LONG SHEATH OR LS, 70 CM, LONG SHEATH OR LS, 80 CM, LONG SHEATH OR LS, 90 CM
K152876 · Tni Manufacturing, Inc. · DQY · Jan 8, 2016 · Cardiovascular
Device Facts
Record ID
K152876
Device Name
LONG SHEATH OR LS, 70 CM, LONG SHEATH OR LS, 80 CM, LONG SHEATH OR LS, 90 CM
Applicant
Tni Manufacturing, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Jan 8, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
Device Story
Long Sheath is a variable stiffness, braid-reinforced catheter designed to provide a conduit for interventional devices into peripheral, coronary, and neuro vasculature. Device features a PTFE-lined lumen, hydrophilic coating, and distal radiopaque Platinum/Iridium marker band. Available in 70, 80, and 90 cm lengths with 0.088-inch inner diameter. Supplied with a Tuohy-Borst hemostasis valve and vessel dilator. Used by clinicians to facilitate access to target sites; inserted through a guide catheter or vascular sheath. Provides a stable, reinforcing pathway for secondary intravascular devices. Single-use, sterile, non-pyrogenic device.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included biocompatibility (ISO 10993-1), radiographic detectability, visual inspection, working length, hub compatibility (ISO 594-1/2), dimensional verification, leak testing (air/liquid), particulate matter (USP 788), catheter burst, tensile strength, corrosion resistance, and packaging integrity (ASTM F1929-12, ASTM F88-09). All tests met acceptance criteria.
Technological Characteristics
Variable stiffness catheter; PEBAX shaft with stainless steel braid reinforcement; PTFE-lined lumen; hydrophilic coating; Platinum/Iridium distal marker band; Polycarbonate hub; Polyolefin strain relief. Dimensions: 0.109" OD, 0.088" ID, 70-90 cm lengths. Sterilized via Ethylene Oxide. Complies with ISO 10555-1, ISO 594-1/2, and ISO 10993-1.
Indications for Use
Indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature for patients requiring such access.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 8, 2016
TNI Manufacturing, Inc. Ms. Marianne Grunwaldt Director, Quality Assurance & Regulatory Affairs 4635 NW 103td Avenue Sunrise. Florida 33351
Re: K152876
Trade/Device Name: Long Sheath Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: December 3, 2015 Received: December 7, 2015
Dear Ms. Grunwaldt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena AS
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K152876
Device Name Long Sheath
Indications for Use (Describe)
The Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasulature.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
### Submitter's Name and Address
TNI Manufacturing, Inc. 4635 NW 103rd Avenue Sunrise, Florida 33351 Telephone: 1-954-742-5988 Facsimile: 1-954-742-5989
#### Contact Information
Marianne Grunwaldt Director, Quality Assurance & Regulatory Affairs InNeuroCo, Inc 4635 NW 103rd Avenue Sunrise, Florida 33351 Telephone: 1-305-495-3883 Facsimile: 1-954-742-5989 E-Mail: Marianne@InNeuroCo.com
### Date Prepared
September 25, 2015
#### Device Trade or Proprietary Name
Long Sheath or LS
## Device Common or Classification Name:
Catheter, Percutaneous, 21CFR870.1250, Class II
### Product Code:
DQY
{4}------------------------------------------------
# Identification of the Legally Marketed Devices to which Equivalence is Being Claimed
| Name of Predicate Device | Name of Manufacturer | 510(k) Number |
|--------------------------|---------------------------------------|---------------|
| NEURON MAX SYSTEM | Penumbra, Inc.<br>Alameda, California | K111380 |
### Device Description
The Long Sheath or LS is a variable stiffness catheter and has a catheter shaft reinforced with a stainless steel braid and has a radiopaque Platinum/Iridium marker band on the distal end. The catheter has a nominal outer diameter of 0.109 inches. It is available with a nominal inner diameter of 0.088 inches. They are available in three working lengths: 70 cm, 80 cm, and 90 cm. The Long Sheath has a PTFE-lined lumen, which is braidreinforced, flexible, and has a hydrophilic coating. The LS are inserted through a guide catheter or vascular sheath, provide access to the target site and once in place, provide a reinforcing conduit for other intravascular devices. Accessories included with the device are a Tuohy-Borst hemostasis valve with an extension luer and a dilator. The LS is supplied sterile, non-pyrogenic, and intended for single use only.
### Indications for Use
The Long Sheath is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
| | Predicate Device<br>Penumbra NEURON<br>MAX SYSTEM | TNI Manufacturing, Inc.<br>Long Sheath |
|----------------|---------------------------------------------------|----------------------------------------|
| 510(k) Number | K111380 | K152876 |
| Classification | 21CFR870.1250, Class II | 21CFR870.1250, Class II |
| Product Code | DQY | DQY |
## Comparison to Predicate Device
{5}------------------------------------------------
| Review Panel | Cardiovascular | Cardiovascular |
|---------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication For Use | The Neuron™ MAX<br>System is indicated for the<br>introduction of<br>interventional devices into<br>the peripheral, coronary,<br>and neuro vasculature. | The Long Sheath is<br>indicated for the<br>introduction of<br>interventional devices into<br>the peripheral, coronary,<br>and neuro vasculature. |
| Components Supplied | Sheath, Vessel Dilator,<br>Hemostasis Valve, | Sheath, Vessel Dilator,<br>Hemostasis Valve |
| Catheter Shaft<br>Material | Nylon, Polyurethane | Polyether Block Amide,<br>(PEBAX) |
| Inner Liner | PTFE | PTFE |
| Hub Material | Polycarbonate | Polycarbonate |
| Strain Relief | Polyolefin | Polyolefin |
| Catheter Shaft<br>Reinforcement | Stainless Steel Braid | Stainless Steel Braid |
| Lubricious Coating | Hydrophilic Coating | Hydrophilic Coating |
| Radiopaque Marker<br>Band | Platinum/ Iridium | Platinum/ Iridium |
| Packaging | Tyvek Pouch,<br>polyethylene support tube,<br>packaging card, SBS<br>carton | Tyvek/Nylon Pouch,<br>polyethylene support tube,<br>packaging card, SBS carton |
| Sterilization | Ethylene Oxide | Ethylene Oxide |
| Pyrogenicity | Nonpyrogenic | Nonpyrogenic |
| Working Lengths | 70, 80, 90, and 100 cm | 70, 80, 90 cm |
| Internal Diameter | 0.088 inches | 0.088 inches |
| Outer Diameter | 0.108 inches | 0.109 inches |
| Shelf Life | 3 years | 3 years |
| Vessel Dilator | Aid in sheath introduction<br>during procedure | Aid in sheath introduction<br>during procedure |
| Hemostasis Valve | Minimizes blood loss | Minimizes blood loss |
| Luer Tapered Hub | Yes | Yes |
{6}------------------------------------------------
### Summary of Non Clinical Data
Biocompatibility tests conducted with the Long Sheath and accessories were selected in accordance with ISO 10993-1 guidelines (Biological Evaluation of Medical Devices) for limited duration (<24 hours), external communicating devices, contacting circulating blood. Studies were conducted pursuant to 21CFR58, Good Laboratory Practices. Biocompatibility testing found the Long Sheath to be biocompatible and non-pyrogenic.
The conclusions drawn from the physical, mechanical, and performance testing of the subject LS and accessories demonstrates that the product is Substantially Equivalent to the legally marketed predicate device for its labeled indications.
| Test | Test Method<br>Summary | Results |
|-------------------------------|-----------------------------------------------------|----------------------------------------------------------------------------------------------------|
| Biocompatibility | Testing completed<br>per ISO 10993-1 | All units tested meet the acceptance<br>criteria. All samples met the acceptance<br>criteria |
| Radiographic<br>Detectability | Testing completed<br>per ISO 10555-1 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
| Visual Inspection | Testing completed<br>per ISO 10555-1 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
| Working Length | Testing completed<br>per ISO 10555-1 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
| Hub compatibility | Testing completed<br>per ISO 594-1 and<br>ISO 594-2 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
| Test | Test Method<br>Summary | Results |
| Outside diameter | Testing completed<br>per ISO 10555-1 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
| Leak – Air | Testing completed<br>per ISO 10555-1 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
| Leak - Liquid | Testing completed<br>per ISO 10555-1 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
| Particulates | Testing completed<br>per USP 788 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
| Catheter Burst | Testing completed<br>per ISO 10555-1 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
| Tensile | Testing completed<br>per ISO 10555-1 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
| Corrosion | Testing completed<br>per ISO 10555-1 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
| Packaging - Dye Leak | Testing completed<br>per ASTM F1929-12 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
| Packaging - Peel | Testing completed<br>per ASTM F88-09 | All units tested must meet the<br>acceptance criteria. All samples met the<br>acceptance criteria |
Long Sheath Performance Testing
{7}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.