaedics, Triverse Total Knee System (K222380)
{5}
K253239 - Page 2 of 4
SignatureOrtho
## Reference Device
- Signature Orthopaedics, Rx Knee System (K243680)
## World Total Knee System - Line Extension
### Primary Predicate
- Signature Orthopaedics, World Total Knee System (K181530)
### Additional Predicates
- Signature Orthopaedics, World Total Knee System (K220737, K223062)
## Reference Device
- Signature Orthopaedics, Rx Knee System (K243680)
## Device Description:
### Active-V Total Knee System
The Active-V Total Knee is a modular total knee system consisting of a femoral component, a meniscal insert, a patella, tibial baseplate with a fin keel and an external tibial augment. The femoral and tibial components are intended for cemented and cementless use. The tibial insert bearing components are available in cruciate retaining (CR) and ultracongruent (UC) variants while the femoral components are available in posterior stabilizing (PS) and CR variants. All components are intended for single use only.
The femoral component is an anatomically accurate designed prosthesis manufactured from cast cobalt chrome (CoCrMo) alloy. The tibial component is a modular symmetrical tray design, manufactured from Ti6Al4V Alloy. Cementless femoral components with plasma sprayed commercially pure titanium and hydroxyapatite coating with the option of titanium nitride coating are included. Tibial components with the same coating specifications as the femoral components are available. The tibial external augment is manufactured from the same Ti6Al4V alloy. The meniscal insert is available in CR and Medialconstrained (MC) variant surfaces, and is manufactured from Vitamin-E stabilized UHMWPE (HXLPE). The patella is a spherical, 3-peg patella design manufactured from Vitamin-E stabilized UHMWPE. Components are available for cemented or cementless use.
### World Total Knee System - Line Extension
The World Knee line extension presented in this 510(k) is addition of the following components:
- Cementless femoral components with new coating specifications (plasma sprayed commercially pure titanium and hydroxyapatite coating with the option of titanium nitride coating)
- Tibial components with the modified coating specifications (same as the femoral components) and tibial tray to fin keel modular design
- Constrained condylar knee (CCK) variant of the all poly tibia component and meniscal insert, and medialconstrained (MC) variant of the meniscal insert
- Tibial external augment
{6}
K253239 - Page 3 of 4
SignatureOrtho
The femoral component is an anatomically accurate designed prosthesis manufactured from cast cobalt chrome (CoCrMo) alloy. The tibial component is a modular symmetrical tray design, manufactured from Ti6Al4V Alloy. The tibial external augment is manufactured from the same Ti6Al4V alloy. The meniscal inserts and all poly tibia are manufactured from conventional Ultrahigh Molecular Weight Polyethylene (UHMWPE) and Vitamin E Stabilized and 100 kGy Crosslinked UHMWPE (Vit-E HXLPE). Components are available for cemented or cementless use.
## Indications for Use:
Both subject devices have the same indications for use.
The patient should be skeletally mature to receive a knee replacement. Patients should have adequate bone stock and size to support and accept the prosthesis.
The patient's needs for knee replacement should be due to one or more of the following conditions:
- Non-inflammatory degenerative joint disease including osteoarthritis, traumatic arthritis or avascular necrosis
- Inflammatory degenerative joint disease including rheumatoid arthritis
- Functional deformity such as varus, valgus or flexion deformities
- Revision procedures where other treatments or device have failed
- Fractures that are unmanageable using other techniques
Signature Orthopaedics' Active-V/World Total Knee components may be intended for cemented and cementless use.
## Comparison of Technological Characteristics:
The subject and predicate devices have nearly identical technological characteristics, and the minor differences do not raise any different issues of safety and effectiveness. Specifically, the following characteristics are similar between the subject and predicates:
- Indications for Use
- Materials of manufacture
- Fixation method
- Principle of operation
- Sizes
- Design
Differences with predicate devices exist with respect to available porous coatings, available constraint options, and tibial augment components.
## Performance Testing:
Both subject devices' pre-clinical testing was planned following the guidance provided by the following FDA Guidance "Knee Joint Patellofemorotibial and Femorotibial Metal/Polymer Porous-coated Uncemented Prostheses". For non-applicable/not required testing, appropriate justification could be provided. Two sets of additional engineering evaluation were conducted to verify that the performances of the subject devices are adequate for anticipated in-vivo use. The following engineering evaluations were carried out on the worst-case sizes of the subject devices:
- Finite Element Analysis (FEA) performed on Active-V Knee Tibial Baseplate
{7}
SignatureOrtho
K253239 - Page 4 of 4
- Range of Motion (ROM) Analysis performed on CCK all poly tibia/meniscal insert and MC meniscal insert
The combination of engineering evaluation results and scientific justification provided demonstrate that the Signature Orthopaedics' Active-V and World Total Knee Systems are sufficient for their intended use and substantially equivalent to the legally marketed predicate device.
## Substantial Equivalence:
Both subject devices have the same intended use and same indications for use as the corresponding predicate devices. The subject devices use the same operating principle, incorporate the same basic design, and is manufactured using the same materials as the predicate devices.
Any differences do not raise different questions of safety and effectiveness as established with performance testing. The subject devices are at least as safe and effective as the legally marketed predicate device.
## Conclusion:
Based on similarities in device design, materials, sizes, intended use, indications for use, packaging, sterilization, labeling, and performance testing to the predicate devices, the Active-V Total Knee System and World Total Knee System should perform in a substantially equivalent manner when used as indicated.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.