ProSeal™ Bag Spike with Additive Port (423370ST, 423370)
K251340 · Epic Medical Pte. , Ltd. · ONB · May 29, 2025 · General Hospital
Device Facts
Record ID
K251340
Device Name
ProSeal™ Bag Spike with Additive Port (423370ST, 423370)
Applicant
Epic Medical Pte. , Ltd.
Product Code
ONB · General Hospital
Decision Date
May 29, 2025
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The ProSeal™ CSTD mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminations into the drug or fluid path for up to 7 days when used as intended.
Device Story
Adaptor between I.V. bags and ProSeal™ CSTD components; facilitates closed system fluid transfer. Input: I.V. bag connection; generic I.V. spikes. Operation: Uses elastomeric double membrane technology for airtight/leak-proof connections; includes spiking port with TPE stopper. Output: Closed fluid path for drug administration. Used in clinical settings by healthcare professionals. Benefits: Minimizes exposure to hazardous drug vapors/aerosols/spills; prevents microbial ingress for up to 7 days.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via leak integrity, tensile strength, flow rate, protective caps, spike penetration force, vapor containment, and 7-day microbial ingress testing per ISO 8536-4, ISO 22413, and ANSI AAMI CN27 standards. Biocompatibility assessed per ISO 10993 series.
Technological Characteristics
Materials: Polyisoprene (membrane), Polypropylene (spike port), Thermoplastic elastomer (spiking port stopper). Principle: Elastomeric double membrane for airtight/leak-proof connection. Form factor: 3-port access device. Sterilization: Ethylene Oxide (EO) to SAL 10^-6. Shelf-life: 3 years. Single-use.
Indications for Use
Indicated for use by adequately trained healthcare professionals in clinical settings for closed system fluid transfer of parenteral drugs.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
May 29, 2025
Epic Medical Pte. Ltd.
Freddie Lee
CEO/MD
105 Cecil Street #20-04, The Octagon
Singapore, SG 069534
Singapore
Re: K251340
Trade/Device Name: ProSeal™ Bag Spike with Additive Port (423370ST, 423370)
Regulation Number: 21 CFR 880.5440
Regulation Name: Intravascular Administration Set
Regulatory Class: Class II
Product Code: ONB
Dated: April 30, 2025
Received: April 30, 2025
Dear Freddie Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251340 - Freddie Lee
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251340 - Freddie Lee
Page 3
Sincerely,

David Wolloscheck, Ph.D.
Assistant Director
DHT3C: Division of Drug Delivery and
General Hospital Devices, and
Human Factors
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251340 | |
| Device Name ProSeal™ Bag Spike with Additive Port (423370ST, 423370) | |
| Indications for Use (Describe) The ProSeal™ CSTD mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminations into the drug or fluid path for up to 7 days when used as intended. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
EPIC
# K251340 – 510(k) Summary
## I. Submitter
Epic Medical Pte. Ltd.
105 Cecil Street #20-04,
The Octagon,
Singapore 069534
Phone: +65 9635 2618 / +66 81 761 5292
Contact Person: Mr. Freddie LEE, Chief Executive Officer/ Managing Director
Date Prepared: May 26, 2025
Content and Format: Prepared in accordance with 21 CFR 807.92
Type of Submission: Special
## II. Subject Device
| 510(k) Number: | K251340 |
| --- | --- |
| Trade/ Device Name: | ProSeal™ Bag Spike with Additive Port (423370ST, 423370) |
| Common/ Usual Name: | Closed Antineoplastic and Hazardous Drug Reconstitution and Transfer System |
| Regulation Number: | Set, Administration, Intravascular |
| Regulation Name: | 21 CFR 880.5440 |
| Regulatory Class: | Class: II |
| Product Code: | ONB |
## III. Predicate
| 510(k) Number: | K241988 |
| --- | --- |
| Trade/ Device Name: | ProSeal™ Closed System Bag Access |
| Common/ Usual Name: | Closed Antineoplastic and Hazardous Drug Reconstitution and Transfer System |
| Regulation Number: | Set, Administration, Intravascular |
| Regulation Name: | 21 CFR 880.5440 |
| Regulatory Class: | Class: II |
| Product Code: | ONB |
Page 1 of 7
{5}
EPIC
## IV. Purpose of Submission and Device Description
This special 510(k) Submission is to comply with FDA requirements, ensuring the ProSeal™ Bag Spike with Additive Port (or “and Seal Tab”) is safe and effective for use in closed system fluid transfer applications.
The ProSeal™ Bag Spike with Additive Port (or “and Seal Tab” (simplified ST model no.)) serves as an adaptor between I.V. bags and ProSeal™ CSTD components, facilitating closed system fluid transfer. The spiking port of this device is compatible with generic I.V. spikes. Additionally, the additive port (injection site) allows medication to be added to the bag using the cleared ProSeal™ Injector or Injector Plus (both Syringe Adaptors).
The injection site of the ProSeal™ Bag Spike with Additive Port (or “and Seal Tab”) and all corresponding interface membranes ensure a dry connection during fluid transfer. Utilizing this component and its appropriate ProSeal™ CSTD connecting component reduces the risk of microbial ingress for up to 168 hours (7 days).
## V. Indications for Use Statement
The ProSeal™ CSTD mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminations into the drug or fluid path for up to 7 days when used as intended.
## VI. Comparison of Intended Use & Technological Characteristics
The Subject device and the Predicate device share the following characteristics:
### Intended Use comparison
1) Indications for use statements
2) Primary product code and regulation number
3) Intended user population/ intended use environment
4) Intended drug type
5) Prescription use or over-the-counter use
### Technological characteristics comparison
### Equivalencies – Technology & Design
The Subject device and the Predicate device share the following design characteristics, and from the evaluation in the comparison table, there was no substantial difference from the Predicate device identified, that would raise a safety or performance issue/ concern:
1) Connection between component devices within the CSTD system
2) Transfer mechanism (role: to facilitate airtight & leak-proof connections to prevent microbial contamination)
3) Sterile barrier packaging
4) Sterilization process
5) Shelf-life validation
6) Reuse or single-use
7) Labeling specifications
Page 2 of 7
{6}
EPIC
# Comparison of intended use characteristics and the technological characteristics
An overview table summarizing the comparison of the key characteristics between the Subject and the Predicate device is provided hereunder:
Comparison of intended use characteristics and the technological characteristics
| Characteristic compared | Predicate Device (K241988) ProSeal™ Closed System Bag Access | Subject Device (K251340) ProSeal™ Bag Spike with Additive Port | Comment/Discussion |
| --- | --- | --- | --- |
| Intended use and Indications for Use statement | The ProSeal™ CSTD system mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminants into the drug or fluid path for up to 7 days when used as intended. | The ProSeal™ CSTD system mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminants into the drug or fluid path for up to 7 days when used as intended. | Same |
| Primary product code and regulation number | ONB 21 CFR 880.5440 | ONB 21 CFR 880.5440 | Same |
| Intended user population/ intended use environment | Adequately trained health care professionals/ clinical setting | Adequately trained health care professionals/ clinical setting | Same |
| Intended drug type | Parenteral drugs | Parenteral drugs | Same |
| Prescription use or over-the-counter use | R only | R only | Same |
{7}
EPIC
Comparison of intended use characteristics and the technological characteristics
| Characteristic compared | Predicate Device (K241988) ProSeal™ Closed System Bag Access | Subject Device (K251340) ProSeal™ Bag Spike with Additive Port | Comment/Discussion |
| --- | --- | --- | --- |
| Principles of operation | The infusion bag is connected with the bag access and, in order to create a fluid path connection between the infusion bag and the bag access, they are connected with the matching end of the ProSeal Injector or Injector Plus (K240171). All system components are sealed with self-resealing membranes incorporating elastomeric double membrane technology that features airtight and leakage-proof connections during the fluid transfer process. Transfer is performed through the connection of an I.V. set with an indwelling needle of the ProSeal Injector (or Injector Plus) connector, e.g. the ProSeal Closed System Administration Set (cleared K230343), into the human veins. | The infusion bag is connected with the bag access and, in order to create a fluid path connection between the infusion bag and the bag access, they are connected with the matching end of the ProSeal Injector or Injector Plus (K240171). All system components are sealed with self-resealing membranes incorporating elastomeric double membrane technology that features airtight and leakage-proof connections during the fluid transfer process. Transfer is performed through the connection of an I.V. set with an indwelling needle of the ProSeal Injector (or Injector Plus) connector. The spiking port of the ProSeal™ Bag Spike with Additive Port (or and Seal Tab (suffixed ST model)) is compatible with generic I.V. spikes connected to I.V. Administration Sets for dispensing into the human veins. | Different see Comment #1 |
| Number of access points | Device has 2 access points: 1. Closed system port (injection site) 2. Spike port, I.V. (bag spike (integrated with injection site MLL)) | Device has 3 access points: 1. Closed system port (injection site) 2. Spike port, I.V. (bag spike (integrated with injection site MLL & spiking/administration port)) 3. Spiking port stopper | Different see Comment #1 |
{8}
EPIC
Comparison of intended use characteristics and the technological characteristics
| Characteristic compared | Predicate Device (K241988)
ProSeal™ Closed System Bag Access | Subject Device (K251340)
ProSeal™ Bag Spike with Additive Port | Comment/
Discussion |
| --- | --- | --- | --- |
| Composition of fluid path materials | • Polyisoprene (IR) (membrane of injection site)
• Polypropylene (PP) (spike port, I.V.) | • Polyisoprene (IR) (membrane of injection site)
• Polypropylene (PP) (spike port, I.V.)
• Thermoplastic elastomer (TPE)
(spiking port stopper) | Different
see Comment #2 |
| Connection between component devices within the CSTD system | Collet-style fitting with elastomeric double membranes | Collet-style fitting with elastomeric double membranes | Same |
| Transfer mechanism (role: to facilitate airtight & leak-proof connections to prevent microbial contamination) | Elastomeric double membrane | Elastomeric double membrane | Same |
| Biocompatibility | Acceptable biological risks established | Acceptable biological risks established | Same
see Comments #2 |
| Sterile barrier packaging | Medical grade paper and medical plastic film, heat sealed | Medical grade paper and medical plastic film, heat sealed | Same |
| Sterilization process | Ethylene Oxide (EO), SAL 10^{-6} | Ethylene Oxide (EO), SAL 10^{-6} | Same |
| Shelf-life validation | 3 years (36 months) | 3 years (36 months) | Same |
| Reuse or single-use | Single use only | Single use only | Same |
| Labeling specifications | Met the requirements specified in 21 CFR 801 | Met the requirements specified in 21 CFR 801 | Same |
Submitter's Comments
Comment #1
Subject devices are different from the Predicate device in having a third access point – the spiking port (with its TPE stopper) for connection to standard I.V. spikes of I.V. Administration Sets. The differences between the Subject devices and Predicate device did not raise different questions of safety and effectiveness as functional and analytical testing have been conducted and data evaluated. These data are summarized in section VII.A, and section VII.B of this 510(k) Summary, and in Comment #2 hereunder. The differences were determined to be insignificant as performance results were determined to have met the intended use
Comment #2
Subject device's spiking port stopper is same subcomponent part (made of the same material and of the same configuration) used in the cleared K223674 eZSURE™ Empty Fluid Container's spiking port stopper
Page 5 of 7
{9}
EPIC
## VII. Performance Data Supporting Substantial Equivalence
### A. Functional Performance
The Subject device in this Summary was evaluated to be in conformance with the following ISO standards document:
- ANSI AAMI CN27:2021, General requirements for Luer activated valves (LAVs) incorporated into medical devices for intravascular applications
- ISO 8536-4: 2019, Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed
- ISO 15747: 2018, Plastic containers for intravenous injections
- ISO 22413:2010, Transfer sets for pharmaceutical preparations — Requirements and test methods
- ISO 80369-7: 2016, Small-bore connectors for liquids and gases in healthcare application - Part 7, Connectors for intravascular or hypodermic applications
- Intravascular-Administration-Sets-Premarket-Notification-Submissions-[510(k)]---Guidance-for-Industry-and-FDA-Staff
Bench performance verifications and validations performed on the Subject device and referred-to from existing devices (K222929 (data of isoprene membrane in cleared Injection Site w/Extended MLL (K240433), the same membrane part used in Predicate K241988) and K223674):
- Leak integrity test (functional) – per ISO 8536-4:2019, paragraph 7.2 and Annex A.3, performed on Subject device
- Tensile strength test (functional) – per ISO 8536-4:2019, paragraph 7.3 and Annex A.4, performed on Subject device
- Flow rate test (functional) – per ISO 8536-4:2019, Annex A.5.1 performed on Subject device
- Protective caps test (functional) – per ISO 8536-4:2019, Section 7.13, performed on Subject device
- IV Bag Spike penetration force test (functional) – per ISO 22413:2021, Annex A.7 on Subject device
- Spiking port penetration ability (functional) on device cleared under K223674
- Spiking port adhesion strength (functional) on device cleared under K223674
- Impermeability to microorganism (Annex C.2 ISO 15747:2018)(functional) on device cleared under K223674
- Vapor containment test per NIOSH CSTD 2016 draft test protocol (functional) from testing data on device cleared under K222929
- Microbial ingress test per FDA guidance and ANSI AAMI CN27:2021 (functional – 7-days) from testing data on devices cleared under K222929 and K223674
### B. Biocompatibility
In accordance with ISO 10993-1:2018, the Subject device is classified as: Externally Communicating Device, Blood Path Indirect, Prolonged Contact (>24hr to 30d). The following testing were conducted on the existing 510(k) cleared referred-to devices (K222929 (data of isoprene membrane in cleared Injection Site w/Extended MLL (K240433)), K223674 (data of polypropylene 2-port sub-assembly for IV bag, as well as for TPE spiking port stopper) and Predicate K241988):
- Cytotoxicity to ISO 10993-5 performed on devices cleared under K222929 and K223674
- Sensitization to ISO 10993-10 performed on devices cleared under K222929 and K223674
- Intracutaneous reactivity to ISO 10993-10 performed on devices cleared under K222929 and K223674
- Acute systemic toxicity to ISO 10993-11 performed on devices cleared under K222929 and K223674
Page 6 of 7
{10}
EPIC
- 14-day subacute/subchronic systemic toxicity to ISO 10993-11 performed on devices cleared under K222929 and K223674
- In-vitro hemolysis assessment to ISO 10993-4 performed on device cleared under K222929 and K223674
- Material mediated pyrogenicity to ISO 10993-11 performed on devices cleared under K222929 and K223674
- Chemical characterization and toxicological risk management to ISO 10993-18 and ISO 10993-17 on device cleared under Predicate K241988 (leveraging data of polypropylene spike port, I.V. as well as for isoprene membrane)
- Chemical requirements testing to ISO15747:2018, Annex B – Chemical requirements performed on device cleared under K223674
- Particulate matter analysis to ISO 8536-4:2019, Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed and USP <788> Particulate Matter in Injections were conducted on devices cleared under K223674 Predicate K241988
## C. Sterility, Shipping, and Shelf-Life
The Subject device complies with sterilization requirements of ISO 11135:2014, Sterilization of Health Care Products – Ethylene Oxide – Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices and the following testing/evaluations:
- Simulated shipping testing per ASTM D 4169-16, Standard Practice for Performance Testing of Shipping Containers and Systems (leveraging data under K151650)
- Package integrity tests per ASTM F1980-21, Standard guide for accelerated aging of sterile barrier systems for medical devices and Sterile Barrier Packaging Testing performed on the proposed device: Seal strength – ASTM F88/F88M-21, Standard test method for seal strength of flexible barrier materials; Dye Penetration – ASTM F1929-23, Standard test method for detecting seal leaks in porous medical device packaging by dye penetration; EN 868-5:2009, Packaging materials and systems for medical devices which are to be sterilized – Part 5: Heat and self-sealable pouches and reels of paper and plastic film construction – Requirements and test methods on Subject device
- Pyrogen tests per ANSI/AAMI ST72/2019, Bacterial endotoxins – Test methods, routing monitoring, and alternatives to batch testing, USP 42-NF 37 <151>, Pyrogen test (USP rabbit test), USP 42-NF 37 <161>, Medical Devices-Bacterial Endotoxin and Pyrogen Tests, USP 42- NF 37 <85>, Bacterial Endotoxins Test under K151650 and testing will be conducted on every lot
- Shelf-life of 3 years has been validated using the FDA recognized standard, ASTM 1980-21, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices on Subject device
## VIII. Clinical Tests
Not applicable
## IX. Conclusion
The difference between the Predicate and the Subject device does not raise any new or different questions of safety or effectiveness. The Subject ProSeal™ Bag Spike with Additive Port (K251340) is substantially equivalent to the Predicate, ProSeal™ Close System Bag Access (K241988), with respect to the indications for use, principles of operation and technological characteristics
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.