K192827 · Medacta International S.A. · LZO · Jan 6, 2020 · Orthopedic
Device Facts
Record ID
K192827
Device Name
Quadra-P
Applicant
Medacta International S.A.
Product Code
LZO · Orthopedic
Decision Date
Jan 6, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3353
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The hip prostheses QUADRA-P and QUADRA-P collared are designed for cementless use in total or partial hip arthroplasty, for primary or revision surgery. The hip prosthesis QUADRA-P cemented is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery. Hip replacement is indicated in the following cases: · Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid polyarthritis or congenital hip dysplasia · Avascular necrosis of the femoral head · Acute traumatic fracture of the femoral head or neck · Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement or total hip arthroplasty.
Device Story
Quadra-P is a line extension of femoral stems for total or partial hip arthroplasty. Cementless stems are titanium-niobium alloy with titanium plasma spray (MectaGrip) and HA coating; cemented stems are high-nitrogen stainless steel. Used by orthopedic surgeons in clinical settings to replace diseased or damaged hip joints. Implants integrate with modular femoral heads and acetabular components to restore joint function. Benefits include pain relief and improved mobility for patients with severe joint degeneration or trauma. Device is implanted surgically; performance relies on mechanical stability and biocompatibility of materials.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including range of motion (EN ISO 21535), fatigue testing (ISO 7206-4, ISO 7206-6), static load testing (ISO 7206-10), and axial disassembly force testing (ASTM F2009).
Technological Characteristics
Femoral stems: Titanium-Niobium alloy (cementless) with MectaGrip plasma spray and HA coating, or high-nitrogen stainless steel (cemented). Modular design. Mechanical testing per ISO 7206 series and ASTM F2009. Biocompatible materials. Sterilized.
Indications for Use
Indicated for patients requiring total or partial hip arthroplasty (primary or revision) due to arthrosis, traumatic arthritis, rheumatoid polyarthritis, congenital hip dysplasia, avascular necrosis of the femoral head, acute traumatic fracture of the femoral head/neck, or failure of previous hip surgeries.
Regulatory Classification
Identification
A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name in a sans-serif font. The logo is simple and professional, conveying the agency's authority and mission.
Medacta International SA % Chris Lussier Director, Quality and Regulatory Medacta USA 3973 Delp Street Memphis, Tennessee 38118
Re: K192827
Trade/Device Name: Quadra-P Regulation Number: 21 CFR 888.3353 Regulation Name: Hip Joint Metal/Ceramic/Polymer Semi-Constrained Cemented Or Nonporous Uncemented Prosthesis Regulatory Class: Class II Product Code: LZO, MEH, KWY, LZY, LPH, JDI Dated: September 30, 2019 Received: October 2, 2019
Dear Chris Lussier:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
January 6, 2020
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
FOR Vesa Vuniqi Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
## K192827
Device Name Quadra-P
### Indications for Use (Describe)
The hip prostheses QUADRA-P and QUADRA-P collared are designed for cementless use in total or partial hip arthroplasty, for primary or revision surgery. The hip prosthesis QUADRA-P cemented is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery. Hip replacement is indicated in the following cases:
· Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid polyarthritis or congenital hip dysplasia
· Avascular necrosis of the femoral head
· Acute traumatic fracture of the femoral head or neck
· Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement or total hip arthroplasty.
Type of Use (Select one or both, as applicable)> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Quadra-P Traditional 510(k)
# 2.0 510(k) Summary
#### I. Submitter
Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
Contact Person: Stefano Baj, Regulatory and Compliance Director, Medacta International SA Date Prepared: September 30, 2019
#### II. Device
| Device Proprietary Name: | Quadra-P |
|--------------------------|--------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Femoral Stems |
| Classification Name: | Hip joint metal/ceramic/polymer semi-constrained cemented<br>or nonporous un-cemented prosthesis |
| Primary Product Code: | LZO |
| Secondary Product Codes: | MEH, KWY, LZY, LPH, JDI |
| Regulation Number: | 21 CFR 888.3353, 21 CFR 888.3390, 21CFR 888.3360, 21<br>CFR 888.3358, |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following devices:
Primary predicate device:
- Quadra-P, K181254, Medacta International SA ●
Secondary predicate device:
- Medacta Total Hip Prosthesis System Quadra C (also referred to as "Quadra C"), K083558, . Medacta International SA
The following devices are referenced:
- AMIStem-H Proximal Coating, K121011, Medacta International SA ●
- AMIStem-H Proximal Coating, AMIStem-P and AMIStem-P Collared, K173794, Medacta . International SA
{4}------------------------------------------------
K192827 Page 2/4
> Quadra-P Traditional 510(k)
#### IV. Device Description
The Quadra-P implants are line extensions to the currently cleared Quadra-P implants (K181254).
The Quadra-P implants in this submission are comprised of the following products:
- Quadra-P STD (Stem size 00); .
- Quadra-P Short Neck (STD Stem sizes 0 10 and LAT Stem sizes 0 10); ●
- . Quadra-P Collared (STD Stem sizes 00 - 10 and LAT Stem sizes 0 - 10); and
- Quadra-P Cemented (STD Stem sizes 0 8 and LAT Stem sizes 0 8). ●
The Quadra-P implants are part of the Medacta Total Hip Prosthesis system. The Medacta Total Hip Prosthesis system consists of femoral stems, modular femoral heads, and acetabular components.
The Quadra-P, Quadra-P short neck, and Quadra-P Collared are cementless stems manufactured from Titanium-Niobium alloy with a Titanium plasma spray coating (MectaGrip) on the proximal area and HA coating on the shaft. The Quadra-P Cemented stems manufactured from high nitrogen stainless steel.
The acetabular components consist of metal cups and liners made of ultra-high molecular weight polyethylene (UHMWPE), or Highcross highly crosslinked ultra-high molecular weight polyethylene (HXUHMWPE). Acetabular components include the Mpact DM (K143453), VersafitCup (K083116 and K092265), VersafitCup CC Trio (K103352, K120521 and K122911), Mpact (K103721 and K132879), Mpact 3D Metal (K171966), and Medacta Bipolar Head (K091967).
The Quadra-P implants can be combined with the CoCr Ball Heads (K072857 and K080885), Endo Head (K111145), or MectaCer BIOLOX® Forte (K073337) or MectaCer BIOLOX® Delta Femoral Heads (K112115). or MectaCer Biolox Option Heads (K131518).
MectaCer Biolox Option Heads (K131518) cannot be combined with Quadra-P cemented stems, but only with the Quadra-P cementless stems.
#### V. Indications for Use
The hip prostheses QUADRA-P and QUADRA-P collared are designed for cementless use in total or partial hip arthroplasty, for primary or revision surgery. The hip prosthesis QUADRA-P cemented is designed for cemented use in total or partial hip arthroplasty in primary or revision surgery. Hip replacement is indicated in the following cases:
- Severely painful and/or disabled joint as a result of arthrosis, traumatic arthritis, rheumatoid ● polyarthritis or congenital hip dysplasia
- 0 Avascular necrosis of the femoral head
- Acute traumatic fracture of the femoral head or neck ●
- . Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, partial hip arthroplasty, hip resurfacing replacement or total hip arthroplasty.
{5}------------------------------------------------
#### VI. Comparison of Technological Characteristics
The Quadra-P implants and the predicate devices share the following characteristics:
- CCD angle;
- materials of construction; ●
- coating;
- biocompatibility; ●
- device usage; ●
- sterility; ●
- shelf life; and
- packaging.
The Quadra-P femoral stems are technologically different from the predicate devices with respect to size and stem lengths.
#### VII. Performance Data
Testing was conducted according to written protocols with acceptance criteria that were based on standards. The following mechanical studies were performed on worst-case stems in support of a substantial equivalence determination:
Non-Clinical Studies:
- Performance Tests
- o range of motion (ROM): EN ISO 21535:2009 Non-Active Surgical Implants Joint Replacement Implants — Specific Requirements for Hip-Joint Replacement Implants (ISO 21535:2007/Amendment 1:2016);
- o fatigue testing: ISO 7206-4 Third Edition 2010-06-15 Implants For Surgery Partial And Total Hip Joint Prostheses - Part 4: Determination Of Endurance Properties And Performance Of Stemmed Femoral Components [Including AMENDMENT 1 (2016)];
- fatigue testing: ISO 7206-6 Second Edition 2013-11-15 Implants For Surgery o Partial And Total Hip Joint Prostheses - Part 6: Determination Of Endurance Properties Of Head And Neck Region Of Stemmed Femoral Components;
- static fatigue testing: ISO 7206-10:2003 Implants For Surgery -- Partial And Total o Hip-Joint Prostheses -- Part 10: Determination Of Resistance To Static Load Of Modular Femoral Heads; and
- pull off force testing: ASTM F2009-00 (Reapproved 2011) Standard Test Method o For Determining The Axial Disassembly Force Of Taper Connections Of Modular Prostheses.
- Pyrogenicity ●
- o Bacterial Endotoxin Test (LAL test) was conducted according to European Pharmacopoeia §2.6.14 (which is equivalent to USP chapter <85>) and pyrogen test according to USP chapter <151> for pyrogenicity determination.
- o the subject devices are not labeled as non-pyrogenic or pyrogen free.
{6}------------------------------------------------
#### K192827 Page 4/4
Quadra-P Traditional 510(k)
Clinical Studies:
- No clinical studies were conducted.
# VIII. Conclusion
The information provided above supports that the Quadra-P implants are substantially equivalent to the identified predicate devices.
Substantial equivalence has been demonstrated through a comparison of intended use, design and technological characteristics, as well as performance evaluations. The Quadra-P implants are as safe and effective as the predicate devices Quadra-P (K181254) and Quadra C (K083558).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.