The indications for use for total hip arthroplasty include: Painful, disabling joint disease of the hip resulting from: degenerative arthritis, post-traumatic arthritis or late stage avascular necrosis. Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure. Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results. Where bone stock is of poor quality or madequate for other reconstructive techniques, such as cementless fixation, as indicated by deficiencies of the acetabulum. The Exeter X3 RimFit Cup is intended for cemented use only. The indications for use for total and hemi hip arthroplasty include: noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis; rheumatoid arthritis; correction of functional deformity; revision procedures where other treatments or devices have failed; and, treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques. The Exeter V40 Femoral Stem is intended for use in total or hemi hip replacement. It is intended for cemented use only.
Device Story
Exeter X3 RimFit Cup and Exeter V40 Femoral Stem are orthopedic implants for cemented total or hemi hip arthroplasty. Devices replace diseased or damaged hip joint components to restore function and reduce pain. Implants are surgically placed by orthopedic surgeons in clinical/hospital settings. Subject devices represent modifications to predicate designs, including alternate terminal sterilization (EtO), alternate PMMA resin for cement spacers, and updated MR Conditional labeling. Implants function as passive mechanical prostheses; no active electronic components or software. Clinical benefit derived from restoration of hip joint mechanics and stability in patients with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including material characterization (ASTM F648, F2565), biocompatibility (ISO 10993), sterilization validation (ISO 11135), and MRI safety/compatibility evaluations per FDA guidance.
Technological Characteristics
Passive hip prostheses. Materials: UHMWPE (ASTM F648), PMMA (Colacryl TS2270). Sterilization: Ethylene Oxide (EtO) or Gas Plasma. Design: Cemented semi-constrained hip joint components. MR Conditional labeling applied. No software or electronic components.
Indications for Use
Indicated for patients requiring total or hemi hip arthroplasty due to painful, disabling joint disease (degenerative/rheumatoid arthritis, avascular necrosis, post-traumatic arthritis), functional deformity, revision of failed prior procedures, or nonunion/fractures of the proximal femur where other techniques are unmanageable.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Trident, Restoration ADM, and MDM X3 UHMWPE Acetabular Inserts (K182468)
Triathlon X3 UHMWPE Tibial Inserts and Patellar Components (K172634)
Exeter Centralizer and Exeter 2.5mm Intramedullary Plug (K191414)
Submission Summary (Full Text)
{0}------------------------------------------------
August 13, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a larger font size than the full name. The full name of the agency is "U.S. Food & Drug Administration".
Howmedica Osteonics Corp Alexandra Kirby Regulatory Affairs Specialist 325 Corporate Drive Mahwah. New Jersey 07430
### Re: K193429
Trade/Device Name: Exeter V40 Femoral Stem, Exeter X3 RimFit Cup Regulation Number: 21 CFR 888.3350 Regulation Name: Hip Joint Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JDI, LZO, JDG, KWY Dated: July 16, 2020 Received: July 17, 2020
#### Dear Alexandra Kirby:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vesa Vuniqi Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K193429
Device Name Exeter® X3® RimFit Cup Exeter® V40™ Femoral Stem
Indications for Use (Describe) Exeter X3 RimFit Cup
The indications for use for total hip arthroplasty include:
· Painful, disabling joint disease of the hip resulting from: degenerative arthritis, post-traumatic arthritis or late stage avascular necrosis.
- · Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
· Clinical management problems where arthrodesis or alternative techniques are less likely to achieve satisfactory results.
· Where bone stock is of poor quality or madequate for other reconstructive techniques, such as cementless fixation, as indicated by deficiencies of the acetabulum.
The Exeter X3 RimFit Cup is intended for cemented use only.
Exeter V40 Femoral Stem
The indications for use for total and hemi hip arthroplasty include:
· noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
• rheumatoid arthritis;
· correction of functional deformity;
· revision procedures where other treatments or devices have failed; and,
· treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The Exeter V40 Femoral Stem is intended for use in total or hemi hip replacement. It is intended for cemented use only.
Type of Use (Select one or both, as applicable)
| <span style="vertical-align:middle;">☑</span> Research Use (Part 21 CFR 201.3 Subject to) | <span style="vertical-align:middle;">☐</span> Over-The-Counter (21 CFR 201.3 Subject to) |
|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Sponsor | Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, NJ 07430 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Alexandra Kirby<br>Regulatory Affairs Specialist<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: (201) 831-5963 |
| Alternate Contact | Krutanjali Shah<br>Manager, Regulatory Affairs<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: (201) 831-5665 |
| Date Prepared: | December 9, 2019 |
| Proprietary Name: | Exeter X3 RimFit Cup<br>Exeter V40 Femoral Stem |
| Common Name: | Total and Hemi Hip Prosthesis |
| Classification Name: | Hip joint metal/polymer semi-constrained cemented prosthesis (21 CFR §888.3350)<br>Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis (21 CFR §888.3353) |
{4}------------------------------------------------
#### K193427 Page 2/7
Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis (21 CFR §888.3360) Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis (21 CFR §888.3390)
Product Codes: JDI, LZO, JDG, KWY
#### Legally Marketed Device to Which Substantial Equivalence is Claimed:
- . Exeter X3 RimFit Cup (K111848)
- Exeter V40 Femoral Stem (K011623, K052718, K110290, K121308, K153345, . K173499)
### Legally Marketed Reference Devices Used to Support Substantial Equivalence:
- Trident®, Restoration ADM®, and MDM® X3 UHMWPE Acetabular Inserts (K182468) 0
- Triathlon® X3 UHMWPE Tibial Inserts and Patellar Components (K172634) 0
- . Exeter Centralizer and Exeter 2.5mm Intramedullary Plug (K191414)
#### Device Description:
The subject Exeter X3 RimFit Cup sterilized by ethylene oxide (EtO) is referred to throughout this submission as the Exeter X3-EtO RimFit Cup. The subject Exeter X3-EtO RimFit Cup is a modified version of the predicate Exeter X3 RimFit Cup and features an alternate terminal sterilization method with EtO, an additional polyethylene resin consolidation method (conventional method which meets the specifications of ASTM F648), an alternate cement spacer material in Acrylic Resin Colacryl® TS2270 PMMA, minor dimensional changes related to flange diameter and blow hole locations, and newly introduced Magnetic Resonance (MR) Conditional labeling. The geometry and design of the subject Exeter X3-EtO RimFit Cup is similar to the predicate Exeter X3 RimFit Cup. There are minor modifications to the flange diameter and blow hole locations between the subject Exeter X3-EtO RimFit Cup and predicate Exeter X3 RimFit Cup. The alternate terminal sterilization method and polyethylene resin consolidation method were previously cleared for Stryker Orthopaedics knee devices in K173849 and Stryker Orthopaedics hip devices in K182468. The alternate cement spacer
{5}------------------------------------------------
material was previously cleared in K191414. The MR Conditional labeling was previously cleared in K171768.
The subject Exeter X3 RimFit Cup sterilized via gas plasma (GP) is referred to throughout this submission as the Exeter X3-GP RimFit Cup and has device designs identical to the predicate Exeter X3 RimFit Cup in terms of design and geometry. In comparison to the predicate Exeter X3 RimFit Cup, the only change to the subject Exeter X3-GP RimFit Cup is that the labeling is updated to add MR Conditional information. The MR Conditional language was previously cleared in K171768.
The subject Exeter V40 Femoral Stem is a modified version of the predicate Exeter V40 Femoral Stem in that the stem component is packaged with two polymethyl methacrylate (PMMA) centralizers that are manufactured from an alternate PMMA material (Acrylic Resin Colacryl® TS2270 PMMA). The alternate PMMA material was previously cleared in K191414. The subject device is identical to the predicate device in terms of design and geometry. The labeling for the subject device maintain the MR Conditional designation that was previously cleared in K171768 and K174399 (certain sizes of Exeter V40 Femoral Stem only).
#### Indications for Use:
The subject Exeter X3 RimFit Cup and Exeter V40 Femoral Stem devices have the same intended use as their respective predicate devices as specified in the 510(k) submissions for the predicate devices and listed as follows.
## Exeter X3 RimFit Cup
The indications for use for total hip arthroplasty include:
- . Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- . Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- . Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
{6}------------------------------------------------
- . Where bone stock is of poor quality or inadequate for other reconstructive techniques. such as cementless fixation, as indicated by deficiencies of the acetabulum.
The Exeter X3 RimFit Cup is intended for cemented use only.
#### Exeter V40 Femoral Stem
The indications for use for total and hemi hip arthroplasty include:
- . noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- rheumatoid arthritis; o
- correction of functional deformity; o
- revision procedures where other treatments or devices have failed; and, ●
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur o with head involvement that are unmanageable using other techniques.
The Exeter V40 Femoral Stem is intended for use in total or hemi hip replacement. It is intended for cemented use only.
# Summary of Technological Characteristics:
The subject Exeter X3-EtO RimFit Cup and Exeter V40 Femoral Stem are identical in terms of intended use, indications for use, and principles of operation to their respective predicate devices (Exeter X3 RimFit Cup and Exeter V40 Femoral Stem, respectively). The subject Exeter X3-EtO RimFit Cup and Exeter V40 Femoral Stem are similar in terms of technological characteristics to the predicate devices. There are noted differences between the subject and predicate devices as described in the Device Description header above. The geometry and design of the subject Exeter X3-EtO RimFit Cup is similar to the predicate Exeter X3 RimFit Cup. There are minor modifications to the flange diameter and blow hole locations between the subject Exeter X3-EtO RimFit Cup and predicate Exeter X3 RimFit Cup.
The subject Exeter X3-GP RimFit Cup is identical in terms of intended use, indications for use, and principles of operation to the predicate Exeter X3 RimFit Cup. The predicate Exeter X3
{7}------------------------------------------------
RimFit Cup has been cleared in K111848. The change to add MR Conditional labeling to the existing Exeter X3 RimFit Cup does not impact the technological characteristics. The subject device has the same design and is manufactured from the same materials as the predicate Exeter X3 RimFit Cup.
#### Non-Clinical Testing:
The following non-clinical laboratory testing was performed to determine substantial eequivalence of the subject devices:
#### Exeter X3-EtO RimFit Cup
- 1) Material testing per ASTM F648, ASTM F2565, and FDA Guidance, "Characterization of Ultrahigh Molecular Weight Polyethylene (UHMWPE) Used in Orthopedic Devices" (April 26, 2019).
- 2) Biocompatibility Evaluation per ISO 10993-1, ISO 10993-7, and FDA Guidance, "Use of International Standard ISO 10993-1, 'Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process' (June 16, 2016)
- 3) Ethylene oxide sterilization validation per ISO 11135:2014
- 4) Bacterial endotoxin testing (BET)
- 5) MRI Analysis per the FDA guidance document "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment - Guidance for Industry and FDA Staff," dated December 11, 2014 was conducted to characterize the compatibility of Stryker Orthopaedics total hip passive implants in the MR environment. FDA guidance "Assessment of Radiofrequency-Induced Heating in the Magnetic Resonance (MR) Environment for Multi-Configuration Passive Medical Devices – Draft Guidance for Industry and FDA Staff", dated March 22, 2016 was also consulted for the heating evaluations performed.
- 6) Accelerated Aging Studies of PMMA coupons
- 7) Effect of Cleaning and Sterilization on the Dimensional Stability of PMMA
#### Exeter X3-GP RimFit Cup
{8}------------------------------------------------
- 1) MRI Analysis per the FDA guidance document "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment - Guidance for Industry and FDA Staff," dated December 11, 2014 was conducted to characterize the compatibility of Stryker Orthopaedics total hip passive implants in the MR environment. FDA guidance "Assessment of Radiofrequency-Induced Heating in the Magnetic Resonance (MR) Environment for Multi-Configuration Passive Medical Devices – Draft Guidance for Industry and FDA Staff", dated March 22, 2016 was also consulted for the heating evaluations performed.
#### Exeter V40 Femoral Stem
- 1) Centralizer Insertion
- 2) Centralizer Retention
- 3) Winged Centralizer Wing Flexibility
- 4) Accelerated Aging Studies of PMMA coupons
- 5) MRI Analysis per the FDA guidance document "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment - Guidance for Industry and FDA Staff," dated December 11, 2014 was conducted to characterize the compatibility of Stryker Orthopaedics total hip passive implants in the MR environment. FDA draft guidance "Assessment of Radiofrequency-Induced Heating in the Magnetic Resonance (MR) Environment for Multi-Configuration Passive Medical Devices – Draft Guidance for Industry and FDA Staff," dated March 22, 2016 was also consulted for the heating evaluations performed.
#### Clinical Testing:
Clinical testing was not required as a basis for substantial equivalence.
#### Conclusion:
The subject Exeter X3 RimFit Cup and Exeter V40 Femoral Stem are substantially equivalent to the respective predicate devices identified in this premarket notification. Device comparison showed that the proposed devices are substantially equivalent in intended use, technological
{9}------------------------------------------------
characteristics, and principles of operation as compared to the respective predicate devices. The proposed modifications do not affect safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.