K162084 · Medacta International S.A. · HSX · Jan 3, 2017 · Orthopedic
Device Facts
Record ID
K162084
Device Name
MOTO Partial Knee System
Applicant
Medacta International S.A.
Product Code
HSX · Orthopedic
Decision Date
Jan 3, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3520
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MOTO™ Partial Knee System is designed for cemented use in partial knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. Partial replacement of the articulating surfaces of the knee is indicated when only one side of the joint is affected due to the compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty.
Device Story
MOTO Partial Knee System; cemented unicompartmental knee prosthesis. Components: Co-Cr-Mo femoral component (ISO 5832-4), UHMWPE tibial insert (ISO 5834-2), Ti-6Al-4V tibial baseplate (ISO 5832-3). Used in partial knee arthroplasty to replace articulating surfaces in one compartment. Implanted by orthopedic surgeons in clinical settings. Provides mechanical stability via fixation pegs (femoral) and keel/peg (tibial). Benefits: restores joint function in patients with localized degenerative disease or trauma. Single-use, sterile.
Clinical Evidence
Bench testing only. Studies included fatigue testing under dynamic physiological loads (ASTM F1800), wear testing (ASTM F2083), sterilization validation (ANSI/AAMI/ISO 11137-1/2, ISO 11135-1), accelerated/real-time aging, packaging validation, and pyrogen limit testing (LAL).
Technological Characteristics
Materials: Co-Cr-Mo (ISO 5832-4), UHMWPE (ISO 5834-2 Type 1), Ti-6Al-4V (ISO 5832-3). Fixed bearing design. Cemented fixation. Sterilization: Gamma irradiation or ethylene oxide. Sizes: 10 femoral, 8 tibial insert/tray. Mechanical stabilization via pegs and keel.
Indications for Use
Indicated for patients requiring partial knee arthroplasty due to compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity, or revision of previous arthroplasty, provided there is sufficient sound bone to seat and support components.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. The text is written in a sans-serif font.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 3, 2017
Medacta International SA % Elizabeth Wheeler Manager, Regulatory Affairs Mapi USA, Inc 2343 Alexandria Drive, Suite 100 Lexington, Kentucky 40504
Re: K162084 Trade/Device Name: Moto Partial Knee System Regulation Number: 21 CFR 888.3520 Regulation Name: Knee Joint Femorotibial Metal/Polymer Non-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: HSX Dated: November 16, 2016 Received: November 17, 2016
Dear Elizabeth Wheeler:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K162084
Device Name
MOTO™ Partial Knee System
#### Indications for Use (Describe)
The MOTO™ Partial Knee System is designed for cemented use in partial knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. Partial replacement of the articulating surfaces of the knee is indicated when only one side of the joint is affected due to the compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Medacta International. The logo consists of two blue triangles on the left, followed by the word "Medacta" in blue. A yellow line is underneath the word "Medacta". The word "International" is in a smaller font below the yellow line, and a Swiss flag is on the right.
#### 1 510(k) Summary
| Applicant/Sponsor: | Medacta International SA |
|--------------------|-----------------------------|
| | Strada Regina |
| | 6874 Castel San Pietro (CH) |
| | Switzerland |
| | Phone (+41) 91 696 60 60 |
| | Fax (+41) 91 696 60 66 |
- Contact Person: Stefano Baj Regulatory Affairs Manager Phone: +41 91 696 60 60 Fax: +41 91 696 60 66 Email: baj@medacta.ch
| Date Prepared: | July 26, 2016 |
|----------------|-------------------|
| Date Revised: | November 16, 2016 |
#### DEVICE INFORMATION
| Trade/Proprietary Name: | MOTO™ Partial Knee System |
|-------------------------|-------------------------------------------------------------------------------|
| Common or Usual Name: | Unicompartmental Knee prosthesis |
| Classification Name: | Knee joint femorotibial metal/polymer/ non-constrained cemented<br>prosthesis |
| Product Code: | HSX |
| Regulation Number: | 21 CFR 888.3520 |
| Device Class: | II |
## PREDICATE DEVICE INFORMATION
## Primary predicates:
| 510(k) | Product | 510(k) Holder |
|---------|---------|----------------|
| K073175 | Journey | Smith & Nephew |
## Additional predicates:
| 510(k) | Product | 510(k) Holder |
|---------|---------|----------------|
| K102069 | Journey | Smith & Nephew |
| K033363 | UNI | Zimmer |
{4}------------------------------------------------
## DEVICE DESCRIPTION
The MOTO™ Partial Knee System is a knee prosthesis designed for cemented use in medial compartment partial knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. Partial replacement of the articulating surfaces of the knee is indicated when only one side of the joint is affected due to the compartmental primary degenerative or post-traumatic degenerative disease, previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty. The femoral component is made of cobalt-chromiummolybdenum (Co-Cr-Mo per ISO 5832-4), and the tibial component consists of an ultra-high molecular weight polyethylene (UHMWPE per ISO 5834-2 Type 1) insert, and a metal baseplate component (Ti-6Al-4V per ISO 5832-3). The MOTO™ Partial Knee System implants are offered sterile (via gamma irradiation or ethylene oxide), are intended for single use only, and may not be re-sterilized.
The Anatomic Femoral Component (cemented) is anatomically shaped and designed with two (2) fixation pegs for sizes 1 to 7 and three (3) fixation pegs for sizes 8 to 10. Available in ten (10) sizes (1 - 10), the femoral component is offered in both Right Medial (RM) and Left Medial (LM) options for each size.
The Medial Fixed Tibial Insert has a fixed design and is available in five (8) sizes (1 – 8). Each size is offered in six (6) levels of thickness (8, 9, 10, 11, 12, 14 mm).
The Medial Fixed Tibial Tray (cemented) has a fixed bearing design with one fixation keel and peg to ensure primary stability. Available in five (8) sizes (1 - 8), the Medial Fixed Tibial Tray are offered in both Right Medial (RM) and Left Medial (LM) options for each size.
#### INDICATIONS FOR USE
The MOTO™ Partial Knee System is designed for cemented use in partial knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. Partial replacement of the articulating surfaces of the knee is indicated when only one side of the joint is affected due to the compartmental primary degenerative or post-traumatic degenerative disease. previous tibial condyle or plateau fractures, deformity or revision of previous arthroplasty.
#### Discussion:
The Indications for Use Statement is similar to the predicate device. The only difference is that the subject device calls out the requirements for "evidence of sufficient sound bone to seat and support the components", and the predicate device cites specific example of degenerative joint disease ("osteoarthritis, traumatic arthritis, or avascular necrosis"). These differences do not create new intended uses for the subject device: the subject and predicate devices are all prostheses intended for use in knee arthroplasty.
{5}------------------------------------------------
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS
The design features, and materials of the subject devices are substantially equivalent to those of the predicate devices. The tables below compare characteristics of the subject and predicate devices.
| Feature | MOTO™ Partial Knee<br>System - Anatomic<br>Femoral Component<br>(cemented) | Smith & Nephew<br>Journey<br>(K073175) | Zimmer UNI<br>(K033363) |
|-----------------------|----------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------|-------------------------|
| Sizes | 10 sizes (Size 1-10) | 7 sizes (Size 1 -7) | 7 sizes (Size A - G) |
| Configuration | RM and LM | RM/LL and LM/RL | RM/LL and LM/RL |
| Material | CoCrMo - ISO 5832-4 | Same<br>(with an oxinium<br>oxidized zirconium<br>coating) | Same |
| Cemented? | Yes | Same | Same |
| Device Usage | Single Use | Same | Same |
| Sterility | Provided sterile via<br>Gamma Radiation | Same | Same |
| Shelf Life | 5 Years | Same | > 5 Years |
| Stabilization<br>pegs | Anatomic Femoral<br>component is cemented<br>with two (2) pegs for<br>sizes 1-7 and three (3)<br>pegs for sizes 8-10 | 2 Pegs | 2 Pegs |
| Feature | MOTO™ Partial Knee System-<br>Fixed Tibial Insert | Zimmer UNI<br>(K033363) |
|--------------|----------------------------------------------------------------------------|-------------------------|
| Sizes | 8 sizes (from 1 to 8) | 6 sizes (from 1 to 6) |
| Thickness | Each size is offered in 6 levels of<br>thickness (8, 9, 10, 11, 12, 14 mm) | Same |
| Material | UHMWPE (ISO 5834 -2) Type 1 | Same |
| Device Usage | Single Use | Same |
| Sterility | Provided sterile via Ethylene Oxide | Same |
| Shelf Life | 5 years | > 5 years |
{6}------------------------------------------------
| Feature | MOTO™ Partial<br>Knee System - Fixed<br>Plus Tibial Tray<br>(cemented) | Smith & Nephew<br>Journey (K102069) | Zimmer UNI<br>(K033363) |
|----------------|------------------------------------------------------------------------|-------------------------------------|--------------------------------|
| Sizes | 8 sizes (from 1 to 8) | 6 sizes | 6 sizes (from 1 to 6) |
| Configurations | RM and LM | RM/LL and LM/RL | Same |
| Material | Ti-6Al-4V (ISO 5832-<br>3) | Same | Ti-6Al-4V with PMMA<br>precoat |
| Cemented | Cemented | Same | Same |
| Device Usage | Single Use | Same | Same |
| Sterility | Provided sterile via<br>Gamma Radiation | Unknown | Same |
| Shelf Life | 5 years | Unknown | > 5 years |
| Stabilization | Stabilized with peg and<br>keel | Same | Same |
| Fixed Bearing? | Yes | Same | Same |
The fundamental scientific technology of the modified devices has not changed relative to the predicate devices. The safety and effectiveness of the subject devices is adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification.
#### PERFORMANCE TESTING
The following studies were performed to support substantial equivalence:
- Fatigue under dynamic physiological loads according to ASTM F1800
- Excessive wear according to ASTM F2083 .
- Sterilization validation in accordance with ANSI/AAMI/ISO 11137-1:2006. ● ANSI/AAMI/ISO 11137-2:2006 and ISO 11135-1:2007
- Accelerated and real time aging
- . Packaging validation
- Limulus Amebocyte Lysate (LAL) testing was evaluated to establish the device meets . pyrogen limit specifications.
#### Conclusion:
Based on the above information, the MOTO™ Partial Knee System can be considered substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.