K161416 · Nobel Biocare AB · NHA · Mar 16, 2017 · Dental
Device Facts
Record ID
K161416
Device Name
Multi-unit Abutment Plus
Applicant
Nobel Biocare AB
Product Code
NHA · Dental
Decision Date
Mar 16, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
The Multi-unit Abutment Plus is a pre-manufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Device Story
Multi-unit Abutment Plus is a transmucosal prosthetic component for multiple unit screw-retained restorations; connects directly to endosseous dental implants with internal conical connections. Available in straight and angled (0°, 17°, 30°) configurations with varying collar heights. Includes a 'Snap Temporary Coping' feature allowing provisional restoration positioning without screws during try-in; final restoration secured with prosthetic screws. Used by dental clinicians in clinical settings. Benefits include facilitated prosthetic rehabilitation and simplified provisional restoration placement. Device is manufactured from titanium vanadium alloy.
Clinical Evidence
Bench testing only. No clinical data provided. Non-clinical evidence includes fatigue testing per modified ISO 14801, sterilization validation (Gamma radiation per ANSI/AAMI/ISO 11137 for abutments; fractionated pre-vacuum/gravity displacement for non-sterile copings), and shelf-life validation via real-time aging.
Technological Characteristics
Materials: Titanium vanadium alloy. Connection: Internal conical connection (NP, RP, WP). Angulations: 0°, 17°, 30°. Collar heights: 1.5, 2.5, 3.5, 4.5 mm. Sterilization: Gamma radiation (abutments); end-user sterilization required for temporary copings. Packaging: Thermoform tray with peel top lid.
Indications for Use
Indicated for patients requiring prosthetic rehabilitation via multiple unit screw-retained restorations connected to endosseous dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 16, 2017
Nobel Biocare AB c/o Mr. Charlemagne Chua Senior Regulatory Affairs Manager Nobel Biocare USA LLC 22715 Savi Ranch Parkway Yorba Linda, CA 92887
Re: K161416
Trade/Device Name: Multi-unit Abutment Plus Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: February 10, 2017 Received: February 13, 2017
Dear Mr. Chua:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| на | PSC Publishing Services (301) 443-6740 | Page 1 of 1 | | Artis For For 1888 (1744 |
|----|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
| | şərarın qurğuna mərkəzi ərazi var. Bu bir və mənisə çərənin mənalı və mənisələr mənalı mən<br>and the may be and sund of the spensive of beteritse as notiermonity on not only not and the sent medium of the | Department of Health and Human Services<br>Paperwork Reduction Act (PRA) Stati<br>Office of Chief Information Officer | this information collection, including suggestions for reducting this burden, to:<br>้าง แต่เวอร์เดอ ย , ว่า ทิตตุลา การเล่น ค.ศ. 2017 ค.ศ. 2017 ค.ศ. 2017 ค.ศ. 2019 ค.ศ. 2019 ค.ศ. 2019 ค.ศ. 2019 ค.ศ. 2019 ค.ศ. 2019 ค.ศ. 2019 ค.ศ. 2019 ค.ศ. 2019 ค.ศ. 2019 ค.ศ<br>A moja a market and source of the on MAO comments on one and<br>sing to to A noticules only the Peninentante Reading Actives sing<br>information unless it significative a currently valid OMB number<br>Food and Drug Administration<br>PRASSISTINGOfdations.gov | |
| | | FOR FDA USE ONLY | Concerters of Centre for Derigoloibes and Radiological (Signing (Signing | |
| | PART PERSELLER | | | |
| | Over-The-Counter Use (21 CFFR 801 Subpart C) | | 2 Presscription Use (Fart 21 CFR 801 Support D | Type of Use (Select one or both, as applicable) |
| | | | | |
| | The Multified button Plus is a pre-ments streng surficedly connected to the endosseous sent | | implant and is intended for use as an aid in proothetic rehabilitation | |
| | | | | Multi-unit Abutunent Plus<br>Indications for Use (Describe,<br>Device Name |
| | | | | 510(k) Number (if known<br>1161416 |
| | ormandes : OMB No. 0910-0120<br>one the parties damusly 31, 2017<br>PRA Statement below | | DEPARTMENT OF FEATH AND HUMA SERVICES<br>ndications for Use<br>Food and Drug Administration | |
| | | | | |
· · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · · ·
11:10:30 %
ຣຸຟິຈິກອີການປັກ
్లాల
{3}------------------------------------------------
I. SUBMITTER
Nobel Biocare AB Vastra Hamngatan 1 Goteborg, SE-411 17 Sweden
Submitted by: Nobel Biocare USA LLC 22715 Savi Ranch Parkwav Yorba Linda, CA 92887
Contact Person: Charlemagne Chua, Senior Regulatory Affairs Manager Phone: (714) 282-4800 x 7830 Fax: (714) 998-9348
Date Prepared: March 15, 2017
II. DEVICE
Name of Device: Multi-unit Abutment Plus Common or Usual Name: Endosseous Dental Implant Abutment Classification Name: Endosseous Dental Implant Abutment (21 CFR 872.3630) Requlatory Class: II Product Code: NHA
III. PREDICATE DEVICE
Primary Predicate: NobelActive Multi Unit Abutment - K072570
Reference Predicate NobelActive Wide Platform (WP) - K133731
IV. DEVICE DESCRIPTION
Multi Unit Abutment Plus
The Multi-unit Abutment Plus are transmucosal abutments used for multiple unit screw retained restorations. The abutment family consists of straight and angled abutments in varying collar heights. The non-angled abutments have a built in screw and the angled abutments are co-packed with a separate abutment screw. The Multi-unit Abutment Plus is compatible with the Nobel Biocare dental implants that have the internal conical connection.
A.4.
{4}------------------------------------------------
The Multi-unit Abutment Plus is available for Internal Conical Connection implants with the Narrow Platform (NP), Regular Platform (RP) and Wide Platforms (WP). The Multi-unit Abutment Plus is available with collar heights of 1.5, 2.5, 3.5, and 4.5 mm and angulations of 0°, 17°, 30°.
### Snap Temporary Coping Multi-Unit Titanium
The Temporary Snap Coping is a premanufactured dental abutment directly connected to the Multi-unit abutment plus. The temporary coping is co-packed with a prosthetic screw. It is used as a temporary aid in prosthetic rehabilitation until the final restoration is attached. Maximum intraoral use is 180-days.
## V. INDICATIONS FOR USE
The Multi-unit Abutment Plus is a pre-manufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
| Technological<br>characteristics | | Subject Device | Predicate |
|----------------------------------|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | Multi-unit Abutment Plus | NobelActive Multi Unit Abutment<br>(K072570) |
| Design Features | Compatible<br>Implant<br>Platform | Nobel Biocare Internal Conical Connection<br>- Narrow Platform (NP)<br>- Regular Platform (RP)<br>- Wide Platform (WP) | Nobel Biocare Internal Conical<br>Connection<br>- Narrow Platform (NP)<br>- Regular Platform (RP) |
| | Device Material | Abutments and screws – Titanium<br>vanadium alloy | Abutments and screws – Titanium<br>vanadium alloy |
| | Abutment collar<br>height | 1.5, 2.5, 3.5, 4.5 mm | 1.5, 2.5, 3.5, 4.5 mm |
| | Abutment width | 4.8 mm | 4.8 mm |
| | Abutment<br>Angulation | 0°, 17°, 30° | 0°, 17°, 30° |
| | Intended use | Multi-unit Abutment Plus in combination<br>with endosseous implants are intended for<br>multiple unit reconstructions when screw<br>retained prosthetics is preferred. | Multi-unit Abutment in combination<br>with endosseous implants are intended<br>for multiple unit reconstructions when<br>screw retained prosthetics is preferred. |
| Indication for Use | | The Multi-unit Abutment Plus is a pre-<br>manufactured prosthetic component<br>directly connected to the endosseous<br>dental implant and is intended for use as<br>an aid in prosthetic rehabilitation. | NobelActive Multi Unit Abutment is a<br>pre-manufactured prosthetic<br>component directly connected to the<br>endosseous dental implant and is<br>intended for use as an aid in prosthetic<br>rehabilitation. |
## VI. Comparison of Technological Characteristics
{5}------------------------------------------------
Analysis of Differences Between Subject Device and Predicate
The Multi-unit Abutment Plus is an evolution of the predicate NobelActive Multi Unit Abutment. It differs from the predicate in the addition of a Wide Platform (WP) abutment (K133731, NobelActive WP, as a reference predicate) and through integration of a snap coping feature. The snap feature, when used with corresponding coping, facilitates try-in by allowing the provisional restoration to be positioned without using screws. Once fitting of the restoration is completed the restoration is secured with prosthetic screws.
## Summary:
The documentation submitted in the premarket notification demonstrates that the Multi Unit Plus is substantially equivalent to the predicate device. Differences in technology were evaluated through comparative performance testing.
# VII. PERFORMANCE DATA
## Summary of Non-Clinical Testing:
Since the subject device does not represent a new worst case, data from the predicate device was leveraged in the following aspects of the 510(k).
- Sterile Device Information
- o The sterilization method for the subject device is the same as the predicate. The sterilization method is Gamma radiation and has been validated in accordance with ANSI/AAMI/ISO 11137. Therefore, no additional testing was required.
- o The Temporary Snap Coping is delivered non-sterile. A sterilization validation was performed to ensure sterility of the subject device when processed by the end user. Sterilization validations were performed using fractionated pre-vacuum and gravity displacement. Sterilization validations were performed with the worst case dental devices.
- Device Packaging -
- The packaging for the subject device is the same as the predicate. O This is a thermoform tray with peel top lid. Therefore, no additional testing was required.
- -Shelf Life
- The packaging for the subject device is the same as the predicate and o is labeled with a 3 year expiration date. Real time aging was used to determine the expiration dating. Therefore, no additional testing was required.
{6}------------------------------------------------
- Biocompatibilitv
- o The subject device is manufactured from the same material as the predicate (K072570, Nobel Biocare), uses the same manufacturing method as the predicate, has the same intended use, and the same patient contact type and duration. Therefore, no additional testing was required.
The fatigue limit of the Multi-unit Abutment Plus was determined using a modified version of ISO 14801. The modifications to ISO 14801 were done to reflect the likely worst-case clinical use of the device. Both the subject and predicate device were tested under identical conditions. The results of the testing were used to address questions related to substantial equivalence based on difference in design between the subiect and predicate devices.
# VIII. CONCLUSIONS
The Multi-unit Abutment Plus was evaluated for substantial equivalence using standard and/or comparative testing. In cases where the Multi-unit Abutment Plus could be shown to not represent a worst-case with respect to the predicates, data from these predicate devices was leveraged to support the subject device. Based on technological characteristics and non-clinical test data included in this submission, the Multi-unit Abutment Plus has been shown to be substantially equivalent to the NobelActive Multi Unit Abutment (K072570).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.