The Intermediate Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
Device Story
Intermediate Catheter is a variable stiffness, single-use, sterile, non-pyrogenic device; functions as a reinforcing conduit for other intravascular devices. Input: manual insertion through guide catheter or vascular sheath by physician. Operation: provides access to target site in peripheral and neuro vasculature. Construction: PTFE-lined lumen; stainless steel braid, stainless steel coil, and Nitinol coil reinforcement; hydrophilic coating; radiopaque Platinum/Iridium marker band. Output: physical access path for interventional devices. Clinical benefit: facilitates delivery of interventional tools to target vascular sites. Accessories include Tuohy-Borst hemostasis valve and split introducer sheath.
Clinical Evidence
Bench testing only. No clinical data. Performance verified via biocompatibility (ISO 10993-1), sterilization (ISO 11135), radiographic detectability, visual inspection, working length, hub compatibility, dimensional analysis, leak testing (air/liquid), particulate testing (USP 788), catheter burst, tensile strength, corrosion resistance, and packaging integrity (ASTM F1929-12, ASTM F88-09). All samples met acceptance criteria.
Technological Characteristics
Variable stiffness catheter; PTFE-lined lumen; stainless steel braid, stainless steel coil, and Nitinol coil reinforcement; hydrophilic coating; Platinum/Iridium marker band. Dimensions: 105/115/125 cm lengths; 0.060" ID; 0.072" distal/0.078" proximal OD. Sterilization: Ethylene Oxide. Standards: ISO 10993-1, ISO 11135, ISO 10555-1, ISO 594-1/2, USP 788, ASTM F1929-12, ASTM F88-09.
Indications for Use
Indicated for the introduction of interventional devices into the peripheral and neuro vasculature in patients requiring such access.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 14, 2015
TNI Manufacturing, Inc. % Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street NW Buffalo, Minnesota 55313
Re: K152202
Trade/Device Name: Intermediate Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY, DYB, DTL Dated: November 19, 2015 Received: November 23, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Carlos L. Pena -S/^
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152202
Device Name Intermediate Catheter
Indications for Use (Describe)
The Intermediate Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
#### Submitter's Name and Address
TNI Manufacturing, Inc. 4635 NW 103rd Avenue Sunrise, Florida 33351 Telephone: 1-954-742-5988 Facsimile: 1-954-742-5989
#### Contact Information
Daniel Sablyak President 4635 NW 103rd Avenue Sunrise, Florida 33351 Telephone: 1-954-742-5988 Facsimile: 1-954-742-5989 E-Mail: Dan@tnimfg.com
### Date Prepared
November 16, 2015
### Device Trade or Proprietary Name
Intermediate Catheter Hemostasis Valve Split Sheath Introducer
### Device Common or Classification Name:
Catheter, Percutaneous, 21CFR870.1250, Class II Hemostasis Valve, 21CFR870.4290, Class II Catheter Introducer, 21CFR870.1340, Class II
#### Product Code:
{4}------------------------------------------------
DQY (Catheter)
DTL (Hemostasis Valve)
DYB (Catheter Dilator)
## Identification of the Legally Marketed Devices to which Equivalence is Being Claimed
| Name of Predicate Device | Name of Manufacturer | 510(k) Number |
|-----------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| 5MAXTM DDC<br>Distal Delivery Catheters | Penumbra, Inc.<br>Alameda, California | K122772 |
| | Predicate Device<br>Penumbra 5MAXTM<br>DDC | Intermediate Catheter |
| 510(k) Number | K122772 | K152202 |
| Classification | 21CFR870.1250, Class II | Same |
| Product Code | DQY | Same |
| Indication For Use | The DDC Catheters are<br>indicated for the<br>introduction of<br>interventional devices into<br>the peripheral and neuro<br>vasculature. | The Intermediate Catheter is<br>indicated for the<br>introduction of<br>interventional devices into<br>the peripheral and neuro<br>vasculature. |
| Inner Lumen | PTFE | PTFE |
| Catheter Shaft<br>Material | Nylon, PTFE,<br>polyurethane,<br>hydrophilic coating | Nylon, PTFE, polyimide,<br>polyether block amide, PET,<br>polycarbonate/urethane<br>hydrophilic coating |
| Hub Material | Polycarbonate | Same |
| Catheter Shaft<br>Reinforcement | Stainless Steel/Nitinol | Same |
| Marker Band | Platinum/Iridium | Same |
| Packaging | Tyvek Pouch,<br>polyethylene support<br>hoop, packaging card,<br>SBS Carton | Same |
| Sterilization | Ethylene Oxide | Same |
| Pyrogenicity | Nonpyrogenic | Same |
| Inside Diameter (ID) | 0.062" | Same |
| Outside Diameter<br>(OD) | 6 Fr | Same |
| Working Lengths (cm) | 115 and 125 cm | 105, 115, and 125 cm |
| | Predicate Device<br>Penumbra 5MAX™<br>DDC | Intermediate Catheter |
| Compatible Guidewire | 0.038" | Same |
### Device Description
The Intermediate Catheter is a variable stiffness catheter and has a catheter shaft reinforced with a stainless steel braid, a stainless steel coil, a Nitinol coil, and has a radiopaque Platinum/Iridium marker band on the distal end. The catheter has a nominal distal outer diameter of 0.072" and a nominal proximal outer diameter of 0.078". It is available with a nominal inner diameter of 0.060". They are available in three working lengths, 105 cm, 115 cm. and 125 cm. The Intermediate Catheters have a PTFE-lined lumen, which is braid and coil reinforced, flexible, and has a hydrophilic coating. The Intermediate Catheters are inserted through a guide catheter or vascular sheath, provide access to the target site and once in place, provides a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst hemostasis valve and a split introducer sheath. The Intermediate Catheter is supplied sterile, non-pyrogenic, and intended for single use only.
### Indications For Use
The Intermediate Catheter is indicated for the introduction of interventional devices into the peripheral and neuro vasculature.
### Comparison to Predicate Device
{5}------------------------------------------------
{6}------------------------------------------------
### Summary of Non Clinical Data
Biocompatibility tests conducted with the Intermediate Catheters and accessories were selected in accordance with ISO 10993-1 guidelines (Biological Evaluation of Medical Devices) for limited duration (<24 hours), external communicating devices, contacting circulating blood. Studies were conducted pursuant to 21CFR58, Good Laboratory Practices. Biocompatibility testing found the Intermediate Catheters to be biocompatible non-pyrogenic. and
The conclusions drawn from the physical, mechanical, and performance testing of the subject Intermediate Catheter and accessories demonstrates that the product is as safe and as effective for its labeled indications and is Substantially Equivalent to the legally marketed predicate device.
| Test | Test Method<br>Summary | Results |
|-------------------------------|-----------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Biocompatibility | Testing completed<br>per ISO 10993-1 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Sterilization<br>Validation | Testing completed<br>per ISO 11135 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Radiographic<br>Detectability | Testing completed<br>per ISO 10555-1 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Visual Inspection | Testing completed<br>per ISO 10555-1 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Test | Test Method<br>Summary | Results |
| Working Length | Testing completed<br>per ISO 10555-1 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Hub compatibility | Testing completed<br>per ISO 594-1 and<br>ISO 594-2 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Outside diameter | Testing completed<br>per ISO 10555-1 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Leak — Air | Testing completed<br>per ISO 10555-1 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Leak — Liquid | Testing completed<br>per ISO 10555-1 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Particulates | Testing completed<br>per USP 788 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Catheter Burst | Testing completed<br>per ISO 10555-1 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Tensile | Testing completed<br>per ISO 10555-1 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Corrosion | Testing completed<br>per ISO 10555-1 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Packaging - Dye<br>Leak | Testing completed<br>per ASTM F1929-<br>12 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
| Packaging - Peel | Testing completed<br>per ASTM F88-09 | All units tested must meet the acceptance<br>criteria. All samples met the acceptance<br>criteria. |
Intermediate Catheter Performance Testing
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.