The Empty Fluid Container is used to hold an admixture of compatible fluids for intravenous administration to a patient. Medication transfer in and out of the container is done using aseptic technique.
Device Story
eZSURE™ Empty Fluid Container (EFC) is a sterile, single-use, non-pyrogenic flexible IV bag (100 mL or 500 mL). Device features two ports: a needle-free additive/filling port with a self-sealing silicone valve and a spiking/administration port for connection to standard IV sets. Used in clinical settings by healthcare professionals or pharmacists to hold and administer parenteral drugs. Filling occurs via syringe or automated device through the additive port; administration occurs via standard spike insertion into the administration port. The device allows for medication addition during administration. Benefits include PVC-free construction and secure, self-sealing additive access, reducing risk of contamination compared to clamped systems. The device is hung on an IV stand during use.
Clinical Evidence
Bench testing only. No clinical data. Conformance demonstrated with ISO 10993-1 (biocompatibility), ISO 15747 (plastic containers), ISO 11135 (EO sterilization), and ANSI/AAMI CN27 (Luer activated valves). Performance tests included leakage, drop resistance, water vapor permeability, spiking port penetration/adhesion, additive port liquid tightness, hanger tensile strength, particulate matter, and microbial ingress.
Technological Characteristics
Flexible polyolefin film (Nexcel® Grade M315) bag; two-port configuration. Additive port: Polypropylene, Polycarbonate, and Liquid Silicone Rubber (USP Class VI). Spiking port: Thermoplastic elastomer. Sterilization: Ethylene Oxide (SAL 10^-6). Connectivity: Standard luer lock and spike interfaces. Non-PVC, non-DEHP, latex-free.
Indications for Use
Indicated for holding admixture of compatible fluids for IV administration to patients. Used by trained healthcare professionals or pharmacists in clinical settings.
Regulatory Classification
Identification
An I.V. container is a container made of plastic or glass used to hold a fluid mixture to be administered to a patient through an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the rest of the name. The seal features an eagle emblem.
September 1, 2023
Epic Medical Pte Ltd Freddie Lee CEO 105 Cecil, 20-01 The Octagon Singapore, Singapore 069534 Singapore
Re: K223674
Trade/Device Name: eZSURE™ Empty Fluid Container (models 426030, 426110) Regulation Number: 21 CFR 880.5025 Regulation Name: I.V. Container Regulatory Class: Class II Product Code: KPE Dated: August 2, 2023 Received: August 2, 2023
Dear Freddie Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K223674
Device Name
eZSURE™ Empty Fluid Container (models 426030, 426040, 426110)
Indications for Use (Describe)
The Empty Fluid Container is used to hold an admixture of compatible fluids for intravenous administration to a patient. Medication transfer in and out of the container is done using aseptic technique.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## e Dil
### K223674- 510(k) Summary
### I. Submitter
Epic Medical Pte. Ltd. 105 Cecil Street #20-04, The Octagon, Singapore 069534. Phone: +65 9635 2618 / +66 81 761 5292
Contact Person: Mr. Freddie LEE, Chief Executive Officer/ Managing Director Date Prepared: September 1, 2023
#### II. Subject Device
| 510(k) Number: | K223674 |
|-------------------------|------------------------------------------------------------------|
| Proprietary/Trade Name: | eZSURE™ Empty Fluid Container (models 426030, 426040,<br>426110) |
| Common/Usual Name: | Empty I.V. bag |
| Regulation Name: | I.V. Container |
| Product Code: | KPE |
| Regulation Number: | 21 CFR 880.5025 |
| Device Class: | Class: II |
#### Predicate Device III.
| 510(k) Number: | K181393 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------|
| Proprietary/Trade Name: | Empty EVA Bag (models FVM0134BP, FVM0135BP,<br>FVM0136BP, FVM0137BP, FVM0138BP, FVM0139BP,<br>FVM0140BP, FVM0141BP) |
| Common/Usual Name: | Empty I.V. bag |
| Regulation Name: | I.V. Container |
| Product Code: | KPE |
| Regulation Number: | 21 CFR 880.5025 |
| Device Class: | Class: II |
{4}------------------------------------------------
### Device Description IV.
The eZSURE™ Empty Fluid Container (EFC) devices are empty single-use, sterile, nonpyrogenic flexible IV container devices/ bags. These are discarded after use.
The Subject EFC device is composed of a flexible plastic film bag and two separate ports with closures, one for injection and another, infusion. An inlet-/ entry-/ additive- port is used for filling one or more compatible fluid(s) into the bag by a transfer set/ syringe without needle, and another port, the spiking/ administration port, is used for accessing the infusate in the bag with a standard bag spike. A transfer device with a male luer lock is used to connect to the filling-/ additive- port for filling. The additive port incorporates a needle-free valve; hence no injection needle/ cannula is needed. The transfer device is removed at the end of the preparation step, and the needle-free self-sealing additive-/ injection-/ filling- port secures the admixture contents until their administration.
For administration to a patient, the device is then connected to an external IV set /IV line, via a bag spike. The IV bag is piped by inserting the spike point of a bag spike into the spiking-/administration- port of the IV bag, doing this by performing a twisting motion. When the bag is already filled, other medications can be added using the additive/ injection/ filling port, even during administration. Medication transfer in and out of the container is done using aseptic technique. The bags range in volume capacity of 100 mL, and 500 mL. The device has a hanger hole so it can be placed on an IV bag holder.
The EFC is made of non-PVC materials and provided in a two-port configuration: The needle free additive port which is used for filling the container and the other, the spiking -/ administration port, which is used for IV therapy administration from the EFC. The EFC sub-components are externally communicating devices with no contact to the blood path. The contact duration is categorized as B-prolonged, (>24h to 30d), per ISO 10993-1 :2018 biocompatibility guidelines.
#### V. Indications for Use Statement
The Empty Fluid Container is used to hold an admixture of compatible fluids for intravenous administration to a patient. Medication transfer in and out of the container is done using aseptic technique.
#### VI. Comparison of Technological Characteristics
The eZSURE™ Empty Fluid Container devices and the Predicate devices share the following characteristics:
- . Same 2-port configuration - an additive/ filling port and a spiking/ administration port
- . Equivalent basic design features - clear, flexible, and empty IV fluid bag that allows for preparation and administration of IV therapies
- Equivalent means of attachment provided for an IV administration set/ line, for the delivery of IV therapy
{5}------------------------------------------------
# GDIC
The overview table hereunder compares parameters between the Subject and Predicate devices.
| Parameter compared | Predicate Device<br>Empty EVA Bag<br>(K181393) | Subject Device<br>eZSURETM<br>Empty Fluid Container<br>(EFC)<br>(K223674) | Comment/ Discussion |
|--------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use<br>statement | "The Empty EVA Bag is an empty<br>container used for administration of<br>intravenous solutions to the patient<br>using an intravascular administration<br>set. Medication transfer in and out of<br>the container is done using aseptic<br>technique." | "The Empty Fluid Container is<br>used to hold an admixture of<br>compatible fluids for<br>intravenous administration to<br>a patient. Medication transfer<br>in and out of the container is<br>done using aseptic technique." | Same as Predicate<br>Both devices are intended for and<br>indicated for the use in holding<br>admixture fluids/IV solutions for the<br>subsequent administration/ transfer of<br>the IV solutions from the container to<br>an external IV administration set/<br>line, working in combination,<br>complementarily, applying aseptic<br>technique |
| Product code and<br>regulation number | KPE,<br>21 CFR 880.5025 | KPE,<br>21 CFR 880.5025 | Same as Predicate |
| Intended user<br>population/intended use<br>environment | Adequately trained<br>health care professionals or<br>pharmacists,<br>clinical setting | Adequately trained<br>health care professionals or<br>pharmacists,<br>clinical setting | Same as Predicate |
| IV drug<br>component bag<br>components/ access<br>components | Empty EVA Bag, has 2 access points:<br>1. Additive port/ filling port<br>2. Spiking/ administration port, IV<br>spike bag access for IV set/ IV line | eZSURETM EFC, has 2 access<br>points:<br>1. Additive port/ filling port<br>2. Spiking/administration port,<br>IV spike bag access for IV<br>set/ IV line | Same as Predicate |
| Intended drug type(s) | Parenteral drugs | Parenteral drugs | Same as Predicate |
| Rx/<br>Prescription Use | R only | R only | Same as Predicate |
| Parameter compared | Predicate Device<br>Empty EVA Bag<br>(K181393) | Subject Device<br>eZSURETM<br>Empty Fluid Container<br>(EFC)<br>(K223674) | Comment/Discussion |
| Principles of operation | The empty bags are filled by connecting<br>to containers containing one or more<br>solutions, e.g., through standard<br>syringes or automated filling devices<br>After filling, the bags are clamped to<br>secure the contents prior to<br>administration.<br>IV set/ IV line is attached through the<br>spiking/ administration port to<br>dispense IV therapy using a standard<br>spike and tubing | The empty bags are filled by<br>connecting to containers<br>containing one or more<br>solutions, e.g., through<br>standard syringes (without<br>needle) or automated filling<br>devices<br>The additive port is a self-<br>sealing needle- free female<br>luer lock connector, no<br>different from standard<br>injection site devices of<br>market-cleared IV sets.<br>Silicone valve of the additive<br>port self- seals/ re-seals when<br>mating component devices<br>are disconnected from one<br>another<br>IV set/ IV line is attached<br>through the spiking/<br>administration port to dispense<br>IV therapy using a standard<br>spike and tubing | Different, but Substantially Equivalent<br>to Predicate:<br>Both the Subject device and the<br>Predicate device comprise two ports –<br>the additive port and the spiking/<br>administration port. Both the Subject<br>devices and Predicate devices are<br>bulk-filled by the user through the<br>additive port. Subject devices allow<br>additional-filling through the additive<br>port, whereas the additive/ filling port<br>on the Predicate device is designed to<br>be used to filled once and clamped<br>with inviolable clamp after filling<br>Both the Subject device and Predicate<br>device comprise a spiking/<br>administration port to connect an IV<br>administration set/ line for the<br>delivery of IV therapy to the patient<br>Both the Subject and Predicate devices<br>provide a mechanism for safe filling<br>of the bag and the administration of<br>IV therapy to the patient using an IV<br>administration set/ line. Performance<br>testing of the subject device<br>demonstrates that the differences do<br>not raise new or different questions of<br>safety and effectiveness. |
| Parameter compared | Predicate Device<br>Empty EVA Bag<br>(K181393) | Subject Device<br>eZSURE™<br>Empty Fluid Container<br>(EFC)<br>(K223674) | Comment/ Discussion |
| Technology and design | The EVA bag device includes an<br>administration port and an additive<br>port. The administration port<br>provides a connection to the IV set/<br>line. The additive port is a connector<br>by which medications can be added to<br>the bag/ container using the<br>medication port before IV<br>administration. A capsule closes the<br>medication port after use. The bags/<br>containers are clamped after filling by<br>the means of inviolable clamps<br>An IV set/ IV line is connected to the<br>fluid container/ bag to dispense IV<br>therapy | The EFC device includes an<br>administration port and an<br>additive port. The<br>administration port provides a<br>connection to the IV set/ line.<br>The additive port, a needle-free<br>female luer lock connector,<br>like the injection sites of other<br>market- cleared IV sets, allows<br>for medication(s) to be added<br>to the bag/container prior to<br>and during IV administration.<br>Silicone valve of the additive<br>port self-seals/ re- seals when<br>mating component devices are<br>disconnected from one another<br>An IV set/ IV line is connected<br>to the fluid container/ bag to<br>dispense IV therapy | Different, but Equivalent to Predicate:<br>Though a difference in securing the<br>fluid container content after being<br>filled – silicone valve of the sterile<br>additive port self-seals/ re-seals when<br>mating component devices are<br>disconnected from one another in the<br>case of the Subject device. This<br>contrasts with the Predicate device in<br>which the user clamps the port after<br>filling by means of an inviolable<br>clamp.<br>Performance testing of the subject<br>device demonstrates that the<br>differences do not raise new or<br>different questions of safety and<br>effectiveness. |
| Technology and design | The bag is manufactured by welding a<br>plastic attachment to the bag body and<br>assembling the bottom with entry and<br>exit connectors/ ports | The bag is manufactured by<br>welding a plastic attachment to<br>the bag body and assembling<br>the bottom with entry and exit<br>connectors/ ports | Same as Predicate |
| Parameter compared | Predicate Device<br>Empty EVA Bag<br>(K181393) | Subject Device<br>eZSURE™<br>Empty Fluid Container<br>(EFC)<br>(K223674) | Comment/ Discussion |
| Material<br>of empty fluid<br>container body | Ethylene-vinyl acetate (EVA) | Sealed Air Corp's<br>Medical Packaging Film, Nexcel®<br>Film,<br>Grade M315, Polyolefin | Different, but Substantially Equivalent<br>to Predicate:<br>Though a difference in the material<br>used to build the fluid container main<br>body exists, it did not raise additional<br>questions of safety or of performance<br>effectiveness. This is because both<br>materials are highly transparent clear,<br>flexible plastic film formulated for its<br>intended use as primary medical film<br>for bag body of medical IV bags -<br>free of PVC, plasticizers, adhesives,<br>latex - that could potentially leach<br>into certain solutions, and both are<br>suitable for EO sterilization<br>Performance, shelf life, and<br>biocompatibility testing of the subject<br>device demonstrates that the differences<br>do not raise new or different questions of<br>safety and effectiveness. |
| Materials of additive/<br>filling port | Polyisoprene, Polypropylene<br>Acrylonitrile Butadiene Styrene | Polypropylene Polycarbonate<br>Liquid Silicone Rubber | Different, but Substantially Equivalent<br>to Predicate:<br>Though a difference in the materials<br>used for the additive port exists, it did<br>not raise additional questions of<br>safety or of performance<br>effectiveness.<br>The valve material, Silicone, QP1-30, is<br>an USP (U.S. Pharmacopoeia) class<br>VI elastomer silicone rubber that does<br>not contain phthalates nor latex. |
| Parameter compared | Predicate Device<br>Empty EVA Bag<br>(K181393) | Subject Device<br>eZSURE™<br>Empty Fluid Container<br>(EFC)<br>(K223674) | Comment/ Discussion |
| | | | additives. It is recommended for<br>medical devices including those<br>implanted in humans for less than 30<br>days besides non-implant applications<br>Performance, shelf life, and<br>biocompatibility testing of the subject<br>device demonstrates that the<br>differences do not raise new or<br>different questions of safety and<br>effectiveness. |
| Material of spiking/<br>administration port | PVC not made with DEHP | Thermoplastic elastomer | Different, but Equivalent to Predicate.:<br>Though a difference in the materials<br>used for the spiking/ administration<br>port exists, it did not raise additional<br>questions of safety or of performance<br>effectiveness. This is because<br>eZSURE™ EFC does not use PVC, not<br>even those not made with plasticizer<br>DEHP - that could potentially leach<br>into certain solutions.<br>Performance, shelf life, and<br>biocompatibility testing of the subject<br>device demonstrates that the<br>differences do not raise new or<br>different questions of safety and<br>effectiveness. |
| Biocompatibility | Acceptable biological risks established by<br>demonstrating that<br>the device meets ISO 10993-1 | Acceptable biological risks<br>established by demonstrating<br>that the device meets ISO 10993-1 | Same as Predicate |
| Sterilization method | Ethylene Oxide, EO, SAL 10-6 | Ethylene Oxide, EO, SAL 10-6 | Same as Predicate |
| Parameter compared | Predicate Device<br>Empty EVA Bag<br>(K181393) | Subject Device<br>eZSURETM<br>Empty Fluid Container<br>(EFC)<br>(K223674) | Comment/ Discussion |
| Shelf life | 3 years (36 months) | 5 years | Different, but Equivalent to Predicate:<br>Device performance and shelf-life<br>testing of the subject device<br>demonstrates that the differences do not<br>raise new or different questions of<br>safety and effectiveness. |
| Reuse or single-use | Single-use only | Single-use only | Same as Predicate |
| Primary package top web | Medical grade paper and medical<br>plastic film, heat sealed | Medical grade paper and<br>medical plastic film, heat sealed | Same as Predicate |
{6}------------------------------------------------
# @DIC
{7}------------------------------------------------
# GDIC
{8}------------------------------------------------
# 집 1 C
{9}------------------------------------------------
# @DIC
{10}------------------------------------------------
# GDIC
{11}------------------------------------------------
### Discussion of differences in technological characteristics
The preceding table summarizes the technological comparisons between the eZSURE™ EFC and the Predicate devices identifying differences in materials of construction: a) empty fluid container body, b) additive/ filling port and c) administration port. Although the Subject devices contain differences in material composition compared to the Predicate devices, these differences between the Subject and Predicate devices were evaluated and do not raise new questions of safety and effectiveness. These differences were further addressed and evaluated for safety and effectiveness through non-clinical performance testing, biocompatibility evaluation and sterilization and shelf-life studies, and summarized in VII below. The outcome of the evaluation and testing demonstrates that data within this submission support that the eZSURE™ Empty Fluid Container devices are substantially equivalent to the Submitter designated Predicate devices in terms of safety and performance effectiveness.
#### Performance Data Supporting Substantial Equivalence VII.
- A. The sterile, single-use, non-pyrogenic flexible IV container devices/ bags described in this Summary were tested and demonstrated to be in conformance with the following ISO and FDA recognized standards:
- . ANSI/AAMI CN27:2021, General requirements for Luer activated valves (LAVs) incorporated into medical devices for intravascular applications
- . ANSI/AAMI ST72/ 2019, Bacterial endotoxins – Test methods, routing monitoring, and alternatives to batch testing
- . ASTM F1929-15, Standard test method for detecting seal leaks in porous medical device packaging by dye penetration
- ASTM F1980-16, Standard guide for accelerated aging of sterile barrier systems for medical devices
- . EN 868-5: 2009. Packaging materials and systems for medical devices which are to be sterilized – Part 5: Heat and self-sealable pouches and reels of paper and plastic film construction – Requirements and test methods
- ISO 15747: 2018, Plastic containers for intravenous injections
- ISO 22413: 2021, Transfer sets for pharmaceutical preparations – Requirements and test methods
- . ISO 8536-4: 2019, Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed
- USP 40 <151>, Pyrogen test (USP rabbit test)
- USP-NF <161>, Medical Devices-Bacterial Endotoxin and Pyrogen Tests
- USP-NF <85>, Bacterial Endotoxins Test ●
{12}------------------------------------------------
Testing done:
- . Resistance to Temperature, Pressure and Leakage
- . Resistance to Dropping
- Water Vapor Impermeability
- Spiking Port Penetration Ability
- Spiking Port Adhesion Strength ●
- Additive Port Liquid Tightness ●
- Hanger Tensile Strength
- Identification Clarity ●
- . Particulate Non-contamination
- Impermeability to Microorganism .
- . 7-day Microbial Ingress Test
### B. Biocompatibility
In accordance with ISO 10993-1: 2018, the ToxiSeal™ Vial Adaptor with External Balloon is classified as: Externally Communicating Device, Blood Path Indirect, Prolonged Contact (>24hr to 30d). The biological safety and biocompatibility were evaluated, and found to have met their respective acceptance criteria.
Conclusion to the biological evaluation of the eZSURE™ Empty Fluid Container was that upon completing the evaluation according to ISO 10993-1:2018, the results demonstrated that the eZSURE™ EFC met the biological safety and biocompatibility requirements when used as intended.
Particulate matter testing was conducted in accordance with ISO 15747: 2018, Plastic containers for intravenous injections and USP <788> Particulate Matter in Injections and found to have met the ISO and USP acceptance criteria.
- C. Sterility, Shipping, and Shelf-Life
The Subject device complies with sterilization requirements of ISO 11135: 2014, Sterilization of Health Care Products - Ethylene Oxide - Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices.
- Package Integrity Test ●
- Sterile Barrier Packaging Testing performed on the proposed device: Seal strength ASTM . F88
- . Shelf-life of 3 years has been validated using the FDA recognized standard, ASTM 1980-16, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices.
{13}------------------------------------------------
### VIII. Conclusion
The differences between the predicate and the subject device do not raise any new or different questions of safety or effectiveness. The eZSURE™ EFC is substantially equivalent to the predicate, Empty EVA Bag, with respect to the indications for use, principles of operation and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.