The hip system devices included in this submission are sterile, single-use devices intended for use in primary and revision total hip arthroplasty to alleviate pain and restore function.
Device Story
Hip system includes femoral heads, acetabular shells, acetabular inserts, modular necks, and modular stems for hip arthroplasty. Devices are sterile, single-use, and implanted by orthopedic surgeons in clinical settings. Components are press-fit or cemented into the proximal femur and acetabulum to replace diseased or damaged joint surfaces, alleviating pain and restoring function. This submission specifically updates labeling to include safety information regarding modular hip stem taper junctions and harmonizes language across Howmedica Osteonics' hip stem labeling. No changes to device design, materials, or operational principles were made; devices remain substantially equivalent to previously cleared systems.
Clinical Evidence
No clinical data provided. This is a labeling-only submission for previously cleared devices.
Technological Characteristics
Modular hip arthroplasty components including femoral heads, acetabular shells, inserts, necks, and stems. Materials include metal, polymer, and ceramic combinations. Designed for cementless (press-fit) or cemented fixation. Sterile, single-use. No software or electronic components.
Indications for Use
Indicated for patients requiring total or hemi-hip replacement due to noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, failed prior treatments, or proximal femur fractures (nonunion, femoral neck, trochanteric) with head involvement. Specific stems (Hipstar) indicated for high-risk dislocation patients (prior dislocation, bone loss, laxity, neuromuscular disease, instability) when used with constrained liners. Restoration Modular system indicated for primary/revision arthroplasty and severe proximal bone loss. Contraindications: None explicitly listed, though specific systems are cementless only.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Submission Summary (Full Text)
{0}------------------------------------------------
K121308 (pg 1/3)
#### 510(k) Summary
Hip Systems
JUL 30 2012
Proprietary Name:
Common Name:
Classification Names And References:
Artificial Hip Replacement Components -Acetabular and Femoral
Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis21 CFR §888.3358
Hip joint metal/polymer constrained cemented or uncemented prosthesis.21 CFR §888.3310
Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis 21 CFR §888.3353
Hip joint metal/polymer semi-constrained cemented prosthesis 21 CFR \$888.3350
Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis 21 CFR 888.3360
Hip joint femoral (hemi-hip) metal/polymer cemented or uncemented prosthesis 21 CFR §888.3390
#### Proposed Regulatory Class: Class II
Product Codes:
LPH - prosthesis, hip, semi-constrained, metal/polymer, porous uncemented
KWZ - prosthesis, hip, constrained, cemented or uncemented, metal/polymer
LZO - prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented
MEH - prosthesis, hip. semi-constrained. uncemented. metal/polymer, non-porous
JDI - prosthesis, hip, semi-constrained, metal/polymer, cemented
LWJ - prosthesis, hip, semi-constrained, metal/polymer, uncemented
KWL - prosthesis, hip, hemi-, femoral, metal
MAY - prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous cemented, osteophilic finish
KWY - prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented
MBL- prosthesis, hip. semi-constrained, uncemented, metal/polymer, porous
{1}------------------------------------------------
K121308 (pg 2/3)
For Information contact:
Karen Ariemma, RAC Senior Strategic Regulatory Affairs Manager Howmedica Osteonics Corp. 325 Corporate Drive Mahwah. NJ 07430 Phone: (201) 831-5718 Fax: (201) 831-4718
Date Prepared:
July 29, 2012
## Description:
The devices included in this submission are femoral heads, acetabular shells, acetabular inserts, modular necks, modular stems and accessory components used in hip arthroplasty procedures. All devices have been previously deemed substantially equivalent in prior 510(k) submissions and are commercially available. The purpose of this submission is to modify the labeling of these devices to include safety information in the Instructions for Use (IFUs) regarding modular hip stem junctions. In addition to this specific update, general revisions have been made to the IFUs to harmonize the language between the Howmedica Osteonics' hip stem labeling
#### Intended Use:
The hip system devices included in this submission are sterile, single-use devices intended for use in primary and revision total hip arthroplasty to alleviate pain and restore function.
### Indications:
The overall indications for use for the subject total and hemi hip replacement prostheses include:
- 1. noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis:
- 2. rheumatoid arthritis:
- correction of functional deformity; 3.
- 4. revision procedures where other treatments or devices have failed; and,
- 5. treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
The Rejuvenate Hip System is intended for cementless use only.
## Additional indications specific to the Hipstar Stem:
The Hipstar Femoral Stem is intended for cementless use only.
When mated with a constrained acetabular liner the Hipstar Stem is indicated for use in primary and revision total hip arthroplasty for patients at high risk of hip dislocation due to a history of prior dislocations, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability. t
{2}------------------------------------------------
K121308(pg 3/3)
# Additional indications specific to the Restoration Modular Hip System:
The Restoration Modular Hip System is intended for primary and revision total hip arthroplasty as well as in the presence of severe proximal bone loss. These femoral stems are designed to be press fit into the proximal femur.
## Substantial Equivalence:
The subject hip system devices have been deemed substantially equivalent to other commercially available hip arthroplasty systems based upon an evaluation of intended use, design, materials, and operational principles in prior 510(k) submissions. The purpose of this submission is solely to revise the labeling to add safety information regarding modular hip stem taper junctions.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal for the Department of Health & Human Services USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. In the center of the seal is a stylized symbol that resembles an abstract human figure or a caduceus, with three curved lines representing the body and a single line representing the head.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUL 30 2012
Howmedica Osteonics Corp. % Ms. Karen Ariemma Senior Strategic Regulatory Affairs Manager 325 Corporate Drive Mahwah, New Jersey 07430
Re: K121308
Trade/Device Name: Hip Systems Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Regulatory Classi Class Class 22, LZO, MEH, JDI, LWJ, KWL, MAY, KWY, MBL Dated: April 30, 2012 Received: May 1, 2012
Dear Ms. Ariemma:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encrosule) (o regars) to regard) (and in the Medical Device Amendments, or to conninered phor to May 20, 1978, the eccordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). alle Cosment Act (Act) that to nev request to the general controls provisions of the Act. The r ou may, merciole, market the device, occurements for annual registration, listing of general controls provisions of tactice, labeling, and prohibitions against misbranding and devices, good mananataring practices, and evaluate information related to contract liability adulteration. Ticase note: "ODIC assess that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (See above) into ting major regulations affecting your device can be filly be subject to additional controls. "Enological of the 200 to 898. In addition, FDA may found in the Code of I cacrai resguing your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualled of a bacession with other requirements of the Act
that FDA has made a determination that your device complies with other must that FDA has made a determination as administered by other Federal agencies. You must
{4}------------------------------------------------
Page 2 – Ms. Karen Ariemma
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Of it Fart 607); accemes (21 CFR 803); good manufacturing practice requirements as set de rior related davers over (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 Jou atttp://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note to for questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the I ou may oounn other geficitional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.lftm.
Sincerely yours,
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
510(k) Number (if known): K121308 (pg 1/1)
Device Name: Stryker Hip Systems
Indications for Use:
The overall indications for use for the subject total and hemi hip replacement prostheses include:
- 1. noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis;
- 2. rheumatoid arthritis;
- correction of functional deformity; 3.
- revision procedures where other treatments or devices have failed; and, 4.
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur ડ. with head involvement that are unmanageable using other techniques.
The Rejuvenate Hip System is intended for cementless use only.
# Additional indications specific to the Hipstar Stem:
The Hipstar Femoral Stem is intended for cementless use only.
When mated with a constrained acetabular liner the Hipstar Stem is indicated for use in primary w len nialed with a combroplasty for patients at high risk of hip dislocation due to a history of prior dislocations, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.
# Additional indications specific to the Restoration Modular Hip System:
The Restoration Modular Hip System is intended for primary and revision total hip arthroplasty as well as in the presence of severe proximal bone loss. These femoral stems are designed to be press fit into the proximal femur.
(Division Sign-Oft)
Division of Surgical, Orthopedic, and Restorative Devices
Over-The-Counter Use X Prescription Use ANDAPO(k) Number LCFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.