Published clinical literature; Post-market surveillance data; Internal routine monitoring data
The sponsor used published clinical literature and post-market surveillance data to evaluate the safety and performance (osseointegration) of the device's surface treatment, comparing it to historical data from implants with the same material and surface treatment.
osseointegration; post-market surveillance; published literature; surface treatment
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature review of bone changes; Retrospective cohort/clinical evaluation; Follow-up/Duration: 11 months to 4 years (average 2.3 years)
Male and female adult patients with implants of varying diameters and lengths; Sample Size: 87 implants (174 surfaces)
Not applicable for this study
Bone level changes (improved, maintained, decreased) and osseointegration success
The Paltop Narrow Implant is indicated for use in surgical and restorative applications for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited by the adjacent teeth and roots, to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. The Paltop Narrow Implant is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
Paltop Narrow Implant is an endosseous, screw-type dental implant system designed for placement in partially edentulous jaws with limited horizontal space. The device consists of titanium alloy (ASTM F136) implants with SLA surface treatment and various abutment configurations (Single-Unit, Multi-Unit, ball, temporary). Implants are placed surgically by dental professionals. The system provides support for prosthetic devices (artificial teeth) to restore chewing function. The device features three thread/body configurations (Advanced, Advanced+, Dynamic) and is compatible with previously cleared Paltop prosthetic components. It supports immediate loading protocols when primary stability is achieved. The device is provided sterile (gamma irradiation) or non-sterile (steam sterilization required). Clinical benefit includes restoration of chewing function and esthetics in patients with limited anatomical space.
Clinical Evidence
No clinical trials were conducted for this submission. Evidence relies on non-clinical bench testing (ISO 14801 fatigue, pull-out, bone-to-implant contact) and a literature review of 87 implants (174 surfaces) using identical materials and surface treatments. Literature results showed 92% of implants maintained or improved bone levels, with 100% achieving successful osseointegration over a 2.3-year average follow-up. Biocompatibility was supported by historical SEM/EDS data, post-market surveillance, and endotoxin testing (USP <161>).
Technological Characteristics
Materials: Titanium Ti-6Al-4V ELI (ASTM F136), PEEK. Surface: SLA (Sand-blasted, Large Grit, Acid-Etched). Design: Endosseous screw-type, internal hex connection, parallel coronal/mid-section with apical taper. Sterilization: Gamma irradiation (sterile components) or steam (non-sterile components). Connectivity: N/A (mechanical dental implant).
Indications for Use
Indicated for partially edentulous patients requiring dental implants in mandibular central/lateral incisor or maxillary lateral incisor regions with limited horizontal space. Supports prosthetic devices to restore chewing function. Indicated for immediate loading if primary stability and occlusal conditions are met.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Paltop Advanced Dental Solutions, Ltd % Chris Brown Manager Aclivi, LLC 3250 Brackley Drive Ann Arbor, Michigan 48105
Re: K210117
Trade/Device Name: Paltop Narrow Implant Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: November 1, 2021 Received: November 3, 2021
Dear Chris Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K210117
Device Name Paltop Narrow Implant
Indications for Use (Describe)
The Paltop Narrow Implant is indicated for use in surgical and restorative applications for placement in the mandbular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited by the adjacent teeth and roots, to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. The Pattop Narrow Implant is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------|--|
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |
| <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K210117 510(k) Summary Paltop Advanced Dental Solutions, Ltd Paltop Narrow Implant
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Paltop Advanced Dental Solutions, Ltd<br>Hashita 5, Industrial Park<br>Caesarea 3088900<br>Israel |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Telephone: | +(972) 4-627 1711 |
| Fax: | +(972) 4-627 5363 |
| Official Contact | Zina Gurgov, Director of QA/RA<br>Email: zgurgov@keystonedental.com |
| Consultant | Chris Brown, BSEE<br>Aclivi, LLC<br>3250 Brackley Drive<br>Ann Arbor, Michigan 48105<br>Telephone: +1 (810) 360-9773<br>E-mail: acliviconsulting@gmail.com |
Date submitted: 11/30/2021
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | Paltop Narrow Implant |
|---------------------------|-------------------------------------------------------------------------------------------|
| Common Name: | Implant, Dental, Endosseous, Root-Form |
| Classification Name: | Endosseous dental implant |
| Classification Regulation | 21 CFR 872.3640 |
| Device Class: | Class II |
| Product Code: | DZE, NHA |
| Review Panel: | Dental |
| Reviewing Branch: | Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1) Dental Devices (DHT1B) |
#### PREDICATE DEVICE INFORMATION
The devices within this submission are substantially equivalent in indications, intended use and design principles to the following predicate and reference devices:
| 510(k) | Predicate Device Name | Company Name |
|---------|-----------------------|----------------------------------|
| K130462 | Paltop Narrow Implant | Paltop Advanced Dental Solutions |
| | Reference Predicate Device Name | |
|---------|----------------------------------------|----------------------------------|
| K102436 | NobelActive 3.0 | Nobel Biocare USA, LLC |
| K112795 | Paltop Advanced Dental Solution System | Paltop Advanced Dental Solutions |
| K131451 | Paltop Dental Sterile Accessories | Paltop Advanced Dental Solutions |
{4}------------------------------------------------
### DEVICE DESCRIPTION
This submission expands the Predicate Narrow Implant device to include 3.0 mm diameter implants, additional implant thread configurations and additional prosthetic components to the previously cleared Paltop Narrow Implant (K130462).
Endosseous dental implants are surgically implanted into a patient's mouth to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. Endosseous dental implant abutments are secured to dental implants with a retaining screw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function.
Paltop Narrow Implant are one- and two-stage endosseous screw type dental implants with associated abutments. The implants, titanium abutment screws are fabricated from a Titanium-6 Aluminum 4 Vanadium ELI titanium alloy which conforms to ASTM F136. The Paltop Narrow Implants are surface treated with SLA (Sand-blasted, Large Grit, Acid Etched).
The implants are available in three thread/body configurations: Advanced, Advanced+, and Dynamic. The Advanced, Advanced+ and Dynamic implants have micro threads at the neck. All implants have a parallel coronal and mid-section area with an apical taper. The families have slight differences in thread profile and either a passive or active apex. The 3.0 mm diameter implants are prosthetically compatible with the previously cleared Paltop Narrow Implant (K130462) prosthetic devices.
This submission introduces Single-Unit Abutments for the Paltop Narrow Implant device. The submission replaces the original straight Multi-Unit Abutments from the K130462 submission and includes additional gingival heights. The submission also introduces an angulated Multi-Unit titanium abutment, all compatible with the Paltop Narrow Implant device. The submission expands the compatible prosthetic components to include new temporary titanium abutments and healing caps compatible with Single-Unit and Multi-Unit titanium implant abutments.
The Subject device implants may be used with the Predicate device abutments previously cleared under K130462, based on non-clinical performance bench testing provided in this submission.
The Subject device abutments may be used with the Predicate device implants previously cleared under K130462, based on non-clinical performance bench testing provided in this submission. The only exception is the implant cover screw (P/N 80-70100) which is specific to the 3.0 mm diameter implants in this submission.
The Subject device Multi-Unit Abutment components such as copings/interfaces, temporary abutments, cylinders, and screws may be used with the Predicate device Multi-Unit Abutments previously cleared under K130462 based on non-clinical performance bench testing provided in this submission.
All implants and prosthetic components are one-time use devices. All Subject devices in this submission are provided sterile and sterilized by gamma irradiation except for Single-Unit and Multi-Unit copings and all replacement screws which are provided non-sterile. Devices provided as non-sterilized by steam.
# INDICATIONS FOR USE
The Paltop Narrow Implant is indicated for use in surgical and restorative applications for placement in the mandibular central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited by the adjacent teeth and roots, to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. The Paltop Narrow Implant is indicated also for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
#### EQUIVALENCE TO MARKETED DEVICE
The Subject device is highly similar to the Predicate device with respect to Indications for Use and technological principles. The Comparison tables below compare the Indications for Use and Technological Characteristics of the Subject and Predicate/Reference devices.
{5}------------------------------------------------
# Comparison of Indications for Use Statements
| Device | Indications for Use Statement |
|--------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device<br>Paltop Narrow<br>Implant | The Paltop Narrow Implant is indicated for use in surgical and restorative applications for placement in the mandibular<br>central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is<br>limited by the adjacent teeth and roots, to provide support for prosthetic devices, such as artificial teeth, in order to<br>restore the patient's chewing function. The Paltop Narrow Implant is indicated also for immediate loading when good<br>primary stability is achieved and with appropriate occlusal loading. |
| Predicate<br>Device<br>Paltop Narrow<br>Implant<br>(K130462) | The Paltop Narrow Implant is indicated for use in surgical and restorative applications for placement in the mandibular<br>central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is<br>limited by the adjacent teeth and roots, to provide support for prosthetic devices, such as artificial teeth, in order to<br>restore the patient's chewing function. The Paltop Narrow Implant is indicated also for immediate loading when good<br>primary stability is achieved and with appropriate occlusal loading. |
| Reference<br>Device<br>NobelActive 3.0<br>(K102436) | The NobelActive 3.0mm implant is indicated for use in the treatment of missing maxillary lateral incisors or the<br>mandibular central and lateral incisors to support prosthetic devices, such as artificial teeth, in order to restore chewing<br>function in partially endentulous patients. The NobelActive 3.0 implants may be put into immediate function provide that<br>stability requirements detailed in the manual are satisfied. |
| Reference<br>Device<br>Paltop Advanced<br>Dental Solutions<br>System (K112795) | The Paltop Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of<br>the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's<br>chewing function. The Paltop Dental Implant System is indicated also for immediate loading when good primary stability<br>is achieved and with appropriate occlusal loading. |
| Reference<br>Device<br>Paltop Dental<br>Sterile Accessories<br>(K134151) | The Paltop Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of<br>the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's<br>chewing function. The Paltop Dental Implant System is indicated also for immediate loading when good primary stability<br>is achieved and with appropriate occlusal loading. |
The Subject and Predicate devices have identical Indications for Use, supporting a finding of substantial equivalence. Minor differences in the wording between the Subject device and the NobelActive Reference device do not affect the intended use of the device and demonstrate the use of a 3.0 mm diameter implant in a previously cleared device. The Paltop K112795 and K131451 Reference devices have a highly similar Indications for Use statement which does not change the intended use of the device to restore a patient's chewing function. All Predicate and Reference devices are intended to provide support for dental prostheses.
{6}------------------------------------------------
### Comparison of Technological Characteristics
| Design<br>Parameter | Subject Device | Predicate Device | Reference Device | Reference Device | Reference Device |
|---------------------------------------|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| | Paltop Narrow Implant<br>Paltop Advanced Dental Solutions | Paltop Narrow Implant (K130462)<br>Paltop Advanced Dental Solutions | NobelActive 3.0 (K102436)<br>Nobel Biocare | Paltop Advanced Dental Solutions<br>System (K112795)<br>Paltop Advanced Dental Solutions | Paltop Dental Sterile<br>Accessories (K131451)<br>Paltop Advanced Dental<br>Solutions |
| Regulation # | 21 CFR 872.3640 | 21 CFR 872.3640 | 21 CFR 872.3640 | 21 CFR 872.3640 | 21 CFR 872.3640 |
| Product Code | DZE, NHA | DZE, NHA | DZE, NHA | DZE, NHA | DZE, NHA |
| Classification | Class II | Class II | Class II | Class II | Class II |
| Materials | Implants/Screws - Titanium Ti-6Al-4V ELI<br>Abutments-<br>Titanium Ti-6Al-4V ELI, PEEK | Titanium Ti-6Al-4V ELI | Implants - CP Titanium<br>Abutments - Titanium<br>Vanadium Alloy, PEEK | Implants/Screws - Titanium Ti-6Al-<br>4V ELI<br>Abutments-<br>Titanium Ti-6Al-4V ELI or PEEK | Abutments-<br>Titanium Ti-6Al-4V ELI or<br>PEEK |
| Reason for<br>Predicate/Refere<br>nce | Not Applicable | Narrow Platform Implant, Implant<br>length, implant design, implant<br>modified surface, prosthetic<br>connection, sterilization,<br>biocompatibility, how provided | 3.0 mm implant diameter,<br>implant design | Multi-unit angulated abutment<br>design | Sterile packaging of<br>prosthetic components |
{7}------------------------------------------------
### Implant Designs
| Design<br>Parameter | Subject Device | Predicate Device | Reference Device |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Paltop Narrow Implant<br>Paltop Advanced Dental Solutions | Paltop Narrow Implant (K130462)<br>Paltop Advanced Dental Solutions | NobelActive 3.0 (K102436)<br>Nobel Biocare |
| Implant Design<br>Advanced | Endosseous screw-type implant with internal<br>hex connection<br><br>Parallel coronal and midsection, micro<br>threads on neck, double leaded "V" shape<br>progressive thread, tapered apically, rounded<br>passive apex.<br><br>Diameter - Ø3.0 mm<br>Total length - 10, 11.5, 13, 16 mm | Endosseous screw-type implant with<br>internal hex connection.<br><br>Advanced - Parallel coronal and<br>midsection, micro threads on neck, double<br>leaded "V" shape progressive thread,<br>tapered apically, rounded passive apex.<br><br>Diameter - Ø3.25 mm<br>Total length - 10, 11.5, 13, 16 mm | Endosseous screw-type implant with<br>internal morse taper/hex connection.<br><br>Parallel midsection with double lead thread<br>tapered apically, active apex.<br><br>Diameter - Ø3.0 mm<br>Total length - 10, 11.5, 13, 15 mm |
| Implant Design<br>Advanced + | Endosseous screw-type implant with internal<br>hex connection<br><br>- Parallel coronal and midsection, micro<br>threads on neck, double leaded "V" shape<br>progressive thread, tapered apically, active<br>apex.<br><br>Diameter - Ø3.0 mm<br>Total length - 10, 11.5, 13, 16 mm | Endosseous screw-type implant with<br>internal hex connection.<br><br>Advanced - Parallel coronal and<br>midsection, micro threads on neck, double<br>leaded "V" shape progressive thread,<br>tapered apically, rounded passive apex.<br><br>Diameter - Ø3.25 mm<br>Total length - 10, 11.5, 13, 16 mm | Endosseous screw-type implant with<br>internal morse taper/hex connection.<br><br>Parallel midsection with double lead thread<br>tapered apically, active apex.<br><br>Diameter - Ø3.0 mm<br>Total length - 10, 11.5, 13, 15 mm |
| Implant Design<br>Dynamic | Endosseous screw-type implant with internal<br>hex connection<br><br>Parallel coronal and midsection, micro<br>threads on neck, reverse buttress thread in<br>mid-section tapering to an active apex.<br><br>Diameter - Ø3.0 mm<br>Total length - 10, 11.5, 13, 16 mm | Endosseous screw-type implant with<br>internal hex connection.<br><br>Dynamic - Parallel coronal and midsection,<br>micro threads on neck. Initial "V" shape<br>thread transitioning to reverse buttress<br>thread in mid-section. Tapering apically to<br>a passive apex.<br><br>Diameter - Ø3.25 mm<br>Total length - 10, 11.5, 13, 16 mm | Endosseous screw-type implant with<br>internal morse taper/hex connection.<br><br>Parallel midsection with double lead thread<br>tapered apically, active apex.<br><br>Diameter - Ø3.0 mm<br>Total length - 10, 11.5, 13, 15 mm |
The new implant diameter of 3.0 mm is highly similar to the 3.25 mm Predicate and supported by the Nobel Active Reference device implant designs are supported by Predicate device implants. Active apex of the Advanced + and Dynamic implants of the Subject device are additionally supported by the K102436 Reference device. Slight differences to mot affect substantial equivalence nor change the intend use of the devices. The 16 mm implant length is supported by means of the Predicate device and must be placed within the clinical in the Subject device Indications for Use statement. The Subject device has been non-chical bench performance testing.
{8}------------------------------------------------
| Design<br>Parameter | Subject Device | Predicate Device | Reference Device | Reference Device | Reference Device |
|----------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| | Paltop Narrow Implant<br>Paltop Advanced Dental Solutions | Paltop Narrow Implant (K130462)<br>Paltop Advanced Dental Solutions | NobelActive 3.0 (K102436)<br>Nobel Biocare | Paltop Advanced Dental Solutions<br>System (K112795)<br>Paltop Advanced Dental Solutions | Paltop Dental Sterile Accessories<br>(K131451)<br>Paltop Advanced Dental<br>Solutions |
| Mode of<br>Operation | Provide support for prosthetic devices,<br>such as artificial teeth, to restore the<br>patient's chewing function. | Provide support for prosthetic devices,<br>such as artificial teeth, to restore the<br>patient's chewing function. | Threaded, root-form<br>dental implants intended for<br>use in the upper and/or lower<br>jaw to support prosthetic<br>devices, such as artificial<br>teeth, in order to restore<br>patient esthetics and chewing<br>function to partially or fully<br>edentulous patients. | Provide support for single and<br>multi-unit prostheses to restore<br>the patient's chewing function. | Provide support for prosthetic<br>devices, such as artificial teeth,<br>to restore the patient's chewing<br>function. |
| Implant Surface<br>Treatment | Sand-blasted, large grit, Acid-Etched<br>(SLA) | Sand-blasted, large grit, Acid-Etched<br>(SLA) | TiUnite | Sand-blasted, large grit, Acid-<br>Etched (SLA) | n/a |
| Abutment<br>Surface<br>Treatment | None | None | Unknown | None | n/a |
| Sterilization<br>Method of sterile<br>components | Gamma Sterilization | Gamma Sterilization | Unknown | Gamma Sterilization | Gamma Sterilization |
| Sterilization<br>Method of non-<br>sterile<br>components | Steam sterilization | Steam sterilization | n/a | Steam sterilization | n/a |
| Implant/<br>Abutment<br>Interface | NP Platform Diameter<br>3.0, 3.25 mm<br>Internal interface | NP Platform Diameter<br>3.25 mm<br>Internal interface | 3.0 mm Platform Diameter<br>3 mm<br>Internal interface | SP Platform Diameter<br>3.75, 4.2 and 5.0 mm<br>Internal interface | SP Platform Diameter<br>3.75, 4.2 and 5.0 mm<br>Internal interface |
| | Subject Device | Predicate Device | Reference Device | Reference Device | |
| Abutment<br>Design | Paltop Narrow Implant<br>Paltop Advanced Dental Solutions | Paltop Narrow Implant (K130462)<br>Paltop Advanced Dental Solutions | Paltop Advanced Dental Solutions<br>System (K112795)<br>Paltop Advanced Dental Solutions | Paltop Dental Sterile Accessories<br>(K131451)<br>Paltop Advanced Dental Solutions | |
| | Straight Single-Unit -<br>1, 2, 3, 4 mm GH | Straight Multi-Unit -<br>1, 2, 3 mm GH | Straight Multi-Unit -<br>1, 2, 3 mm GH | Straight Single-Unit -<br>1, 2, 3, 4 mm GH | |
| Straight Single-<br>Unit<br>w/Interface<br>Coping | Angulation/Angle Correction<br>0 degrees<br><br>Top-Portion<br>Single-Unit Interface Coping<br><br>SUA Post Height (incl. Top Portion)<br>5.5 mm<br><br>SUA Prosthetic/Gingival Diameter<br>4.5 mm | Single Straight –<br>1, 2, 3 mm GH<br><br>Angulation/Angle Correction<br>0 degrees<br><br>Top-Portion<br>Multi-Unit Ti-Abutment<br><br>MUA Post Height (incl. Top Portion)<br>4.5 mm<br><br>MUA Prosthetic/Gingival Diameter<br>4.5 mm | Single Straight –<br>0, 1, 2, 3 mm GH<br><br>Top-Portion<br>Multi-Unit Ti-Abutment<br><br>MUA Post Height (incl. Top Portion)<br>4.5 mm<br><br>MUA Prosthetic/Gingival Diameter<br>4.5 mm | Top-Portion<br>Single-unit Cylinder<br><br>Post Height (incl. Top Portion)<br>10 mm<br><br>SUA Prosthetic/Gingival Diameter<br>4.5 mm | |
| | Straight Single-Unit -<br>1, 2, 3, 4 mm GH | Straight Multi-Unit -<br>1, 2, 3 mm GH | Straight Multi-Unit -<br>1, 2, 3 mm GH | Straight Single-Unit -<br>1, 2, 3, 4 mm GH | |
| Straight Single-<br>Unit<br>w/Temp<br>Cylinder | Angulation/Angle Correction<br>0 degrees<br><br>Top-Portion<br>Single-Unit Cylinder<br><br>Post Height (incl. Top Portion)<br>5 mm (minimum)<br><br>SUA Prosthetic/Gingival Diameter<br>4.5 mm | Angulation/Angle Correction<br>0 degrees<br><br>Top-Portion<br>Multi-Unit Ti-Abutment<br><br>Min Post Height (incl. Top Portion)<br>Not specified<br><br>MUA Prosthetic/Gingival Diameter<br>4.5 mm | Top-Portion<br>Multi-Unit Ti-Abutment<br><br>Min Post Height (incl. Top Portion)<br>Not specified<br><br>MUA Prosthetic/Gingival Diameter<br>4.5 mm | Top-Portion<br>Single-Unit Cylinder<br><br>Min Post Height (incl. Top Portion)<br>Not specified<br><br>SUA Prosthetic/Gingival Diameter<br>4.5 mm | |
| | Straight Single-Unit -<br>1, 2, 3, 4 mm GH | Straight Multi-Unit -<br>1, 2, 3 mm GH | Straight Multi-Unit -<br>1, 2, 3 mm GH | Straight Single-Unit -<br>1, 2, 3, 4 mm GH | |
| Straight Single-<br>Unit<br>w/Healing Cap | Angulation/Angle Correction<br>0 degrees<br><br>Top-Portion<br>Healing Cap for Single Unit<br><br>SUA Post Height (incl. Top Portion)<br>3.5 mm<br><br>SUA Prosthetic/Gingival Diameter<br>4.5 mm | Angulation/Angle Correction<br>0 degrees<br><br>Top-Portion<br>Healing Cap for Multi-Unit<br><br>Post Height (incl. Top Portion)<br>3.3 mm<br><br>MUA Prosthetic/Gingival Diameter<br>4.5 mm | Top-Portion<br>Healing Cap for Multi-Unit<br><br>Post Height (incl. Top Portion)<br>3.3 mm<br><br>MUA Prosthetic/Gingival Diameter<br>4.5 mm | Top-Portion<br>Healing Cap for Single Unit<br><br>Post Height (incl. Top Portion)<br>3.3 mm<br><br>SUA Prosthetic/Gingival Diameter<br>4.5 mm | |
# Mode of Operation, Modified Surfaces, Sterilization, Implant/Abutment Interface
The Mode of Operation, Modified Surfaces, Sterilization Methods and Implant/Abutment interface of the Subject device is the sponsor's Predicate and Reference devices. The "NP" Restorative with both 3.0 and 3.25 Implant Platform Diameters. Recent confirmatory performance testing has been done to validate sterilization methods.
{9}------------------------------------------------
### Straight Single-Unit Abutments
{10}------------------------------------------------
The Subject device Straight Single-Unit abutments are supported by the Straight Single-Unit abutments in the K131451 Reference device. Furthermor, they combine the features of the Single straight (one part) and Straight Multi-Unit (two-part) abutments, which are both part of the K130462 Predicate and K112795 Reference device, into a single two-part abutment. The use of Interface copings, temporary of two-part abutments are supported by the sponsor's Predicate and encompass similar dimensions.
{11}------------------------------------------------
### Multi-Unit Abutments
| Abutment<br>Design | Subject Device<br>Paltop Narrow Implant<br>Paltop Advanced Dental Solutions | Predicate Device<br>Paltop Narrow Implant (K130462)<br>Paltop Advanced Dental Solutions | Reference Device<br>Paltop Advanced Dental Solutions<br>System (K112795)<br>Paltop Advanced Dental Solutions | Reference Device<br>Paltop Dental Sterile Accessories<br>(K131451)<br>Paltop Advanced Dental Solutions |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Straight Multi-<br>Unit<br>w/Interface<br>Coping | Straight Multi-Unit -<br>1, 2, 3, 4, 5 mm GH<br>Angulation/Angle Correction<br>0 degrees<br>Top-Portion<br>Retentive Multi-Unit Interface Coping<br>Non-retentive Multi-Unit Interface<br>Coping<br>MUA Post Height (incl. Top Portion)<br>4.5, 5.5 mm<br>MUA Prosthetic/Gingival Diameter<br>5 mm | Straight Multi-Unit -<br>1, 2, 3 mm GH<br>Angulation/Angle Correction<br>0 degrees<br>Top-Portion<br>Multi-Unit Ti-Abutment<br>MUA Post Height (incl. Top Portion)<br>4.5 mm<br>MUA Prosthetic/Gingival Diameter<br>5 mm | n/a | Straight Multi-Unit -<br>1, 2, 3, 4, 5 mm GH<br>Top-Portion<br>Multi-Unit Ti-Abutment<br>MUA Post Height (incl. Top Portion)<br>10 mm<br>MUA Prosthetic/Gingival Diameter<br>5 mm |
| Straight Multi-<br>Unit<br>w/Temp<br>Cylinder | Straight Multi-Unit -<br>1, 2, 3, 4, 5 mm GH<br>Angulation/Angle Correction<br>0 degrees<br>Top-Portion<br>Multi-Unit Cylinder<br>MUA Post Height (incl. Top Portion)<br>5 mm (minimum)<br>MUA Prosthetic/Gingival Diameter<br>5 mm<br>Duration of Use | Straight Multi-Unit -<br>1, 2, 3 mm GH<br>Angulation/Angle Correction<br>0 degrees<br>Top-Portion<br>Multi-Unit Ti-Abutment<br>MUA Post Height (incl. Top Portion)<br>Not specified<br>MUA Prosthetic/Gingival Diameter<br>5 mm<br>Duration of Use<br>Not specified | n/a | Straight Multi-Unit -<br>1, 2, 3, 4, 5 mm GH<br>Top-Portion<br>Multi-Unit Ti-Abutment<br>MUA Post Height (incl. Top Portion)<br>Not specified<br>MUA Prosthetic/Gingival Diameter<br>5 mm<br>Duration of Use<br>Not specified |
| Straight Multi-<br>Unit<br>w/Healing Cap | Straight Multi-Unit -<br>1, 2, 3, 4, 5 mm GH<br>Angulation/Angle Correction<br>0 degrees<br>Top-Portion<br>Multi-Unit Healing Cap<br>Multi-Unit Healing Cap, Tapered<br>MUA Post Height (incl. Top Portion)<br>4.5 mm<br>MUA Prosthetic/Gingival Diameter<br>5 mm | Straight Multi-Unit -<br>1, 2, 3 mm GH<br>Angulation/Angle Correction<br>0 degrees<br>Top-Portion<br>Healing Cap for Multi-Unit<br>MUA Post Height (incl. Top Portion)<br>4.5 mm<br>MUA Prosthetic/Gingival Diameter<br>5 mm | n/a | Straight Multi-Unit -<br>1, 2, 3, 4, 5 mm GH<br>Top-Portion<br>Healing Cap for Multi-Unit<br>MUA Post Height (incl. Top Portion)<br>3.3 mm<br>MUA Prosthetic/Gingival Diameter<br>5 mm |
| 17° Multi-Unit<br>w/Interface<br>Coping | 17° Multi-Unit -<br>3 mm GH<br>MUA Top-Portion<br>Retentive Multi-Unit Interface Coping<br>Non-retentive Multi-Unit Interface<br>Coping<br>MUA Post Height (incl. Top Portion)<br>4.5 mm<br>MUA Prosthetic/Gingival Diameter<br>5 mm | n/a | Straight Multi-Unit -<br>1, 2, 3 mm GH<br>Angulated 25° Abutment<br>1, 2, 3 mm GH<br>MUA Top-Portion<br>Multi-Unit Ti-Abutment<br>MUA Post Height (incl. Top Portion)<br>10 mm<br>MUA Prosthetic/Gingival Diameter<br>5 mm | n/a |
| 17° Multi-Unit<br>w/Temp<br>Cylinder | 17° Multi-Unit<br>3 mm GH<br>MUA Top-Portion<br>Multi-Unit Cylinder<br>Min MUA Post Height (incl. Top<br>Portion)<br>5 mm<br>MUA Prosthetic/Gingival Diameter<br>5 mm<br>Duration of Use | n/a | Straight Multi-Unit -<br>1, 2, 3 mm GH<br>Angulated 25° Abutment<br>1, 2, 3 mm GH<br>MUA Top-Portion<br>Multi-Unit Ti-Abutment<br>Min MUA Post Height (incl. Top<br>Portion)<br>Not specified<br>MUA Prosthetic/Gingival Diameter<br>5 mm<br>Duration of Use<br>Not specified | n/a |
| | 17° Multi-Unit<br>3 mm GH | | Straight Multi-Unit –<br>1, 2, 3 mm GH<br><br>Angulated 25° Abutment<br>1, 2, 3 mm GH | |
| 17° Multi-Unit<br>w/Healing Cap | MUA Top-Portion<br>Multi-Unit Healing Cap<br>Multi-Unit Healing Cap, Tapered<br><br>MUA Post Height (incl. Top Portion)<br>4.5 mm<br><br>MUA Prosthetic/Gingival Diameter<br>5 mm | n/a | MUA Top-Portion<br>Healing Cap for Multi-Unit<br><br>MUA Post Height (incl. Top Portion)<br>3.3 mm<br><br>MUA Prosthetic/Gingival Diameter<br>5 mm | n/a |
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
The Subject device Straight Multi-Unit abutments are supported by the Straight Multi-Unit abutments in the K130462 Predicate device. The increased Gingival Height for the straight abutments is supported by the K131451 Reference device and non-clinical performance testing. The temporary cylinder configuration is supported by both the Multi-Unit Ti-Abutment and Gold-Based Abutment of the sponsor's K130462 and K131451 previously cleared devices which can be customized for individual patient occlusion and encompass similar dimensions. Duration of use of temporary cylinders is supported by non-clinical performance testing.
The Subject device 17° Angled Multi-Unit abutments are supported by the combination of the technology of the previously cleared Straight Multi-Unit and the 25° angulated abutments in the K112795 Reference device. The 17° Angulation of the Subject device Multi-unit abutments is also supported by non-clinical performance testing. The temporary cylinder configuration is supported by the sponsor's K112795 Multi-Unit Ti-Abutment is a multi-unit (2-part) abutment accessory which can be customized for individual patient occlusion. Duration of use of temporary cylinders is supported by non-clinical performance testing.
The use of Interface copings, temporary cylinders, and healing caps as part of two-part abutments are supported by the sponsor's Predicate and Reference devices.
| Temporary Abutments (direct to implant) | | | |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Abutment<br>Design | Subject Device<br>Paltop Narrow Implant<br>Paltop Advanced Dental Solutions | Predicate Device<br>Paltop Narrow Implant (K130462)<br>Paltop Advanced Dental Solutions | Reference Device<br>NobelActive 3.0 (K102436)<br>Nobel Biocare |
| Temporary<br>(Engaging,<br>Non-<br>Engaging) | Temporary (Engaging, Non-<br>Engaging)<br>0.5, 2, 3 mm GH<br>Post Height<br>5 mm minimum<br>Prosthetic/Gingival Diameter<br>4.25 mm<br>Duration of Use<br>90 days | n/a | Temporary Abutment Engaging<br>1.5m GH<br>Post Height<br>5 mm*<br>Prosthetic/Gingival Diameter<br>Not defined*<br>Duration of Use<br>Not defined* |
| Temporary<br>Titanium<br>Immediate | Temporary Titanium Immediate<br>1.5, 3 mm GH<br>Post Height<br>5 mm<br>Prosthetic/Gingival Diameter<br>3.5 mm<br>Duration of Use<br>30 days | n/a | Immediate Temporary Abutment<br>1.5, 3 mm GH<br>Post Height<br>5 mm*<br>Prosthetic/Gingival Diameter<br>Not defined*<br>Duration of Use<br>Not defined* |
# Temporary Abutments (direct to implant)
*Information not provided in 510(k) Summary document; information obtained from Reference device labeling
The Subject device Temporary Abutments are similar to the K102436 Reference device. Slight differences in gingival height dimension and duration of use are supported by the results of non-clinical performance testing.
The Subject device Temporary Titanium Intermediate abutments are highly similar to the K102436 Reference device and encompass the same gingival height and post height dimensions. Duration of use is supported by the results of non-clinical performance testing.
{15}------------------------------------------------
| Abutment<br>Design | Subject Device | Predicate Device | Reference Device |
|-------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| | Paltop Narrow Implant<br>Paltop Advanced Dental Solutions | Paltop Narrow Implant<br>(K130462)<br>Paltop Advanced Dental<br>Solutions | NobelActive 3.0 (K102436)<br>Nobel Biocare |
| Straight ball | Straight ball –<br>1, 2, 3 mm GH<br><br>Post Height<br>2.8 mm<br><br>Prosthetic/Gingival Diameter<br>4 mm | Straight ball –<br>1, 2, 3 mm GH<br><br>Post Height<br>2.8 mm<br><br>Prosthetic/Gingival Diameter<br>4 mm | n/a |
| Healing Caps –<br>Integral Thread<br>Concave<br>Emergence,<br>Straight<br>Emergence | Healing Caps – Integral Thread<br>Concave Emergence<br>4, 5, 6, 7 mm GH<br>Image: Healing Cap | Healing Caps – Integral Thread<br>Concave Emergence<br>1, 2, 3 mm GH<br>Image: Healing Cap | Healing Abutments – Integral<br>Thread<br>3, 5, 7 mm GH |
| | Prosthetic/Gingival<br>Diameter<br>4 mm<br><br>Straight Emergence<br>2, 3, 5 mm GH<br>Image: Straight Emergence | Prosthetic/Gingival Diameter<br>4 mm | Prosthetic/Gingival Diameter<br>3.2, 3.8 mm<br>Image: Healing Abutments |
| | Prosthetic/Gingival Diameter<br>4 mm<br><br>Post Height<br>1 mm | Post Height<br>1 mm | Post Height<br>Not specified |
| Cover Screw | Cover Screw<br>Ø3 mm Implants only<br><br>Prosthetic/Gingival Diameter<br>3 mm | Cover Screw<br>Ø3.25 mm Implants only<br><br>Prosthetic/Gingival Diameter<br>3.25 mm | Cover Screw<br>Ø3 mm Implants<br><br>Prosthetic/Gingival Diameter<br>3 mm |
### Ball Abutments, Healing Caps, Cover Screws
The Subject device Ball abutments are the K130462 Predicate device Ball abutments.
The Subject Healing Caps are highly similar to the K130462 Predicate device with highly similar gingival height dimensions. The difference in emergence profile of the healing abutments is supported by non-clinical performance testing. The higher gingival height dimensions are supported by the K102436 Reference device.
Minor differences in the implant external thread designs, dimensions, and correction angles do not affect substantial equivalence.
Differences between the Subject and Predicate device dimensions or designs are supported by Reference devices. Any differences in implant abutment designs or dimensions have been mitigated and demonstrated to be suitable for intended use through non-clinical bench performance testing.
# NON-CLINICAL PERFORMANCE TEST DATA
Fatigue testing was performed according to the requirements of ISO 14801:2016, Dentistry – Implants – Dynamic loading test for Endosseous Dental Implants. The worst-case scenario was chosen based on the FDA Guidance, Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments.
Bone to implant contact area comparison and pull- out testing was performed to compare the Subject, Predicate and Reference devices.
{16}------------------------------------------------
The Subject devices have the identical nature of body contact duration, material formulation and sterilization methods compared to the sponsor's Predicate and Reference devices.
Test results and Biological Evaluation performed on the sponsor's Predicate and Reference devices to demonstrate suitable biocompatibility of the Subject device. The evaluation included review of published literature, internal routine monitoring data related to implant modified surface treatment and post-market surveillance data on Paltop implants subject to the same modified surface treatment and cleaning process as the Subject device.
The published literature evaluated bone changes on patients implants made from the same material and with the same surface treatment as the Subject device. Implants of varying diameters and lengths were placed in male and female adult population patients. Placement locations varied and included placement in extraction sockets, healed bone, and grafted bone locations. Bone levels were evaluated at time of abutment placement and final loading and a follow-up evaluation time from final abutment insertion which ranged from 11 months to 4 years, with an average of (2.3 years). Mesial and distal surfaces were examined and graded as bone improved, bone maintained, and bone decreased. A total of 174 surfaces were graded (87 implants). Results demonstrated bone level improvement or maintenance for 92% of the implants. Bone level decreases identified were less than 1 mm, well within accepted levels. 100% of the implants met criteria for successful implant osseointegration.
Review of historical SEM/EDS data was performed on Paltop implants made from the same material and surface treatment process, and same lot numbers as the published literature. Data included SEM images to quantify residual particles over a portion of each implant which was then extrapolated to estimate total particles over the entire length of the implant.
Review of post market surveillance data and was performed to identify any significant trends in osseointegration failures. The results indicated failure rates below industry levels.
The Biological Evaluation and review of published literature, internal in-process monitoring, and post-market surveillance data support a conclusion that residual aluminum particle levels, if present, as demonstrated from the manufacturer's specific surface treatment and included assessment, do not have negative impact on osseointegration and implant survival.
Confirmatory endotoxin testing was performed according the acceptance criteria defined in USP <161>.
Test results and sterilization validations performed for the sponsor's Predicate and Reference devices demonstrate suitable sterilization of the Subject device sterile components.
A cleaning validation, and a sterilization according to ISO 17665-1, were performed for the Subject device non-sterile components and sterile components which may be modified and require subsequent sterilization.
The results of the non-clinical testing demonstrate conformance with testing requirements and support a finding of substantial equivalence with respect to the Subject and Predicate device.
# CONCLUSION
Overall, the Indications for Use statements for the Subject and Predicate devices are substantially equivalent.
Overall, the same or highly similar Technological Characteristics of the Subject device support a finding of substantial equivalence. Any differences between the Subject and Predicate dimensions or designs are supported by Reference devices. Any differences in implant abutment designs or
{17}------------------------------------------------
dimensions have been mitigated and demonstrated to be suitable for intended use through non-clinical bench performance testing.
Overall, the Subject device and Predicate devices have been demonstrated to be Substantially Equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.