GMK-SPHERE Tibial inserts FLEX Tibial Insert CR and Resurfacing Patella made of E-CROSS (Vitamin E Highly Crosslinked UHMWPE)
Applicant
Medacta International S.A.
Product Code
JWH · Orthopedic
Decision Date
Sep 18, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GMK knee prosthesis is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components. This knee replacement system is indicated in the following cases: Severely painful and/or disabled joint as a result of arthritis, traumatoid arthritis, rheumatoid arthritis. Avascular necrosis of femoral condyle. Post traumatic loss of joint configuration. Primary implantation failure. GMK Sphere can be implanted using a kinematic alignment approach. When a kinematic alignment approach is utilized, this knee replacement system is indicated in the following cases: Severely painful and/or disabled joint as a result of arthritis, traumatoid arthritis or polyarthritis. Collagen disorders, and/or avascular necrosis of the femoral condyle. Moderate valgus, varus, or flexion deformities. Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement. The screwed tibial augments are for screwed fixation to the tibial baseplate. In case a semi-constrained liner is used, an extension stem must be implanted both on the femoral components. In case a GMK Revision tibial tray is used, an extension stem must be implanted.
Device Story
GMK-SPHERE Tibial inserts and Resurfacing Patella are line extensions to the GMK Sphere Total Knee System. These components are manufactured from E-CROSS (Vitamin E Highly Crosslinked UHMWPE). The device is used in total knee arthroplasty to replace damaged joint surfaces. It is implanted by orthopedic surgeons in a clinical/hospital setting. The system functions as a mechanical bearing surface within the knee joint, providing stability and articulation. The use of Vitamin E-stabilized UHMWPE is intended to improve wear resistance compared to standard UHMWPE. The device is compatible with previously cleared GMK Sphere femoral components and tibial baseplates. No new instrumentation is required for implantation. The output is a restored joint surface, which benefits the patient by reducing pain and improving joint function.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including constraint (A/P draw, M/L shear, rotary laxity), wear testing in a knee simulator, dynamic endurance testing, and contact pressure/area analysis. Material characterization of E-CROSS was performed. Pyrogenicity was evaluated via LAL test (European Pharmacopoeia 2.6.14/USP <85>) and USP <151>.
Technological Characteristics
Components made of E-CROSS (Vitamin E Highly Crosslinked UHMWPE). Compatible with Co-Cr-Mo (ISO 5832-4) femoral components and tibial trays. Tibial inserts utilize Ti6Al4V (ISO 5832-3) where applicable. Mechanical sensing/actuation is not applicable; device is a passive orthopedic implant. Sterilization method is consistent with predicate devices.
Indications for Use
Indicated for patients with severely painful/disabled joints due to arthritis (traumatic, rheumatoid, polyarthritis), avascular necrosis of femoral condyle, post-traumatic joint loss, primary implantation failure, or collagen disorders. Suitable for patients requiring total knee arthroplasty with sufficient bone stock for component seating, including those undergoing kinematic alignment with moderate valgus, varus, or flexion deformities.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
Medacta International SA, GMK Sphere Extension (K140826)
Medacta International SA, GMK Knee Prosthesis- GMK Sphere Tibial Insert Flex (K162035)
Medacta International SA, GMK Sphere CR Tibial Inserts (K181635)
Medacta International SA, GMK Total Knee System (K090988)
Medacta International SA, GMK Resurfacing Patella Size 4 (K113571)
Reference Devices
Zimmer-Biomet, Vivacit-E Vitamin E Highly Crosslinked Polyethylene Liners (K120370)
Submission Summary (Full Text)
{0}------------------------------------------------
September 18, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and "ADMINISTRATION" in a smaller font size below.
Medacta International SA % Chris Lussier Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis, Tennessee 38118
Re: K202022
Trade/Device Name: GMK-SPHERE Tibial inserts FLEX Tibial Insert CR and Resurfacing Patella made of E-CROSS (Vitamin E Highly Crosslinked UHMWPE) Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: July 21, 2020 Received: July 22, 2020
Dear Chris Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
## 510(k) Number (if known) K202022
### Device Name
GMK-SPHERE Tibial inserts FLEX Tibial Insert CR and Resurfacing Patella made of E-CROSS (Vitamin E Highly Crosslinked UHMWPE)
## Indications for Use (Describe)
GMK knee prosthesis is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components.
This knee replacement system is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthritis, traumatoid arthritis, rheumatoid arthritis.
- Avascular necrosis of femoral condyle.
- · Post traumatic loss of joint configuration.
- · Primary implantation failure.
GMK Sphere can be implanted using a kinematic alignment approach. When a kinematic alignment approach is utilized, this knee replacement system is indicated in the following cases:
- · Severely painful and/or disabled joint as a result of arthritis, traumatoid arthritis or polyarthritis.
- · Collagen disorders, and/or avascular necrosis of the femoral condyle.
- · Moderate valgus, varus, or flexion deformities.
Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement. The screwed tibial augments are for screwed fixation to the tibial baseplate.
In case a semi-constrained liner is used, an extension stem must be implanted both on the femoral components. In case a GMK Revision tibial tray is used, an extension stem must be implanted.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510(k)
K202022 page 1 of 7
Medacta International SA
# 3.0 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Applicant Correspondent: Chris Lussier, Director of Quality and Regulatory, Medacta USA Date Prepared: July 20, 2020
#### II. Device
| Device Proprietary Name: | GMK-SPHERE Tibial inserts FLEX Tibial Insert CR and<br>Resurfacing Patella made of E-CROSS (Vitamin E Highly<br>Crosslinked UHMWPE) |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Tibial Inserts and Resurfacing Patella |
| Classification Name: | Knee joint patellofemorotibial polymer/metal/polymer<br>semiconstrained cemented prosthesis. |
| Primary Product Code: | JWH |
| Regulation Number: | 21 CFR 888.3560 |
| Device Classification | II |
#### Predicate Device III.
Substantial equivalence is claimed to the following device:
- Medacta International SA, GMK Sphere K121416
- Medacta International SA. GMK Sphere Extension, K140826
- Medacta International SA, GMK Knee Prosthesis- GMK Sphere Tibial Insert Flex, K162035
- Medacta International SA, GMK Sphere CR Tibial Inserts, K181635
- o Medacta International SA, GMK Total Knee System, K090988
- Medacta International SA, GMK Resurfacing Patella Size 4, K113571
{4}------------------------------------------------
K202022 page 2 of 7 Traditional 510(k)
Medacta International SA
Reference predicate:
- Zimmer-Biomet, Vivacit-E Vitamin E Highly Crosslinked Polyethylene Liners, K120370 ●
#### Device Description IV.
The GMK Sphere Flex insert E-CROSS (Vitamin E Highly Crosslinked UHMWPE) , GMK Sphere CR E-CROSS (Vitamin E Highly Crosslinked UHMWPE) inserts and the Resurfacing Patella (Vitamin E Highly Crosslinked UHMWPE) are a line extension to the GMK Sphere Total Knee System and are:
- o Tibial insert Fixed Sphere FLEX: Left and Right, Sizes 1-6, Thicknesses 10-11-12-13-14-17-20 mm, E-cross-Vitamin-E Highly Crosslinked UHMWPE;
- Tibial Insert Fixed SPHERE CR: Left and Right, Sizes 1-6, Thicknesses 10-11-12-13-● 14mm, E-cross-Vitamin-E Highly Crosslinked UHMWPE;
- Patella resurfacing E-Cross: Sizes 1-4 E-Cross Vitamin-E Highly Crosslinked ● UHMWPE
GMK Sphere implants are sterile implantable devices designed for tricompartmental replacement of the natural knee joint.
They are a line extension to Medacta previously cleared implants: GMK Sphere (K121416), GMK Sphere Extension, (K140826), GMK Knee Prosthesis- GMK Sphere Tibial Insert Flex (K162035), GMK Sphere CR Tibial Inserts (K181635) and GMK Total Knee System (K0909889) and GMK Resurfacing Patella Size 4 (K113571).
Liners and Resurfacing patella subject of this submission are implants made of E-Cross (Vitamin-E Highly Crosslinked UHMWPE).
The introduction of the subject items do not require additional instrumentation needed during the surgical procedure and do not alter the indented use or outcomes.
The following components of the GMK Sphere have been cleared previously under K121416 and K140826 predicate device:
{5}------------------------------------------------
Traditional 510(k)
K202022 page 3 of 7
Medacta International SA
- . Femoral Component Left and Right, Sizes 1-7 Co-Cr-Mo (ISO 5832-4 Third Edition 2014-09-15 Implants for Surgery - Metallic Materials - Part 4: Cobalt-Chromium-Molybdenum Casting Alloy);
- Femoral Component Left and Right. Sizes 1+ to 6+ (intermediate sizes) Co-Cr-Mo (ISO . 5832-4 Third Edition 2014-09-15 Implants for Surgery - Metallic Materials - Part 4: Cobalt-Chromium-Molybdenum Casting Alloy);
- Tibial tray fixed cemented Left and Right, 4 intermediate sizes Co-Cr-Mo (ISO 5832-4 . Third Edition 2014-09-15 Implants for Surgery - Metallic Materials - Part 4: Cobalt-Chromium-Molybdenum Casting Alloy);
- 0 Tibial Insert Fixed Flex, Left and Right, Sizes 1-6. 10mm -20mm UHMWPE (ISO 5834 2:2011 Implants for Surgery - Ultra-High-Molecular-Weight Polyethylene - Part 2: Moulded Forms) Type 1, Ti6Al5V (ISO 5832-3:1996 Implants for Surgery - Metallic Materials - Part 3 : Wrought Titanium 6-Aluminium 4-Vanadium Alloy);
- o Tibial Insert Fixed Flex. Left and Right, Sizes 1-6. 11mm and 13mm UHMWPE (ISO 5834-2:2011 Implants for Surgery - Ultra-High-Molecular-Weight Polyethylene - Part 2: Moulded Forms) Type 1. Ti6Al4V (ISO 5832-3:1996 Implants for Surgery - Metallic Materials – Part 3: Wrought Titanium 6-Aluminium 4-Vanadium Alloy; and
- Instrumentation. .
The following components of the GMK Sphere have been cleared previously under the Medacta GMK Total Knee System:
- Resurfacing patella Sizes 1-4 (K090988 and K113571); o
- Tibial tray fixed cemented Left and Right, Sizes 1-6 (K090988); and ●
- Primary extension stem Ø 11mm/ L 65mm (K090988) and L 30mm (K133630). o
In detail, the liners and the resurfacing patella subject of the current submission are compatible with the all the GMK Sphere tibial baseplates and femurs previously cleared.
#### V. Indications for Use
GMK knee prosthesis is designed for cemented use in total knee arthroplasty, if there is evidence of sufficient sound bone to seat and support the components.
This knee replacement system is indicated in the following cases:
· Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis.
- · Avascular necrosis of femoral condyle.
- · Post traumatic loss of joint configuration.
- · Primary implantation failure.
GMK Sphere can be implanted using a kinematic alignment approach. When a kinematic alignment approach is utilized, this knee replacement system is indicated in the following cases:
{6}------------------------------------------------
Traditional 510(k)
K202022 page 4 of 7
Medacta International SA
· Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid arthritis or polyarthritis.
- · Collagen disorders, and/or avascular necrosis of the femoral condyle.
- · Moderate valgus, varus, or flexion deformities.
Tibial wedges cemented are to be attached to the tibial baseplate with both the fixing cylinders and bone cement.
The screwed tibial augments are for screwed fixation to the tibial baseplate.
In case a semi-constrained liner is used, an extension stem must be implanted both on the tibial and on the femoral components. In case a GMK Revision tibial tray is used, an extension stem must be implanted.
#### VI. Comparison of Technological Characteristics
- Tibial insert O
- o Tibial insert Fixed Sphere FLEX E-CROSS (Vitamin E Highly Crosslinked UHMWPE) implants and the Medacta predicate devices GMK SPHERE (K121416, K140826, K162035 and K181635) share the following characteristics:
- Indications for Use: ●
- Sizes: ●
- Packaging, ●
- Device Usage;
- Biocompatibility
- Shelf Life; and
- Sterilization Method.
Tibial insert Fixed Sphere FLEX E-CROSS (Vitamin E Highly Crosslinked UHMWPE) implants are technologically different from the predicate device as follows:
- Materials ●
- Design
- Tibial Insert Fixed SPHERE CR E-CROSS (Vitamin E Highly Crosslinked UHMWPE) ● implants and the predicate devices GMK SPHERE (K121416, K140826, K162035 and K181635) share the following characteristics:
- Indications for Use; 0
- 0 Sizes;
- Packaging ●
- Device Usage: ●
- Biocompatibility; ●
{7}------------------------------------------------
Traditional 510(k)
K202022 page 5 of 7
Medacta International SA
- Shelf Life; and
- Sterilization Method. ●
Tibial insert Fixed SPHERE CR E-CROSS (Vitamin E Highly Crosslinked UHMWPE)implants are technologically different from the predicate device as follows:
- Materials ●
- Patella Resurfacing E-CROSS (Vitamin E Highly Crosslinked UHMWPE) implants and O the predicate devices GMK SPHERE (K090988, K113571, K121416, K140826, K162035 and K181635) share the following characteristics:
- Design;
- Sizes: ●
- Device usage: ●
- Biocompatibility: ●
- Sterility; ●
- Shelf life; ●
- Packaging: .
Patella Resurfacing E-CROSS (Vitamin E Highly Crosslinked UHMWPE)implants are technologically different from the predicate device as follows:
- Materials
#### Performance Data VII.
A review of the mechanical data on the subject and predicate devices indicates that the subject devices are equivalent to devices currently cleared for use and do not alter the intended surgical outcomes. The purpose of this submission is to introduce new GMK Sphere inserts and Resurfacing patella made of Vitamin E Highly Crosslinked UHMWPE. The introduction of the subject items do not require additional instrumentation needed during the surgical procedure. The new inserts and resurfacing patella design introduction was evaluated bv risk analysis to identify any new risks associated. Based on the risk analysis, additional tests have been performed in order to demonstrate that the new E-Cross inserts and new E-Cross Resurfacing patella are as safe as effective as the predicate devices.
Testing was conducted according to written protocols with acceptance criteria that were based on standards. The following mechanical studies were performed in support of a substantial equivalence determination:
{8}------------------------------------------------
Traditional 510(k)
K202022 page 6 of 7
Medacta International SA
Non-Clinical Studies:
- PERFORMANCE TESTING O
- "GMK® SPHERE TIBIAL INSERTS MADE OF E-CROSS (VITAMIN E HIGHLY ● CROSSLINKED UHMWPE): CONSTRAINT A/P DRAW TEST - M/L SHEAR TEST - ROTARY LAXITY TEST, Test Report A.1
- GMK®-SPHERE TIBIAL INSERTS E-CROSS (VITAMIN E HIGHLY ● CROSSLINKED UHMWPE): WEAR TEST IN KNEE SIMULATOR MACHINE WITH LOAD CONTROL, test Report A.2
- GMK® SPHERE TIBIAL INSERT FLEX MADE OF E-CROSS (VITAMIN ● EHIGHLY CROSSLINKED UHMWPE): MAXIMUM POSTERO-ANTERIOR LOAD THAT CAN OCCUR ON THE TIBIAL INSERT DURING ACTIVITIES OF DAILY LIVING, Test Report A.3
- GMK® SPHERE TIBIAL INSERT FLEX MADE OF ● E-CROSS (VITAMIN EHIGHLY CROSSLINKED UHMWPE): DYNAMIC ENDURANCE TEST UNDER PA LOAD, Test Report A.4
- GMK-SPHERE RESURFACING PATELLA MADE OF E-CROSS (VITAMIN E ● HIGHLY CROSSLINKED UHMWPE): CONTACT PRESSURE AND AREAS, Test Report A.5
- GMK®-SPHERE E-CROSS TIBIAL INSERTS CONTACT PRESSURE AND ● AREAS: COMPARATIVE ANALYSIS BETWEEN E-CROSS and UHMWPE DEVICES, Test Report A.6
- MATERIAL CHARACTERIZATION OF MEDACTA'S VITAMIN E HIGHLY CROSSLINKED UHMWPE (E-CROSS)
- PYROGENICITY O
- Bacterial Endotoxin Test (LAL test) was conducted according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>)
- Pyrogen test according to USP chapter <151> for pyrogenicity determination
- o The subject devices are not labeled as non-pyrogenic or pyrogen free
# Clinical Studies
- No clinical studies were conducted
----------------------------This space intentionally left blank---------------------------------------------------------------------------------------------------------------
{9}------------------------------------------------
Traditional 510(k)
K202022 page 7 of 7
Medacta International SA
## VIII. Conclusion
Based on the above information, the GMK Sphere Flex insert E-CROSS (Vitamin E Highly Crosslinked UHMWPE), GMK Sphere CR E-CROSS (Vitamin E Highly Crosslinked UHMWPE) inserts and the E-CROSS Resurfacing Patella (Vitamin E Highly Crosslinked UHMWPE) are substantially equivalent to the identified predicate devices.
Substantial equivalence has been demonstrated through a comparison of intended use, design, and technological characteristics, as well as performance evaluations.
The GMK Sphere Extension implants are as safe and effective as the predicate devices
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.