The intended use of the World Total Knee System is for total knee replacement procedures in skeletally mature patients with structural joint damage.
Device Story
The World Total Knee System is a modular knee prosthesis comprising a femoral component, meniscal insert, patella, and tibial baseplate (or all-polyethylene tibia). Components are available in posterior-stabilized or cruciate-retaining variants, with femoral components offered in anatomic asymmetrical and symmetrical designs. The system is intended for total knee replacement in patients with structural joint damage. This modification introduces an optional titanium nitride (TiN) coating applied via physical vapor deposition (PVD) to the cemented femoral components and includes a minor design update to the largest size of the asymmetrical femoral component. The device is implanted by orthopedic surgeons in a clinical/hospital setting. The prosthesis replaces damaged joint surfaces to restore function and reduce pain. The TiN coating is intended to provide surface characteristics suitable for the intended orthopedic application. The system is provided sterile.
Clinical Evidence
Bench testing only. Verification activities included cytotoxicity testing (ISO 10993-5), bioburden testing (ISO 11737-1), coating pullout strength verification, and 5-million-cycle wear testing per ISO 14243-1:2009. Design verification confirmed that the updated femoral component geometry matches the predicate. No clinical data was provided.
Technological Characteristics
Modular knee prosthesis. Materials: Cast CoCr alloy (ASTM F75, ISO 5832-4), Wrought Ti6Al4V ELI (ASTM F136-13), UHMWPE, and Vit-E HXLPE. Femoral components feature optional TiN coating applied via PVD. Sterilization: Sterile (SAL 10^-6). Connectivity: None. Design: Posterior-stabilized or cruciate-retaining; anatomic asymmetrical or symmetrical.
Indications for Use
Indicated for skeletally mature patients requiring knee replacement due to non-inflammatory degenerative joint disease (osteoarthritis, traumatic arthritis, avascular necrosis), inflammatory degenerative joint disease (rheumatoid arthritis), functional deformity (varus, valgus, flexion), revision procedures, or unmanageable fractures. Requires adequate bone stock.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Signature Orthopaedics World Knee Total Knee System (K181530)
Signature Orthopaedics World Knee Total Knee System (K190577)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 4, 2022
Signature Orthopaedics Pty Ltd. Declan Brazil Managing Director 7 Sirius Road Lane Cove, NSW 2113 AUSTRALIA
Re: K223062
Trade/Device Name: World Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: September 20, 2022 Received: September 30, 2022
Dear Declan Brazil:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ting Song -S
Ting Song, Ph.D. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K223062
Device Name World Total Knee System
#### Indications for Use (Describe)
The patient should be skeletally mature to receive a knee replacement. Patients should have stock and size to support and accept the prosthesis.
The patient's need for knee replacement should be due to one or more of the following conditions:
- · Non-inflammatory degenerative joint disease including osteoarthritis, or avascular nerosis.
- · Inflammatory degenerative joint disease including rheumatoid arthritis.
- · Functional deformity such as varus, valgus or flexion deformities.
- · Revision procedures where other treatments or devices have failed.
- · Fractures that are unmanageable using other techniques.
Signature Orthopaedic's World Knee replacement components may be intended for cemented or cementless use. Please verify whether the particular component is intended for cementless use by checking the package label.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY
| Manufacturer: | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| | Signature Orthopaedics Europe Ltd<br>Unit A, IDA Business & Technology Park Garrycastle<br>Athlone Westmeath N37 DY26<br>IRELAND |
| Device Trade<br>Name: | World Total Knee System |
| Common Name: | Total Knee Prosthesis |
| Contact: | Dr. Declan Brazil<br>Managing Director of Signature Orthopaedics |
| Prepared By: | Signature Orthopaedics Pty Ltd<br>7 Sirius Road<br>Lane Cove, NSW 2066<br>Australia<br>Phone: +61 (2) 9428 5181<br>Fax: +61 (2) 8456 6065 |
| Date Prepared: | September 16th, 2022 |
| Classification: | Knee joint patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis (JWH, 21CFR 888.3560) |
| Predicate Devices: | Primary Predicate<br>• Signature Orthopaedics World Knee Total Knee System<br>(K181530) |
| | Additional Predicate<br>• Signature Orthopaedics World Knee Total Knee System<br>(K190577)<br>• OMNI TiN Coated Apex Knee System (K191765) |
#### Device Description:
The World Total Knee System is a modular knee system consisting of a femoral component, meniscal insert, a patella and a tibial baseplate or all polyethylene tibia. The femoral component and tibial baseplate components are intended for use with or without bone cement. The femoral component, meniscal inserts and all polyethylene
{4}------------------------------------------------
tibias are available as posterior stabilised or cruciate retaining variants. The femoral components are also available as anatomic asymmetrical and symmetrical design variants. The femoral component may be used with modular pegs.
The primary purpose of this Special 510(k) Device Modification to the World Knee System is to notify the FDA of the addition of titanium nitride (TiN) material coating to the femoral components (posterior stabilised{PS}, cruciate retaining{CR}) of the cemented variant. This 510(k) also notifies the FDA of minor design update to the femoral component implant.
Materials: Cast Cobalt Chromium (CoCr) alloy (ASTM F75 and ISO 5832-4) for the femoral component and tibial baseplate, Wrought Titanium-6Aluminium-4Vanadium ELI Alloy (Ti6Al4V ELI, ASTM F136-13) for the femoral peg component and tibial baseplate variant, Ultra-High-Molecular-Weight Polyethylene (UHMWPE) for all variants of the patellar implants, Ultra-High-Molecular-Weight Polyethylene (UHMWPE) or Vit-E HXLPE for the All-poly tibial baseplate components and modular meniscal tibial inserts, and titanium nitride (TiN) for coating cemented femoral components.
### Indications for Use:
The patient should be skeletally mature to receive a knee replacement. Patients should have adequate bone stock and size to support and accept the prosthesis.
The patient's need for knee replacement should be due to one or more of the following conditions:
- . Non-inflammatory degenerative joint disease including osteoarthritis, traumatic arthritis, or avascular necrosis.
- Inflammatory degenerative joint disease including rheumatoid arthritis.
- Functional deformity such as varus, valgus or flexion deformities.
- . Revision procedures where other treatments or devices have failed.
- . Fractures that are unmanageable using other techniques.
Signature Orthopaedic's World Knee replacement components may be intended for cemented or cementless use. Please verify whether the particular component is intended for cemented or cementless use by checking the package label.
#### Intended Use:
The intended use of the World Total Knee System is for total knee replacement procedures in skeletally mature patients with structural joint damage. This is the same intended use as previously cleared devices for the World Total Knee System, K181530.
#### Comparison of Technological Characteristics:
The World Total Knee System described in this Special 510(k) Device Modification is essentially the same device as the primary predicate device cleared in K181530. The technological characteristics that remain the same for the World Total Knee System are as follows:
{5}------------------------------------------------
- . Clinical Indications for Use. Intended Use and Surgical Techniques for the World Knee variant and the World Total Knee System are identical.
- o The main components of the World Total Knee System use the same materials during the manufacturing process. This includes:
- O Cast Cobalt Chromium Alloy (ASTM F75 and ISO 5832-4) for the femoral component,
- o Wrought Titanium-6Aluminium-4Vanadium ELI Alloy (Ti6Al4V EELI, ASTM F136-13) for the femoral peg and tibial baseplate components,
- O Ultra-High-Molecular-Weight Polyethylene (UHMWPE) for all variants of the patellar implant,
- o Ultra-High-Molecular-Weight Polyethylene (UHMWPE) option for Allpoly tibial baseplates and modular meniscal tibial inserts.
- Processes including cleaning, passivation, packaging, sterilisation and transport are the same as the World Total Knee System processes.
- The World Knee variant has the same geometry and fundamental design as the World Total Knee system component. All interconnections between components are the same.
- o The World Knee variant has the same body contact as the World Total Knee System and as such has the same contact stresses.
- . All implants are provided sterile with SAL of 106 as seen in the World Total Knee System.
The primary differences between the subject and primary predicate World Knee devices are as follows:
- Titanium nitride (TiN) coating was added as an option for the femoral . components (PS and CR) of the cemented variant only; and
- . The manufacturing process for the World Total Knee femoral component variant in question includes an additional step of TiN coating applied by Physical Vapor Deposition (PVD) method. The TiN coating is applied to fully machined parts.
- . Minor design change limited to the largest size of asymmetrical variant of the femoral component in accordance with the device intended use and function.
Note that for all other implants of the World Total Knee System described in this 510(k) notification, there are no changes in the manufacturing processes in comparison to the predicate devices K181530, K190577 and K191765.
## Performance Testing:
Engineering evaluations were conducted to verify that the performance of the World Knee TiN coated femoral components is equal to and/or better than the predicate device and therefore adequate for the anticipated in-vivo use. The following V&V activities were conducted:
- Cytotoxicity testing of TiN coating on CoCr alloy implant as per . ANSI/AAMI/ISO 10993-5
- Bioburden testing of TiN coated CoCr alloy implant as per ANSI/AAMI/ISO 11737-1:2018
- . Verification of substantial equivalence of coating pullout strength of the TiN coating on CoCr alloy components.
- . Wear testing for 5 million cycles was done in accordance with ISO 14243-
{6}------------------------------------------------
1:2009 and reported in master file MAF 1413 of the TiN coating applied by PVD method to the subject device.
- . An overlay of the largest size femoral component of asymmetrical and symmetrical variants showed that the edge radii of both components were matching, hence the geometry of the updated design change is the same as predicate devices.
- . Risk Analysis and Design Control Review found no new or changed risks relative to the Indications for Use and efficacy of the Subject Device.
- The addition of TiN coating on CoCr alloy implants was verified by Signature Orthopaedics by reviewing the list of contents of MAF1413 for Medthin 01 TiN provided by IHI Ionbond AG.
#### Substantial Equivalence Conclusion:
The results of the V&V testing, associated engineering review, risk analysis and design control activities demonstrated substantial equivalence of the subject World Total Knee System to the primary predicate cited herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.