K213071 · Medacta International S.A. · HSX · Nov 16, 2021 · Orthopedic
Device Facts
Record ID
K213071
Device Name
MOTOTM Partial Knee & MOTO PFJ Systems Extension
Applicant
Medacta International S.A.
Product Code
HSX · Orthopedic
Decision Date
Nov 16, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3520
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
MOTO™ Partial Knee System The MOTO™ Partial Knee System is designed for cemented use in partial knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. Partial replacement of the articulating surfaces of the knee is indicated when only one compartment of the joint is affected due to the compartmental primary degenerative or post-traumatic decenerative disease, previous tibial condule or plateau fractures, deformity or revious arthroplasty. o MOTO™ PFJ System The MOTO™ PFJ is designed for cemented use in partial knee arthroplasty, if there is evidence of enough sound bone to seat and support the components. Patellofemoral replacement is indicated in the following cases: - · Osteoarthritis, post-traumatic arthritis, severe chondrocalcinosis of the patellofemoral joint. - · Previously failed surgical attempts (i.e. arthroscopy, lateral release, tibial tubercle elevation, cartilage transplantation). - · History of patellar dislocation or fracture, resulting in cartilage degeneration of the patellofemoral joint. - · Degeneration induced by dysplasia. If the surgeon evaluates an unequivocal indication for replacement of the patellofemoral joint, with or without a patella resurfacing, which outweighs the risks associated with the surgery, PFJ replacement may be considered, particularly for young patients.
Device Story
MOTO Partial Knee & MOTO PFJ Systems Extension are sterile, cemented, implantable orthopedic prostheses for partial knee arthroplasty. Components include TiNbN-coated Co-Cr-Mo femoral components and E-Cross (Vitamin-E highly crosslinked UHMWPE) tibial inserts and patellar components. Devices are used by orthopedic surgeons in clinical settings to replace articulating surfaces of the knee joint. The extension expands compatibility of the MOTO Patella with existing MOTO PFJ, MOTO Partial Knee, and GMK total knee systems. Implants provide mechanical stability and articulation to restore joint function in patients with degenerative or post-traumatic conditions. Clinical benefit is derived from pain relief and improved joint mobility through surgical replacement of damaged cartilage and bone surfaces.
Clinical Evidence
No clinical data. Evidence is based on non-clinical bench testing, including mechanical performance, wear behavior, ion release, contact pressure/area, and constraint testing (jump height) per ASTM F2083, alongside biocompatibility and pyrogenicity assessments.
Technological Characteristics
Materials: Co-Cr-Mo alloy (ISO 5832-4) with TiNbN coating; E-Cross (Vitamin-E highly crosslinked UHMWPE). Components: Femoral, tibial inserts, patellofemoral joint, patella. Fixation: Cemented. Testing standards: ASTM F2083 for contact pressure/area and clipping system. Sterilization: Sterile.
Indications for Use
Indicated for patients requiring partial knee or patellofemoral joint arthroplasty due to primary degenerative or post-traumatic degenerative disease, previous fractures, deformity, failed prior surgical attempts, or dysplasia, provided sufficient sound bone exists to support components.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 16, 2021
Medacta USA % Chris Lussier Senior Director, Quality and Regulatory 3973 Delp Street Memphis, Tennessee 38118
Re: K213071
Trade/Device Name: MOTO™ Partial Knee & MOTO PFJ Systems Extension Regulation Number: 21 CFR 888.3520 Regulation Name: Knee Joint Femorotibial Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSX, KRR, JWH, NPJ Dated: November 11, 2021 Received: November 12, 2021
Dear Chris Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
### Device Name
MOTO™ Partial Knee & MOTO PFJ Systems Extension
### Indications for Use (Describe)
### o MOTO™ Partial Knee System
The MOTO™ Partial Knee System is designed for cemented use in partial knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. Partial replacement of the articulating surfaces of the knee is indicated when only one compartment of the joint is affected due to the compartmental primary degenerative or posttraumatic decenerative disease, previous tibial condule or plateau fractures, deformity or revious arthroplasty.
### o MOTO™ PFJ System
The MOTO™ PFJ is designed for cemented use in partial knee arthroplasty, if there is evidence of enough sound bone to seat and support the components. Patellofemoral replacement is indicated in the following cases:
- · Osteoarthritis, post-traumatic arthritis, severe chondrocalcinosis of the patellofemoral joint.
- · Previously failed surgical attempts (i.e. arthroscopy, lateral release, tibial tubercle elevation, cartilage transplantation).
- · History of patellar dislocation or fracture, resulting in cartilage degeneration of the patellofemoral joint.
- · Degeneration induced by dysplasia.
If the surgeon evaluates an unequivocal indication for replacement of the patellofemoral joint, with or without a patella resurfacing, which outweighs the risks associated with the surgery, PFJ replacement may be considered, particularly for young patients.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Senior Director of Quality and Regulatory, Medacta USA Date Prepared: September 22, 2021 Date Revised: November 12, 2021
#### II. Device
| Device Proprietary Name: | MOTO™ Partial Knee & MOTO PFJ Systems Extension |
|--------------------------|------------------------------------------------------------------------------|
| Common or Usual Name: | Prosthesis, Knee, Femorotibial, Non-Constrained, Cemented,<br>Metal/Polymer |
| Classification Name: | Knee joint femorotibial metal/polymer non-constrained<br>cemented prosthesis |
| Primary Product Code | HSX |
| Secondary Product Code: | KRR, JWH, NPJ |
| Regulation Number: | 21 CFR 888.3520, 21 CFR 888.3540 and 21 CFR 888.3560 |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following devices:
Primary predicate device
- > MOTO™ Partial Knee System, K162084, Medacta International SA
Additional predicate devices
- > MOTO™ Lateral Partial Knee System, K183029, Medacta International SA
- > MOTO™ PFJ System, K200122, Medacta International SA
Additional reference devices
- > GMK Sphere E-cross, K202022, Medacta International SA
- > GMK Total Knee System TiNbN Coating, K202684, Medacta International SA
{4}------------------------------------------------
#### IV. Device Description
The purpose of this submission is to gain clearance for the MOTO™ Partial Knee & MOTO™ PFJ Systems Extension and to extend the compatibility of the already cleared MOTO Patella. Specifically, both the new MOTO Patella and the already cleared one (K200122) may be used in the following combinations:
- . with the MOTO PFJ
- with the MOTO PFJ and with the MOTO Partial Knee System (K162084, K183029), GMK UNI ● (K162084).
- alone (without MOTO PFJ) with the GMK Primary or GMK Sphere devices (K090988, ● K120790, K121416, K122232, K140826, K142069 and K202684) or GMK Spherika (K211004).
The MOTO™ Partial Knee & MOTO™ PFJ Systems is a line extension to provide a larger product offering. The subject devices are sterile implantable devices designed for cemented use in partial knee arthroplasty procedures. The MOTO™ Partial Knee & MOTO™ PFJ Systems extension includes:
- Medial femoral components TiNbN coated, left medial, sizes from 1 to 10; ●
- Lateral femoral components TiNbN coated, sizes from 1 to 7; ●
- Medial tibial inserts fixed E-cross, left medial and right medial, sizes from 1 to 8 and 6 thicknesses ● from 8 to 14;
- . Lateral tibial inserts fixed E-cross, sizes from 1 to 8 and 6 thicknesses from 8 to 14;
- . Patello Femoral Joint TiNbN coated, left and right, sizes from 1 to 6;
- MOTO Patella E-cross, 6 sizes. ●
The subject femoral components, as well as the patello femoral joint, are manufactured from cobaltchromium-molybdenum alloy (Co-Cr-Mo) according to ISO 5832-4 with Titanium Nitride (TiNbN) coating.
The subject tibial inserts, as well as the MOTO patella, are made of E-Cross (Vitamin-E Highly Crosslinked UHMWPE).
#### Indications for Use V.
## MOTO™ Partial Knee System
The MOTOM Partial Knee System is designed for cemented use in partial knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. Partial replacement of the articulating surfaces of the knee is indicated when only one compartment of the joint is affected due to the compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty.
## o MOTO™ PFJ System
The MOTO™ PFJ is designed for cemented use in partial knee arthroplasty, if there is evidence of enough sound bone to seat and support the components. Patellofemoral replacement is indicated in the following cases:
{5}------------------------------------------------
- . Osteoarthritis, post-traumatic arthritis, severe chondrocalcinosis of the patellofemoral joint.
- Previously failed surgical attempts (i.e. arthroscopy, lateral release, tibial tubercle elevation, ● cartilage transplantation).
- History of patellar dislocation or fracture, resulting in cartilage degeneration of the patellofemoral . joint.
- Degeneration induced by dysplasia. ●
If the surgeon evaluates an unequivocal indication for replacement of the patellofemoral joint, with or without a patella resurfacing, which outweighs the risks associated with the surgery, PFJ replacement may be considered, particularly for young patients.
#### Comparison of Technological Characteristics VI.
The MOTO™ Partial Knee & MOTO™ PFJ Systems Extension implants and the predicate devices (K162084, K183029 and K200122) share the following characteristics:
- indications for use; ●
- sizes: ●
- shape and design: ●
- fixation and stability;
- biocompatibility; ●
- . device usage:
- sterility; ●
- shelf-life; and ●
- packaging.
The MOTO™ Partial Knee & MOTO™ PFJ Systems Extension implants differ from the predicate devices (K162084, K183029 and K200122) as follow:
- MOTO tibial inserts and MOTO Patella material; and .
- femoral components and MOTO PFJ coating. .
## Discussion
Medacta International SA has not made any change to the indications for use, sizes, general design and shape, fixation and stability features, device usage, biocompatibility, sterility, shelf life, and packaging of the subject devices respect to the predicate devices.
Based on the comparison of technological characteristics and performance data provided within this submission, the data supports the substantial equivalence of the MOTO™ Partial Knee & MOTO™ PFJ Systems Extension implants to the identified predicate devices.
#### VII. Performance Data
Based on the risk analysis, performance testing were conducted to written protocols. The following tests and rationales are provided in support of the substantial equivalence determination: Non-Clinical Studies
{6}------------------------------------------------
- PERFORMANCE TESTING .
- TiNbN coated MOTO Femoral component Mechanical performances о
- TiNbN coated MOTO Femoral component Wear behavior o
- TiNbN coating Excessive Ions release o
- TiNbN coated PFJ implant Mechanical performances o
- TiNbN coated PFJ implant Wear behavior O
- E-Cross MOTO medial and lateral tibial inserts Wear behavior. ROM and constraints O rationale
- MOTO Partial knee system E-Cross medial and lateral tibial inserts: static ML SHEAR and O AP draw test of the insert-tray clipping system according to ASTM F2083
- MOTO E-Cross tibial inserts Contact pressures and areas according to ASTM F2083 о
- Contact pressure and areas: MOTO Patella articulating on GMK-Sphere & GMK-Primary o femoral component
- Constraint comparison within Medacta MOTO Patella & Medacta EVOLIS Patella in terms O of jump height
- E-Cross MOTO Patella Wear behavior, ROM and constraints rationale o
- Constraints equivalence of GMK Resurfacing and MOTO Patella articulating with GMK o Spherika femoral component and GMK Spherika anatomical femoral component and the predicated device
- Contact areas equivalence of GMK Resurfacing and MOTO Patella articulating with GMK о Spherika femoral component and GMK Spherika anatomical femoral component and the predicated device
- PYROGENICITY ●
- Bacterial endotoxin test (LAL test) according to European Pharmacopoeia \$2.6.14 (which is o equivalent to USP chapter <85>)
- o Pyrogen test according to USP chapter <151> for pyrogenicity determination
- The subject devices are not labeled as non-pyrogenic or pyrogen free. o
- BIOCOMBATIBILITY evaluation .
- SHELF-LIFE evaluation ●
## Clinical Studies:
- No clinical studies were conducted. ●
{7}------------------------------------------------
## VIII. Conclusion
The information provided above supports that the MOTO™ Partial Knee & MOTO™ PFJ Systems Extension implants are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.