The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved. - Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only. - Ø5.0 mm and larger implants are intended for delayed loading in the molar region. All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center.
Device Story
DIMPLO Implant System comprises endosseous dental fixtures, abutments, and attachments. Fixtures (Grade 4 titanium, ASTM F67) feature SLA surface treatment; implanted into maxilla/mandible by dental surgeons. Abutments (Ti-6Al-4V ELI, ASTM F136) connect to fixtures via internal hex interface to support prosthetic restorations. System includes Ti-Base (titanium base with zirconia superstructure) and pre-milled blanks for CAD/CAM custom fabrication at validated centers. Fixtures supplied sterile (gamma); abutments/attachments non-sterile (end-user steam sterilization). Device provides mechanical foundation for artificial teeth; restores chewing function; supports fixed/removable prosthetics. Clinical benefit: tooth replacement in edentulous patients.
Clinical Evidence
No clinical data included. Bench testing only: sterilization validation (ISO 11137, ISO 17665), shelf-life (ASTM F1980), biocompatibility (ISO 10993), surface characterization (SEM/EDS), and mechanical fatigue testing (ISO 14801). MRI safety evaluated via scientific rationale.
Technological Characteristics
Materials: Titanium Grade 4 (ASTM F67) for fixtures; Ti-6Al-4V ELI (ASTM F136) for abutments; Nylon for retention males. Surface: SLA (Sandblasted, Large grit, Acid-etched). Interface: Internal hex. Sterilization: Gamma (fixtures), steam (abutments). CAD/CAM components: Zirconia (InCoris Zi) bonded with RelyX Unicem 2. Connectivity: None (mechanical).
Indications for Use
Indicated for partially or fully edentulous patients (maxilla/mandible) requiring single or multiple-unit restorations (cemented, screw-retained, overdenture) or fixed bridgework support. Ø3.3 mm implants limited to maxillary lateral/mandibular incisors (delayed loading). Ø5.0 mm+ implants for molar regions (delayed loading).
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 22, 2025
DIMPLO Ltd.
Jiyeon Sung
2F, 16, Hwangsaeal-ro, Yeonje-gu
Busan, 47507
REPUBLIC OF KOREA
Re: K251605
Trade/Device Name: DIMPLO Implant System
Regulation Number: 21 CFR 872.3640
Regulation Name: Endosseous Dental Implant
Regulatory Class: Class II
Product Code: DZE, NHA
Dated: April 30, 2025
Received: December 5, 2025
Dear Jiyeon Sung:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251605 - Jiyeon Sung
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
{2}
K251605 - Jiyeon Sung
Page 3
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Andrew I. Steen -S
Andrew I. Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (7/17)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251605 | |
| Device Name DIMPLO Implant System | |
| Indications for Use (Describe) The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved. - Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only. - Ø5.0 mm and larger implants are intended for delayed loading in the molar region. All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
510(k) Summary
K251605
Date Prepared: December 19, 2025
I. SUBMITTER
DIMPLO Ltd.
Jiyeon Sung
2F, 16, Hwangsaeal-ro, Yeonje-gu,
Busan, Republic of Korea
Phone: +82-51-503-2885
Fax: +82 51 504 2885
II. DEVICE
- Trade Name: DIMPLO Implant System
- Common Name: Endosseous dental implant, Endosseous dental implant abutment
- Classification Name: Abutment, Implant, Dental, Endosseous
- Product Code: DZE, NHA
- Panel: Dental
- Regulation Number: 21 CFR 872.3640
- Device Class: Class II
III. PREDICATE DEVICE
1. Primary Predicate (Submission Number / Applicant / Trade Name)
- K181138 / Neobiotech Co., Ltd. / IS-III active System
2. Predicate/Reference Device (Submission Number / Applicant / Trade Name)
- K161987 / UF(II) Narrow Implant System / DIO Corporation
- K141457 / DENTIUM CO. LTD. / DENTIUM IMPLANTIUM AND SUPERLINE ABUTMENTS
- K231426 / HOOWON EDI Co., Ltd. / 8palnt Implant System
- K161689 / OSSTEM Implant Co., Ltd. / OSSTEM Implant System - Abutment
- K223924 / Ossvis Co., Ltd. / LW Implant System
- K231753 / Oneday Biotech Co., Ltd. / Oneday Implant Abutment
- K231411 / Cowellmedi Co., Ltd. / INNO SLA Submerged Hybrid Ti-Base System
- K220253 / DIO Corporation / Eco Abutment, Multiunit Abutment
- K210117 / Paltop Advanced Dental Solutions, Ltd / Paltop Narrow Implant
- K231395 / Cowellmedi Co., Ltd. / INNO SLA Submerged Narrow Implant System
- K201323 / Cowellmedi Co., Ltd. / Sonator Abutment for INNO SLA Submerged Implant System
- K161713 / Dentium Co., Ltd / Dentium CAD/CAM Abutments
Page 1 / 14
{5}
Page 2 / 14
# IV. DEVICE DESCRIPTION
The DIMPLO Implant System consists of fixtures, abutments, and attachments designed for dental implant procedures. Fixture: Made of unalloyed titanium (Grade 4, ASTM F67), these dental implants are available in a range of diameters and lengths to accommodate various clinical applications in the maxilla and mandible. The surface is treated with SLA (Sandblasted, Large grit, and Acid-etched) and provided sterile via gamma irradiation. Available in two platform sizes: Mini and Regular.
Abutment: These components aid in prosthetic restorations. This category includes Cover Screws, Healing Abutments, Temporary Abutments, Cemented Abutments, Angled Abutments, Ti-Base, Pre-milled Blanks, Screws, Multi-unit Abutments, Healing Caps, and Cylinders.
Attachments: Includes Locatus Male Cap and Locatus Retention Male. The Locatus Male Cap is made of titanium alloy (Ti-6Al-4V ELI), while the Locatus Retention Male is made of nylon.
Fixtures and cover screws are supplied sterile. All other abutments and attachments are provided non-sterile and must be sterilized by the end user before use.
Ti-Base consists of a two-piece abutment that includes a titanium base at the bottom and a zirconia superstructure (CAD/CAM patient-specific superstructure) at the top, composing the final abutment.
* Superstructure and Bonding Materials
- Zirconia material: InCoris Zi (ZrO₂) (K123664, Sirona Dental Systems GmbH)
- Bonding agent: RelyX Unicem 2 Automix (K100756, 3M ESPE)
Design Envelope for Zirconia superstructure
| Dimensional Characteristics | Range/Limit |
| --- | --- |
| Post Angle (°) | 0~20 |
| Cuff Height (mm) | 0.5~5.0 |
| Abutment Post Height (mm) | 4.0~6.0 |
| Diameter (Ø, mm) | 5.0~8.0 |
| Wall Thickness (mm) | 0.4~ |
- Design Envelope for Pre-milled Blank
| Dimensional Characteristics | Range/Limit |
| --- | --- |
| Post Angle (°) | 0~30 |
| Cuff Height (mm) | 0.5~6.0 |
| Abutment Post Height (mm) | 4.0~7.0 |
| Diameter (Ø, mm) | 3.5~6.5 |
| Wall Thickness (mm) | 0.4~ |
* Note: Abutment Post Height is defined as the minimum cementable post height for single-unit restoration, measured as the height above the restorative margin.
{6}
V. INDICATIONS FOR USE
The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center.
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
Fixture
| | Subject Device | Primary Predicate Device | Reference Predicate Device |
| --- | --- | --- | --- |
| 510(k) No. | K251605 | K181138 | K161987 |
| Applicant | DIMPLO Ltd. | Neobiotech Co., Ltd. | DIO Corporation |
| Device Name | DIMPLO Implant System | IS-III active System | UF(II) Narrow Implant System |
| Classification Name | Endosseous Dental Implant, Fixture (872.3640) | Endosseous Dental Implant, Fixture (872.3640) | Endosseous Dental Implant, Fixture (872.3640) |
| Product Code | DZE | DZE | DZE |
| Material | Titanium Grade 4 (ASTM F67) | Titanium Grade 4 (ASTM F67) | Titanium Grade 4 (ASTM F67) |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | The IS-III active System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. IS-III active System is dedicated for two stage surgical procedures and for immediate loading when there is good primary stability and an appropriate occlusal load. Also, implants with diameters larger than 5mm are indicated for molar regions. | The UF(II) Narrow Implant System is intended for two stage surgical procedures in the following situations and with the following clinical protocols:
- The intended use for the 3.0mm, 3.3mm diameter UF(II) Narrow Implant is limited to the replacement of maxillary lateral incisors and mandibular incisors.
- Immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge.
- It is intended for delayed loading. |
| Diameter x Lengths(mm) | Ø3.3 X 8.5, 10, 11.5, 13, 15
Ø3.75 X 8.5, 10, 11.5, 13, 15
Ø4.25 X 8.5, 10, 11.5, 13, 15
Ø4.6 X 8.5, 10, 11.5, 13, 15
Ø5.05 X 8.5, 10, 11.5, 13, 15
Ø5.45 X 8.5, 10, 11.5, 13
Ø5.95 X 8, 9.5, 11, 12.5
Ø6.8 X 8, 9.5, 11, 12.5 | Ø3.5 X 8.5, 10, 11.5, 13, 15
Ø4.0 X 7.3, 8.5, 10, 11.5, 13, 15
Ø4.5 X 7.3, 8.5, 10, 11.5, 13, 15
Ø5.0 X 7.3, 8.5, 10, 11.5, 13, 15
Ø5.5 X 7.3, 8.5, 10, 11.5, 13, 15
Ø6.0 X 7.3, 8.5, 10, 11.5, 13
Ø7.0 X 7.3, 8.5, 10, 11.5, 13 | Ø3.0 X 8.5, 10, 11.5, 13, 15
Ø3.3 X 8.5, 10, 11.5, 13, 15 |
| Connection Interface | Internal Hex | Internal Hex | Internal Hex |
Page 3 / 14
{7}
Cover Screw
| Surface Treatment | SLA | SLA | SLA |
| --- | --- | --- | --- |
| Sterile | Gamma sterilization | Gamma sterilization | Gamma sterilization |
| Shelf Life | 5 years | 5 years | 5 years |
| Substantial Equivalence | The subject device has the same intended use, material, abutment-implant interface, surface treatment, sterilization method, and shelf life as the primary predicate device(K181138), and the reference predicate(K161987). The implant diameters and lengths of the subject device are within or comparable to those of the predicate devices and do not raise new questions regarding safety or effectiveness. Therefore, the subject device is substantially equivalent to the predicate devices. | | |
Hearing Abutment
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | - | K141457 |
| Applicant | DIMPLO Ltd. | DENTIUM CO. LTD. |
| Device Name | DIMPLO Implant System | DENTIUM IMPLANTIUM AND SUPERLINE ABUTMENTS-Cover Screw |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | Dentium Prosthetics are intended for use as an aid in prosthetic rehabilitation. |
| Diameter(mm) | 3.03–3.88 | 3.6 |
| Angulation | 0 | 0 |
| Sterile | Non-sterile | Non-sterile |
| Substantial Equivalence | The Subject Device and the predicate device(K141457) have equivalent characteristics, including Indications for Use, material, angulation, and sterilization method. The difference in diameter size does not impact the product's fundamental technology. Therefore, the Subject Device and the predicate device are substantially equivalent. | |
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | K251605 | K231426 |
| Applicant | DIMPLO Ltd. | HOOWON EDI Co., Ltd. |
| Device Name | DIMPLO Implant System | 8palnt Implant System |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | 8 palnt Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Implant bodies with a diameter of 5mm or more are intended to be used in the molar region. |
Page 4 / 14
{8}
Temporary Abutment
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | K251605 | K161689 |
| Applicant | DIMPLO Ltd. | OSSTEM Implant Co., Ltd. |
| Device Name | DIMPLO Implant System | OSSTEM Implant System - Abutment |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Surface Treatment | None | None |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | The OSSTEM Implant System-Abutment is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. |
| Diameter(mm) | 4.1, 4.6 | 4.0, 4.5 |
| G/H(mm) | 1.0, 3.0 | 1.0, 3.0 |
| Post Height(mm) | 10 | 10 |
| Angulation(°) | 0 | 0 |
| Abutment-Implant Interface | Internal Connection | Internal Connection |
| Sterile | Non-sterile | Non-sterile |
| Maximun duration | Less than 6 months | Less than 6 months |
| Principle of operation | It is dental Abutments designed to serve as a temporary dental prosthesis during the healing process until a | It is dental Abutments designed to serve as a temporary dental prosthesis during the healing process until a |
Page 5 / 14
{9}
| | permanent crown is made. | permanent crown is made. |
| --- | --- | --- |
| Substantial Equivalence | The subject device and the predicate device (K161689) have equivalent material, abutment-implant interface, surface treatment, sterilization state, and principle of operation. Both devices are intended for temporary clinical use during the healing process until a permanent prosthesis is placed. The maximum recommended duration of use for the subject device is the same as the predicate device, limited to less than 6 months. Therefore, the subject device is substantially equivalent to the identified predicate device. | |
Cemented Abutment
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | K251605 | K223924 |
| Applicant | DIMPLO Ltd. | Ossvis Co., Ltd. |
| Device Name | DIMPLO Implant System | LW Implant System |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Surface Treatment | None | None |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | The LW Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screw retained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. The LW Implant System is dedicated for two stage surgical procedures and is intended for delayed loading. Also, implants with diameters larger than 5mm are indicated for molar regions. |
| Diameter(mm) | 4.6, 5, 6, 7 | 4.5, 4.6, 5.0, 6.0, 7.0 |
| G/H(mm) | 1, 2, 3, 4, 5, 6, 7 | 1, 2, 3, 4, 5, 6, 7 |
| Post(mm) | 4.0, 5.5, 7.0 | 4.0, 5.5, 7.0 |
| Angulation(*) | 0 | 0 |
| Abutment-Implant Interface | Internal Connection | Internal Connection |
| Sterile | Non-sterile | Non-sterile |
| Principle of operation | It is indicated for screw-retained single tooth or cement retained single tooth and bridge restorations. | It is indicated for screw-retained single tooth or cement retained single tooth and bridge restorations. |
| Substantial Equivalence | The subject device and the predicate device (K223924) have equivalent indications for use as both screw-retained and cement-retained abutments in partially or fully edentulous patients. Both devices have the same material, abutment-implant interface, sterilization state, and principle of operation. The subject device does not include the 4.5 mm diameter available in the predicate device; all other diameters are equivalent. Therefore, the subject device is substantially equivalent to the predicate device. | |
Angled Abutment
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | - | K231753 |
| Applicant | DIMPLO Ltd. | Oneday Biotech Co., Ltd. |
| Device Name | DIMPLO Implant System | Oneday Implant Abutment |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
Page 6 / 14
{10}
Page 7 / 14
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| --- | --- | --- |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | Oneday Implant Abutment is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Implants with diameters larger than 5 mm are intended to be used in the molar region. |
| Diameter(mm) | 4, 4.5, 5, 6 | 4, 4.5, 5, 5.5, 6, 7, 8 |
| Length(mm) | 12.5~14.5 | 9.7, 10.7, 11.7, 12.7 |
| Angulation(°) | 17 | 15, 25 |
| Abutment-Implant Interface | Internal Connection | Internal Connection |
| Sterile | Non-sterile(End User Sterilization) | Non-sterile(End User Sterilization) |
| Principle of operation | This product is a superstructure which is connects with the fixtures. It replaces the functions of the missing teeth as a dental abutment. | This product is a superstructure which is connects with the fixtures. It replaces the functions of the missing teeth as a dental abutment. |
| Substantial Equivalence | The subject device and the predicate device(K231753) have equivalent indications for use, material, abutment-implant interface, sterilization, and principle of operation. Differences in diameter (within the predicate's range), length, and angulation (17° for the subject device versus 15° and 25° for the predicate) do not impact equivalence. Therefore, the subject device is substantially equivalent to the predicate device. | |
Ti-Base
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | - | K231411 |
| Applicant | DIMPLO Ltd. | Cowellmedi Co., Ltd. |
| Device Name | DIMPLO Implant System | INNO SLA Submerged Hybrid Ti-Base System |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
| Material | Ti-6Al-4V ELI (ASTM F136)
Zirconia Oxide | Ti-6Al-4V ELI (ASTM F136)
Zirconia Oxide |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | INNO SLA Submerged Hybrid Ti-Base System is intended for use in conjunction with the fixture in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit cement retained restorations. Digitally All digitally designed zirconia superstructures for use with the INNO SLA Submerged Hybrid Ti-Base System are intended to be sent to a Cowellmedi validated milling center for manufacture. |
| Design Limits | - Post Angle (°): 0~20
- Cuff Height (mm): 0.5~5.0
- Abutment Post Height (mm): 4.0~6.0 | Maximum Angulation 20°
Maximum Cuff Height 5.0mm
Minimum Gingival Height, 0.5mm |
{11}
Page 8 / 14
| | - Diameter (Ø, mm): 5.0~8.0
- Thickness (mm): 0.4~
* Note: Abutment Post Height is defined as the minimum cementable post height for single-unit restoration, measured as the height above the restorative margin. | Minimum Diameter Ø 5.0mm
Minimum Thickness 0.4mm
Minimum Abutment Post Height 4.0mm |
| --- | --- | --- |
| Abutment-Implant Interface | Internal Connection | Internal Connection |
| Sterile | Non-sterile | Non-sterile |
| Substantial Equivalence | Both the subject device and predicate device(K231411) share equivalency in indications for use, material, abutment-implant interface, sterilization, design limit and restoration compatibility. Accordingly, the subject device is substantially equivalent to the predicate device. | |
## Multi-unit Abutment
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | - | K220253 |
| Applicant | DIMPLO Ltd. | DIO Corporation |
| Device Name | DIMPLO Implant System | Eco Abutment, Multiunit Abutment |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | The Multiunit Abutment is a premanufactured prosthetic component directly connected to an endosseous implant and it is intended for use in prosthetic rehabilitation. |
| Diameter(mm) | 4.8, 4.9 | 4.8 |
| Gingival Height (mm) | 1.5, 2.5, 3.5, 4.5, 5.5, 6.5, 7.5 | 2.5, 3.0, 3.5, 4.5, 5.5, 6.5, 7.5 |
| Angulation(*) | 0, 17, 30 | 0, 20, 30 |
| Abutment-Implant Interface | Internal Connection | Internal Connection |
| Sterile | Non-sterile | Non-sterile |
| Principle of operation | It is indicated for screw-retained single tooth or cement retained single tooth and bridge restorations. | It is indicated for screw-retained single tooth or cement retained single tooth and bridge restorations. |
| Substantial Equivalence | The subject device and predicate device(K220253) are equivalent in indications for use, material, abutment-implant interface, sterilization, and restoration. Variations in diameter, gingival height, and angulation do not impact the device's equivalence. Hence, the subject device and predicate device are substantially equivalent. | |
## Healing Cap
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | - | K210117 |
| Applicant | DIMPLO Ltd. | Paltop Advanced Dental Solutions, Ltd |
| Device Name | DIMPLO Implant System | Paltop Narrow Implant-w/Healing Cap |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
{12}
Page 9 / 14
| Product Code | NHA | NHA |
| --- | --- | --- |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | The Paltop Narrow Implant is indicated for use in surgical and restorative applications for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited by the adjacent teeth and roots, to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. The Paltop Narrow Implant is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading. |
| Diameter(mm) | 4.8 | 5 |
| Height (mm) | 4.6 | 1,2,3,4,5 |
| Angulation(*) | 0 | 0 |
| Sterile | Non-sterile | Non-sterile |
| Substantial Equivalence | The subject device and predicate device(K210117) are equivalent in indications for use, material, sterilization, and angulation. The differences in diameter and height do not affect substantial equivalence. Therefore, the subject device and predicate device are substantially equivalent. | |
Cylinder Screw
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | - | K231395 |
| Applicant | DIMPLO Ltd. | Cowellmedi Co., Ltd. |
| Device Name | DIMPLO Implant System | INNO SLA Submerged Narrow Implant System |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | The INNO SLA Submerged Narrow Implant System is intended for two stage surgical procedures in the following situations and with the following clinical protocols:
- The intended use for the 3.3mm, 3.5mm diameter Narrow Implant is limited to the replacement of maxillary lateral incisors and mandibular incisors.
- Immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge.
- It is intended for delayed loading. |
| Diameter(mm) | 2.15 | 2.25 |
| Length(mm) | 3.9 | 5 |
| Sterile | Non-sterile | Non-sterile |
| Principle of operation | Multi-unit Cylinder Screw is a premanufactured prosthetic component directly connected to the Multi-unit Abutment and is intended for use as an aid in prosthetic rehabilitation. | Multi-unit Cylinder Screw is a premanufactured prosthetic component directly connected to the Multi-unit Abutment and is intended for use as an aid in prosthetic rehabilitation. |
| Substantial | The subject device and predicate device(K231395) are equivalent in indications for use, material, sterilization, and principle of | |
{13}
Page 10 / 14
Equivalence
operation. Differences in diameter and length do not affect the equivalency of the devices. As such, the subject device is substantially equivalent to the predicate device.
Temporary Cylinder
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | - | K231395 |
| Applicant | DIMPLO Ltd. | Cowellmedi Co., Ltd. |
| Device Name | DIMPLO Implant System | INNO SLA Submerged Narrow Implant System |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | The INNO SLA Submerged Narrow Implant System is intended for two stage surgical procedures in the following situations and with the following clinical protocols:
- The intended use for the 3.3mm, 3.5mm diameter Narrow Implant is limited to the replacement of maxillary lateral incisors and mandibular incisors.
- Immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge.
- It is intended for delayed loading. |
| Diameter(mm) | 4.8, 5.3 | 4.5 |
| Height (mm) | 12 | 9 |
| Angulation(°) | 0 | 0 |
| Sterile | Non-sterile | Non-sterile |
| Principle of operation | This is designed to serve as a temporary dental prosthesis during the healing process until a permanent are made. | This is designed to serve as a temporary dental prosthesis during the healing process until a permanent are made. |
| Duration of use | 90days | 90days |
| Substantial Equivalence | Equivalent characteristics between the subject device and predicate device(K231395) include indications for use, material, sterilization, principle of operation, angulation, and duration of use. Variations in diameter and height do not alter substantial equivalence. Therefore, the subject device is substantially equivalent to the predicate device. | |
Ti Cylinder
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | - | K231395 |
| Applicant | DIMPLO Ltd. | Cowellmedi Co., Ltd. |
| Device Name | DIMPLO Implant System | INNO SLA Submerged Narrow Implant System |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved. | The INNO SLA Submerged Narrow Implant System is intended for two stage surgical procedures in the following situations and with the following clinical protocols:
- The intended use for the 3.3mm, 3.5mm diameter Narrow Implant is limited to the replacement of maxillary lateral incisors and mandibular incisors.
- Immediate placement in extraction sites and in situations with a partially or completely healed alveolar ridge.
- It is intended for delayed loading. |
{14}
Page 11 / 14
Locatus
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | - | K201323 |
| Applicant | DIMPLO Ltd. | Cowellmedi Co., Ltd. |
| Device Name | DIMPLO Implant System | Sonator Abutment for INNO SLA Submerged Implant System |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | The INNO SLA Submerged Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple-unit restorations including; cemented retained, screwretained, or overdenture restorations, and terminal or intermediate abutment support for fixed bridgework. |
| Diameter(mm) | 3.86 | 3.87, 4.5 |
| Gingival Height (mm) | 1, 2, 3, 4, 5, 6, 7 | 1, 1.5, 2, 3, 4, 5, 6 |
| Angulation(*) | 0, | 0, 15 |
| Abutment-Implant Interface | Internal Connection | Internal Connection |
| Sterile | Non-sterile | Non-sterile |
| Principle of operation | This component is used for implantretained muscosa-supported restorations, such as dentures. It supports implant retained mucosa supported over dentures. | This component is used for implantretained muscosa-supported over dentures. It supports implant retained mucosa supported over dentures. |
| Substantial Equivalence | Equivalent characteristics between the subject device and predicate device(K201323) include indications for use, material, sterilization, and principle of operation. Variations in diameter, gingival height, and angulation do not impact equivalence. Therefore, the subject device is substantially equivalent to the predicate device. | |
Pre-milled Blank
| | Subject Device | Predicate Device |
| --- | --- | --- |
{15}
Page 12 / 14
| 510(k) No. | - | K161713 |
| --- | --- | --- |
| Applicant | DIMPLO Ltd. | Dentium Co., Ltd |
| Device Name | DIMPLO Implant System | Dentium CAD/CAM Abutments |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | Dentium abutments are intended for use on endosseous dental implants in the edentulous or partially edentulous maxilla or mandible, as an aid in prosthetic rehabilitation.
All digitally designed abutments for use with Dentium CAD/CAM Abutments are intended to be sent to a Dentium-validated milling center for manufacture. |
| Platform Diameter(mm) | 3.2 - 6.8 | 3.6 - 5.0 |
| Design Limits | - Post Angle (°): 0~30
- Cuff Height (mm): 0.5~5.0
- Abutment Post Height (mm): 4.0~6.0
- Diameter (Ø, mm): 5.0~8.0
- Thickness (mm): 0.4~
* Note: Abutment Post Height is defined as the minimum cementable post height for single-unit restoration, measured as the height above the restorative margin. | Up to 30° |
| Abutment-Implant Interface | Internal Connection | Internal Connection |
| Sterile | Non-sterile | Non-sterile |
| Substantial Equivalence | Equivalent characteristics between the subject device and predicate device(K201323) include indications for use, material, sterilization, and principle of operation. Variations in diameter, gingival height, and angulation do not impact equivalence.
Therefore, the subject device is substantially equivalent to the predicate device. | |
Abutment Screw
| | Subject Device | Predicate Device |
| --- | --- | --- |
| 510(k) No. | - | K231753 |
| Applicant | DIMPLO Ltd. | Oneday Biotech Co., Ltd. |
| Device Name | DIMPLO Implant System | Oneday Implant Abutment |
| Classification Name | Endosseous Dental Implant Abutments (872.3630) | Endosseous Dental Implant Abutments (872.3630) |
| Product Code | NHA | NHA |
| Material | Ti-6Al-4V ELI (ASTM F136) | Ti-6Al-4V ELI (ASTM F136) |
| Indications for use/Intended use | The DIMPLO Implant System is intended for use in the maxilla or mandible of partially or fully edentulous patients to support single or multiple-unit restorations, including cemented, screw-retained, or overdenture restorations, as well as terminal or intermediate abutment support for fixed bridgework. It is intended for two-stage surgical procedures and may be used for immediate loading when good primary stability and appropriate occlusal loading conditions are achieved.
- Ø3.3 mm implants are for replacing maxillary lateral incisors and mandibular incisors with delayed loading only.
- Ø5.0 mm and larger implants are intended for delayed loading in the molar region. | Oneday Implant Abutment is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, screw retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Implants with diameters larger than 5 mm are intended to be used in the molar region. |
{16}
| | loading in the molar region.
All digitally designed custom components (including Ti-Base and Pre-milled Blanks) for use with the DIMPLO Implant System are to be manufactured at a DIMPLO validated milling center. | |
| --- | --- | --- |
| Diameter(mm) | 2.2, 2.3 | 2.3 |
| Lenght(mm) | 10.2, 8.35 | 8.3 |
| Sterile | Non-sterile | Non-sterile |
| Substantial Equivalence | The subject device and predicate device(K231753) have equivalent characteristics in indications for use, material, and sterilization. Variations in diameter and length do not influence the substantial equivalency of the devices. Thus, the subject device and the predicate device are substantially equivalent. | |
Page 13 / 14
{17}
Page 14 / 14
# VII. NON-CLINICAL PERFORMANCE DATA
Non-clinical testing was performed to evaluate the safety and performance of the DIMPLO Implant System.
1. Sterilization and Shelf-Life Testing
Sterilization validation was conducted for gamma-sterilized components (e.g., fixtures) in accordance with ISO 11137-1 and ISO 11137-2. The tests to validate the Shelf-Life of the device were conducted using the accelerated aging method in accordance to ASTM F1980Shelf-life testing was conducted in accordance with the guidance provided in FDA's Shelf Life of Medical Devices Guidance Document, and endotoxin testing was conducted following ANSI/AAMI ST72. Components provided non-sterile (e.g., abutments and attachments) are intended to be steam sterilized by the end user, with sterilization validation conducted in accordance with ISO 17665-1, ISO 17665-2, and ANSI/AAMI ST79. and validated accordingly per ISO 17665 and ANSI/AAMI ST79 guidance.
2. Biocompatibility Testing
Biocompatibility testing including cytotoxicity, sensitization, and irritation testing were conducted according to ISO 10993-5, 10993-10, and 10993-23 respectively. Biocompatibility testing was conducted following ISO 10993. Specific testing was conducted per ISO 10993-5 (Cytotoxicity), ISO 10993-10 (Sensitization), and ISO 10993-10 (Irritation).
3. Material and Surface Characterization
The implant body surface was characterized using Scanning Electron Microscopy (SEM) and Energy Dispersive X-ray Spectroscopy (EDS). Characterization was also performed on the anodized surface of the devices.
4. Mechanical Testing
Mechanical testing, including static and fatigue testing, was performed per ISO 14801:2016.
5. MR Environment Condition
Non-Clinical worst-case MRI review was performed to evaluate the DIMPLO Implant System devices in the MRI environment using scientific rationale and published literature (e.g., Terry O. Woods, Jana Delfino, & Sunder Rajan. (2019). Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices. Journal of Testing and Evaluation 49.2, 783-795), based on the entire system including all variations (all compatible implant bodies, dental abutments, and fixation screws) and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
No clinical data are included in this submission.
# VIII. CONCLUSIONS
The subject device has the same intended use and similar technological characteristics as the predicate and reference devices. Any differences in design or materials do not raise new questions of safety or effectiveness.
The data provided in this submission, including non-clinical performance testing, support that the subject device is substantially equivalent to the predicate and reference devices.
Specifically, the subject device:
- has the same intended use,
- operates on the same principles,
- incorporates similar basic designs,
- utilizes the same or similar materials, and
- has comparable packaging and sterilization methods.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.