The Genesis Implant System is intended for use in single-stage or two-stage surgical procedures in all types of bone in partially or fully edentulous mandibles and maxillae. The Genesis Implant System supports single or multiple-unit restorations to re-establish patient chewing function and esthetics. Genesis implants are intended for placement following natural tooth loss or for immediate placement into an extraction socket. Immediate function may be achieved when good primary stability is established and appropriate occlusal loading is applied.
Device Story
Genesis Implant System comprises endosseous dental implants and associated surgical/restorative components. Implants serve as replacement tooth roots providing foundation for restorations. Used in single-stage or two-stage surgical procedures by dental professionals. Implants feature tapered or straight cylindrical designs with internal indexing connections. Surface treatments include BioSpark (hydrophilic, calcium/phosphorous-enriched nano-topography) to promote osseointegration and AnaTite (pink-colored anodization) on collars/abutments for gingival esthetics. System supports immediate function if primary stability is achieved. Clinical benefit includes re-establishment of chewing function and esthetics.
Clinical Evidence
No human clinical data provided. Evidence consists of bench testing (fatigue testing per ISO 14801), in vitro studies (cell metabolic activity, osteoblast proliferation, fibronectin absorption), and in vivo animal studies (sheep model, BIC analysis). BioSpark surface demonstrated hydrophilic properties and enhanced osseointegration markers compared to machined/blasted titanium controls.
Indicated for partially or fully edentulous patients requiring single or multiple-unit dental restorations in the mandible or maxilla. Suitable for placement following natural tooth loss or immediate placement into extraction sockets.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
PrimaConnexTM Internal Connection Implant System (K051614)
{0}------------------------------------------------
## 510(k) Summary
## OCT 2 8 2010
## Administrative Information
| Manufacturer Name: | Keystone Dental, Inc.<br>144 Middlesex Turnpike<br>Burlington, MA 01803<br>Telephone: (781) 328-3300<br>Fax: (781) 328-3400 |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Official Contact: | John A. Capone, RAC |
| Date Prepared: | October 27, 2010 |
| Device Name and Classification | |
| Trade/Proprietary Name: | Genesis Implant System |
| Common Name: | Dental Implants and Abutments |
| Classification Information: | Endosseous Dental Implant<br>(21 CFR 872.3640), Class II<br>Endosseous Dental Implant Abutment<br>(21 CFR 872.3630), Class II |
| Product Codes | DZE, NHA |
### Device Classification Panel
The Classification Panel for these devices is the Dental Products Panel. Premarket Notifications are reviewed by FDA's Dental Devices Branch.
| Device Name | Manufacturer | 510(k) Number(s) |
|--------------------------------------------------|-----------------|------------------|
| PrimaConnexTM Internal Connection Implant System | Keystone Dental | K051614 |
| Straumann® Dental Implant System | Straumann | K053088, K061176 |
| NanoTiteTM Dental Implants | Biomet 3i | K062432 |
| TiUnite® NobelActiveTM Implants | Nobel Biocare | K050705, K071370 |
## Legally Marketed Devices to Which Equivalence is Claimed (Predicate Devices)
## Device Description
The Genesis Implant System includes implants, and associated surgical, restorative and dental laboratory components. Genesis implants are surgically inserted into the upper and/or lower jawbone and serve as a replacement tooth root, which provides a stable foundation for restorations.
Genesis implants are manufactured from Grade 4 titanium, have a tapered or straight cylindrical design with an internal indexing connection, and are available in various platform diameters and lengths (see table below).
{1}------------------------------------------------
| Implant Type | Implant Diameter | Implant Lengths |
|------------------|------------------|------------------------------------|
| Straight Implant | 3.8mm | 8.5, 10, 11.5, 13, 14.5, 16 & 18mm |
| | 4.5mm | 8.5, 10, 11.5, 13, 14.5, 16 & 18mm |
| Tapered Implant | 3.8mm | 8.5, 10, 11.5, 13, 14.5, 16 & 18mm |
| | 4.5mm | 8.5, 10, 11.5, 13, 14.5, 16 & 18mm |
| | 5.5mm | 8.5, 10, 11.5, 13, 14.5 & 16mm |
| | 6.5mm | 8.5, 10, 11.5 & 13mm |
The implants have a macro-, micro- and nano-topography and are treated with the BioSpark process, which results in a hydrophilic surface enriched with calcium and phosphorous ions. The implant collar is micro-roughened and treated with the AnaTite" process, which results in a pink color for enhanced esthetics.
The majority of Genesis abutments are manufactured from Grade 5 titanium and are treated with the AnaTite process. Other Genesis abutments are made from Grade 5 titanium/plastic or gold alloy/plastic. Abutments intended for fixed restorations utilize a Grade 5 titanium screw for attachment to the implant. Genesis abutments and associated restorative components are manufactured in a variety of sizes and configurations to be compatible with the implant platforms (see table below).
| Abutment Type | Abutment Material | Abutment Angulations |
|------------------------------|---------------------------|----------------------|
| Healing Abutment | Grade 5 titanium | N/A |
| Temporary Abutment | Plastic (PMMA) | N/A |
| Ti Temporary Abutment | Grade 5 titanium | N/A |
| Esthetic Contour Ti Abutment | Grade 5 titanium | Straight and 15° |
| UCLA Gold/Plastic Sleeve | Gold Alloy/ Plastic (POM) | N/A |
| SureConnect Abutment | Grade 5 titanium | N/A |
| Locator Abutment | Grade 5 titanium | N/A |
| Snap Abutment | Grade 5 titanium | N/A |
## Indications for Use
The Genesis Implant System is intended for use in single-stage or two-stage surgical procedures in all types of bone in partially or fully edentulous mandibles and maxillae. The Genesis Implant System supports single or multiple-unit restorations to re-establish patient chewing function and esthetics. Genesis implants are intended for placement following natural tooth loss or for immediate placement into an extraction socket. Immediate function may be achieved when good primary stability is established and appropriate occlusal loading is applied.
## Summary of Technological Characteristics
Predicate devices for the Genesis Implant System were selected based on similar intended use and technological characteristics. All devices are intended for replacing missing teeth and supporting single or multiple-unit restorations in the mandible and/or maxilla. Genesis implants are similar in design and materials to the predicates in that all are threaded, rootform implants.
{2}------------------------------------------------
#### 510(k) Premarket Notification
Genesis, SLActive and TiUnite implants are all constructed of biocompatible Grade 4 titanium and are sterilized using gamma radiation. Genesis implants are macro- and microroughened as are PrimaConnex, SLActive and TiUnite implants. Genesis implants are further treated with the BioSpark process, which results in a nano-rough topography. NanoTite implants are similar in this regard, as they also feature a nano-rough surface. In addition to the nano-topography, the BioSpark process produces a hydrophilic surface. These surface characteristics are also found on SLActive implants. The BioSpark surface incorporates calcium and phosphorus ions to create an enriched surface, a characteristic shared by NanoTite and TiUnite implants.
Genesis abutments are made from biocompatible materials used in the predicate PrimaConnex, SLActive, NanoTite and TiUnite abutments, including Grade 5 titanium, gold alloy and plastic. The Genesis system has the same implant/abutment TiLobe" internal connection as the PrimaConnex system. Genesis titanium alloy abutments and implant collars receive the AnaTite surface treatment, which produces a pink-colored surface intended to resemble the appearance of gingival tissue. The color is produced by anodization, a common process for treating titanium, which is also used for components of the Straumann and PrimaConnex implant systems.
### Summary of Non-Clinical Testing
The performance of the Genesis Implant System has been demonstrated through non-clinical testing and published literature. The following is a summary of the relevant in vivo and in vitro studies and bench testing.
Genesis implants with angled abutments were fatigue tested in accordance with ISO 14801 and the FDA's Special Controls Guidance on root-form endosseous dental implants and abutments. Genesis instruments, implant threads, and implantation procedures have been designed to optimize insertion torque in order to limit the pressure applied to bone while ensuring good primary stability. In a bench study using simulated bone material, moderate insertion torque was shown when Genesis implants were placed utilizing appropriate surgical instrumentation.
The AnaTite treatment produces a pink coloration on the surface of the Genesis implant collar and abutments. An in vivo study has shown that pink coloration most closely resembles natural gingival tissue.
The BioSpark treatment produces chemical and physical implant surface characteristics that promote osseointegration. These characteristics have been evaluated by various methods. Contact angle (wettability) testing has shown the Genesis implant surface to be hydrophilic. Energy dispersive spectroscopy and sputter mass spectroscopy have confirmed the presence of calcium and phosphorus ions on BioSpark treated surfaces. Mineralization potential of the BioSpark surface has been reported in published literature. Surface roughness at the macro-, micro-, and nano-level has been demonstrated through scanning electron microscopy. These combined characteristics of the BioSpark surface facilitate faster osseointegration, as demonstrated by significant bone-to-implant-contact (BIC) in the cortical bone of sheep at 4 weeks compared to machined or blasted titanium, and as demonstrated by in vitro studies, which vielded the following individual results:
{3}------------------------------------------------
- Significant increase in cellular metabolic activity over 1 to 7 days in simulated body . fluid on cp-Ti disks treated with BioSpark when compared to machined titanium
- . Significant increase in osteoblast proliferation over 1 to 3 days on cp-Ti disks treated with BioSpark when compared to machined or acid etched titanium
- Early fibronectin absorption and enhanced adhesion, proliferation, and differentiation . of osteoblasts at 24, 48 and 72 hours on cp-Ti disks treated with BioSpark when compared to other anodic spark deposition (ASD) surfaces
Results in animal studies are not necessarily predictive of human clinical results.
## Conclusion
The Genesis Implant System utilizes the same materials and exhibits similar mechanical properties and surface characteristics as the predicate implant systems. The information provided in this 510(k) premarket notification supports the claim of substantial equivalence.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Mr. John A. Capone Director, Regulatory Affairs and Quality Assurance Keystone Dental, Inc. 144 Middlesex Turnpike Burlington, Massachusetts 01803
OCT 2 8 2010
Re: K101545
Trade/Device Name: Genesis Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: October 22, 2010 Received: October 25, 2010
Dear Mr. Capone:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a preemarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Mr. Capone
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act), 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Wh for
Anthony Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K101545
Device Name: Genesis Implant System
#### Indications for Use:
The Genesis Implant System is intended for use in single-stage or two-stage surgical procedures in all types of bone in partially or fully edentulous mandibles and maxillae. The Genesis Implant System supports single or multiple-unit restorations to re-establish patient chewing function and esthetics. Genesis implants are intended for placement following natural tooth loss or for immediate placement into an extraction socket. Immediate function may be achieved when good primary stability is established and appropriate occlusal loading is applied.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Susan Kumer
Concurrence of CDRH, Office of Deflivision Bignom (f(ODE)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Numbe
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.