K200417 · Rist Neurovascular, Inc. · QJP · May 2, 2020 · Cardiovascular
Device Facts
Record ID
K200417
Device Name
RIST Radial Access Catheter
Applicant
Rist Neurovascular, Inc.
Product Code
QJP · Cardiovascular
Decision Date
May 2, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The RIST Radial Access Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
Device Story
Flexible, single-lumen percutaneous catheter; 0.070" OD, 0.040" ID; 120/130 cm lengths. Features radiopaque shaft reinforced with stainless steel; luer connector at proximal end; Simmons 2 or Berenstein distal tip shapes. Used for transradial access to deliver interventional devices to target sites in peripheral, coronary, and neuro vasculature. Operated by physicians in clinical settings. Input: guidewires (0.035"/0.038"). Output: conduit for interventional devices and fluid infusion. Benefits: provides access to target vasculature for diagnostic or therapeutic procedures.
Clinical Evidence
Bench testing only. Includes design verification (tensile, torque, kink, burst, leak, particulate, dimensional) and validation (simulated-use study, radiopacity, usability). Biocompatibility testing per ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity). All acceptance criteria met.
Indicated for introduction of interventional devices into peripheral, coronary, and neuro vasculature. No specific age or gender restrictions provided.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Predicate Devices
Penumbra, Inc. Benchmark Intracranial Access System (5F Select Catheter) (K142321)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font size below.
May 2, 2020
RIST Neurovascular, Inc. Vyoma Chikara Design Quality/RA 11611 Interchange Circle South Miramar, Florida 33025
Re: K200417
Device Name: RIST Radial Access Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: OJP, DOY Dated: February 19, 2020 Received: February 20, 2020
Dear Vyoma Chikara:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng, Ph.D., M.S. Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200417
Device Name RIST Radial Access Catheter
Indications for Use (Describe)
The RIST Radial Access Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Date Summary Prepared: | April 23, 2020 |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | RIST Neurovascular, Inc.<br>11611 Interchange Circle S<br>Miramar, FL 33025 |
| Primary Submission Contact: | Vyoma Upadhya Chikara<br>Design Quality/RA<br>RIST Neurovascular Inc.<br>11611 Interchange Circle South<br>Miramar, FL 33025<br>Telephone: 1-954-559-1323<br>Facsimile: 1-954-742-5989<br>E-Mail: Vyoma@InNeuroCo.com |
| Secondary Submission Contact: | Marianne Grunwaldt<br>Director, Quality Assurance & Regulatory Affairs<br>RIST Neurovascular Inc.<br>11611 Interchange Circle South<br>Miramar, FL 33025<br>Telephone: 1-305-495-3883<br>Facsimile: 1-954-742-5989<br>E-Mail: Marianne@InNeuroCo.com |
| Trade Name: | RIST Radial Access Catheter |
| Regulation Number: | 21 CFR 870.1250 |
| Device Common or<br>Classification Name: | Percutaneous Catheter |
| Product Class: | Class II |
| Product Panel: | Cardiovascular |
| Product Code: | QJP, DQY |
| Predicate Device: | Penumbra, Inc. Benchmark Intracranial Access<br>System (5F Select Catheter), #K142321 |
| Reference Predicate: | RIST Cath Radial Access Long Sheath, #K191551 |
{4}------------------------------------------------
#### 1. Device Description
The RIST Radial Access Catheter is a flexible, single lumen catheter compatible with 0.035 inch and 0.038-inch guidewires. It is designed to deliver interventional devices into the peripheral, coronary, and neuro vasculature and is intended to provide access to the target site via transradial access. The RIST Radial Access Catheter consists of a radiopaque catheter shaft reinforced with stainless steel and has a luer connector at its proximal end for the attachment of accessories and the infusion of fluids. The distal tip of the catheter is shaped to have a smooth, rounded tip and is offered in two different distal segment shapes, namely Simmons 2 (SIM2) and Berenstein (BER). The catheter has a nominal outer diameter of 0.070 inches and a nominal inner diameter of 0.040 inches. It is available in two working lengths: 120 cm and 130 cm. The RIST Radial Access Catheter is compatible with the RIST Cath Radial Access Long Sheath (K191551) and off-theshelf accessories. The RIST Radial Access Catheter is supplied sterile, non-pyrogenic, and intended for single use only.
#### 2. Indications for Use
The RIST Radial Access Catheter is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
The indication for use for the RIST Radial Access Catheter is identical to the indication for use of the predicate device.
#### 3. Technological Characteristics and Basis for Substantial Equivalence
The RIST Radial Access Catheter, subject of this 510(k) submission, is substantially equivalent in its intended use/indications for use, technology/principal of operation, materials, sterilization method and performance to the predicate device, the Penumbra, Inc. 5F Select Catheter that was cleared as part of the Benchmark Intracranial Access System, and the reference predicate device, the RIST Cath Radial Access Long Sheath.
A comparison of the technological characteristics of the subject device, the predicate, and the reference predicate devices, is summarized in Table 1.
{5}------------------------------------------------
| Parameter | Subject Device<br>RIST Radial Access<br>Catheter | Predicate Device<br>Penumbra Benchmark<br>Access System (5F Select<br>Catheter)<br>510(k) # K142321 | Reference Predicate Device<br>RIST Cath Radial Access<br>Long Sheath<br>510(k) #K191551 |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The RIST Radial Access<br>Catheter is indicated for the<br>introduction of interventional<br>devices into the peripheral,<br>coronary, and neuro<br>vasculature. | The Benchmark Intracranial<br>Access System is indicated<br>for the introduction of<br>interventional devices into<br>the peripheral, coronary, and<br>neuro vasculature. | The RIST Cath Radial<br>Access Long Sheath is<br>indicated for the introduction<br>of interventional devices into<br>the peripheral, coronary, and<br>neuro vasculature. |
| Product Code | QJP, DQY | DQY | DQY |
| Regulation No. | 21 CFR 870.1250 | 21 CFR 870.1250 | 21 CFR 870.1250 |
| Classification | Class II | Class II | Class II |
| Catheter Material | Polyether Block Amide<br>(PEBAX) with BaSO4 | Commonly used medical<br>grade plastics | Chronoflex/Polyblend (distal<br>most section)<br>Chronoflex<br>Polyether Block Amide<br>(PEBAX)<br>Vestamid (proximal most<br>section) |
| Hub Material | Polycarbonate (Makrolon) | Not Available | Polycarbonate (Makrolon) |
| Catheter Shaft<br>Reinforcement | Stainless Steel | Stainless Steel | Stainless Steel Coil<br>(proximal)<br>Nitinol Wire Coil (distal) |
| Working Length | 120, 130 cm | 123, 131.5 cm | 95, 100 cm |
| Inner Diameter | 0.040 inches Minimum | 0.043 inches Maximum | 0.079 inches |
| Outer Diameter | 0.070 inches Maximum | 0.069 inches Maximum | 0.093 inches |
| Tip Shapes | Simmons 2<br>Berenstein | Simmons<br>Berenstein<br>HI | N/A |
| Packaging | Tyvek/Nylon pouch,<br>polyethylene support tube,<br>packaging card, SBS carton | Commonly used medical<br>device packaging materials | Tyvek/Nylon pouch,<br>polyethylene support tube,<br>packaging card, SBS carton |
| Sterilization | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| Pyrogenicity | Nonpyrogenic | Nonpyrogenic | Nonpyrogenic |
| Number of Uses | Single Use | Single Use | Single Use |
Table 1 – Comparison Between the RIST Radial Access Catheter and Predicate Device
{6}------------------------------------------------
#### 4. Performance Data
Design verification and validation were performed to ensure that the RIST Radial Access Catheter meets its performance specifications and demonstrates substantial equivalence to the predicate device. There are no known performance standards for this device. A list of the performance testing conducted on the RIST Radial Access Catheter is presented in Table 2. Some of the tests were also conducted on the predicate device to help establish substantial equivalence. All predetermined acceptance criteria were met.
In some cases, verification test data were leveraged from data that had been generated from testing done on previously cleared devices. These tests are included in Table 3.
The data demonstrates that the RIST Radial Access Catheter is substantially equivalent to the predicate device.
| Test Performed | Test Method | Results | Discussion of How This<br>Test Supports a Finding<br>of Substantial<br>Equivalence | | | | | | | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|-------------------|----------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------|
| Design Verification Testing | | | | | | | | | | | |
| Tensile Strength | Testing was completed per<br>ISO 10555-1. Using a force<br>gauge, test samples were<br>pulled until failure. | All units met the Tensile<br>Strength acceptance<br>criteria. | This test demonstrates<br>that the RIST Radial<br>Access Catheter is<br>structurally sound after<br>passing it through a<br>model which simulates<br>actual anatomy.<br>Simulating actual use<br>supports the indications<br>for use which are the<br>same in the RIST Radial<br>Access Catheter and in<br>the predicate device. | | | | | | | | |
| Torque Strength | The distal end of the unit was<br>held rigid while the proximal<br>end was turned until failure. | All units met the Torque<br>Strength acceptance<br>criteria. | This test demonstrates the<br>ability of the unit to<br>function in a clinical<br>setting and therefore<br>supports the indications<br>for use which is the same<br>in the RIST Radial Access<br>Catheter and in the<br>predicate device. | | | | | | | | |
| Particulates | Particulate testing was<br>performed in a tortuous<br>model and particulate count<br>was evaluated per USP<br><788>. | The acceptance criteria<br>for Particulate testing was<br>met. | This test quantifies the<br>number of particles<br>generated during<br>simulated use and<br>demonstrate that the<br>particle quantities are<br>within expected<br>parameters which do not<br>raise any new questions | | | | | | | | |
| Test Performed | Test Method | Results | Discussion of How This<br>Test Supports a Finding<br>of Substantial<br>Equivalence | | | | | | | | |
| | | | of safety and efficacy.<br>The test supports the<br>indications for use<br>which are the same in<br>the RIST Radial Access<br>Catheter and in the<br>predicate device. | | | | | | | | |
| Dimensional<br>Verification - ID | The ID was measured to<br>ensure the acceptance criteria<br>was met. | The acceptance criteria<br>for ID Dimensional<br>Verification was met. | This test verifies<br>functional characteristics<br>of the RIST Radial<br>Access Catheter. The<br>internal diameter of the<br>RIST Radial Access<br>Catheter and the<br>predicate are minimally<br>different, but this does<br>not affect safety or<br>efficacy. The test<br>supports the indications<br>for use which are the<br>same in the RIST Radial<br>Access Catheter and in<br>the predicate device. | | | | | | | | |
| Dimensional<br>Verification -OD | The OD was measured to<br>ensure the acceptance criteria<br>was met. | The acceptance criteria<br>for OD Dimensional<br>Verification was met. | This test verifies<br>functional characteristics<br>of the RIST Radial<br>Access Catheter. The<br>outer diameter of the<br>RIST Radial Access<br>Catheter and the<br>predicate are minimally<br>different, but this does<br>not affect safety or<br>efficacy. The test<br>supports the indications<br>for use which are the<br>same in the RIST Radial<br>Access Catheter and in<br>the predicate device. | | | | | | | | |
| Dimensional<br>Verification - Working<br>Length | The working length was<br>measured to ensure the<br>acceptance criteria were met. | The acceptance criteria<br>for Working Length<br>Dimensional Verification<br>was met. | This test verifies the<br>functional characteristics<br>of working length. The<br>RIST Radial Access<br>Catheter is offered in<br>similar working lengths<br>as the predicate device.<br>The working length of<br>the RIST Radial Access<br>Catheter and the | | | | | | | | |
| Test Performed | Test Method | Results | Discussion of How This<br>Test Supports a Finding<br>of Substantial<br>Equivalence | | | | | | | | |
| | | | predicate are minimally<br>different, but this does<br>not affect safety or<br>efficacy. The test<br>supports the indications<br>for use which are the<br>same in the RIST Radial<br>Access Catheter and in<br>the predicate device. | | | | | | | | |
| Dimensional<br>Verification - Tip<br>Length | The tip length was measured<br>to ensure the acceptance<br>criteria were met. | The acceptance criteria<br>for Tip Length<br>Dimensional Verification<br>were met. | Both the RIST Radial<br>Access Catheter and the<br>predicate device have<br>similar functional<br>characteristics. This test<br>demonstrates the<br>integrity of the RIST<br>Radial Access Catheter<br>under predetermined<br>conditions. The test<br>supports the indications<br>for use which are the<br>same in the RIST Radial<br>Access Catheter and in<br>the predicate device. | | | | | | | | |
| Kink Resistance | Test units were wrapped<br>around progressively smaller<br>diameter pegs and/or<br>mandrels until a kink was<br>observed. | All units met the Kink<br>Resistance acceptance<br>criteria. | This test demonstrates the<br>ability of the unit to<br>maintain structural<br>integrity and therefore<br>supports the indications<br>for use which is the same<br>in the RIST Radial Access<br>Catheter and in the<br>predicate device. | | | | | | | | |
| Visual Inspection (Tip<br>Taper/Transitions) | Samples were visually<br>inspected to ensure the<br>acceptance criteria were met. | All units met the Tip<br>Taper/Transition Visual<br>Inspection acceptance<br>criteria | Both the RIST Radial<br>Access Catheter and the<br>predicate device have<br>similar functional<br>characteristics. This test<br>demonstrates the integrity<br>of the RIST Radial<br>Access Catheter under<br>predetermined conditions.<br>The test supports the<br>indications for use which<br>are the same in the RIST<br>Radial Access Catheter<br>and in the predicate<br>device. | | | | | | | | |
| Test Performed | Test Method | Results | Discussion of How This<br>Test Supports a Finding<br>of Substantial<br>Equivalence | | | | | | | | |
| Shape Retention<br>(Catheter Tip Shape) | Samples were individually<br>inspected to ensure that the<br>shape met the specified<br>specification. | All units met the Shape<br>Retention Acceptance<br>criteria. | Both the RIST and the<br>predicate device have<br>similar functional<br>characteristics. This test<br>demonstrates the RIST<br>Radial Access Catheter<br>tip will retain its shape.<br>The test supports the<br>indications for use which<br>are the same in the RIST<br>Radial Access Catheter<br>and in the predicate<br>device. | | | | | | | | |
| Burst Test | Testing was completed per<br>ISO 10555-1 by clamping the<br>distal end and pressurizing<br>the assembly was pressurized<br>and peak pressure was<br>recorded. | All units met the Catheter<br>Burst acceptance criteria. | Both the RIST Radial<br>Access Catheter and the<br>predicate device have<br>similar functional<br>characteristics. This test<br>demonstrates the integrity<br>of the RIST Radial<br>Access Catheter under<br>predetermined conditions.<br>The test supports the<br>indications for use which<br>are the same in the RIST<br>Radial Access Catheter<br>and in the predicate<br>device. | | | | | | | | |
| Liquid Leak Test | Testing was completed per<br>ISO 10555-1 by connecting<br>the catheter to test<br>equipment, sealing the distal<br>end of the catheter,<br>pressurizing the catheter,<br>holding the pressure, and<br>ensuring there was no<br>leakage. | All units met the Liquid<br>Leak acceptance criteria. | Both the RIST Radial<br>Access Catheter and the<br>predicate device have<br>similar functional<br>characteristics. This test<br>demonstrates the<br>functional integrity of the<br>RIST Radial Access<br>Catheter under<br>predetermined conditions.<br>The test supports the<br>indications for use which<br>are the same in the RIST<br>Radial Access Catheter<br>and in the predicate<br>device. | | | | | | | | |
| Air Leak Test | Testing was conducted per<br>ISO 80369-7 to ensure no air<br>leaks into the product<br>assembly. | All units met the Air Leak<br>Test acceptance criteria. | Both the RIST Radial<br>Access Catheter and the<br>predicate device have<br>similar functional<br>characteristics. This test | | | | | | | | |
| Test Performed | Test Method | Results | Discussion of How This<br>Test Supports a Finding<br>of Substantial<br>Equivalence | | | | | | | | |
| | | | demonstrates the<br>functional integrity of the<br>RIST Radial Access<br>Catheter under<br>predetermined conditions.<br>The test supports the<br>indications for use which<br>are the same in the RIST<br>Radial Access Catheter<br>and in the predicate<br>device. | | | | | | | | |
| Chemical Compatibility | Samples of catheter and<br>accessories were exposed to<br>saline, dextrose, heparin, and<br>radiocontrast and then<br>inspected for any signs of<br>degradation and ensure the<br>ID had no obstruction. | All units met the<br>Chemical Compatibility<br>acceptance criteria. | The materials of the<br>predicate are mostly the<br>same as those of the RIST<br>Radial Access Catheter.<br>This test confirms that the<br>materials of the RIST<br>Radial Access Catheter<br>can be used with<br>chemicals typically used<br>in a clinical setting. The<br>test supports the<br>indications for use which<br>are the same in the RIST<br>Radial Access Catheter<br>and in the predicate<br>device. | | | | | | | | |
| Corrosion | Testing was conducted per<br>ISO 10555-1 Annex A. | All units met the<br>Corrosion acceptance<br>criteria. | This test demonstrates the<br>ability of the stainless-<br>steel reinforcement to<br>resist deterioration. This<br>test supports substantial<br>equivalence because the<br>RIST Radial Access<br>Catheter and the predicate<br>device have the same<br>stainless steel<br>reinforcement. | | | | | | | | |
| Packaging - Visual<br>Inspection | Packaging was visually<br>inspected to verify the<br>integrity of the pouch and<br>verify the product was free<br>from damage prior to, and<br>when removed from the<br>packaging. | All units met the<br>Packaging Visual<br>Inspection. | The packaging of the<br>RIST Radial Access<br>Catheter is identical to<br>the reference predicate<br>device. This test verifies<br>the integrity of the<br>pouched unit which<br>helps support the<br>packaging comparison.<br>The test supports the<br>indications for use | | | | | | | | |
| Test Performed | Test Method | Results | Discussion of How This<br>Test Supports a Finding<br>of Substantial<br>Equivalence | | | | | | | | |
| | | | which are the same in<br>the RIST Radial Access<br>Catheter and in the<br>predicate device. | | | | | | | | |
| Packaging - Pouch<br>Leak | Testing was conducted per<br>ASTM F-1929-15. The<br>pouch was dipped into a dyed<br>solution and visually<br>inspected for dye penetration<br>through the seal. | All units met the Pouch<br>Leak test | The packaging of the<br>RIST Radial Access<br>Catheter is identical to<br>the reference predicate<br>device.…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.