The Exeter X3 RimFit Acetabular Cup is a sterile, single-use device intended for use in primary and revision total hip arthroplasty to alleviate pain and restore function. This device is intended to be used with any currently available compatible Howmedica Osteonics' acetabular components. Compatibility with the femoral heads includes: V40 and C-Taper (LFITM, CoCr, Biolox delta, Alumina, and Orthinox).
Device Story
Exeter X3 RimFit Acetabular Cup is a sterile, single-use orthopedic implant for primary and revision total hip arthroplasty. Manufactured from modified sequentially crosslinked and annealed X3 UHMWPE. Designed for cemented fixation in the acetabulum. Used by orthopedic surgeons in clinical/hospital settings to replace diseased hip joints, alleviate pain, and restore function. Compatible with various Howmedica Osteonics femoral heads (V40, C-Taper, CoCr, Biolox delta, Alumina, Orthinox). Device provides a bearing surface for the femoral head component. Clinical benefit includes improved joint mobility and pain reduction for patients with degenerative or traumatic hip conditions.
Clinical Evidence
No clinical data. Substantial equivalence established via non-clinical bench testing, including lever-out post fatigue testing and pull-out testing of cement spacers, per FDA guidance.
Technological Characteristics
Acetabular cup manufactured from X3 UHMWPE (modified sequentially crosslinked and annealed). Designed for cemented fixation. Compatible with V40 and C-Taper femoral heads. Evaluated via Finite Element Analysis and mechanical bench testing (lever-out fatigue, pull-out strength).
Indications for Use
Indicated for patients with painful, disabling hip joint disease (degenerative, rheumatoid, post-traumatic arthritis, or late-stage avascular necrosis), revision of unsuccessful prior hip procedures, or clinical management where alternative techniques are less satisfactory. Indicated for patients with poor bone stock inadequate for cementless fixation. Intended for cemented use only.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
Trident All Poly Cup
Trident X3 Acetabular Insert
Exeter Hip System With V40 Taper
Opera Cup (Smith & Nephew)
Submission Summary (Full Text)
{0}------------------------------------------------
# KIII 848
| Proprietary Name: | Exeter X3 RimFit Acetabular Cup |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Hip prosthesis |
| Classification Name and Reference: | Hip joint metal/polymer semi-constrained cemented<br>prosthesis 21 CFR §888.3350 |
| Regulatory Class: | Class II |
| Product Codes: | 87 JDI - prosthesis, hip, semi-constrained, metal/polymer,<br>cemented |
| For Information contact: | Estela Celi, Regulatory Affairs Associate<br>Howmedica Osteonics Corp.<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Phone: (201) 831-6461 Fax: (201) 831-3461 |
| Date Prepared: | August 16, 2011 |
### 510(k) Summary of Safety and Effectiveness
#### Description:
This Traditional 510(k) submission is being supplied to the U.S. FDA to provide authorization to market the new Exeter X3 RimFit Acetabular Cup manufactured from a modified sequentially crosslinked and annealed Ultra High Molecular Weight Polyethylene (UHMWPE) material. The proprietary name of the subject polyethylene material is X3® UHMWPE.
### Intended Use:
The Exeter X3 RimFit Acetabular Cup is a sterile, single-use device intended for use in primary and revision total hip arthroplasty to alleviate pain and restore function. This device is intended to be used with any currently available compatible Howmedica Osteonics' acetabular components. Compatibility with the femoral heads includes: V40 and C-Taper (LFITM, CoCr, Biolox delta, Alumina, and Orthinox).
Page 1 of 2
OCT - 5 2011
{1}------------------------------------------------
# KIII 848
### Indications:
The indications for use of total hip replacement prostheses include:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, ● rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or . other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive . techniques are less likely to achieve satisfactory results.
- · Where bone stock is of poor quality or inadequate for other reconstructive techniques, such as cementless fixation, as indicated by deficiencies of the acetabulum
Stryker's Exeter X3 RimFit Acetabular Cup is intended for cemented use only.
# Summary of Technologies:
Device comparison showed that the proposed device is substantially equivalent in intended, use, materials and performance characteristics to the predicate device. The predicate devices used for comparison to the proposed device are the following; The Trident All Poly Cup, Trident X3 Acetabular Insert, Exeter Hip System With V40 Taper and the Opera Cup (Smith & Nephew).
## Non-Clinical Testing:
Non-clinical laboratory testing was performed for the acetabular cup to determine substantial equivalence. Non-clinical testing was provided as outlined in the FDA Guidance Document entitled "Class II Special Controls Guidance Document: Hip Joint Metal/Polymer Constrained Cemented or Uncemented Prosthesis (April 30, 2002)". Lever-out post fatigue testing was conducted on the worst-case size determined by Finite Element Analysis. Pull-out testing was performed on the cement spacers in order to evaluate their mechanical strength. The testing demonstrated that the Exeter X3 RimFit Acetabular Cup is substantially equivalent to devices currently cleared for marketing.
Clinical Testing: Clinical testing was not required for this submission.
Conclusion: The Exeter X3 RimFit Acetabular Cup is substantially equivalent to the predicate devices identified in this premarket notification.
Page 2 of 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling three human figures connected by flowing lines, creating a sense of unity and movement.
Food and Drug Administration 19903 New Hampshire Avenue Document Control Room -- W()66-G609 Silver Spring, MD 20903-0002
Howmedica Osteonics Corporation % Ms. Estela Celi Regulatory Affairs Associate 325 Corporate Drive Mahwah. New Jersev 07430
- 5 2011
Re: K111848
Trade/Device Name: X3 RimFit Acetabular Cup Regulation Number: 21 CFR 888.3350 Regulation Name: Hip joint metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JDI Dated: August 16. 2011 Received: August 17, 2011
Dear Ms. Celi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Estela Celi
forth in the quality systems (QS) regulation (21 CFR Part,820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/Reportal?roblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.hum.
Sincerely yours.
for Tho
Nro. Luv
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known): K111848
Device Name: Exeter X3 RimFit Acetabular Cup
Indications for Use:
The indications for use of the total hip arthoplasty include:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, . rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or . other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive . techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or inadequate for other reconstructive techniques, . such as cementless fixation, as indicated by deficiencies of the acetabulum
Stryker's Exeter X3 RimFit Acetabular Cup is intended for cemented use only.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mitchel Thomas for MXM
(Division Sign-Oft) Division of Surgical, Orthopedic, and Resiorative Devices
510(k) Number K111848
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.