K221381 · Keystone Dental, Inc. · NHA · Aug 10, 2022 · Dental
Device Facts
Record ID
K221381
Device Name
KDG Abutments
Applicant
Keystone Dental, Inc.
Product Code
NHA · Dental
Decision Date
Aug 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
KDG Abutments are pre-manufactured prosthetic components for direct connection to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation. The KDG-Osteon Precision Milled Suprastructure is indicated for attachment to KDG Abutments in the treatment of partially or fully edentulous jaws for the purpose of restoring chewing function. The KDG-Osteon Precision Milled Suprastructure is intended for attachment to a minimum of two (2) abutments.
Device Story
KDG Abutments are pre-manufactured prosthetic components (abutments, screws, and bar suprastructures) connecting to endosseous dental implants. The system includes straight and angled (17°, 30°) multi-unit abutments and straight single-unit abutments. The KDG-Osteon Precision Milled Suprastructure is a patient-specific overdenture bar fabricated via CAD/CAM technology from scans of patient impressions. Used in dental clinics by dentists/prosthodontists, the device supports fixed or removable prostheses. The suprastructure is screw-retained to at least two abutments. By providing a stable interface between implants and prosthetic teeth, the device restores chewing function and aesthetics for edentulous patients.
Clinical Evidence
Bench testing only. Fatigue testing per ISO 14801:2016 performed on worst-case constructs. Biocompatibility evaluated per ISO 10993. Sterilization validated per ISO 11137-2 and ISO 17665-1. MRI compatibility assessed via scientific rationale and published literature.
Technological Characteristics
Materials: Ti-6Al-4V ELI titanium alloy (Grade 23) per ASTM F136. Design: CAD/CAM milled suprastructures; multi-unit and single-unit abutments. Connectivity: Mechanical connection to endosseous implants. Sterilization: Gamma irradiation (sterile components) or steam (non-sterile components).
Indications for Use
Indicated for partially or fully edentulous adult patients requiring prosthetic rehabilitation to restore chewing function via endosseous dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Keystone Dental Inc. % Chris Brown Manager Aclivi, LLC 3250 Brackley Drive Ann Arbor, Michigan 48105
Re: K221381
Trade/Device Name: KDG Abutments Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: May 11, 2022 Received: May 13, 2022
Dear Chris Brown:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K221381
Device Name KDG Abutments
Indications for Use (Describe)
KDG Abutments are pre-manufactured prosthetic components for direct connection to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation.
The KDG-Osteon Precision Milled Suprastructure is indicated for attachment to KDG Abutments in the treatment of partially or fully edentulous jaws for the purpose of restoring chewing function. The KDG-Osteon Precision Milled Suprastructure is intended for attachment to a minimum of two (2) abutments.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K221381 Keystone Dental Inc KDG Abutments
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Keystone Dental Inc.<br>154 Middlesex Turnpike<br>Burlington, MA 01803 |
|-------------------|-----------------------------------------------------------------------------------|
| Telephone: | +1 (781) 328-3490 |
| Fax: | +1 (781) 328-3400 |
| Official Contact | Nancy DeAngelo, Regulatory Affairs Manager<br>Email: NDeAngelo@keystonedental.com |
| Date submitted: | 8/3/2022 |
### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name: | KDG Abutments |
|---------------------------|----------------------------------------------------------------------------------------------|
| Classification Name: | Abutment, Implant, Dental, Endosseous |
| Classification Regulation | 21 CFR 872.3630 |
| Device Class: | Class II |
| Product Code: | NHA |
| Review Panel: | Dental |
| Reviewing Branch: | Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices (OHT1)<br>Dental Devices (DHT1B) |
### PREDICATE DEVICE INFORMATION
The Subject device abutments are substantially equivalent in Indications for Use, and technological/design principles to the following legally marketed Predicate device with Reference devices leveraged for technology, implant connection and design principles. The Subject device KDG-Osteon Precision Milled Suprastructure is substantially equivalent to the K212676 Reference device.
| 510(k) | Predicate Device Name | Company Name |
|---------|--------------------------|------------------|
| K161416 | Multi-unit Abutment Plus | Nobel Biocare AB |
| 510(k) | Reference Predicate Device Name | Company Name |
|----------|----------------------------------------------------------|--------------------------------------------|
| K210117 | Paltop Narrow Implant | Paltop Advanced Dental Solutions |
| K051614 | Lifecore Primaconnex™ Internal Connection Implant System | Lifecore Biomedical, Inc. / Keystone |
| K101545 | Genesis Implant System | Keystone Dental, Inc. |
| K112795 | Paltop Advanced Dental Solution System | Paltop Advanced Dental Solutions |
| K212676 | Osteon Precision Milled Suprastructure | Implant Solutions PTY LTD (Osteon Medical) |
| K220200* | Paltop Conical Implant System | Paltop Advanced Dental Solutions |
*(compatible abutment accessories)
{4}------------------------------------------------
# DEVICE DESCRIPTION
Endosseous dental implants are surgically implanted into a patient's mouth to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. Endosseous dental implant are secured to dental implants with a retaining screw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's chewing function. Prosthetic devices used with the dental implant abutments in this submission may be screw-retained or cemented.
The purpose of this submission is the marketing clearance for KDG Abutments which comprises endosseous rootform dental abutments, abutment screws, and other associated components for single-unit, and overdenture restorations. The Subject device abutment components introduce 30° post correction angle multi-unit abutments compatible with the sponsor's previously cleared implants.
The Keystone family of abutments are compatible with the Keystone implants which have a hex-lobe internal connection. The Paltop family of abutments are compatible with the Paltop implants which have hex-wall internal connection.
| | | | Abutment Design | | | Bar |
|-------------------------------------|-----------------------------|------------------------------|---------------------|-------------------|----------------------|--------------------|
| Implant Family | Implant<br>Diameter<br>(mm) | Platform<br>Diameter<br>(mm) | Straight Multi-Unit | Angled Multi-Unit | Straight Single-Unit | Bar Suprastructure |
| Keystone<br>PrimaConnex | 3.3, 3.5 | 3.3 (SD) | X | X | n/a | X |
| PrimaConnex | 4.0, 4.1 | 3.75 (RD) | X | X | n/a | X |
| | 5.0 | 4.5 (WD) | X | X | n/a | X |
| Keystone<br>Genesis | 3.8 | 3.6 (SD) | X | X | n/a | X |
| | 4.5 | 3.9 (RD) | X | X | n/a | X |
| | 5.5, 6.5 | 4.9 (WD) | X | X | n/a | X |
| Paltop<br>Advanced Dental<br>System | 3.75 | | X | X | X | X |
| | 4.2 | 3.70 (SP) | X | X | X | X |
| | 5.0 | | X | X | X | X |
| | 6.0 | 4.5 (WP) | X | X | X | X |
| | | Interface | Non-Indexed | Indexed | Non-Indexed | Non-Indexed |
| | | Material | Grade 23 Titanium | Grade 23 Titanium | Grade 23 Titanium | Grade 23 Titanium |
# KDG Abutments - Abutment Designs
The KDG-Osteon Precision Milled Suprastructure is an overdenture bar which is compatible with the Subject device abutments. The overdenture bar is dental restorative device that is intended for screw-retained attachment to dental abutments to aid in the treatment of partial and totally edentulous patients for the purpose of restoring their chewing function. The KDG-Osteon Precision Milled Suprastructure is fabricated by means of CAD/CAM technology and is used to facilitate the attachment of both fixed and removable prostheses.
The Osteon Precision Milled Suprastructure is designed for an individual patient from scans of the patient's impression. The suprastructure is manufactured in biocompatible Titanium alloy with the aid of Computer Aided Design (CAD) and Computer Aided Manufacturing (CAM) technology. All CAD/CAM fabrication is performed by Osteon Medical, within our premises. The abutment-born Suprastructure is only indicated for straight placement and is not to exceed the maximum angulation of the connected multi-unit abutments.
KDG-Osteon Precision Milled Suprastructures facilitate the attachment of both removable and fixed dental prosthesis and hence categorized as type B. The design specifications are listed in the table below.
Type A: Intended to act as a supporting structure to facilitate the attachment of removable dental prosthesis and include: Primary Bar and Nexus Removable Bar.
{5}------------------------------------------------
Type B: Intended to act as a supporting structure to facilitate the attachment of fixed dental prosthesis and include: Melbourne Bar and Nexus Fixed Bar.
| Description | Type A<br>(facilitates Removable Prosthesis) | | Type B<br>(facilitates Fixed Prosthesis) | |
|--------------------------------------------|----------------------------------------------|---------|------------------------------------------|---------|
| | Minimum | Maximum | Minimum | Maximum |
| Total Cylinders | 2 | 10 | 2 | 10 |
| Suprastructure Span Between Cylinders (mm) | 1 | 30 | 1 | 30 |
| Suprastructure Height (mm) | 3 | 12 | 3 | 22 |
| Suprastructure Width (mm) | 3.4 | 12 | 3.4 | 12 |
| Distal Cantilever Section (mm) | 0 | 15 | 0 | 15 |
| Cylinder Height (mm) | 0 | 4.6 | 0 | 4.6 |
| Cylinder Diameter (mm) | 4.5 | 8 | 4.5 | 8 |
The table below presents the design specifications for two categories of suprastructures:
The Subject device abutments, abutment screws, and bar suprastructure are fabricated from Ti-6Al-4V ELI titanium alloy (Grade 23) which conforms to ASTM F136, Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications (UNS R56401).
All Subject device components are one-time use device components are provided sterile and sterilized by gamma irradiation except for the Single-Unit and Multi-Unit copings, the KDG-Osteon Precision Milled Suprastructure and all replacement screws which are provided non-sterile. Devices provided as non-sterile are sterilized by steam.
# INDICATIONS FOR USE
KDG Abutments are pre-manufactured prosthetic components for direct connection to endosseous dental implants and are intended for use as an aid in prosthetic rehabilitation.
The KDG-Osteon Precision Milled Suprastructure is indicated for attachments in the treatment of partially or fully edentulous jaws for the purpose of restoring function. The KDG-Osteon Precision Milled Suprastructure is intended for attachment to a minimum of two (2) abutments.
# EQUIVALENCE TO MARKETED DEVICE
The Subject device abutments are highly similar to the Predicate device with respect to Indications for Use and technological principles. The Subject device suprastructure is highly similar to the K212676 Reference device with respect to Indications for Use and technological principles. The Comparison tables below compare the Indications for Use and Technological Characteristics of the Subject and Predicate/Reference devices.
# Comparison of Indications for Use Statements
The Subject and Predicate devices have highly similar Indications for Use Statements (IFUS) with respect to the abutment components, differing only in device name and grammar. The Subject and the sponsor's K212676 Reference device have similar wording the use of the bar suprastructure, differing only in the list of compatible implant/abutment systems. The IFUSs express equivalent intended use to facilitate dental prosthetic rehabilitation, and the indications are expressed equivalently using different specific wording.
Similarly, the differences between the Subject device IFUS and that of each Reference device are related to the specific device names and specific use of implants which are part of the Reference devices. None of these minor differences impact substantial equivalence with the Predicate device because all IFUS express equivalent intended use to facilitate dental prosthetic rehabilitations are expressed equivalently using different specific wording.
{6}------------------------------------------------
# Comparison of Technological Characteristics
Overall, the Subject device abutments are highly similar to the Predicate and Reference device abutments. Abutment designs are the same in principle. Critical abutment dimensions, such as the Gingival Height, Prosthetic Diameter, Post Correction Angle, and Post Heights are highly similar between the Subject, Predicate and Reference device abutment designs. Subject device abutments and screws are fabricated from the same materials as the Predicate device.
Sterilization and packaging of the sterile Subject device components are the Sponsor's K210117, K101545 and K201334 Reference devices. Cleaning and sterile Subject device abutment components are the same as and leveraged from the K210117 Reference device. Packaging and sterilization of the KDG-Osteon Precision Milled Suprastructure is the same as and leveraged from the K212676 Reference device.
# Multi-Unit Abutments
The Subject device Straight Multi-Unit Abutments are the highly similar to the Predicate device Straight Multi-Unit Abutments, differing in the implant connection. Subject device Straight Multi-Unit Abutment implant connections, Gingival Heights and Prosthetic Diameters are supported by the sponsor's Reference devices.
The Subject device 17° Angulated Multi-Unit Abutments are highly similar to the Predicate device 17° Angulated Multi-Unit Abutments differing in the implant connection. Subject device 17° Angulated Multi-Unit Abutment implant connections, Gingival Heights and Prosthetic Diameters are supported by the sponsor's Reference devices.
The Subject device 30° Angulated Multi-Unit Abutments are highly similar to the Predicate device 30° Angulated Multi-Unit Abutments differing in the implant connection. Subject device 30° Angulated Multi-Unit Abutment implant connections, Gingival Heights and Prosthetic Diameters are supported by the sponsor's Reference devices.
Multi-Unit abutment accessories which include interface copings, temporary cylinders, and healing caps as part of two-part abutments are supported by the sponsor's K210117 Reference device. The use of Multi-Unit Abutment accessories including limitations for hand modification of cylinders is the same as the K210117 Reference device.
# Straight Single-Unit Abutments
The Subject device Paltop family Straight Single-Unit Abutments are highly similar to the sponsor's K210117 Reference device Straight Single-Unit Abutments, adding additional implant connections. Single-Unit abutment accessories which include interface copings, temporary cylinders, and healing caps as part of two-part abutments and are supported by the sponsor's K210117 Reference device. The use of Single-Unit Abutment accessories including limitations for hand modification of cylinders is the same as the K210117 Reference device.
Overall, minor differences in the Subject device abutment designs, correction angles and implant connection do not affect substantial equivalence. The Gingival Height dimensions of the Subject device abutment designs (1.0-5.0 mm) are highly similar to Predicate and Reference devices (1.0-5.0 mm). The Prosthetic diameters of the Subject device abutment components (4.8-5.0 mm) are highly similar and encompassed by the sponsor's Reference devices (4.5-5.0 mm).
Any differences in implant abutment designs or dimensions have been mitigated and demonstrated to be suitable for intended use through non-clinical bench performance testing.
### Bar Suprastructure
The Subject device KDG-Osteon Precision Milled Suprastructure is the highly similar as the K212676 Reference device, differing only in the list of compatible implant/abutment systems. The Subject device is the same in terms of design parameters and requirements, as the K212676 Reference device.
{7}------------------------------------------------
Compatibility of the K212676 Reference device was previously demonstrated for the K051614, K101545 and K112795 Reference devices. Compatibility with 30° abutments is supported by the K212676 Reference device and through the non-clinical performance testing of the Subject device.
# NON-CLINICAL PERFORMANCE TEST DATA
Fatigue testing was performed for each family of abutments according to the requirements of ISO 14801:2016, Dentistry – Implants – Dynamic loading test for Endosseous Dental Implants. The worst-case scenario for each family was chosen based on the FDA Guidance, Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments. Fatigue testing of the sponsor's K212676 Reference device is leveraged for the KDG-Osteon Precision Milled Suprastructure.
The Subject device components have the same nature of body contact duration, material formulation and sterilization methods compared to the sponsor's Reference device.
Test results and Biological Evaluation per ISO 10993 performed for the sponsor's K210117 Reference device are leveraged to demonstrate suitable biocompatibility of the Subject device abutments.
Test results and Biological Evaluation per ISO 10993 performed for the sponsor's K212676 Reference device are leveraged to demonstrate suitable biocompatibility of the Subject device KDG-Osteon Precision Milled Suprastructure.
Test results and Sterilization Validations per ISO 11137-2 and ISO/TS 13004 performed for the sponsor's K210117 Reference device are leveraged to demonstrate suitable sterilization of the Subject device sterile components.
Cleaning validation and sterilization per ISO 17665-1, for non-sterile and sterile abutment components which may be modified and require subsequent sterilization is leveraged from the sponsor's K210117 Reference device.
Sterilization validation per ISO 17665-1 and ISO/TS 17665-2 for the KDG-Osteon Precision Milled Suprastructure is leveraged from the sponsor's K212676 Reference device.
Sterile device sterile barrier shelf-life packaging integrity is leveraged from the sponsor's K101545 and K210117 Reference devices.
Non-clinical worst-case MRI review was performed to evaluate the Subject device including all compatible implant fixtures, abutments, bars, and fixation screws and their respective material composition in the MRI environment using scientific rationale and published literature (e.g., Woods, Terry O., Jana G. Delfino, and Sunder Rajan, "Assessment of Maqnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795). Rationale addressed parameters per the FDA guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment," including magnetically induced displacement force and torque.
The results of the non-clinical testing demonstrate conformance with testing requirements and support a finding of substantial equivalence with respect to the Subject and Predicate device.
# CONCLUSION
Overall, the Indications for Use statements for the Subject and Predicate devices are highly similar.
Overall, the Technological Characteristics, mode of operation and materials of the Subject device are the same or highly similar to that of the Predicate device. Slight differences in abutment design and compatible implant connection do not affect the intended use of the device and are mitigated or supported through non-clinical
{8}------------------------------------------------
performance testing results. ISO 14801 mechanical performance testing performed on worst-case constructs of the Subject device to demonstrate suitability for intended use of the Subject device implant platform, gingival height, and post correction angles combinations.
Overall, the Subject and Predicate devices have been demonstrated to be Substantially Equivalent.
The basis for the belief that the Subject device is substantially equivalent to the Predicate device and incorporates technology of the Reference devices is summarized in the following Indication for Use and Technological Characteristics comparison tables.
{9}------------------------------------------------
# Comparison Table - Indications for Use Statements
| Device | Indications for Use Statement | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device<br>KDG Abutments<br>Keystone Dental, Inc. | KDG Abutments are pre-manufactured prosthetic components for direct connection to endosseous dental implants and are<br>intended for use as an aid in prosthetic rehabilitation. | |
| | The KDG-Osteon Precision Milled Suprastructure is indicated for attachment to KDG Abutments Multi-Unit abutments in the<br>treatment of partially or fully edentulous jaws for the purpose of restoring chewing function. The KDG-Osteon Precision<br>Milled Suprastructure is intended for attachment to a minimum of two (2) abutments. KDG Abutments Multi-Unit<br>abutments have a prosthetic diameter of 4.8 mm or 5.0 mm and may have up to 30° abutment post correction angle. | |
| Predicate Device<br>Multi-unit Abutment Plus<br>(K161416)<br>Nobel Biocare AB | The Multi-unit Abutment Plus is a pre-manufactured prosthetic component directly connected to the endosseous dental<br>implant and is intended for use as an aid in prosthetic rehabilitation. | |
| Reference Device<br>Paltop Narrow Implant (K210117)<br>Paltop Advanced Dental Solutions | The Paltop Narrow Implant is indicated for use in surgical and restorative applications for placement in the mandibular<br>central, lateral incisor and maxillary lateral incisor regions of partially edentulous jaws where the horizontal space is limited<br>by the adjacent teeth and roots, to provide support for prosthetic devices, such as artificial teeth, in order to restore the<br>patient's chewing function. The Paltop Narrow Implant is indicated also for immediate loading when good primary stability is<br>achieved and with appropriate occlusal loading. | |
| Reference Device<br>Lifecore Primaconnex™ Internal<br>Connection Implant System<br>(K051614)<br>Lifecore Biomedical, Inc. /<br>Keystone Dental, Inc | Lifecore Biomedical Dental Implant System implants are intended for use in partially or fully edentulous mandibles and<br>maxillae, in support of single or multiple-unit restorations including; cement retained, screw retained, or overdenture<br>restorations, and terminal or intermediate abutment support for fixed bridgework.<br><br>The PrimaConnex Internal Connection Implant is a threaded internal connection implant. The PrimaConnex Internal<br>Connection Implant is intended for immediate placement, where immediate implant placement is defined by the<br>International Congress of Oral Implantologists (ICOI) as the placement of an implant at the time of tooth extraction, into the<br>extraction socket.<br><br>The PrimaConnex Internal Connection Implant is intended for immediate provisionalization, non-occlusal load. Immediate<br>Provisionalization is defined by the International Congress of Oral Implantologists (ICOI) as a clinical protocol for the<br>placement of an interim prosthesis with or without occlusal contact with the opposing dentition, at the same clinical visit of<br>implant placement. The PrimaConnex Internal Connection Implant can be restored with a temporary prosthesis in single<br>tooth and multiple tooth applications with good quality bone. | |
| Reference Device<br>Genesis Implant System<br>(K101545)<br>Keystone Dental, Inc. | The Genesis Implant System is intended for use in single-stage or two-stage surgical procedures in all types of bone in<br>partially or fully edentulous mandibles and maxillae. The Genesis Implant System supports single or multiple-unit<br>restorations to re-establish patient chewing function and esthetics. Genesis implants are intended for placement following<br>natural tooth loss or for immediate placement into an extraction socket. Immediate function may be achieved when good<br>primary stability is established and appropriate occlusal loading is applied. | |
| Reference Device<br>Paltop Advanced Dental Solutions<br>System (K112795)<br>Paltop Advanced Dental Solutions | The Paltop Dental Implant System is indicated for use in surgical and restorative applications for placement in the bone of<br>the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, in order to restore the patient's<br>chewing function. The Paltop Dental Implant System is indicated also for immediate loading when good primary stability is<br>achieved and with appropriate occlusal loading. | |
| Reference Device<br>Osteon Precision Milled<br>Suprastructure (K212676)<br>Implant Solutions PTY LTD (trading<br>as Osteon Medical) | • Keystone Multi Unit Abutment, 4.8mm, 0° | The Osteon Precision Milled Suprastructure is indicated for attachment to dental abutments in the treatment of partially or<br>fully edentulous jaws for the purpose of restoring chewing function. The Osteon Precision Milled Suprastructures are<br>intended for attachment to a minimum of two (2) abutments.<br><br>The Osteon Milled Suprastructure is indicated for compatibility with the following abutment systems:<br>• Nobel Biocare Multi Unit Abutment Plus, 4.8mm, max 30°<br>• Nobel Biocare Xeal Abutments, 4.8mm, max 30°<br>• Nobel Biocare Multi Unit Abutment, 4.8mm, max 30°<br>• MIS Multi-unit Abutments, 4.8mm<br>o C1 Conical Connection Implant System, max 30°<br>o V3 Conical Connection Implant System, max 30°<br>o Internal Hex Implant System, max 30°<br>o Conical Connection, max 30°<br>• Southern Compact Conical Abutments, 4.8mm<br>o MAX Implant System, 0°<br>o Provata Implant System, max 30°<br>o Deep Conical (DC) Implants, 0°<br>o Piccolo Implants, 0°<br>o External Hex Implants, max 30°<br>• Astra Tech Implant System® Multi Base Abutment EV, 4.8mm, max 30° |
| • Neodent GM Mini Conical Abutment, 4.8mm, max 30° | | |
| • Implant Direct GPS® Angled Abutment, 5.0mm, max 30° | | |
| • Dentium SuperLine® Abutments, 4.5-5.5mm, max 30° | | |
| • Zimmer Angled Tapered Abutments, 4.5mm, max 30° | | |
| • Paltop Multi Unit Abutment, 5.0mm, max 17° | | |
{10}------------------------------------------------
# Technological Characteristics Comparison Table - Abutments
| Comparison | Subject Device<br>KDG Abutments<br>Keystone Dental, Inc. | Predicate Device<br>Multi-unit Abutment Plus (K161416)<br>Nobel Biocare AB | Reference Device<br>Paltop Narrow Implant (K210117)<br>Paltop Advanced Dental Solutions | Reference Device<br>Lifecore Primaconnex™ Internal Connection Implant<br>System (K051614)<br>Lifecore Biomedical, Inc. / Keystone Dental, Inc | Comparison | Subject Device<br>KDG Abutments<br>Keystone Dental, Inc. | Reference Device<br>Genesis Implant System (K101545)<br>Keystone Dental, Inc. | Reference Device<br>Paltop Advanced Dental Solutions System<br>(K112795)<br>Paltop Advanced Dental Solutions | |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------|
| Product Code | NHA | NHA | DZE, NHA | DZE, NHA | Product Code | NHA | DZE, NHA | DZE, NHA | |
| Regulation | 872.3630 | 872.3630 | 872.3640, 872.3630 | 872.3640, 872.3630 | Regulation | 872.3630 | 872.3640, 872.3630 | 872.3640, 872.3630 | |
| Classification | Class II | Class II | Class II | Class II | Classification | Class II | Class II | Class II | |
| Intended Use | Functional and esthetic rehabilitation of the<br>edentulous maxilla and mandible | Functional and esthetic rehabilitation of the<br>edentulous maxilla and mandible | Functional and esthetic rehabilitation of the<br>edentulous maxilla and mandible | Functional and esthetic rehabilitation of the edentulous<br>maxilla and mandible | Intended Use | Functional and esthetic rehabilitation of the<br>edentulous maxilla and mandible | Functional and esthetic rehabilitation of the<br>edentulous maxilla and mandible | Functional and esthetic rehabilitation of the<br>edentulous maxilla and mandible | |
| Reason for Predicate/Reference | n/a | Indications for Use, abutment designs | Abutment designs, material, sterilization,<br>biocompatibility, abutment accessories | Implant/Interface compatibility | Reason for Predicate/Reference | n/a | Implant/Interface compatibility | Implant/Interface compatibility | |
| Sterile components Sterilization | Gamma Sterilization…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.