In general, these devices are intended for use in primary or revision hip arthroplasty. Specific indications appear below: [List of specific indications for various components including Trident shells, femoral stems, liners, and accessories].
Device Story
Stryker Orthopaedics Total Hip Systems comprise various femoral stems, acetabular shells, liners, femoral heads, and fixation accessories. These passive orthopedic implants replace diseased or damaged hip joints. The system is used in surgical settings by orthopedic surgeons to perform total or hemi-hip arthroplasty. The devices provide structural support and articulation for the hip joint, restoring mobility and reducing pain. This submission updates labeling to include MR compatibility information and removes a contraindication for selected devices; no changes were made to the device designs, materials, or manufacturing processes.
Clinical Evidence
No clinical data was required or presented; substantial equivalence is based on non-clinical bench testing for MR compatibility.
Technological Characteristics
Passive orthopedic implants (femoral stems, acetabular shells, liners, heads, screws). Materials and design remain unchanged from predicates. MR compatibility testing performed per ASTM F2052-06/14, ASTM F2213-06, ASTM F2119-07, and ASTM F2182-11a.
Indications for Use
Indicated for patients with painful, disabling hip joint disease (degenerative, rheumatoid, post-traumatic arthritis, avascular necrosis), revision of failed previous procedures, or clinical management problems where arthrodesis/reconstruction is less likely to succeed. Also indicated for femoral neck/trochanteric fractures, nonunions, and dislocation risks in primary or revision patients.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing to the right, arranged in a way that they appear to be interconnected. The profiles are rendered in a simple, line-art style. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 25, 2016
Stryker Orthopaedics Ms. Margaret Klippel Principal Regulatory Affairs Specialist 325 Corporate Drive Mahwah, New Jersey 07430
Re: K153345 Trade/Device Name: Stryker Orthopaedics Total Hip Systems Labeling Update Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis. Regulatory Class: Class II Product Code: LPH, KWZ, LZO, MAY, MEH, JDI, KWL, LWJ, KWY, HWC, MBL Dated: April 20, 2016 Received: April 22, 2016
Dear Ms. Klippel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## 510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Trident PSL Shell, Trident Tritanium Shell, Trident Hemispherical Shell
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
- o Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques ● are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
The Trident shells are intended for cementless fixation within the prepared acetabulum.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
## 510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Tritanium Acetabular Shell System
- . Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other ● procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques ● are less likely to achieve satisfactory results.
- . Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
The Tritanium Acetabular Shell System is intended for cementless use only.
Over-The-Counter Use Prescription Use AND/OR X (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
## 510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Accolade II Femoral Stem, Secur-Fit Advanced Femoral Stem, Anato Femoral Stem
The indications for use for total hip arthroplasty with stems include:
- . noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis:
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and,
- nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Additional indication specific to use of the Femoral Stem with compatible Howmedica Osteonics Constrained Liners:
- When the stem is to be used with compatible Howmedica Osteonics Constrained Liners. ● the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
Accolade II, Secur-Fit Advanced and Anato Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{5}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Exeter® V40™ Hip System (includes Orthinox V40 Femoral heads)
The Exeter® V40™ Hip System is intended for use in total hip replacement. It is intended for cemented use only.
The Exeter® Hip is indicated for:
- noninflammatory degenerative joint disease, including osteoarthritis and avascular ● necrosis;
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and, ●
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{6}------------------------------------------------
## 510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Restoration Anatomic Shell
Indications for Use
- . Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, posttraumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- . Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques ● as indicated by deficiencies of the acetabulum.
When used with MDM Liners
- o Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
- Dislocation risks ●
When used with Constrained Liner:
- The Trident Constrained Acetabular Insert is indicated for use in primary and revision o patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
The Restoration® Anatomic Shell is indicated for cementless use only.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{7}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Omnifit HFX Femoral Stems
## Indications:
For use as a Bipolar Hip Replacement:
- Femoral head/neck fractures or non-unions. ●
- Aseptic necrosis of the femoral head. ●
- Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular . involvement or distortion.
Other Considerations:
- Pathological conditions or age considerations which indicate a more conservative acetabular ● procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty.
- Femoral neck fractures.
For use as a Total Hip Replacement:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid ● arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other . procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are ● less likely to achieve satisfactory results.
- Clinical circumstances which require an altered femoral resection level due to a proximal o fracture, bone loss or calcar lysis.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{8}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Omnifit EON Cemented Femoral Stems
## Indications
For use as a Bipolar Hip Replacement:
- . Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head. ●
- o Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
## Other Considerations:
- Pathological conditions or age considerations which indicate a more conservative acetabular ● procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty. ●
For use as a Total Hip Replacement:
- o Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are ● less likely to achieve satisfactory results.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{9}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Secur-Fit Max and Secur-Fit Max Plus Hip Stems
Secur-Fit Max and Secur-Fit Max Plus Hip stems are single use devices and are intended for cementless fixation within the prepared femoral canals of patients requiring hip arthroplasty.
Indications for use as a Total Hip Replacement include:
- noninflammatory degenerative joint disease, including osteoarthritis and avascular ● necrosis;
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and, ●
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{10}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Accolade C Femoral Stems
The Accolade C Femoral Stems are single-use devices intended for cemented fixation.
Indications for use as a Total Hip Replacement include:
- noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- correction of functional deformity; ●
- revision procedures where other treatments or devices have failed; and, ●
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{11}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Accolade HFx Femoral Stems
- noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and, ●
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur ● with head involvement that are unmanageable using other techniques.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{12}------------------------------------------------
## 510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Accolade TMZF and Accolade TMZF Plus Femoral Stems
The subject hip stem is a single-use device intended for use in total hip replacement. It is intended for the reconstruction of the head and neck of the femoral joint. This hip stem is intended for primary reconstruction of the proximal femur or revision total hip arthroplasty. This device is intended for use with any currently available Howmedica Osteonics acetabular component and V40™ femoral heads that can be mated with a TMZF 5° 40' trunnion.
## Indications:
- Cementless primary hip surgery in cases of non-inflammatory degenerative joint disease including osteoarthritis, avascular necrosis, rheumatoid arthritis, and correction of functional deformity.
- Treatment of nonunion, and femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
- Revision procedures where other treatments or devices have failed.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 807 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{13}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Restoration Acetabular Wedge Augments
Indications for Use:
The indications for use of the Restoration Acetabular Wedge Augments:
General Indications for Total Hip Replacement Components:
- . Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other ● procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Indications Specific to the Acetabular Wedges:
- As an alternative to structural allograft in cases of superior and superior/posterior segmental o acetabular deficiencies.
Acetabular Augments are intended for cementless use only to the bone interface, and are affixed to the mating cup using bone cement
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{14}------------------------------------------------
## 510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Trident AD Acetabular shells
Indications for Use:
The Trident AD Acetabular shells are single use devices. The shell is intended for cemented or cementless fixation within the prepared acetabulum. The Trident AD Acetabular Component System is compatible with any appropriately selected Howmedica Osteonics hip stemoral head combination.
Indications:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{15}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Trident T Acetabular shells
Indications for Use:
The Trident Acetabular shells are single-use devices intended for cementless fixation within the prepared acetabulum. They are compatible with Trident polyethylene acetabular bearing insert. If additional fixation is desired, the dome screw holes, if present, have been designed to accept Stryker Orthopaedics 6.5mm or 5.5mm bone screws.
General Indications for Total Hip Replacement Components:
- o Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{16}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Restoration ADM System, Modular Dual Mobility (MDM) Liner and X3 Acetabular Insert
Indications for Use:
The indications for use of the total hip arthroplasty prostheses include:
- Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis;
- Correction of functional deformity; ●
- Revision procedures where other treatments or devices have failed:
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur ● with head involvement that are unmanageable using other techniques.
- . Dislocation risks.
These devices are intended for cementless use only
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{17}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Acetabular Dome Hole Plug
The Dome Hole Plug is an optional device which is available to seal the Howmedica Osteonics Acetabular Shell components during cemented or cementless applications of the acetabular cup. The Howmedica Osteonics Done Hole Plug is threaded into the dome hole of the shell.
## Indications
- . In cemented or cementless hip arthroplasty, when an acetabular shell plug is thought to be advantageous.
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{18}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Trident X3, Trident Crossfire, Trident N2Vac and Trident X3/Crossfire Elevated Rim Acetabular Liners
Indications for Use:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are ● less likely to achieve satisfactory results.
- . Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{19}------------------------------------------------
## 510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Trident® Constrained Acetabular Insert
Indications for Use:
The Trident® Constrained Acetabular Insert is intended for use as a component of a total hip prosthesis in primary or revision patients at a high risk of hip dislocation due to a history of dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{20}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
BIOLOX Delta Ceramic Heads (C-Taper, Universal Taper, V40 to Universal Taper Adapter Sleeve, and C-Taper to Universal Taper Adapter Sleeve)
The femoral heads are intended for mechanical fixation to their mating hip stems, and can be used in cemented or cementless hip arthroplasty procedures.
For Use as a Total Hip Replacement:
- Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid ● arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous cup arthroplasty or other procedures
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions. ●
- Aseptic necrosis of the femoral head. ●
- Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular ● involvement or distortion.
- Pathological considerations or age considerations which indicate a more conservative ● acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty ●
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{21}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
V40 BIOLOX Delta Ceramic Heads
The femoral heads are intended for mechanical fixation to their mating hip stems, and can be used in cemented or cementless hip arthroplasty procedures.
For Use as a Total Hip Replacement:
- Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid ● arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous cup arthroplasty or other procedures
- Clinical management problems where arthrodesis or alternative reconstructive techniques are ● less likely to achieve satisfactory results.
For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions. ●
- Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular o involvement or distortion.
Other considerations:
- Pathological considerations or age considerations which indicate a more conservative ● acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- o Salvage of failed total hip arthroplasty
Over-The-Counter Use Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{22}------------------------------------------------
## 510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Alumina C-Taper Ceramic Heads
## For Use as a Total Hip Replacement:
- Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage vascular necrosis.
- Revision of previous cup arthroplasty or other procedures
- . Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
## For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions. ●
- Aseptic necrosis of the femoral head. ●
- Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
## Other Considerations:
- Pathological considerations or age considerations which indicate a more conservative ● acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{23}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Alumina V40 Ceramic Heads
Indications:
- Painful, disabling joint disease of the hip resulting from: Non-inflammatory degenerative arthritis (osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, pelvic fracture, failed fracture fixation, or diastrophic variant);
- Rheumatoid arthritis ●
- Correction of functional deformity
- Where bone stock is of poor quality or inadequate for other reconstructive techniques as indicated by the deficiencies of the acetabulum;
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques;
- Revision procedures where other treatments or devices have failed.
Over-The-Counter Use Prescription Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
{24}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
C-Taper CoCr Femoral Heads (LFIT and non-LFIT)
For use as a Bipolar Hip Replacement:
- l. Femoral head/neck fractures or non-unions
- 2. Aseptic necrosis of the femoral head.
- 3. Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
Other Considerations:
- Pathological conditions or age considerations which indicate a more conservative 1. approach to the acetabulum and the avoidance of the use of bone cement in the acetabulum.
- 2. Salvage of failed total hip arthroplasty.
Indications for use as part of a Total Hip Replacement include:
- l. Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- 2. Revision of previous unsuccessful femoral head replacement, cup arthroplasty, or other procedure.
- 3. Clinical management situations where arthrodesis or alternate reconstructive techniques are less likely to achieve satisfactory results.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
{25}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
V40 CoCr Femoral Heads (LFIT and non-LFIT)
Indications for Use:
- Non-inflammatory degenerative joint disease, including osteoarthritis and avascular . necrosis;
- Rheumatoid arthritis
- Correction of functional deformity
- Revision procedures where other treatments or devices have failed
- o Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{26}------------------------------------------------
### 510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Unitrax Endoprosthesis, Unitrax V40 Adapter Sleeve, Unitrax C-Taper Adapter Sleeve
The Howmedica Osteonics Unitrax Endoprosthesis, and the Unitrax V40 Modular Adaptor sleeves are used as a hemiarthroplasty device for the following indications: femoral neck fractures, idiopathic avascular necrosis, and non-unions. The Unitrax C-taper sleeves are intended for use as a Hemi-Hip Replacement with the following indications: femoral head/neck fractures or non-unions, aseptic necrosis of the femoral head/neck and osteo- and post traumatic arthritis. The patient's acetabular bone stock must be adequate to support articulation with the head of the endoprosthesis.
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{27}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Universal Distal Spacer
Howmedica Osteonics Corp.'s accessory products for cemented arthroplasty are optional devices intended to assist in the preparation, implantation and/or positioning of a femoral implant intended for cemented application.
## Indications
For cement spacers, mid-shaft restrictors and Cement Plugs:
- In cemented hip arthroplasty, when the cement spacer, restrictor and/or plug is thought to be . advantageous.
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{28}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
Torx Screws, GAP Plate Screws and Osteolock Bone Screws
- Howmedica Osteonics Torx Cancellous Bone Screws are intended for supplemental fixation ● of associated Howmedica Osteonics cementless Acetabular Shells or Howmedica Osteonics Tibial Tray components.
- Howmedica Osteonics Osteolock Bone Screws are intended for supplemental fixation of ● associated cementless Trident Tritanium Multihole Acetabular shells, Restoration Acetabular Augments, and Restoration Anatomic Shell.
- Howmedica Osteonics Restoration GAP Plate Screws are intended only for fixation of the ● dome and iliac plates of the associated Howmedica Osteonics Restoration GAP Acetabular Shells, Trident Tritanium Hemispherical Multihole Acetabular Shells, Restoration Acetabular Augments and Restoration Anatomic Shell.
Prescription Use X Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{29}------------------------------------------------
510(k) Number (if known): K153345
Device Name: Stryker Orthopaedics Total Hip System Labeling Update
V40 to C-Taper Adapter Sleeve
## For Use as a Total Hip Replacement
- Painful disabling joint disease of the hip resulting from: degenerative arthritis. o rheumatoid arthritis, post-traumatic arthritis or late stage vascular necrosis.
- Revision of previous cup arthroplasty or other procedures. ●
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
## For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions. ●
- Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular ● involvement or distortion.
## Other Considerations:
- . Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty. ●
Prescription Use Over-The-Counter Use X AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
{30}------------------------------------------------
## 510(k) Summary
| Sponsor | Stryker Orthopaedics<br>325 Corporate Drive<br>Mahwah, NJ 07430 |
|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Margaret Klippel<br>Senior Regulatory Affairs Project Manager<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Telephone: 201-831-5559<br>Fax: 201-831-4559 |
| Alternate Contact | Patricia Setti-LaPerch<br>Manager, Regulatory Affairs<br>Howmedica Osteonics Corp<br>325 Corporate Drive<br>Mahwah, NJ 07430<br>Telephone: 201-831-5938<br>Fax: 201-831-4938 |
| Date Prepared: | May 24, 2016 |
| Proprietary Name: | Stryker Orthopaedics Total Hip Systems Labeling Update |
| Common Name: | Artificial Hip Replacement Components -- Acetabular and Femoral |
| Classification Name:<br>And References: | Hip joint metal/polymer/metal semi-constrained<br>porous-coated uncemented prosthesis 21 CFR §888.3358 |
| | Hip joint metal/polymer constrained cemented<br>or uncemented prosthesis 21 CFR §888.3310 |
| | Hip joint metal/ceramic/polymer semi-constrained cemented or<br>nonporous uncemented prosthesis 21 CFR §888.3353 |
| | Hip joint metal/polymer semi-constrained cemented prosthesis 21<br>CFR §888.3350 |
| | Hip joint femoral (hemi-hip) metallic cemented or uncemented<br>prosthesis 21 CFR §888.3360 |
| | Hip joint femoral (hemi-hip) metal/polymer cemented or<br>uncemented prosthesis 21 CFR §888.3390 |
| | Smooth or threaded metallic bone fixation fastener 21 CFR<br>§888.3040 |
| Product Codes: | LPH - prosthesis, hip, semi-constrained, metal/polymer, porous<br>uncemented<br>KWZ - prosthesis, hip, constrained, cemented or uncemented,<br>metal/polymer<br>LZO - prosthesis, hip, semi-constrained, metal/ceramic/polymer,<br>cemented or non-porous, uncemented<br>MEH - prosthesis, hip, semi-constrained, uncemented,<br>metal/polymer, non-porous<br>JDI - prosthesis, hip, semi-constrained, metal/polymer, cemented<br>LWJ - prosthesis, hip, semi-constrained, metal/polymer, |
| | uncemented<br>KWL - prosthesis, hip, hemi-, femoral, metal<br>KWY - prosthesis, hip, hemi-, femoral, metal/polymer, cemented<br>or uncemented<br>MAY - prosthesis, hip, semi-constrained, metal/ceramic/polymer,<br>cemented or non-porous cemented, osteophilic finish<br>MBL-prosthesis, hip, semi-constrained, uncemented,<br>metal/polymer, porous<br>HWC - screw fixation bone |
{31}------------------------------------------------
## Legally Marketed Device to Which Substantial Equivalence is Claimed:
| Femoral Stems | |
|-----------------------------------------------------------------------|---------------------------------------|
| Exeter Femoral Stems (V40 taper –<br>includes Orthinox femoral heads) | K011623, K031730, K110290,<br>K121308 |
| SecurFit Max Femoral Stems | K041170, K051738 |
| SecurFit Max Plus Femoral Stems | |
| Accolade C Femoral Stems | K022555, K121308 |
| Accolade TMZF Femoral Stems | K994366, K020572 |
| Accolade TMZF Plus Femoral Stem | K994366, K023102 |
| Accolade HFx Femoral Stem | K051741, K121308 |
| Accolade II Femoral Stems | K103479, K120578, K143085 |
| SecurFit Advanced Femoral Stems | K122853, K143085 |
| Anato Femoral Stems | K123604, K143085 |
| OmniFit HFx Femoral Stems | K031744 |
| OmniFit EON Femoral Stems | K983226 |
| Acetabular Shells/Augments/Plug | |
| Trident Hemispherical Shells (multiple<br>configurations) | K013676, K143085 |
| Tritanium Shells (multiple styles) | K081171, K143085 |
| Restoration Anatomic Acetabular Shell | K142462, K151264 |
| Tritanium Multi-Hole Shell | K010170, K143085 |
| Trident AD Acetabular Shell (multiple styles) | K983502 |
| Trident HA PSL Shell (multiple styles) | K983382, K143085 |
| Trident T Shell | K040412 |
| Restoration ADM Acetabular Shell | K072020 |
| Restoration Acetabular Augments | K102019 |
| Acetabular Dome Hole Plug | K920868 |
| | K942809 |
| Acetabular Liners | |
| MDM Liners | K103233 |
| ADM/MDM X3 inserts | K093644 |
| ADM/MDM Duration Inserts, OD 42mm-52mm | K072020 |
| Trident X3 Liners (multiple sizes) | K033716 |
| | K061434 |
| | K062419 |
| Trident Crossfire Liners (multiple sizes) | K983502 |
| | K991952 |
| | K021911 |
| | K020497 |
| Trident Constrained Liner | K061654 |
| Trident All Polyethylene Constrained Liner | |
| Trident 10 degree Constrained Liner | P960047 |
| Trident Liners N2/Vac(multiple sizes) | K983502 |
| Femoral heads and Adapter Sleeves | |
| BIOLOX Delta Heads (Various sizes, styles) | K052718 |
| | K070885 |
| | K041940 |
| | K051588 |
| Co Cr Heads (various styles/sizes) | K993601 |
| | K010757, K022077, K061434 |
| | K900836, K993601, K021310 |
| | K910988, K061434 |
| | K121308 |
| Alumina C-Taper and V40 Heads (various sizes) | K023901, K971409 |
| | K003391, K991952 |
| Unitrax Endo Head (38, 40-56, 58, 61) | K902365, K014226 |
| Unitrax V40 Adapter Sleeve | K954077, K992570 |
| Unitrax C-Taper Adapter Sleeve | K121308 |
| Adapter Sleeves (various styles) | K003379 |
| | K051737 |
| | K070885 |
| Bone Screws and Distal Spacer | |
| Torx Screws | K894124 |
| GAP Plate Screws | K943549 |
| Osteolock Bone Screws | K903362 |
| Universal Distal Spacer | K914406 |
{32}------------------------------------------------
{33}------------------------------------------------
## Device Description:
All of the subject devices have been found substantially equivalent in previous premarket indications. The purpose of this submission is to modify the labeling to add MR compatibility to the labeling for the devices indicated above, and to remove a contraindication for selected devices. There have been no changes made to the devices requiring 510(k) clearance - only the labeling is being modified.
## Intended Use:
In general, these devices are intended for use in primary or revision hip arthroplasty. Specific indications appear below:
## Trident PSL Shell, Trident Tritanium Shell, Trident Hemispherical Shell
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid o arthritis, posttraumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less ● likely to achieve satisfactory results.
- . Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
The Trident shells are intended for cementless fixation within the prepared acetabulum.
### Tritanium Acetabular Shell System
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid ● arthritis, posttraumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure. ●
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
The Tritanium Acetabular Shell System is intended for cementless use only.
### Accolade II Femoral Stem, Secur-Fit Advanced Femoral Stem, Anato Femoral Stem
The indications for use for total hip arthroplasty with stems include:
- noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; ●
- rheumatoid arthritis;
- correction of functional deformity; ●
- revision procedures where other treatments or devices have failed; and, ●
{34}------------------------------------------------
- nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head ● involvement that are unmanageable using other techniques.
Additional indication specific to use of the Femoral Stem with compatible Howmedica Osteonics Constrained Liners:
- . When the stem is to be used with compatible Howmedica Osteonics Constrained Liners, the device is intended for use in primary or revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, soft tissue laxity, neuromuscular disease, or intra-operative instability.
Accolade II, Secur-Fit Advanced and Anato Femoral Stems are intended for cementless use only and are intended for total and hemiarthroplasty.
### Exeter® V40TM Hip System (includes Orthinox V40 Femoral heads)
The Exeter® V40™ Hip System is intended for use in total hip replacement. It is intended for cemented use only.
The Exeter® Hip is indicated for:
- noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- o rheumatoid arthritis:
- o correction of functional deformity;
- revision procedures where other treatments or devices have failed; and, ●
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head ● involvement that are unmanageable using other techniques.
### Restoration Anatomic Shell
Indications for Use
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid ● arthritis, posttraumatic arthritis or late stage avascular necrosis.
- . Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- o Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
### When used with MDM Liners
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head ● involvement that are unmanageable using other techniques.
- Dislocation risks ●
When used with Constrained Liner:
{35}------------------------------------------------
- The Trident Constrained Acetabular Insert is indicated for use in primary and revision patients at high risk of hip dislocation due to a history of prior dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease, or intraoperative instability.
The Restoration® Anatomic Shell is indicated for cementless use only.
### Omnifit HFX Femoral Stems
### Indications:
For use as a Bipolar Hip Replacement:
- Femoral head/neck fractures or non-unions. ●
- Aseptic necrosis of the femoral head. ●
- Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or ● distortion.
### Other Considerations:
- Pathological conditions or age considerations which indicate a more conservative acetabular . procedure and an avoidance of the use of bone cement in the acetabulum.
- . Salvage of failed total hip arthroplasty.
- Femoral neck fractures. ●
For use as a Total Hip Replacement:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, . post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure. ●
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Clinical circumstances which require an altered femoral resection level due to a proximal fracture, bone loss or calcar lysis.
### Omnifit EON Cemented Femoral Stems
### Indications
For use as a Bipolar Hip Replacement:
- . Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head.
- o Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement or distortion.
### Other Considerations:
- Pathological conditions or age considerations which indicate a more conservative acetabular ● procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty. ●
For use as a Total Hip Replacement:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, . post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure. ●
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less . likely to achieve satisfactory results.
{36}------------------------------------------------
### Secur-Fit Max and Secur-Fit Max Plus Hip Stems
Secur-Fit Max and Secur-Fit Max Plus Hip stems are single use devices and are intended for cementless fixation within the prepared femoral canals of patients requiring hip arthroplasty.
Indications for use as a Total Hip Replacement include:
- noninflammatory degenerative joint disease. including osteoarthritis and avascular necrosis:
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and,
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
### Accolade C Femoral Stems
The Accolade C Femoral Stems are single-use devices intended for cemented fixation.
Indications for use as a Total Hip Replacement include:
- noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; ●
- rheumatoid arthritis;
- correction of functional deformity;
- revision procedures where other treatments or devices have failed; and, ●
- o treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
### Accolade HFx Femoral Stems
- noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; ●
- o rheumatoid arthritis:
- correction of functional deformity; ●
- revision procedures where other treatments or devices have failed; and, ●
- treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head ● involvement that are unmanageable using other techniques.
### Accolade TMZF and Accolade TMZF Plus Femoral Stems
The subject hip stem is a single-use device intended for use in total hip replacement. It is intended for the reconstruction of the head and neck of the femoral joint. This hip stem is intended for primary reconstruction of the proximal femur or revision total hip arthroplasty. This device is intended for use with any currently available Howmedica Osteonics acetabular component and V40™ femoral heads that can be mated with a TMZF 5° 40' trunnion.
### Indications:
- Cementless primary hip surgery in cases of non-inflammatory degenerative joint disease including ● osteoarthritis, avascular necrosis, rheumatoid arthritis, and correction of functional deformity.
{37}------------------------------------------------
- Treatment of nonunion, and femoral neck and trochanteric fractures of the proximal femur with head ● involvement that are unmanageable using other techniques.
- . Revision procedures where other treatments or devices have failed.
### Restoration Acetabular Wedge Augments
The indications for use of the Restoration Acetabular Wedge Augments:
General Indications for Total Hip Replacement Components:
- o Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies of the acetabulum.
Indications Specific to the Acetabular Wedges:
- . As an alternative to structural allograft in cases of superior and superior/posterior segmental acetabular deficiencies.
Acetabular Augments are intended for cementless use only to the bone interface, and are affixed to the mating cup using bone cement
### Trident AD Acetabular shells
The Trident AD Acetabular shells are single use devices. The shell is intended for cemented or cementless fixation within the prepared acetabulum. The Trident AD Acetabular Component System is compatible with any appropriately selected Howmedica Osteonics hip stem/femoral head combination.
Indications:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure. ●
- . Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated ● by deficiencies of the acetabulum.
### Trident T Acetabular shells
The Trident Acetabular shells are single-use devices intended for cementless fixation within the prepared acetabulum. They are compatible with Trident polyethylene acetabular bearing insert. If additional fixation is desired, the dome screw holes, if present, have been designed to accept Stryker Orthopaedics 6.5mm or 5.5mm bone screws.
{38}------------------------------------------------
General Indications for Total Hip Replacement Components:
- . Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure. ●
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less ● likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated ● by deficiencies of the acetabulum.
### Restoration ADM System, Modular Dual Mobility (MDM) Liner and X3 Acetabular Insert
The indications for use of the total hip arthroplasty prostheses include:
- Noninflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- Rheumatoid arthritis:
- Correction of functional deformity; .
- Revision procedures where other treatments or devices have failed;
- Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head ● involvement that are unmanageable using other techniques.
- Dislocation risks.
These devices are intended for cementless use only
### Acetabular Dome Hole Plug
The Dome Hole Plug is an optional device which is available to seal the Howmedica Osteonics Acetabular Shell components during cemented or cementless applications of the acetabular cup. The Howmedica Osteonics Done Hole Plug is threaded into the dome hole of the shell.
Indications
- · In cemented or cementless hip arthroplasty, when an acetabular shell plug is thought to be advantageous.
### Trident X3, Trident Crossfire, Trident X2Vac and Trident X3/Crossfire Elevated Rim Acetabular Liners
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, ● post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less . likely to achieve satisfactory results.
- Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated ● bv deficiencies of the acetabulum.
Trident® Constrained Acetabular Insert
The Trident® Constrained Acetabular Insert is intended for use as a component of a total hip prosthesis in primary or revision patients at a high risk of hip dislocation due to a history of dislocation, bone loss, joint or soft tissue laxity, neuromuscular disease or intraoperative instability.
{39}------------------------------------------------
### BIOLOX Delta Ceramic Heads (C-Taper, Universal Taper, V40 to Universal Taper Adapter Sleeve, and C-Taper to Universal Taper Adapter Sleeve)
The femoral heads are intended for mechanical fixation to their mating hip stems, and can be used in cemented or cementless hip arthroplasty procedures.
For Use as a Total Hip Replacement:
- Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous cup arthroplasty or other procedures
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less ● likely to achieve satisfactory results.
- . Where bone stock is of poor quality or is inadequate for other reconstructive techniques as indicated by deficiencies in the acetabulum.
For Use as a Bipolar Hip Replacement
- . Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or ● distortion.
- Pathological considerations or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty ●
### V40 BIOLOX Delta Ceramic Heads
The femoral heads are intended for mechanical fixation to their mating hip stems, and can be used in cemented or cementless hip arthroplasty procedures.
For Use as a Total Hip Replacement:
- Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- Revision of previous cup arthroplasty or other procedures
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less ● likely to achieve satisfactory results.
For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions. .
- Aseptic necrosis of the femoral head.
- Osteo-. rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or distortion.
### Other considerations:
- Pathological considerations or age considerations which indicate a more conservative acetabular ● procedure and an avoidance of the use of bone cement in the acetabulum.
- . Salvage of failed total hip arthroplasty
## Alumina C-Taper Ceramic Heads
{40}------------------------------------------------
For Use as a Total Hip Replacement:
- Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, ● post-traumatic arthritis or late stage vascular necrosis.
- Revision of previous cup arthroplasty or other procedures ●
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less likely to achieve satisfactory results.
For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular involvement or ● distortion.
### Other Considerations:
- Pathological considerations or age considerations which indicate a more conservative acetabular ● procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty .
## Alumina V40 Ceramic Heads
- Painful, disabling joint disease of the hip resulting from: Non-inflammatory degenerative arthritis ● (osteoarthritis, avascular necrosis, traumatic arthritis, slipped capital epiphysis, pelvic fracture, failed fracture fixation, or diastrophic variant);
- Rheumatoid arthritis ●
- Correction of functional deformity ●
- Where bone stock is of poor quality or inadequate for other reconstructive techniques as indicated ● by the deficiencies of the acetabulum;
- . Treatment of nonunion, femoral neck and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques;
- Revision procedures where other treatments or devices have failed.
## C-Taper CoCr Femoral Heads (LFIT and non-LFIT)
For use as a Bipolar Hip Replacement:
- Femoral head/neck fractures or non-unions 1.
- 2. Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal acetabular involvement 3. or distortion.
Other Considerations:
- 1. Pathological conditions or age considerations which indicate a more conservative approach to the acetabulum and the avoidance of the use of bone cement in the acetabulum.
- 2. Salvage of failed total hip arthroplasty.
Indications for use as part of a Total Hip Replacement include:
- Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid 1. arthritis, post-traumatic arthritis or late stage avascular necrosis.
- 2. Revision of previous unsuccessful femoral head replacement, cup arthroplasty, or other procedure.
{41}------------------------------------------------
- 3. Clinical management situations where arthrodesis or alternate reconstructive techniques are less likely to achieve satisfactory results.
V40 CoCr Femoral Heads (LFIT and non-LFIT)
- Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; ●
- Rheumatoid arthritis
- Correction of functional deformity
- Revision procedures where other treatments or devices have failed
- Nonunions, femoral neck fractures, and trochanteric fractures of the proximal femur with head involvement that are unmanageable using other techniques.
### Unitrax Endoprosthesis, Unitrax V40 Adapter Sleeve, Unitrax C-Taper Adapter Sleeve
- The Howmedica Osteonics Unitrax Endoprosthesis, and the Unitrax V40 Modular Adaptor sleeves are used as a hemiarthroplasty device for the following indications: femoral neck fractures, idiopathic avascular necrosis, and non-unions. The Unitrax C-taper sleeves are intended for use as a Hemi-Hip Replacement with the following indications: femoral head/neck fractures or non-unions, aseptic necrosis of the femoral head/neck and osteo- and post traumatic arthritis. The patient's acetabular bone stock must be adequate to support articulation with the head of the endoprosthesis.
### Universal Distal Spacer
Howmedica Osteonics Corp.'s accessory products for cemented arthroplasty are optional devices intended to assist in the preparation, implantation and/or positioning of a femoral implant intended for cemented application.
### Indications
For cement spacers, mid-shaft restrictors and Cement Plugs:
- In cemented hip arthroplasty, when the cement spacer, restrictor and/or plug is thought to be . advantageous.
### Torx Screws, GAP Plate Screws and Osteolock Bone Screws
- o Howmedica Osteonics Torx Cancellous Bone Screws are intended for supplemental fixation of associated Howmedica Osteonics cementless Acetabular Shells or Howmedica Osteonics Tibial Tray components.
- Howmedica Osteonics Osteolock Bone Screws are intended for supplemental fixation of associated ● cementless Trident Tritanium Multihole Acetabular shells, Restoration Acetabular Augments, and Restoration Anatomic Shell.
- Howmedica Osteonics Restoration GAP Plate Screws are intended only for fixation of the dome and ● iliac plates of the associated Howmedica Osteonics Restoration GAP Acetabular Shells, Trident Tritanium Hemispherical Multihole Acetabular Shells, Restoration Acetabular Augments and Restoration Anatomic Shell.
V40 to C-Taper Adapter Sleeve For Use as a Total Hip Replacement
{42}------------------------------------------------
- Painful disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage vascular necrosis.
- Revision of previous cup arthroplasty or other procedures.
- Clinical management problems where arthrodesis or alternative reconstructive techniques are less o likely to achieve satisfactory results.
For Use as a Bipolar Hip Replacement
- Femoral head/neck fractures or non-unions. ●
- Aseptic necrosis of the femoral head.
- Osteo-, rheumatoid, and post traumatic arthritis of the hip with minimal acetabular ● involvement or distortion.
### Other Considerations:
- Pathological considerations or age considerations which indicate a more conservative 0 acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty.
## Summary of Technological Characteristics:
There have been no changes requiring 510(k) clearance to the technological characteristics of the Stryker Total Hip systems as a result of the revision to the labeling. The subject devices have the same design and are manufactured from the same materials as the predicate devices.
### Non-Clinical Testing:
The following non-clinical laboratory testing was performed to determine substantial equivalence:
Non-clinical testing as outlined in the FDA guidance document "Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment - Guidance for Industry and FDA Staff", dated December 11, 2014 was conducted to characterize the compatibility of Stryker Orthopaedics total hip passive implants in the MR environment. FDA draft guidance "Assessment of Radiofrequency-Induced Heating in the Magnetic Resonance (MR) Environment for Multi-Configuration Passive Medical Devices – Draft Guidance for Industry and FDA Staff", dated June 29, 2015 was also consulted for the heating evaluations performed. Testing was performed according to the standards listed below:
- Magnetically Induced Displacement Force performed per ASTM F2052-06 and ASTM ● F2052-14, Standard Test Method for Measurement of Magnetically Induced Displacement Force on Medical Devices in the MR Environment
- 0 Magnetically Induced Torque - performed per ASTM F2213-06 (Reapproved 2011), Standard Test Method for Measurement of Magnetically Induced Torque on Medical Devices in the MR Environment
- Image Artifact performed per ASTM F2119-07 (Reapproved 2013), Standard Test ● Method for Evaluation of MR Image Artifacts from Passive Implants
{43}------------------------------------------------
- Heating by RF Fields per ASTM F2182-11a, Standard Test Method for Measurement of 0 Radio Frequency Induced Heating near Passive Implants during MR Imaging
The labeling has been modified to include the MR conditional symbol, and to provide the parameters under which a patient who has the device can be safely scanned. Additionally, for certain devices, a contraindication was removed.
Clinical Testing: Clinical testing was not required as a basis for substantial equivalence.
Conclusion: The Stryker Orthopaedics Hip System devices are substantially equivalent to the predicate devices identified in this premarket notification.
Device comparisons showed that the subject devices have the same intended use, and substantially equivalent design, materials and operational principles to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.