K142321 · Penumbra, Inc. · DQY · Jan 26, 2015 · Cardiovascular
Device Facts
Record ID
K142321
Device Name
Benchmark Intracranial Access System
Applicant
Penumbra, Inc.
Product Code
DQY · Cardiovascular
Decision Date
Jan 26, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Indications for Use
The Benchmark™ Intracranial Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
Device Story
Benchmark Intracranial Access System consists of a delivery catheter (guide catheter) and a select catheter (microcatheter). Used by physicians in clinical settings to access target vasculature during interventional procedures. The system supports diagnostic or therapeutic devices, facilitating navigation to target vascular locations and delivery of agents. Compatible with off-the-shelf accessories. Provides physician convenience through various lengths and distal shapes. Benefits include improved access to complex neurovascular anatomy, aiding in the delivery of necessary interventional tools.
Clinical Evidence
Bench testing only. No clinical or animal data required. Biocompatibility testing per EN ISO 10993-1 (cytotoxicity, sensitization, irritation, systemic toxicity, pyrogenicity, hemocompatibility, thrombogenicity) passed. Mechanical bench testing (simulated use in anatomical models, coating integrity, particulate testing, air aspiration, pressure testing, friction force, tensile strength, bond strength, elongation) confirmed performance specifications met.
Technological Characteristics
Catheters constructed from medical-grade plastics and stainless steel. Features include various lengths and distal shapes. Sterilization via Ethylene Oxide (EtO). Compatible with off-the-shelf accessories. Biocompatibility verified per EN ISO 10993-1.
Indications for Use
Indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature in patients requiring interventional procedures.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 26, 2015
Penumbra, Inc. Ms. Michaela Mahl Senior Manager, Regulatory Affairs 1351 Harbor Bay Parkway Alameda, California 94502
Re: K142321
> Trade/Device Name: Benchmark Intracranial Access System Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: December 15, 2014 Received: December 16, 2014
Dear Ms. Michaela Mahl,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Carlos L. Pena - SD/△
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K142321
Device Name Benchmark Intracranial Access System
Indications for Use (Describe)
The Benchmark™ Intracranial Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------|
| <table><tr><td><div> <input checked="true" type="checkbox"/> Research Use (Per 21 CFR 201.66 and .68) </div></td><td><div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 201.66 and .68) </div></td></tr></table> | <div> <input checked="true" type="checkbox"/> Research Use (Per 21 CFR 201.66 and .68) </div> | <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 201.66 and .68) </div> |
| <div> <input checked="true" type="checkbox"/> Research Use (Per 21 CFR 201.66 and .68) </div> | <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 201.66 and .68) </div> | |
> Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) SUMMARY
## (as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra Inc. is providing the summary of Substantial Equivalence for the Benchmark™ Intracranial Access System.
#### 7.1 Sponsor/Applicant Name and Address
Penumbra, Inc. 1351 Harbor Bay Parkway Alameda, CA 94502, USA
#### 7.2 Sponsor Contact Information
Michaela Mahl Senior Manager, Regulatory Affairs Phone: (510) 748-3288 FAX: (510) 217-6414 Email: michaela.mahl@penumbrainc.com
#### 7.3 Date of Preparation of 510(k) Summary
January 20, 2015
#### 7.4 Device Trade or Proprietary Name
Benchmark™ Intracranial Access System
#### 7.5 Device Classification
| Regulatory Class: | ll |
|------------------------------------------|-----------------------------------------------|
| Classification Panel: Cardiovascular | |
| | Classification Name: Percutaneous Catheter |
| Regulation Number: | 21 CFR § 870.1250 |
| Product Code: | DOY |
#### 7.6 Predicate Devices
| 510(k) Number /<br>Clearance Date | Name of Predicate Device | Name of<br>Manufacturer |
|-----------------------------------|------------------------------|-------------------------|
| K082290 / 31Oct2008 | Neuron Delivery Catheter 070 | Penumbra, Inc. |
| K083125 / 21Nov2008 | Neuron Select Catheter 070 | Penumbra, Inc. |
7
{4}------------------------------------------------
#### 7.7 Predicate Comparison
| System Name | Neuron Intracranial Access System | Benchmark Intracranial Access System |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Neuron 070 Delivery Catheter | Benchmark Delivery Catheter |
| 510(k) No. | K082290 | K142321 |
| Classification | Class II, DQY | SAME |
| Indication | The Neuron Intracranial Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. | The Benchmark Intracranial Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. |
| Catheter Materials | Commonly used medical grade plastics & stainless steel | SAME |
| Outer Dimension | 6F (0.081 in -0.083 in) | SAME |
| Inner Dimension | 0.070 in Min | SAME |
| Effective Length | 95 cm, 105 cm | 95 cm, 105 cm, 115 cm |
| Tip Shapes | Straight & Multi-Purpose | SAME |
| Packaging Materials | Commonly used medical device packaging materials | SAME |
| Sterilization | EtO | SAME |
| Shelf-Life | 36 Months | 12 Months [36 Months in-process] |
| System Name | Neuron Intracranial Access System | Benchmark Intracranial Access System |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Neuron Select Catheter 070 | 5F Select Catheter |
| 510(k) No. | K083125 | K142321 |
| Classification | Class II, DQY | Class II, DQY |
| Indication | The Neuron Intracranial Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. | The Benchmark Intracranial Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature. |
| Catheter Materials | Commonly used medical grade plastics & stainless steel | SAME |
| Outer Dimension | 5F (0.069in Max) | SAME |
| Inner Dimension | 0.043 in Max | SAME |
| Effective Length | 123 cm & 131.5cm | SAME |
| Tip Shapes | Berenstein, H1 & Simmons | SAME |
| Packaging Materials | Commonly use medical device packaging materials | SAME |
| Sterilization | EtO | SAME |
| Shelf-Life | 36 Months | SAME |
{5}------------------------------------------------
#### 7.8 Device Description
The Benchmark™ Intracranial Access System is designed to aid the physician in accessing the target vasculature during interventional procedures. Benchmark Intracranial Access System is composed of a Delivery Catheter (guide catheter) used to support other diagnostic or therapeutic devices and a corresponding Select Catheter (microcatheter). Use of the Benchmark Intracranial Access System facilitates navigation to the target vascular location and delivery of the necessary diagnostic and/or therapeutic agents. The Benchmark Intracranial Access System devices are compatible with off-the-shelf accessories. Various lengths and distal shapes of both the Benchmark Delivery Catheter and 5F Select Catheter are provided for physician convenience.
#### 7.9 Indications for Use
The Benchmark™ Intracranial Access System is indicated for the introduction of interventional devices into the peripheral, coronary, and neuro vasculature.
### 7.10 Summary of Non-Clinical Data
As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding safety and effectiveness of the device follows.
Included in this section are descriptions of the testing's. which substantiates the safe and effective performance of the Benchmark™ Intracranial Access System as well as its substantial equivalence to the predicate devices:
- Biocompatibility .
- Design Verification (Bench-Top Testing) .
The subject Benchmark™ Intracranial Access System met all established requirements.
### 7.10.1 Biocompatibility Testing
Biocompatibility tests conducted on the materials of the BenchmarkTM Intracranial Access System were selected in accordance with EN ISO 10993 -1 guidelines (Biological Evaluation of Medical Devices) for limited duration (<24 hours), external communicating devices, contacting circulating blood. All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices. In summary, non-clinical testing found the Benchmark™ Intracranial Access System to be biocompatible according to the requirements of EN ISO 10993 requirements. The following tests were performed and all tests passed successfully:
{6}------------------------------------------------
| Test | Acceptance Criteria | Results | Pass / Fail |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| <i>In Vitro</i> Cytotoxicity | Sample extracts must yield cell<br>lysis grade 2 or lower | Grade 0-1: None to Slight | Pass |
| Sensitization | Test Group shall yield Grade <<br>1 score on Magnusson and<br>Kligman scale (provided<br>control Grade < 1) | Grade 0: No visible change | Pass |
| Acute Intracutaneous<br>Reactivity (Irritation) | The difference in the mean test<br>article and mean control score<br>must be grade 1.0 or lower | Grade $\le$ 1.0 difference<br>between mean test article<br>and mean control score | Pass |
| Systemic Toxicity | | | |
| | Acute Systemic Toxicity | Sample extracts must not cause<br>the following:<br>• > 10% weight loss in 3 or<br>more test animals<br>• Mortality of 2 or more test<br>animals<br>• Abnormal behavior in 2 or<br>more test animals | No evidence of systemic<br>toxicity from sample<br>extracts<br>• No weight loss<br>• No death<br>• All test animals appeared<br>normal |
| Rabbit Pyrogen Study | Sample Extracts must not<br>cause a total rise in body<br>temperature of $\ge$ 0.5°C | Non-pyrogenic: No<br>evidence of material-<br>mediated pyrogenicity; no<br>single animal had a total<br>body temperature rise of<br>$\ge$ 0.5°C | Pass |
| Hemocompatibility | | | |
| | <i>In Vitro</i> Hemolysis<br>[Extract Methode] | Sample extracts must be non-<br>hemolytic ( $\le$ 2% hemolytic<br>index) | Non-hemolytic:<br>Hemolytic Index = 0.70% -<br>1.04%<br>Corrected Hemolytic index<br>= 0.00% - 0.23% |
| Complement Activation | The concentrations of C3a and<br>SC5b-9 in the test samples are<br>statistically similar to the<br>predicate (Exposure Control &<br>Ref Material) control and<br>statistically lower than the<br>positive control for all<br>exposure times | The test sample<br>concentrations of C3a and<br>SC5b-9 were statistically<br>similar or lower than the<br>predicate control sample<br>concentrations, and<br>statistically lower than the<br>positive control sample<br>concentration | Pass |
| Dog Thrombogenicity | The device must be non-<br>thrombogenic after 4 hours in<br><i>vivo</i> when compared to a<br>control device | Test Device was non-<br>thrombogenic after 4 hours<br>in <i>vivo</i> when compared to a<br>control device | Pass |
{7}------------------------------------------------
## 7.10.2 Bench-top Testing
The physical and mechanical properties of the Benchmark™ Intracranial Access System were assessed using standard test methods and pre-determined acceptance criteria. The following tests were performed and all tests passed successfully:
| Attribute | Sample<br>Size | Specification | Acceptance<br>Criteria | Results |
|----------------------------------------------------------------------------------------------------------|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|-----------------------------------------------------------------------------------------------|
| Dimensional/<br>Visual Inspection | | These evaluations confirm that the units used in this Design<br>Verification testing meet all inspection criteria for release of<br>finished goods (clinically acceptable) product. | | Pass |
| Simulated Use<br>[Intracranial<br>Access & Vessel<br>Access Entry<br>Performance] | | Simulated use testing of the Benchmark Delivery Catheter and<br>Benchmark Select Catheter was performed with accessory<br>devices in an anatomical model which simulated the tortuosity<br>of the neurovasculature. | | Pass |
| Catheter Coating<br>Integrity | N=30 | Coating has not delaminated,<br>peeled, or flaked after simulated<br>use | 100% Must<br>meet<br>Specification | 100% Pass |
| | N=10 | Coating has not further<br>delaminated, peeled, or flaked<br>from coating integrity baseline | 100% Must<br>meet<br>Specification | 100% Pass |
| Particulate<br>Testing (Catheter<br>Hydrophilic<br>Coating) | N=10 | The maximum number of<br>particles:<br><= 6000 particles ≥ 10 $μ$ m<br><= 600 particles ≥ 25 $μ$ m<br><= 600 particles ≥ 75 $μ$ m<br>0 particles ≥ 125 $μ$ m | 100% Must<br>meet<br>Specification | 10 $μ$ m - 100% Pass<br>25 $μ$ m - 100% Pass<br>75 $μ$ m - 100% Pass<br>125 $μ$ m - 100% Pass |
| Hub/Catheter Air<br>Aspiration | N=30 | When negative pressure is<br>pulled, no air may leak into hub | 100% Must<br>meet<br>Specification | 100% Pass |
| Pressure Test | N=30 | 45 psi for 30 sec MIN | 100% Must<br>meet<br>Specification | 100% Pass |
| Benchmark<br>Delivery<br>Catheter / Sheath<br>compatibility<br>(Friction Force) | N=30 | Maximum value per<br>specification | 100% Must<br>meet<br>Specification | 100% Pass |
| Benchmark<br>Delivery<br>Catheter /<br>Benchmark<br>Select Catheter<br>compatibility<br>(Friction Force) | N=30 | Maximum value per<br>specification | 100% Must<br>meet<br>Specification | 100% Pass |
| Catheter Shaft<br>Tensile Strength<br>(includes all<br>joints) | N=30 | Minimum value per specification | 100% Must<br>meet<br>Specification | 100% Pass |
| Attribute | Sample<br>Size | Specification | Acceptance<br>Criteria | Results |
| Hub to Shaft &<br>Hub to Hypotube<br>Bond Strength | N=30 | Minimum value per specification | 100% Must<br>meet<br>Specification | 100% Pass |
| Elongation to<br>Failure –<br>Catheter | N=30 | % Elongation ≥ 5% | 100% Must<br>meet<br>Specification | 100% Pass |
{8}------------------------------------------------
The results of the tests appropriately address the physical and mechanical performance expectations of the device. Based on these overall results, the physical and mechanical properties of the Benchmark™ Intracranial Access System are acceptable for the intended use and substantially equivalent to the predicate device.
## 7.10.3 Animal Study
No animal study was required.
## 7.10.4 Summary of Substantial Equivalence
The Benchmark™ Intracranial Access System is substantially equivalent to the predicate device with regard to intended use, operating principle, design concept, materials, shelflife, packaging and sterilization processes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.