K231426 · Hoowon Edi Co., Ltd. · DZE · May 24, 2024 · Dental
Device Facts
Record ID
K231426
Device Name
8plant Implant System
Applicant
Hoowon Edi Co., Ltd.
Product Code
DZE · Dental
Decision Date
May 24, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Indications for Use
8plant Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Implant bodies with a diameter of 5mm or more are intended to be used in the molar region.
Device Story
8plant Implant System comprises titanium dental implants (fixtures), cover screws, and various abutments (healing, rigid, locator, transfer, angled, temporary). Fixtures feature SLA surface; internal hex connection. Used by dental clinicians to support prosthetic restorations (crowns, bridges, overdentures) in edentulous patients. Implants surgically placed in jawbone; abutments attached via screws to provide prosthetic interface. System facilitates osseointegration; provides stable foundation for dental prosthetics. Benefits include restoration of chewing function and aesthetics. Device is non-sterile or sterile (gamma) depending on component; requires end-user sterilization for some parts. Clinician selects appropriate fixture/abutment based on patient anatomy and restoration plan.
Clinical Evidence
Bench testing only. Evaluations included mechanical fatigue testing (ISO 14801), surface analysis (SEM/EDS), sterilization validation (ISO 11137, ISO 17665), shelf-life/packaging (ASTM F88, F1929), MRI safety assessment, bacterial endotoxin testing (USP <85>), and biocompatibility (ISO 10993).
Technological Characteristics
Materials: Pure Titanium Grade 4 (ASTM F67) for fixtures; Ti 6Al 4V ELI (ASTM F136) for abutments/screws. Surface: SLA (Sandblasted, Large-grit, Acid-etched). Connection: Internal hex. Sterilization: Gamma or end-user (autoclave). Dimensions: Various diameters/lengths. Connectivity: None (mechanical system).
Indications for Use
Indicated for partially or fully edentulous patients (mandibles and maxillae) requiring single or multiple unit restorations (cemented, screw-retained, or overdenture) or fixed bridgework support. Intended for delayed loading. Implants ≥5mm diameter are for molar region use.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
HOOWON EDI Co., Ltd % Sanglok Lee Manager Wise Company Inc. #507, #508, 166 Gasan digital 2-ro, Geumcheon-gu Seoul. Seoul 08503 REPUBLIC OF KOREA
May 24, 2024
Re: K231426
Trade/Device Name: 8plant Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: April 24, 2024 Received: April 24, 2024
Dear Sanglok Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products: and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K231426
Device Name 8plant Implant System
#### Indications for Use (Describe)
8plant Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Implant bodies with a diameter of 5mm or more are intended to be used in the molar region.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 10pt;">Prescription Use (Part 21 CFR 801 Subpart D)</span> </span> </div> | <div> <span> <span style="font-size: 10pt;">Over-The-Counter Use (21 CFR 801 Subpart C)</span> </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510(k) Summary
## Date revised: May 24, 2024
## 01. Applicant / Submitter
HOOWON EDI Co., Ltd 20-13, Seogimhaesandan-ro, Gimhae-si, Gyeongsangnam-do, Republic of Korea, 50967 TEL: +82 51 714 7240 Email: who1edi@gmail.com
## 02. Submission Correspondent
Sanglok, Lee Wise COMPANY Inc. #1107, #1108, 204 Gasan digital 1-ro, Geumcheon-gu, Seoul, Korea TEL: +82 2 831 3615 Email: info@wisecompany.org
## 03. Proposed Device Identification
Trade Name: 8plant Implant system Classification Name: Implant, Endosseous, Root-Form Common Name: Endosseous Dental Implant Device Class: Class II Requlation Number: 21 CFR 872.3640 Primary Product Code: DZE Secondary Product Code: NHA Review Panel: Dental
## 04. Indication for use
8plant Implant System is indicated for use in partially or fully edentulous mandibles and maxillae, in support of single or multiple unit restorations including; cemented retained, or overdenture restorations, and final or temporary abutment support for fixed bridgework. It is intended for delayed loading. Implant bodies with a diameter of 5mm or more are intended to be used in the molar region.
## 05. Predicate Device Identification
Primary Predicate Device
| K Number | Manufacturer | Trade Name |
|----------|-----------------------|----------------------|
| K211090 | Izenimplant Co., Ltd. | ZENEX Implant System |
#### Reference Devices
| K Number | Manufacturer | Trade Name |
|----------|---------------------------|----------------------------------------|
| K122231 | MegaGen Implant Co., Ltd. | Xpeed AnyRidge Internal Implant System |
| K210134 | Dentis Co., Ltd. | Dentis s-Clean s-Line |
| K210161 | MegaGen Implant Co., Ltd. | AnyOne Onestage Implant System |
{5}------------------------------------------------
## 06. Device Description
8plant Implant System consists of SF Fixtures, Cover Screw, Abutment, and Abutment Screw.
- 1) 8plant Implant System (SF Fixtures and Cover Screw)
This titanium dental implant features a SLA (Sandblasted with Large-grit and Acid-etched) surface with the alveolar bone and is designed to support single or multiple-unit restorations in partially or fully edentulous patients. It is secured to the upper structure using an internal hex fastening system.
#### 2) 8plant Abutment System (Abutment and Abutment Screw)
8plant Abutment System is compatible with the SF Fixtures. 8plant Abutment System is intended for use with a dental implant to provide support for prosthetic restorations such as crowns, bridges, or overdentures. The 8plant Abutment System consists of the following;
| Healing Abutment | Healing is used during the healing period prior to restorations and maintain the<br>shape of the gum. |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Rigid Abutment | It is a one-body product that consists of a support column and a screw integrally<br>and the implant fixture and the 11° taper angle body taper face meet and are used |
| Locator Abutment | Locator Abutment is a product that is located inside and acts as a support when<br>manufacturing Overdenture through an implant fixture. |
| Transfer Abutment | It is an abutment used when manufacturing screw retaining prosthesis in multiple<br>cases. |
| Angled Abutment | It is an abutment used when manufacturing screw retaining prosthesis in multiple<br>cases. |
| Temporary Abutment | Temporary Abutment is used by removing the healing abutment as an abutment for<br>making temporary prostheses. |
| Abutment Screw | It is used to fix the abutment to the fixture. |
{6}------------------------------------------------
## 7. Substantial Equivalence Comparison
Dental Implant System has the same intended uses, principle of operation and similar technicality as predicate device and reference devices.
## 1) Fixture and Cover Screw
## 1-1) Fixture
| | Subject Device | Predicate Device | Reference Device |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | HOOWON EDI Co., Ltd. | Izenimplant Co., Ltd. | MegaGen Implant Co., Ltd. |
| Device Name | 8palnt Implant System | ZENEX Implant System | Xpeed AnyRidge Internal Implant System |
| 510(k) Number | K231426 | K211090 | K122231 |
| Product Code / | DZE / 21CFR 872.3640 | DZE / 21CFR 872.3640 | DZE / 21CFR 872.3640 |
| Regulation No. | | | |
| Classification | Class II | Class II | Class II |
| Indications for Use | 8palnt Implant System is indicated for use<br>in partially or fully edentulous mandibles<br>and maxillae, in support of single or<br>multiple unit restorations including;<br>cemented retained, screw retained, or<br>overdenture restorations, and final or<br>temporary abutment support for fixed<br>bridgework. It is intended for delayed<br>loading. Implant bodies with a diameter of<br>5mm or more are intended to be used in<br>the molar region. | ZENEX Implant System is indicated for use in<br>partially or fully edentulous mandibles and<br>maxillae, in support of single or multiple unit<br>restorations including; cemented retained,<br>screw retained, or overdenture restorations,<br>and final or temporary abutment support for<br>fixed bridgework. It is intended for delayed<br>loading. Wide Fixture System is intended to be<br>used in the molar region. | The Xpeed AnyRidge Internal Implant System<br>is intended to be surgically placed in the<br>maxillary or mandibular molar areas for the<br>purpose providing prosthetic support for dental<br>restorations (Crown, bridges, and<br>overdentures) in partially or fully edentulous<br>individuals. It is used to restore a patient's<br>chewing function. Smaller implants (less than Ø<br>6.0 mm) are dedicated for immediate loading<br>when good primary stability is achieved and<br>with appropriate occlusal loading. Larger<br>implants are dedicated for the molar region and<br>are indicated for delayed loading. |
| Structure | - Internal Hex connected<br>- Submerged Fixture<br>- Straight/Taper body shape | - Internal Hex connected<br>- Submerged Fixture<br>- Straight/Taper body shape | - Internal Hex connected<br>- Submerged Fixture<br>- Straight/Taper body shape |
| Body Diameter(Ø)<br>and Length (mm) | Ø3.75 x L 8.5, 10, 11.5, 13<br>Ø4.25 x L7, 8.5, 10, 11.5, 13<br>Ø4.6 x L7, 8.5, 10, 11.5, 13<br>Ø5.05 x L7, 8.5, 10, 11.5, 13<br>Ø5.6 x L7, 8.5, 10, 11.5, 13<br>Ø6.2 x L7, 8.5, 10, 11.5, 13<br>Ø7.1 x L7, 8.5, 10, 11.5, 13 | Mini<br>Ø 3.75 x L8.5, 10, 11.5, 13, 15<br>Regular<br>Ø 4.25 x L7, 8.5, 10, 11.5, 13,15<br>Ø 4.6 x L7, 8.5, 10, 11.5, 13, 15<br>Wide<br>Ø 5.05 x L7, 8.5, 10, 11.5, 13, 15<br>Ø 5.4 x L7, 8.5, 10, 11.5, 13<br>Ø 5.9 x L7, 8.5, 10, 11.5, 13 | Normal thread<br>Ø4.0 X 7.7, 9.2, 10.7, 12.2, 14.2, 17.2<br>Ø4.4 X 7.7, 9.2, 10.7, 12.2, 14.2, 17.2<br>Ø4.9 X 7.7, 9.2, 10.7, 12.2, 14.2, 17.2<br>Ø5.4 X 7.7, 9.2, 10.7, 12.2, 14.2, 17.2<br>Ø5.9 X 7.7, 9.2, 10.7, 12.2, 14.2, 17.2<br>Deep thread<br>Ø6.4 X 7.9, 9.4, 10.9, 12.4, 14.4<br>Ø6.9 X 7.9, 9.4, 10.9, 12.4, 14.4 |
{7}------------------------------------------------
| | | Ø 6.75 x L7, 8.5, 10, 11.5, 13 | Ø7.4 X 7.9, 9.4, 10.9, 12.4, 14.4<br>Ø7.9 X 7.9, 9.4, 10.9, 12.4, 14.4<br>Ø8.4 X 7.9, 9.4, 10.9, 12.4, 14.4 |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|-------------------------------------------------------------------------------------------------------------|
| | | | |
| Material | Pure Titanium Grade 4 (ASTM F67) | Pure Titanium Grade 4 (ASTM F67) | Pure Titanium Grade 4 (ASTM F67) |
| Surface Treatment | Sandblasted and acid-etched | Sandblasted and acid-etched | Sandblasted and acid-etched |
| Sterilization | Gamma | Radiation | Radiation |
| Shelf life | 2 years | 5 years | 5 years |
| Similarities | The subject device has the same device characteristics with the predicate device such as diameters, length, intended use, material, principle<br>of operation, connection design, design, structure and sterilization method. | | |
| Differences | The diameter of the subject device includes Ø 7.1 that is slightly bigger than the primary predicate. To support this difference, K122231 is<br>added as a reference device to support the difference in diameter.<br>As the subject device doesn't categorize the fixture size into 'Mini, Regular, and Wide' like the predicate device, we have removed the phrase<br>related to the Wide Fixture System and revised it as follows: Implant bodies with a diameter of 5mm or more are intended to be used in the<br>molar region. | | |
# 1-2) Cover Screw
| | | Subject Device | Predicate Device |
|-------------------------------|------------|---------------------------|---------------------------|
| Manufacturer | | HOOWON EDI Co., Ltd. | Izenimplant Co., Ltd. |
| Device Name | | 8paint Implant System | ZENEX Implant System |
| 510(k) Number | | K231426 | K211090 |
| Product Code / Regulation No. | | DZE / 21CFR 872.3640 | DZE / 21CFR 872.3640 |
| Classification | | Class II | Class II |
| Design | | | |
| Connection Type | | Non-Hex | Non-Hex |
| Dimension | D(ø) | 3.1<br>3.58 | 3.0 ~ 3.9 |
| | Length(mm) | 5.1 ~ 5.9 | 5 ~ 7.3 |
| | Angle(°) | 0 | 0 |
| Material | | Ti 6Al 4V ELI (ASTM F136) | Ti 6Al 4V ELI (ASTM F136) |
| Surface Treatment | | No Treatment | No Treatment |
{8}------------------------------------------------
# K231426
| Sterilization | End User Sterilization | Gamma |
|---------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Shelf life | Non-shelf Life | 5 years |
| Similarities | The subject device has the same device characteristics with the predicate device such as diameters, length, intended use, material, principle of operation, connection design, design, structure and surface treatment method. | |
| Differences | The subject device's sterilization is End User Sterilization, which is different from the predictive Device. Sterilization validation has been conducted to support this difference. | |
## 2) Abutment and Abutment Screw
# 2-1) Healing Abutment
| | | Subject Device | Predicate Device |
|-------------------------------|----------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Manufacturer | | HOOWON EDI Co., Ltd. | Izenimplant Co., Ltd. |
| Device Name | | 8palnt Implant System | ZENEX Implant System |
| 510(k) Number | | K231426 | K211090 |
| Product Code / Regulation No. | | DZE / 21CFR 872.3640 | DZE / 21CFR 872.3640 |
| Classification | | Class II | Class II |
| Design | | Image: Subject Device Design | Image: Predicate Device Design |
| Connection Type | | Non-Hex | Non-Hex |
| | D(ø) | 4.3,4.8,5.3,5.8,6.3,7.3,8.3 | 4.3 ~ 9.0 |
| Dimension | G/H(mm) | 1.0 ~ 12.0 | 2.0 ~ 9.0 |
| | Angle(°) | 0 | 0 |
| Material | | Ti 6Al 4V ELI (ASTM F136) | Ti 6Al 4V ELI (ASTM F136) |
| Surface Treatment | | No Treatment | No Treatment |
| Sterilization | | End User Sterilization | Gamma |
| Shelf life | | Non-shelf Life | 5years |
| Similarities | | The subject device has the same device characteristics with the predicate device such as intended use, material, connection<br>design, design, structure and surface treatment method. | |
| Differences | | G/H of subject device is slightly larger than the Predictive Device. This discrepancy is not critical to the device.<br>The subject device's sterilization is End User Sterilization, which is different from the predictive Device. Sterilization validation<br>has been conducted to support this difference. | |
{9}------------------------------------------------
# K231426
# 2-2) Rigid Abutment
| | | Subject Device | Reference Device |
|-------------------------------|----------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Manufacturer | | HOOWON EDI Co., Ltd. | Dentis Co., Ltd. |
| Device Name | | 8palnt Implant System | Dentis s-Clean s-Line |
| 510(k) Number | | K231426 | K210134 |
| Product Code / Regulation No. | | DZE / 21CFR 872.3640 | DZE / 21CFR 872.3640 |
| Classification | | Class II | Class II |
| Design | | | |
| Connection Type | | Non-Hex | Non-Hex |
| | D(ø) | 4.0~7.0 | 4.5~6.5 |
| Dimension | P/H(mm) | 4.0,5.5,7.0 | - |
| | Angle(°) | 0 | 0 |
| Material | | Ti 6Al 4V ELI (ASTM F136) | Ti 6Al 4V ELI (ASTM F136) |
| Surface Treatment | | No Treatment | No Treatment |
| Sterilization | | End User Sterilization | End User Sterilization |
| Similarities | | The subject device has the same device characteristics with the predicate device such as diameters, Length, intended use,<br>material, principle of operation, connection design, design, structure, surface treatment and sterilization method. | |
| Differences | | The subject's P/H contains P/Hs that are not supported by the reference device, K210134. This discrepancy is not critical to the<br>device. | |
# 2-3) Locator Abutment
| | Subject Device | Reference Device |
|-------------------------------|-----------------------|--------------------------------|
| Manufacturer | HOOWON EDI Co., Ltd. | MegaGen Implant Co., Ltd. |
| Device Name | 8palnt Implant System | AnyOne Onestage Implant System |
| 510(k) Number | K231426 | K210161 |
| Product Code / Regulation No. | DZE / 21CFR 872.3640 | DZE / 21CFR 872.3640 |
| Classification | Class II | Class II |
{10}------------------------------------------------
| Design | Image: Dental implant | Image: Dental implant | |
|-------------------|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------|
| Connection Type | Non-Hex | Non-Hex | |
| Dimension | D(ø) | 3.87,4.0 | 3.89 |
| | G/H(mm) | 0.84~5.84 | 0.3,0.8,1.8,2.8,3.8,4.8,5.8,6.8 |
| | Angle(°) | 0 | Up to 20° |
| Material | | Ti 6Al 4V ELI (ASTM F136) | Ti 6Al 4V ELI (ASTM F136) |
| Surface Treatment | | No Treatment | Partial TiN coated |
| Sterilization | | End User Sterilization | Non-Sterile |
| Similarities | | The subject device has the same device characteristics with the predicate device such as diameters, length, intended use,<br>material, principle of operation, connection design, design, structure and method. | |
| Differences | | The diameter of the subject device is slightly larger than that of the reference device, K210161. This difference does not affect<br>the basic functionality of the device. Sterilization validation has been conducted. | |
# 2-4) Transfer Abutment
| | Subject Device | Predicate Device | Reference Device |
|-------------------------------|----------------------------|----------------------------|----------------------------|
| Manufacturer | HOOWON EDI Co., Ltd. | Izenimplant Co., Ltd. | Dentis Co., Ltd. |
| Device Name | 8palnt Implant System | Cemented Abutment | Dentis s-Clean s-Line |
| 510(k) Number | K231426 | K211090 | K210134 |
| Product Code / Regulation No. | DZE / 21CFR 872.3640 | DZE / 21CFR 872.3640 | DZE / 21CFR 872.3640 |
| Classification | Class II | Class II | Class II |
| Design | Image: Gray dental implant | Image: Gold dental implant | Image: Gray dental implant |
| Connection Type | Hex | Hex | Hex |
{11}------------------------------------------------
| Dimension | D(ø) | 4.0,4.5,5.0,5.5,6.0,6.5,7.0 | 4.5 ~ 6.5 | 4.5, 5.5, 6.5 |
|-------------------|----------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|---------------------------|
| | P/H(mm) | 4.0,5.5,7.0 | 4.0~7.0 | - |
| | Angle(°) | 0 | 0 | 0 |
| Material | | Ti 6Al 4V ELI (ASTM F136) | Ti 6Al 4V ELI (ASTM F136) | Ti 6Al 4V ELI (ASTM F136) |
| Surface Treatment | | No Treatment | Partial TiN coated in upper | No Treatment |
| Sterilization | | End User Sterilization | Non-Sterile | End User Sterilization |
| Similarities | | The subject device has the same device characteristics with the predicate device such as diameters, length, intended use, material, principle of operation, connection design, design, structure and method. | | |
| Differences | | The diameters of the subject device include Ø 4.0 and Ø 7.0 which are slightly smaller or larger than the predicate device. This difference does not affect the basic functionality of the device. Sterilization validation has been conducted. | | |
# 2-5) Angled Abutment
| | | Subject Device | Predicate Device |
|-------------------------------|----------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Manufacturer | | HOOWON EDI Co., Ltd. | Izenimplant Co., Ltd. |
| Device Name | | 8palnt Implant System | Angled Abutment |
| 510(k) Number | | K231426 | K211090 |
| Product Code / Regulation No. | | DZE / 21CFR 872.3640 | DZE / 21CFR 872.3640 |
| Classification | | Class II | Class II |
| Design | | Image: Subject Device Design | Image: Predicate Device Design |
| Connection Type | | Hex | Hex, Non-Hex |
| | D(ø) | 4.0,4.5,5.0,6.0 | 4.5, 5.5, 6.5 |
| Dimension | G/H(mm) | 0.8,1.8,2.8,3.8 | 2.8, 3.8 |
| | Angle(°) | 17 | 15, 25 |
| Material | | Ti 6Al 4V ELI (ASTM F136) | Ti 6Al 4V ELI (ASTM F136) |
| Surface Treatment | | No Treatment | Partial TiN coating |
| Sterilization | | End User Sterilization | Non-Sterile |
| Similarities | | The subject device has the same device characteristics with the predicate device such as diameters, length, intended use, material, principle of operation, connection design, design, structure and sterilization method. | |
| Differences | | The diameter of the subject contains Ø4.0, slightly smaller than the reference device, K211090 This is discrepancy is not critical to | |
{12}------------------------------------------------
#### the device. Sterilization validation has been conducted.
# 2-6) Temporary Abutment
| | | Subject Device | Predicate Device |
|-------------------------------|----------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Manufacturer | | HOOWON EDI Co., Ltd. | Izenimplant Co., Ltd. |
| Device Name | | 8palnt Implant System | ZENEX Implant System |
| 510(k) Number | | K231426 | K211090 |
| Product Code / Regulation No. | | DZE / 21CFR 872.3640 | DZE / 21CFR 872.3640 |
| Classification | | Class II | Class II |
| Design | | Image: Subject Device Design | Image: Predicate Device Design |
| Connection Type | | Hex | Hex, Non-Hex |
| | D(ø) | 4.0, 4.6 | 4.0,4.5 |
| Dimension | P/H(mm) | 10 | 10 |
| | Angle(°) | 0 | 0 |
| Material | | Ti 6Al 4V ELI (ASTM F136) | Ti 6Al 4V ELI (ASTM F136) |
| Surface Treatment | | No Treatment | Non-Coating |
| Sterilization | | End User Sterilization | Non-Sterile |
| Similarities | | The subject device has the same device characteristics with the predicate device such as diameters, length, intended use,<br>material, principle of operation, connection design, structure and method. | |
| Differences | | The diameter of the subject is slightly larger than that of the predict Device. This difference does not affect the basic<br>functionality of the device. Sterilization validation has been conducted. | |
# 2-7) Abutment Screw
| | Subject Device | Reference Device |
|-------------------------------|-----------------------|-----------------------|
| Manufacturer | HOOWON EDI Co., Ltd. | Dentis Co., Ltd. |
| Device Name | 8paInt Implant System | Dentis s-Clean s-Line |
| 510(k) Number | K231426 | K210134 |
| Product Code / Requlation No. | DZE / 21CFR 872.3640 | DZE / 21CFR 872.3640 |
| Classification | Class II | Class II |
{13}------------------------------------------------
| Design | | | |
|-------------------|------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Connection Type | | Hex | Hex |
| Dimension | D(ø) | 2.1,2.2,2.3,2.5,2.9 | 2.32 |
| | Length(mm) | 3.75,8.0,8.35,10.2 | 9.4 |
| | Angle(°) | 0 | 0 |
| Material | | Ti 6Al 4V ELI (ASTM F136) | Ti 6Al 4V ELI (ASTM F136) |
| Surface Treatment | | No Treatment | Non-Coating |
| Sterilization | | End User Sterilization | End User Sterilization |
| Similarities | | The Subject device has the same device characteristics with the predicate device such as diameters, length, intended use,<br>material, principle of operation, connection design, design, structure, surface treatment and sterilization method. | |
| Differences | | The diameter of the subject device includes Ø 2.5,2.9 that is slightly bigger than the reference device, K210134. This<br>discrepancy is not critical to the device. | |
## 08. Non-clinical Test
Bench tests were conducted to verify that the subject device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The results of the below tests have met the standards and demonstrated the substantial equivalence with the predicate device. The test results demonstrated that the subject device complies with the following standards:
#### · Mechanical Performance
- Fatigue testing for 8plant Implant system was conding to the "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseus Dental Implants and "SO 14001:2016 Dentisty - Fatigue test for Endosseous dental implants" under the worst case scenario.
#### · Surface Analysis
Surface modification testing, utilizing EDS and SEM evaluations was conducted across multiple lots to consistency and appropriateness of the S.L.A. treated implant surface according to the "Guidance for industry and FDA staff Class II Special Controls Guidance Document Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments."
#### · Sterilization, Shelf-life and Packaqing for Sterile Product
- Gamma sterilization validation according to ISO 11137-1: 2006+A2:2018, ISO 11137-2:2013 and ISO 11137-3:2017
{14}------------------------------------------------
## K231426
- End User Sterilization Validation according to ISO 11138-1:2017. ISO 17665-1:2006. ISO 17665-2:2009. ISO 11737-1:2018 and ISO 11737-2:2019
- Real time aging test according to ASTM F88, ASTM F1929
#### = MR Environment Condition
Non-clinical worst-case MR review was performed to evice components in the MRI environment using scientific rationale and published iterature (e.g., Woods, Terry O., Jana G. Delfin, and Sunder Rajan." Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices," Journal of Testing and Eraluation 49.2 (2019): 783-795), based on the entire (al compatible implant bodies, abutments, and material composition. Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetically including magnetically induced displacement force and torque.
#### · Bacterial Endotoxin
The subject device will not be labeled as "non-pyrogenc", and the endotoxin testing will be the subject device with the testing limit of below 0.5 EU/mL in accordance with the USP <85> "Bacterial Endotoxins Test.
#### · Biocompatibility
Biocompatibility evaluations according to ISO 10993-1 and Biocompatibility according to ISO 10993-12.
## 09. Substantial Equivalence Conclusion
The subject device has the same indications for use and technological characteristics, as the above identified predicate device for each difference was presented, similarities and differences were addressed.
Based upon the comparison and information described above, HOOWON EDI Co., Ltd. has concluded that the Splantially equivalent to the discussed primary predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.