K243680 · Olea Medical S.A.S. · LLZ · Aug 25, 2025 · Radiology
Device Facts
Record ID
K243680
Device Name
Neurovascular Insight V1.0
Applicant
Olea Medical S.A.S.
Product Code
LLZ · Radiology
Decision Date
Aug 25, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Neurovascular Insight V1.0 is an optional user interface for use on a compatible technical integration environment and designed to be used by trained professionals with medical imaging education including, but not limited to, physicians. Neurovascular Insight V1.0 is intended to: - Display and, if necessary, export neurological DICOM series and outputs provided by compatible processing docker applications, through the technical integration environment. - Allow the user to edit and modify parameters that are optional inputs of aforementioned applications. These modified parameters are provided by the technical integration environment as inputs to the docker application to reprocess the outputs. When available, Neurovascular Insight V1.0 display can be updated with the reprocessed outputs. - If requested by an application, allow the user to confirm information before displaying associated outputs and export them. The device does not alter the original image information and is not intended to be used as a diagnostic device. The outputs of each compatible application must be interpreted by the predefined intended users, as specified in the application's own labeling. Moreover, the information displayed is intended to be used in conjunction with other patient information and based on professional judgment, to assist the clinician in the medical imaging assessment. It is not intended to be used in lieu of the standard care imaging. Trained professionals are responsible for viewing the full set of native images per the standard of care.
Device Story
Neurovascular Insight V1.0 is a user interface for medical image management; operates within a compatible technical integration environment. Inputs: neurological DICOM series and outputs from compatible docker processing applications. Function: displays images/outputs; allows user to edit/modify parameters for re-processing by docker applications; enables user confirmation of information before display/export. Output: updated visualizations and exported data. Used in clinical/hospital environments by trained professionals (physicians, technicians). Does not perform calculations or contain algorithms; does not alter original image data. Not a diagnostic device; intended to assist clinicians in assessment alongside standard care imaging. Benefits: provides standardized interface for viewing and interacting with third-party processing application outputs.
Clinical Evidence
Bench testing only. Validation included product risk assessment, software module verification, and software validation testing. Usability testing conducted to evaluate human factors. No clinical data or performance metrics (sensitivity/specificity) reported as device performs no automated calculations or diagnostic functions.
Technological Characteristics
User interface for medical image management. Features: MR/CT image loading, 3D MIP visualization, manual AIF/VOF selection, manual ASPECTS region selection. Connectivity: operates within a compatible technical integration environment (docker-based). No internal algorithms or calculation features. Software-based interface.
Indications for Use
Indicated for trained professionals with medical imaging education (e.g., physicians) for the display, export, and parameter modification of neurological DICOM series and outputs from compatible docker applications within a technical integration environment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
Olea Medical S.A.S.
August 25, 2025
% John Smith
Partner
Hogan Lovells US LLP
Columbia Square 555 Thirteenth Street, NW
Washington, DC 20004
Re: K243680
Trade/Device Name: Neurovascular Insight V1.0
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: July 21, 2025
Received: July 21, 2025
Dear John Smith:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K243680 - John Smith
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K243680 - John Smith
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb, PhD
Assistant Director
Imaging Software Team
DHT8B: Division of Radiological Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
Submission Number (if known)
K243680
Device Name
Neurovascular Insight V1.0
Indications for Use (Describe)
Neurovascular Insight V1.0 is an optional user interface for use on a compatible technical integration environment and designed to be used by trained professionals with medical imaging education including, but not limited to, physicians. Neurovascular Insight V1.0 is intended to:
- Display and, if necessary, export neurological DICOM series and outputs provided by compatible processing docker applications, through the technical integration environment.
- Allow the user to edit and modify parameters that are optional inputs of aforementioned applications. These modified parameters are provided by the technical integration environment as inputs to the docker application to reprocess the outputs. When available, Neurovascular Insight V1.0 display can be updated with the reprocessed outputs.
- If requested by an application, allow the user to confirm information before displaying associated outputs and export them.
The device does not alter the original image information and is not intended to be used as a diagnostic device. The outputs of each compatible application must be interpreted by the predefined intended users, as specified in the application's own labeling. Moreover, the information displayed is intended to be used in conjunction with other patient information and based on professional judgment, to assist the clinician in the medical imaging assessment. It is not intended to be used in lieu of the standard care imaging.
Trained professionals are responsible for viewing the full set of native images per the standard of care.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}
| 510(k) #: K243680 | 510(k) Summary | Prepared on: 2025-07-18 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | Olea Medical S.A.S. | |
| Applicant Address | 93 Avenue des Sorbiers ZI ATHELIA IV La Ciotat 13600 France | |
| Applicant Contact Telephone | +33(0)442712420 | |
| Applicant Contact | Mrs. Nathalie Palumbo | |
| Applicant Contact Email | qa-ra@olea-medical.com | |
| Correspondent Name | Hogan Lovells US LLP | |
| Correspondent Address | Columbia Square 555 Thirteenth Street, NW Washington DC 20004 United States | |
| Correspondent Contact Telephone | +12026373638 | |
| Correspondent Contact | Mr. John J. Smith | |
| Correspondent Contact Email | john.smith@hoganlovells.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | Neurovascular Insight V1.0 | |
| Common Name | Medical image management and processing system | |
| Classification Name | System, Image Processing, Radiological | |
| Regulation Number | 892.2050 | |
| Product Code(s) | LLZ, N/A | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K223532 | Olea S.I.A. Neurovascular V1.0 | LLZ |
| K233342 | CINA-ASPECTS | POK |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| Neurovascular Insight V1.0 is an optional user interface for use on a compatible technical integration environment and designed to be used by trained professionals with medical imaging education including, but not limited to, physicians and medical technicians. It is worth noting that Neurovascular Insight V1.0 is an evolution of the FDA cleared medical device Olea S.I.A. Neurovascular V1.0 (K223532). Neurovascular Insight V1.0 does not contain any calculation feat ure or any algorithm (deterministic or AI). | | |
{5}
Intended Use/Indications for Use
21 CFR 807.92(a)(5)
Neurovascular Insight V1.0 is an optional user interface for use on a compatible technical integration environment and designed to be used by trained professionals with medical imaging education including, but not limited to, physicians. Neurovascular Insight V1.0 is intended to:
- Display and, if necessary, export neurological DICOM series and outputs provided by compatible processing docker applications, through the technical integration environment.
- Allow the user to edit and modify parameters that are optional inputs of aforementioned applications. These modified parameters are provided by the technical integration environment as inputs to the docker application to reprocess the outputs. When available, Neurovascular Insight V1.0 display can be updated with the reprocessed outputs.
- If requested by an application, allow the user to confirm information before displaying associated outputs and export them.
The device does not alter the original image information and is not intended to be used as a diagnostic device. The outputs of each compatible application must be interpreted by the predefined intended users, as specified in the application's own labeling. Moreover, the information displayed is intended to be used in conjunction with other patient information and based on professional judgment, to assist the clinician in the medical imaging assessment. It is not intended to be used in lieu of the standard care imaging.
Trained professionals are responsible for viewing the full set of native images per the standard of care.
Indications for Use Comparison
21 CFR 807.92(a)(5)
Both the subject Neurovascular Insight V1.0 and predicate Olea S.I.A Neurovascular V1.0 are used for the visualization of MR and CT studies and the edition of parameters. Both systems are for use in clinical/hospital environments by any trained professionals.
The minor differences in the indications for use between the two devices are that:
- Neurovascular Insight V1.0 is for use on a compatible technical integration environment, whereas Olea S.I.A Neurovascular V1.0 is solely to be used on the Olea Medical technical integration platform Olea S.I.A. V1.0, which is a specific compatible technical integration environment.
- Neurovascular Insight V1.0 allows the user to confirm information before displaying associated outputs and export them if requested by an application, while Olea S.I.A Neurovascular V1.0 does not. However, this difference in indications for use is largely a reflection of a difference in technological characteristics, which is addressed in the section below.
- Neurovascular Insight V1.0 is intended to be used by trained professionals while Olea S.I.A Neurovascular V1.0 by trained radiologists and surgeons, which are a subset of trained professionals.
Therefore, both devices have substantially equivalent indications for use regarding visualization of MR and CT studies and edition of parameters.
Technological Comparison
21 CFR 807.92(a)(6)
{6}
Both the subject Neurovascular Insight V1.0 and predicate Olea S.I.A. Neurovascular V1.0 have similar technological characteristics as they both provide the same tools:
- MR and CT images loading and visualization
- Standard Viewing tools
- 3D MIP visualization
- Reset layout
- Manual side selection
- Manual AIF/VOF selection
- Dedicated report.
Neurovascular Insight V1.0 and Olea S.I.A. Neurovascular V1.0 have essentially equivalent features. Both are used to view MR and CT images and edit parameters. Note that the modification of both parameters, side selection and AIF/VOF selection, is identically managed in the subject device and in the predicate device.
The minor difference in the technological characteristics between the two devices is that Neurovascular Insight V1.0 allows the user to manually select ASPECTS region and confirm information before displaying associated outputs and export them if requested by an application, while Olea S.I.A Neurovascular V1.0 does not. However, this user feature to confirm outputs is supported by CINA-ASPECTS UI Agent, included in the reference CINA-ASPECTS device which was cleared for the same intended use with this same confirmation feature.
Therefore, this minor difference in the technological characteristics between both devices does not raise different questions of safety and effectiveness.
Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
{7}
Olea Medical has conducted validation testing of the Neurovascular Insight V1.0. Internal verification and validation testing confirms that the product specifications are met, and support of the substantial equivalence of the intended use and technological characteristics to the predicate device.
Neurovascular Insight V1.0 has been validated to ensure that the system, as a whole, provides all the capabilities necessary to operate according to its intended use and in a manner substantially equivalent to the predicate device.
The following performance evaluations were conducted:
- Product risk assessment;
- Software modules verification tests;
- Software validation test.
Based on the performance testing, the Neurovascular Insight V1.0 has a safety and effectiveness profile that is similar to the predicate device.
Neurovascular Insight V1.0 provides no output. Therefore, the comparison to predicate was based on the comparison of features available within both devices.
No performance feature requires a qualitative or quantitative comparison and validation. Each feature has been tested during the verification phases of Neurovascular Insight V1.0 as described in the software test description (STD) and successfully performed as reported in the software test report (STR). Specific features highlighted by the risk analysis have been additionally tested during the usability process of Neurovascular Insight V1.0 to consider human factor as mentioned in the use specification (see Appendix 'OM-DP-STK_HMI_MM-10-021-Use specification-V02_AINN'). All tests were executed in accordance with the User Guide and the operators reported no issue, without occurrence of any clinically blocking bug and no incident during processing.
For more details, please refer to Appendix 'OM-DP-STK_HMI_MM-10-018-Comparison and Validation-V02_AINN'.
Neurovascular Insight V1.0 subject device has substantially equivalent indications for use, technological characteristics, and principles of operation as Olea S.I.A Neurovascular V1.0 predicate device. The minor technological difference between Neurovascular Insight V1.0 and its predicate device raise no new questions of safety or effectiveness, as the user confirmation feature is included in the reference device which was cleared for the same intended use. The methods for verification and validation testing of the subject device are well-supported in this regulation and by the predicate and reference devices' clearances, and data from such testing demonstrates the device's safety and performance.
Thus, Neurovascular Insight V1.0 is substantially equivalent to predicate Olea S.I.A. Neurovascular V1.0.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.