Dentium Prosthetics are intended for use as an aid in prosthetic rehabilitation.
Device Story
Dentium Implantium and SuperLine Abutments are prosthetic components used in dental implant procedures; they serve as the interface between an endosseous dental implant fixture and the final dental restoration. The device line includes angled screw abutments, cover screws, and various cylinders (titanium, gold, metal). These components are supplied non-sterile and must be autoclaved by the clinician prior to use. The abutments feature a universal conical connection designed to mate securely with Dentium Implantium/SuperLine fixtures. By providing a stable connection, these abutments facilitate the attachment of prosthetic teeth, aiding in the restoration of oral function and aesthetics for the patient. The device is used in a clinical setting by dental professionals.
Clinical Evidence
Bench testing only. Performance evaluated via ISO 14801 static compressive and fatigue testing (passed 5,000,000 cycles without fracture, cracks, or severe distortion). Biocompatibility testing performed per ISO 10993-1, including cytotoxicity, systemic toxicity, pyrogenicity, sensitization, and intracutaneous reactivity, confirming materials are nontoxic and non-sensitizing. Additional testing included corrosion and adaptation accuracy.
Indicated for patients requiring prosthetic rehabilitation via endosseous dental implants. No specific age or gender restrictions provided.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting community and support. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 12,2014
Dentium Company Limited C/O Ms. Sheryl Higgins US Agent 6761 Katella Avenue Cypress, CA 90630
Re: K141457
Trade/Device Name: Dentium Implantium® and SuperLine® Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: October 14, 2014 Received: October 15, 2014
Dear Ms. Higgins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Ms. Higgins
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno, DDS, mA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number: K141457
Dentium Implantium® / SuperLine® Abutments Device Name:
Indications for Use:
Dentium Prosthetics are intended for use as an aid in prosthetic rehabilitation.
Prescription Use 区 (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use □ (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# 510(k) Summary
# 1. Company
| | Submitter |
|----------------|-----------------------------------------------------------------------------------------|
| Name | Dentium Co., Ltd. |
| Address | 440, Changnyong-daero, lui-dong, Yeongtong-gu, Suwon-si,<br>Gyeonggi-do, Korea, 443-270 |
| Phone / Fax | +82 31 207 2200 / +82 31 207 3883 |
| Contact person | Sheryl Higgins / US Agent<br>714-226-0229 x107<br>shiggins@dentiumusa.com |
## 2. Device Name
| Proprietary name: | Dentium Implantium® / SuperLine® Abutments |
|----------------------|---------------------------------------------------------------|
| Common name: | Abutment, Prosthetic Device |
| Classification name: | Abutment, Implant, Dental, Endosseous<br>NHA, 21 CFR 872.3630 |
#### 3. Predicated Device
k041368 - Dentium Co. Ltd., Implantium k052823 - Implantium Abutments k052957 - Implantium Prosthetics k112045 - SimpleLine II Abutment System
#### 4. Description
Dentium Implantium and SuperLine Abutments are intended for use an aid in prosthetic rehabilitation. The abutments were originally cleared under 510(k) k041358. The modified designs proposed in this submission consist of the
COMPANY CONFIDENTIAL
{4}------------------------------------------------
Angled Screw Abutment, Cover Screw, and Screw Abutment Cylinders. Abutments are supplied non-sterile and autoclaved by the end user.
| Name | Class | Material | Dia<br>(mm) | G/H<br>(mm) | Connection | Angle | Figure | Model Number |
|-----------------------------|-------|---------------------------------------------------------------|------------------|------------------|--------------------|------------------|------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Angled<br>Screw<br>Abutment | II | Pure Ti-G4<br>(ASTM F67)<br>TiN Coated | 4.5<br>to<br>5.5 | 1.0<br>to<br>1.5 | Hex and<br>Non-hex | 15°<br>to<br>30° | Image: Angled screw abutment | ASA45151018H<br>ASA45151018N<br>ASA45301018H<br>ASA45301018N<br>ASA55151518H<br>ASA55151518N<br>ASA55301518H<br>ASA55301518N |
| Screw | II | Ti-6Al-4V ELI<br>alloy<br>(ASTM F136) | 1.8<br>to<br>2.3 | n/a | n/a | n/a | Image: Screw | ASASC2023<br>SRS18T |
| Cover<br>Screw | II | Ti-6Al-4V ELI<br>alloy<br>(ASTM F136)<br>Annozided | 3.6 | n/a | n/a | n/a | Image: Cover screw | CS36 |
| Cylinder -<br>Titanium | II | Pure Titanium,<br>Grade 2<br>(ASTM F67-00) | 4.5<br>to<br>5.5 | n/a | Hex and<br>Non-hex | n/a | Image: Titanium cylinder | STC45SG<br>STC45BG<br>STC55SG<br>STC55BG |
| Cylinder -<br>Gold | II | Gold Alloy | 4.5<br>to<br>5.5 | n/a | Hex and<br>Non-hex | n/a | Image: Gold cylinder | SGC45SL<br>SGC45BL<br>SGC55SL<br>SGC55BL |
| Cylinder -<br>Metal | II | Cobalt-28<br>Chromium-6<br>Molybdenum<br>Alloy (ASTM<br>F799) | 4.5<br>to<br>5.5 | n/a | Hex and<br>Non-hex | n/a | Image: Metal cylinder | SGC45CSL<br>SGC45CBL<br>SGC55CSL<br>SGC55CBL |
{5}------------------------------------------------
#### 5. Indication for Use
Dentium Prosthetics are intended for use as an aid in prosthetic rehabilitation.
## 6. Identification of the Risk
Risk analysis was performed according to ISO 14971:2007, "Medical devices - Application of risk management to medical devices". All risks identified have been mitigated and any residual risk is within acceptable levels.
# 7. Device Characteristics
Dentium Implantium / SuperLine Abutments are supplied in many different shapes and sizes to meet the patient specific needs of our customers. All abutment are made with Dentium's universal conical connection so that thev securely mate with any Dentium Implantium / SuperLine Fixture.
# 8. Performance Testing
Biocompatibility testing on the proposed Implantium & SuperLine Abutments have been completed. Requirements for biological evaluation of the purposed device were based on the ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process." The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are nontoxic and non-sensitizing to biological tissues consistent with its intended use. The following biocompatibility tests were completed:
| -ISO Cytotoxicity | -ISO Systemic toxicity |
|--------------------------------|------------------------|
| -ISO Pyrogenicity | -ISO Sensitization |
| -ISO Intracutaneous reactivity | |
The proposed Implantium & SuperLine Abutments were evaluated using the following performance bench testing to confirm the performance characteristics:
| -ISO Static compressive | -ISO Fatigue |
|-------------------------|----------------------|
| -Corrosion Testing | -Adaptation Accuracy |
| -Modified Surface | |
Static Compressive mean and Fatigue Limit were evaluated per ISO 14801. Fracture or cracks or severe distortion of any parts were not detected at the fatigue limit and passed 5,000,000 cycles.
COMPANY CONFIDENTIAL
{6}------------------------------------------------
#### 9. Reliance on Standards
All testing was performed in compliance with Guidance for Industry and FDA Staff - Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments.
#### 10. Review
All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the Implantium & SuperLine Abutments meet the established specifications necessary for consistent performance according to its intended use.
Implantium & SuperLine Abutments have been subjected to safety, performance, and product validations prior to release. Safety tests including biocompatibility have been performed to ensure the devices comply with the applicable International and US regulations.
#### 11. Conclusions
All of the data consistent with the recommendations in the FDA guidance document Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments demonstrated that the Dentium Implantium & SuperLine Abutments are substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.