TOTAL HIP PROSTHESIS SYSTEM QUADRA S + COCRMO FEMORAL BALL HEAD + APRICOT
K072857 · Medacta International S.A. · JDI · Feb 4, 2008 · Orthopedic
Device Facts
Record ID
K072857
Device Name
TOTAL HIP PROSTHESIS SYSTEM QUADRA S + COCRMO FEMORAL BALL HEAD + APRICOT
Applicant
Medacta International S.A.
Product Code
JDI · Orthopedic
Decision Date
Feb 4, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Medacta Total Hip Prosthesis System is intended for cementless use in total or partial hip arthroplasty and in primary or revision surgery. Hip replacement is indicated in the following cases: Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia Avascular necrosis of the femoral head Acute traumatic fracture of the femoral head or neck Failure of previous hip surgery, joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement
Device Story
Medacta Total Hip Prosthesis System comprises Quadra S femoral stem, CoCrMo femoral ball heads, and Tri-Plus cup system. Components designed for cementless total or partial hip arthroplasty; primary or revision surgery. Femoral stem (titanium alloy) features sandblasted surface and mirror-polished neck. CoCrMo femoral heads (28-32 mm) provide mechanical fixation to stem. Tri-Plus acetabular shell (hemispherical, plasma-sprayed) uses locking mechanism with direct-compression molded polyethylene liner to reduce wear debris. System used by orthopedic surgeons in clinical settings to replace damaged hip joints, reduce pain, and restore patient mobility. Device performance relies on mechanical fit and material biocompatibility.
Clinical Evidence
Bench testing only. Design verification and validation performed per 21 CFR 820.30 and FDA-recognized standards for mechanical testing. No clinical data provided.
Technological Characteristics
Materials: Quadra S Stem (Titanium alloy, ISO 5832-11), CoCrMo Head (ISO 5832-12), Tri-Plus Liner (Direct-compression molded polyethylene). Design: Cementless, hemispherical acetabular shell with plasma spray coating. Sterilization: Radiation (ISO 11137-1, ISO 11137-2).
Indications for Use
Indicated for patients requiring total or partial hip arthroplasty (primary or revision) due to severe pain/disability from arthritis (traumatic, rheumatoid, or congenital dysplasia), avascular necrosis of the femoral head, acute traumatic fracture of the femoral head/neck, or failure of previous hip surgeries.
Regulatory Classification
Identification
A hip joint metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and an acetabular resurfacing component made of ultra-high molecular weight polyethylene and is limited to those prostheses intended for use with bone cement (§ 888.3027).
Predicate Devices
SL-PLUS and SLR-PLUS Stems, Plus Orthopedics AG (K001942)
SL-PLUS Lateralized Stem, Plus orthopedics AG (K021178)
{0}------------------------------------------------
# 672857 (pg 1/4)
# Premarket Notification 510(k) Summary
#### l ) Submitter Information
- b) Company Name and Address:
MEDACTA International SA Strada Regina 6874 Castel San Pictro (CH) - CE 0546 Switzerland Phone: + 41 91 696 60 60 Fax: + 41 91 696 60 66
- c) Contact Name:
Pamela J. Weagraff, Principal Consultant Quintiles Consulting 18 Bridie Lane Norfolk, MA 02056
508-528-1745 Telephone: Facsimile: 978-752-1225 E-mail: pamela.weagraff(@quintiles.com
- d) Date Prepared: February 1, 2008
### 2) Name of Device
- b) Trade Name: Medacta Total Hip Prosthesis System (Quadra S® Femoral Stem, CoCrMo Femoral Ball Heads and Tri-PlusTM Cup System)
- c) Common Name: Total Hip Prosthesis System
- d) Classification Name and Reference: Part 888.3350 Hip joint metal/polymer semiconstrained cemented prosthesis; Product Code JDI
### 3) Substantial Equivalence Claimed to Predicate Device
| Total Hip Prosthesis<br>System Component | Predicate Device /Manufacturer | K # |
|------------------------------------------|------------------------------------------------------------------|---------|
| Quadra® S | SL-PLUS® and SLR-PLUS® Stems, Plus Orthopedics AG | K001942 |
| Quadra® S | SL-PLUS® Lateralized Stem, Plus orthopedics AG | K021178 |
| CoCrMo Femoral Ball<br>Head | Portland Cobalt Chrome Femoral Head, Portland Orthopaedics, Ltd. | K063278 |
{1}------------------------------------------------
K072857 (pg 2/4)
| Total Hip Prosthesis<br>System Component | Predicate Device /Manufacturer | K # |
|------------------------------------------|------------------------------------------------------------------|---------|
| Tri-Plus™ Cup System | Not applicable, previously cleared acetabular<br>cups and liners | K042565 |
#### 4) Device Description
The components of the Medacta Total Hip Prosthesis System are intended for single, cementless use, only, and are described below:
#### a) Quadra® S Femoral Stem
The Quadra® S Femoral Stem is intended for cementless use, and is available in four versions, standard or lateral stem, both in regular or short neck version. The Quadra® S Femoral Stem is intended, as a component of the Mcdacta Total Hip Prosthesis System, for use in total or partial hip arthroplasty and in primary or revision surgery.
The Quadra® S Femoral Stem is manufactured from a titanium alloy, according to ISO 5832-11, 1994, Implants for surgery -- Metallic materials -- Part 11: Wrought titanium 6aluminium 7-niobium alloy. The stem is sandblasted and the neck has a mirror-polished finish.
The Quadra® S Femoral Stem may be used with the Tri-Plus™ Cup System and CoCrMo Femoral Ball Heads.
#### b) Cobalt Chromium Molybdenum (CoCrMo) Femoral Ball Head
The CoCrMo femoral ball head is intended for mechanical fixation to a mating hip stem such as the Quadra® S Femoral Stem and for use in total or partial hip arthroplasty to provide increased patient mobility and reduced pain by replacing the damaged hip joint, in primary or revision surgery.
The CoCrMo femoral ball heads are manufactured from Cobalt Chromium Molybdenum, according to ISO 5832-12, 2007, Implants for surgery -- Metallic materials -- Part 12: Wrought cobalt-chromium-molybdenum alloy.
The CoCrMo femoral balls heads range in diamcter from 28 mm to 32 mm and may be used with the Quadra® S Femoral Stem and Tri-Plus™ Cup System.
#### c) Tri-plus™ Cup System
The Tri-plus™ Acetabular shell is a true hemispherical design featuring plasma spray coating which allows the substrate to retain its fatigue strength characteristics. The shell and liner utilize a secure locking mechanism designed to decrease the potential for polyethylenc wear debris. The liner is manufactured from direct-compression molded polyethylene and incorporates anti-rotation features. The locking mechanism allows for
{2}------------------------------------------------
secure fixation between the shell and liner. Superior surface tabs designed into liner provides anti-rotation stability and proper liner indexing. The shell is a true hemisphere design available in No-hole, Cluster hole, and Multi-hole options. The liners are available in Neutral and Hooded (10°) options. The Tri-plus™ liners are manufactured from DCM polyethylene. The Tri-plus™ Cup System may be used with the Quadra® S Femoral Stem and CoCrMo femoral ball heads.
#### 5) Intended Use / Indications for Use
The Medacta Total Hip Prosthesis System is intended for ccmentless use in total or partial hip arthroplasty and in primary or revision surgery.
Hip replacement is indicated in the following cases:
- 프 Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia
- 트 Avascular necrosis of the femoral head
- Acute traumatic fracture of the femoral head or neck
- 트 Failure of previous hip surgery, joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement
#### 6) Predicate Device Comparison of Indications for Use / Intended Use and Technical Characteristics
The comparison of the Medacta Total Hip Prosthesis System was based on a review of relevant Design Control documentation, which are included in the company's 510(k) Premarket Notification, and information concerning the predicate devices that was obtained from the FDA web site and the predicate device manufacturers' web sites. The comparison considered technological characteristics and the indications for use / intended use.
#### 7) Performance Data
Note: No performance standards applicable to this device have been adopted under Section 514 of the Food Drug and Cosmetic Act, with respect to the applicable classifications or product codes.
The materials used to manufacture Medacta Total Hip Prosthesis System components conform to the following FDA recognized standards:
- a) Materials:
Quadra® S Femoral Stem: ISO 5832-11, 1994, Implants for surgery -- Metallic materials -- Part 11: Wrought titanium 6-aluminium 7-niobium alloy
CoCrMo Femoral Ball Head: ISO 5832-12, 2007, Implants for surgery -- Metallic
{3}------------------------------------------------
K072857 (pg 4/4)
materials -- Part 12: Wrought cobalt-chromium-molybdenum alloy
- b) Sterility:
ISO 11137-1:2006, Sterilization of health care products -- Radiation -- Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
ISO 11137-2, 2006, Sterilization of health care products -- Radiation -- Part 2: Establishing the sterilization dose
- c) Performance Testing:
Design verification and design validation, e.g., bench testing was performed according to FDA's Design Control Requirements, Title 21 Code of Federal Regulations, Part 820.30, and in accordance to FDA recognized standards for mechanical testing.
#### 8) Conclusion:
The information and data provided in this 510(k) Premarket Notification establish that the Medacta Total Hip Prosthesis System is substantially equivalent to the afore-mentioned predicate devices with respect to indications for use/intended use, and technical characteristics.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Re:
FEB 4 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MEDACTA International SA % Quintiles Consulting Ms. Pamela J. Weagraff Principle Consultant 18 Birdie Lane Norfolk, MA 02056
> K072857 Trade/Device Name: Medacta Total Hip Prosthesis Regulation Number: 21 CFR 888.3350 Regulation Name: Hip joint metal/polymer/metal semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JDI Dated: January 7, 2008 Received: January 8, 2008
Dear Ms. Weagraff:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Pamela J. Weagraff
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Mullison
Mark N. Melkerson i Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# K072857 (pg 1/1)
# ODE Indications Statement
#### 510(k) Number (if known): Unknown
Device Name: Medacta Total Hip Prosthesis System
#### Indications for Use:
The Medacta Total Hip Prosthesis System is intended for cementless use in total or partial hip arthroplasty and in primary or revision surgery.
Hip replacement is indicated in the following cases:
- 미 Severely painful and/or disabled joint as a result of arthritis, traumatic arthritis, rheumatoid polyarthritis, or congenital hip dysplasia
- l Avascular necrosis of the femoral head
- 트 Acute traumatic fracture of the femoral head or neck
- 트 Failure of previous hip surgery, joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement
X ____________________________________________________________________________________________________________________________________________________________________________ Prescription Use: __ AND/OR (21 CFR 801 Subpart D)
Over-the-Counter Use: (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Choubara huchum
Division of General, Restorative, and Neurological Devices
**510(k) Number** K072657
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.