K240171 · Epic Medical Pte. , Ltd. · ONB · Mar 28, 2024 · General Hospital
Device Facts
Record ID
K240171
Device Name
ProSeal™ Injector Plus (Model no. 421050)
Applicant
Epic Medical Pte. , Ltd.
Product Code
ONB · General Hospital
Decision Date
Mar 28, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5440
Device Class
Class 2
Indications for Use
The ProSeal™ Closed System drug Transfer Device (CSTD) mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminations into the drug or fluid path for up to 7 days, when used as intended.
Device Story
ProSeal™ Injector Plus is a sterile, single-use component of the ProSeal™ CSTD system; connects to standard Luer lock syringes for closed-system transfer of hazardous drugs. Device features a stainless steel cannula shielded by a non-latex elastomeric membrane; incorporates a clamp mechanism for system engagement. During use, a double-membrane septum system allows liquid transfer while preventing environmental contaminant entry and drug vapor/aerosol escape. Modifications from predicate include geometric changes to needle hub and housing, transition from UV glue to over-molding for needle bonding, change of housing material to polyoxymethylene, and addition of a protective cap. Used by clinicians/pharmacists to minimize exposure to hazardous drugs and prevent microbial contamination during drug preparation and administration.
Clinical Evidence
Bench testing only. No clinical data. Performance verified against ISO 7864:2016 (needles), ISO 8536-4:2019 (infusion sets), ISO 23908:2011 (sharps protection), and ISO 80369-7:2016 (connectors). Testing included positive/sub-atmospheric pressure leakage, stress cracking, separation resistance, drop/flexural/tensile/compression force, needle bonding strength, and cap removal force. Biocompatibility evaluated per ISO 10993-1:2018; sterility per ISO 11135:2014.
Indicated for use by trained healthcare professionals or pharmacists in clinical settings for the preparation, reconstitution, compounding, and administration of parenteral antineoplastic and hazardous drugs.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
March 28, 2024
Epic Medical Pte. Ltd. Freddie Lee CEO 105 Cecil , 20-01 The Octagon Singapore, Singapore 069534 Singapore
Re: K240171
Trade/Device Name: ProSeal™ Injector Plus (Model no. 421050) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB Dated: March 1, 2024 Received: March 1, 2024
Dear Freddie Lee:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
David Walloschek
David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K240171
Device Name ProSeal™ Injector Plus (Model no. 421050)
### Indications for Use (Describe)
The ProSeal™ Closed System drug Transfer Device (CSTD) mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminants into the drug or fluid path for up to 168 hours or 7 days, when used as intended.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## K240171 – 510(k) Summary
### I. Submitter
Epic Medical Pte. Ltd. 105 Cecil Street #20-04, The Octagon, Singapore 069534. Phone: +65 9635 2618 / +66 81 761 5292
Contact Person: Mr. Freddie LEE, Chief Executive Officer/ Managing Director Date Prepared: March 15, 2024 Content and Format: Prepared in accordance with 21 CFR 807.92 Type of Submission: Special
### II. Subject Device
| 510(k) Number: | K240171 |
|---------------------|--------------------------------------------------------------------------------|
| Trade/ Device Name: | ProSeal™ Injector Plus (Model no. 421050) |
| Common/ Usual Name: | Closed Antineoplastic And Hazardous Drug Reconstitution And<br>Transfer System |
| Regulation Number: | 21 CFR 880.5440 |
| Regulation Name: | Intravascular administration set |
| Regulatory Class: | Class: II |
| Product Code: | ONB |
{5}------------------------------------------------
### III. Predicate
| 510(k) Number: | K192075 |
|---------------------|----------------------------------------------------------------------------------|
| Trade/ Device Name: | ProSeal™ CSTD, Injector,<br>(model 421010) (submitted by Epic Medical Pte. Ltd.) |
| Common/ Usual Name: | Closed Antineoplastic And Hazardous Drug Reconstitution And<br>Transfer System |
| Regulation Number: | 21 CFR 880.5440 |
| Regulation Name: | Intravascular administration set |
| Regulatory Class: | Class: II |
| Product Code: | ONB |
### IV. Device Description
The purpose of this Special 510(k) Submission is to seek FDA clearance prior to the introduction in the US, of a modified design of a ProSeal™ CSTD component device - the Subject ProSeal™ Injector Plus (adapter to the Luer Lock tip syringe). This device is an addition to the ProSeal™ CSTD system of devices. The existing system component device, and also the nominated Predicate device, is the ProSeal™ Injector.
The ProSeal™ CSTD (K192075) is a sterile, single-use CSTD for the preparation, reconstitution, compounding, and administration of antineoplastic and hazardous drugs, intended for use in clinical settings by trained health care professionals and/or pharmacists.
The ProSeal™ Injector Plus is a component of the ProSeal™ CSTD system intended for connection to a standard Luer lock syringe. It is one variant of the cleared ProSeal™ Injector, a component device of the ProSeal™ CSTD system. Just like the ProSeal™ Injector, a nonlatex elastomeric membrane covers the liquid transfer cannula tip of the ProSeal™ Injector Plus. It incorporates a clamp mechanism that reversibly connects to the other component devices of the ProSeal™ CSTD system. The ProSeal™ Injector Plus, like the ProSeal™ Injector includes a stainless steel cannula that is always shielded: during preparation, use and disposal. The ProSeal™ Injector Plus is supplied with a protective cap.
{6}------------------------------------------------
The closed transfer of liquid that takes place with the use of the ProSeal™ CSTD system (K192075) as follows:
- A double membrane septum utilizing self-sealing elastomeric membranes tightly fits together . when the system components engage. A cannula within the ProSeal™ Injector Plus housing perforates the double membrane for the transfer of liquid. When the cannula is retracted, the membranes seal off the transfer of environmental contaminants into the system and/or escape of drug or vapor concentrations outside the system, thereby minimizing the individual and environmental exposure to drug vapor, aerosols, and also minimizing the risk of microbial contamination.
The proposed modifications are:
- Geometric change of needle hub design
- Geometric change of housing design ●
- Needle bonding method, from UV glue to over-molding ●
- Change of housing material from polycarbonate to polyoxymethylene
- Addition of protective cap to Injector Plus
### V. Indications for Use Statement
The ProSeal™ Closed System drug Transfer Device (CSTD) mechanically prohibits environmental contaminants from entering the system and the escape of drug or vapor concentrations from the system, thereby minimizing individual and environmental exposure to drug vapor, aerosols, and spills. The ProSeal™ system also prevents the introduction of microbial contaminations into the drug or fluid path for up to 7 days, when used as intended.
{7}------------------------------------------------
# קמונו
## VI. Comparison of Technological Characteristics
The ProSeal™ Injector Plus, a system component of the ProSeal™ CS7D and the predicate system component of interest, the ProSeal™ Injector, share the following key characteristics:
- same closed system technology
- same interface between system components ●
- same Luer lock connection to fit to syringe .
- . same sharps protection features
- same sterile barrier packaging .
- same sterilization method .
- | same shelf-life
An overview table summarizing the comparison between of key characteristics between the Subject ProSeal™ Injector Plus and the Predicate ProSeal™ CSTD - Injector is provided hereunder.
| Characteristic<br>compared | Predicate Device<br>ProSeal™ CSTD (K192075)<br>ProSeal™ Injector (model 421010) | Subject Device<br>ProSeal™ CSTD (K240171)<br>ProSeal™ Injector Plus (model 421050) | Comment/ Discussion |
|------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|
| Indications for<br>Use statement | "The ProSeal™ Closed System drug Transfer<br>Device (CSTD) mechanically prohibits<br>environmental contaminants from entering<br>the system and the escape of drug or vapor<br>concentrations from the system, thereby<br>minimizing individual and environmental<br>exposure to drug vapor, aerosols, and spills.<br>The ProSeal™ system also prevents the<br>introduction of microbial contaminations into<br>the drug or fluid path for up to 7 days, when<br>used as intended." | "The ProSeal™ Closed System drug Transfer<br>Device (CSTD) mechanically prohibits<br>environmental contaminants from entering<br>the system and the escape of drug or vapor<br>concentrations from the system, thereby<br>minimizing individual and environmental<br>exposure to drug vapor, aerosols, and spills.<br>The ProSeal™ system also prevents the<br>introduction of microbial contaminations into<br>the drug or fluid path for up to 7 days, when<br>used as intended." | Same |
| Product code and<br>regulation<br>number | ONB<br>21 CFR 880.5440 | ONB<br>21 CFR 880.5440 | Same |
| Characteristic<br>compared | Predicate Device<br>ProSeal™ CSTD (K192075)<br>ProSeal™ Injector (model 421010) | Subject Device<br>ProSeal™ CSTD (K240171)<br>ProSeal™ Injector Plus (model 421050) | Comment/ Discussion |
| Intended user<br>population/<br>intended use<br>environment | Adequately trained<br>health care professionals or pharmacists/<br>clinical setting | Adequately trained<br>health care professionals or pharmacists/<br>clinical setting | Same |
| Intended<br>drug type | Parenteral drugs | Parenteral drugs | Same |
| R/<br>Prescription use | R only | R only | Same |
| Closed system<br>technology | Physical barrier to prevent all mass from<br>crossing the system boundary | Physical barrier to prevent all mass from<br>crossing the system boundary | Same |
| Interface between<br>system<br>components | Elastomeric<br>resealing double membrane | Elastomeric<br>resealing double membrane | Same |
| Fitting to external<br>Luer Lock<br>syringe | Female Luer Lock<br>conforms to ISO 80369-7:2016 | Female Luer Lock<br>conforms to ISO 80369-7:2016 | Same |
| System<br>component<br>devices | ProSeal Injector (model 421010)<br>ProSeal Connector (422010)<br>ProSeal Vial Adaptor (models 420020, 420030,<br>420040 for different neck sizes and pressure<br>equalization)<br>ProSeal Assembly Fixture (Fixture for ProSeal<br>Vial Adaptor assembly, model 424010) | ProSeal Injector Plus (model 421050)<br>ProSeal Connector (422010)<br>ProSeal Vial Adaptor (models 420020, 420030,<br>420040 for different neck sizes and pressure<br>equalization)<br>ProSeal Assembly Fixture (Fixture for ProSeal<br>Vial Adaptor assembly, model 424010) | Different<br>difference discussed in<br>Comments #1 and #2 |
| Materials of<br>construction in<br>the fluid path of<br>Injector | Polypropylene (PP),<br>stainless steel (SUS 304),<br>& thermoplastic elastomer (TPE) | Polypropylene (PP),<br>stainless steel (SUS 304),<br>& thermoplastic elastomer (TPE) | Same |
| Characteristic<br>compared | Predicate Device<br>ProSeal™ CSTD (K192075)<br>ProSeal™ Injector (model 421010) | Subject Device<br>ProSeal™ CSTD (K240171)<br>ProSeal™ Injector Plus (model 421050) | Comment/ Discussion |
| Cannula-to-<br>hub assembly | UV adhesive for medical assembly | Hub over-molding<br>onto cannula without adhesive | Different,<br>See Comment #1 |
| Sharp<br>protection feature | The sharps are inaccessible to users by design<br>and had been verified according to ISO<br>23908: 2011 | The sharps are inaccessible to users by design<br>and had been verified according to ISO<br>23908: 2011 | Same |
| Extended FLL<br>shield collar<br>feature and<br>protective cap | Image: without extended FLL shield collar nor protective cap | Image: incorporated with extended FLL shield collar and provided with removable protective cap | Different,<br>See Comment #2 |
| Primary package<br>top web | Medical grade paper and medical plastic film,<br>heat sealed | Medical grade paper and medical plastic film,<br>heat sealed | Same |
| Sterilization<br>method | Ethylene Oxide, EO, SAL 10-6 | Ethylene Oxide, EO, SAL 10-6 | Same |
| Validated<br>Shelf life | 3 years (36 months) | 3 years (36 months) | Same |
| Reuse<br>or single-use | Single-use only | Single-use only | Same |
| Labeling<br>specifications | Meets the requirements<br>specified in 21 CFR 801 | Meets the requirements<br>specified in 21 CFR 801 | Same |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for CDIC. The logo is in black and white. The letters are stylized and connected.
## Discussion of differences in technological characteristics
## Comment #1
The difference in how the cannula is attached to the hub during the production process has been evaluated through needle pull strength testing specified in ISO 7864: 2016, the same test method and acceptance criteria as ProSeal™ CSTD (K192075), and found to have also met the functional performance acceptance criteria. Details are summarized in section VII.A.
## Comment #2
The differences in design - the addition of a shield collar, and a protective cap - have been evaluated through ISO 80369-7: 2021 testing and ISO 8536-4: 2019 testing respectively. The Luer lock connection and the protective cap were found to have met the functional performance acceptance criteria therein. Further details are summarized in section VII.A.
### VII. Performance Data Supporting Substantial Equivalence
### A. Functional Performance
The ProSeal™ Injector Plus described in this Summary, with modifications listed in section IV, were tested and demonstrated to be in conformance with the following ISO and FDA recognized standards:
- ISO 7864: 2016, Sterile hypodermic needles for single use Requirements and test methods. ●
- ISO 8536-4: 2019, Infusion equipment for medical use Part 4: Infusion sets for single use, . gravity feed
- ISO 23908: 2011, Sharps injury protection Requirements and test methods Sharps ● protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling
- ISO 80369-7: 2016, Small-bore connectors for liquids and gases in healthcare application -● Part 7, Connectors for intravascular or hypodermic applications
Bench performance verifications and validations conducted:
- Positive pressure fluid leakage test
- Sub-atmospheric pressure air leakage test ●
- Stress cracking test ●
- Resistance to separation from axial load test .
- . Resistance to separation from unscrewing test
- . Resistance to overriding
- Drop test
- . Flexural force test
- Tensile force test
- . Compression force test
- Testing access to the sharp in safe mode
- Device leakage integrity test
- Needle bonding strength ●
- Sharps injury protection test ●
- Cap removal force test .
{11}------------------------------------------------
- B. Biocompatibility
In accordance with ISO 10993-1: 2018, the ProSeal™ Injector Plus, just like the existing device, is classified as: Externally Communicating Device, Blood Path Indirect, Prolonged Contact (>24hr to 30d). The following testing are referenced from testing on the existing devices:
- . Cytotoxicity
- Sensitization ●
- . Intracutaneous Reactivity
- . Acute Systemic Toxicity
- 14-day Subacute/ Subchronic Acute Systemic Toxicity .
- In-vitro Hemolysis Assessment
- Material Mediated Pyrogenicity ●
- Chemical Characterization as per ISO 10993 Part-18: 2020(E)/AMD-1: 2022 and ISO 10993-● 12: 2021 (E).
Particulate matter testing was conducted in accordance with ISO 8536-4: 2019, Infusion equipment for medical use - Part 4: Infusion sets for single use, gravity feed, Annex A.2 and USP <788> Particulate Matter in Injections.
### C. Sterility, Shipping, and Shelf-Life
The Subject device, like the existing device, complies with sterilization requirements of ISO 11135: 2014, Sterilization of Health Care Products - Ethylene Oxide - Part 1: Requirements for Development, Validation and Routine Control of a Sterilization Process for Medical Devices and the following testing/ evaluations:
- AAMI TIR 28:2016, Product Adoption and process equivalence for ethylene oxide sterilization ●
- USP <85>, Bacterial endotoxin test ●
- ASTM F1980-16, Standard guide for accelerated aging of sterile barrier systems for medical ● devices
- . ASTM F88/ F88M-21, Sterile barrier packaging Testing performed on the existing device: Seal strength
- ISO 10993-7:2008/ Amd 1:2019, Biological evaluation of medical devices -- Part 7: Ethylene oxide sterilization residuals; Amendment 1: Applicability of allowable limits for neonates and infants
### VIII. Conclusion
The differences between the Predicate and the Subject device do not raise any new or different questions of safety or effectiveness. The ProSeal™ CSTD - Injector Plus is substantially equivalent to the Predicate, the ProSeal™ CSTD - Injector, with respect to the indications for use, principles of operation and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.