K143453 · Medacta International · LPH · Mar 2, 2015 · Orthopedic
Device Facts
Record ID
K143453
Device Name
Mpact Double Mobility System
Applicant
Medacta International
Product Code
LPH · Orthopedic
Decision Date
Mar 2, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature. The patient's condition should be due to one or more of: • Severely painful and/or disabled joint: as a result of osteoarthritis, rheumatod arthritis or psoriactic arthritis, congenital hip dysplasia, ankylosing spondylitis. - · Avascular necrosis of the femoral head. · Acute traumatic fracture of the femoral head or neck. · Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present. • Dislocation risks
Device Story
Mpact Double Mobility System is a cementless acetabular hip prosthesis for total hip arthroplasty. System comprises acetabular shells (42-66mm) and modular liners (UHMWPE/HighCross UHMWPE). Shells feature TiGrowth porous coating for bone ingrowth. Device used in primary or revision surgery by orthopedic surgeons in clinical settings. System provides articulation via double mobility design to reduce dislocation risk. Performance verified through risk analysis and bench testing comparing against predicate systems. Output is a stable, semi-constrained hip joint reconstruction.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing including coating adhesion, chemical characterization, friction testing (ASTM G115-04, ASTM D4518-94), ion release/corrosion testing (ASTM F746-04), wear analysis, range of motion (ROM) assessment, and modular connection stability testing.
Technological Characteristics
Acetabular shells: High Nitrogen StSt (ISO 5832-9) with TiGrowth porous coating (ASTM F1580). Liners: UHMWPE (ISO 5834-2 Type 1) and HighCross highly crosslinked UHMWPE. Design: Hemispherical, cementless, semi-constrained, double mobility. Sizes: 42mm-66mm shells; 22.2mm femoral head compatibility.
Indications for Use
Indicated for skeletally mature patients requiring primary or revision total hip arthroplasty due to osteoarthritis, rheumatoid/psoriatic arthritis, congenital hip dysplasia, ankylosing spondylitis, avascular necrosis, acute traumatic fracture of femoral head/neck, or failure of previous hip surgery (e.g., reconstruction, fixation, arthrodesis, hemiarthroplasty, surface replacement, or total hip replacement) where sufficient bone stock exists, or for patients with dislocation risks.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 2, 2015
Medacta USA Mr. Adam Gross Director of Regulatory, Quality and Compliance 1556 West Carroll Avenue Chicago, Illinois 60607
Re: K143453 Trade/Device Name: Mpact Double Mobility System Regulation Number: 21 CFR 888.3358 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Regulatory Class: Class II Product Code: LPH, LZO, MEH Dated: December 22, 2014 Received: December 30, 2014
Dear Mr. Gross:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
Page 2 - Mr. Adam Gross
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143453
#### Device Name Mpact Double Mobility System
#### Indications for Use (Describe)
The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature.
The patient's condition should be due to one or more of:
· Severely painful and/or disabled joint: as a result of osteoarthritis, rheumatod arthritis or psoriactic arthritis, congenital hip dysplasia, ankylosing spondylitis.
- · Avascular necrosis of the femoral head.
· Acute traumatic fracture of the femoral head or neck.
· Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty,
surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
• Dislocation risks
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
K143453 Page 1 of 3
Image /page/3/Picture/1 description: The image shows the logo for Medacta International. The logo consists of two blue triangles, the word "Medacta" in blue, a yellow line underneath the word, and the word "International" in gray. There is also a Swiss flag to the right of the word "International".
## 510(k) Summary
Applicant/Sponsor: Medacta International SA Strada Regina 6874 Castel San Pietro (CH) Switzerland Phone (+41) 91 696 60 60 Fax (+41) 91 696 60 66
- Contact Person: Adam Gross Director of Regulatory, Quality and Compliance Medacta USA 1556 W Carroll Ave Chicago, IL 60607 Phone: (805) 910-6511 Fax: (805) 437-7553 Email: AGross@medacta.us.com
Date Prepared: November 19, 2014
### DEVICE INFORMATION
Trade/Proprietary Name: Mpact Double Mobility System Classification Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis Class II 21 CFR 888.3358, Product Code LPH 21 CFR 888.3353, Product Code LZO, MEH
Predicate Device(s):
| 510(k) | Product | 510(k) Holder | Clearance Date |
|---------|-----------------------------------|-----------------------|----------------|
| K083116 | Versafitcup Double Mobility | Medacta International | 4/7/2009 |
| K092265 | Versafitcup DM Highly Crosslinked | Medacta International | 3/12/2010 |
| K103721 | Mpact Acetabular System | Medacta International | 3/21/2011 |
| K132879 | Mpact Extension | Medacta International | 1/22/2014 |
| K131458 | Mpact DM Converter | Medacta International | 10/1/2013 |
{4}------------------------------------------------
### Product Description
The Mpact Double Mobility System acetabular shells are designed to be used with the Medacta Total Hip Prosthesis System. The Mpact Double Mobility System consists of acetabular shells (size 42mm to 66mm in 2mm increments) that have a TiGrowth porous coating (Ti, ASTM F1580) and hemispherical shape that is identical to the K103721 and K132879 predicate devices. The material of the shells (High Nitrogen StSt - ISO 5832-9) and the double mobility design is the same as the K083116 predicate device.
The shells are used with the Versafitcup Double Mobility Liners that are registered with K083116, K092265, and K131458 predicate devices in addition to one new size (22.2/DMA) of the Medacta Double Mobility liner in both UHMWPE (ISO 5834-2 Type 1) and HighCross highly crosslinked UHMWPE which are included in this submission. These liners are identical to the liners in the K083116, K092265, and K131458 submissions but have a smaller external diameter than the predicates. These new liners can only be coupled with a 22.2mm femoral head and have a minimum thickness of 5 mm.
## Indications for Use
The hip prosthesis is designed for cementless use in total hip arthroplasty in primary or revision surgery. The patient should be skeletally mature.
The patient's condition should be due to one or more of:
• Severely painful and/or disabled joint: as a result of osteoarthritis, post-traumatic arthritis, rheumatoid arthritis or psoriactic arthritis, congenital hip dysplasia, ankylosing spondylitis.
- · Avascular necrosis of the femoral head.
- · Acute traumatic fracture of the femoral head or neck.
· Failure of previous hip surgery: joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or total hip replacement where sufficient bone stock is present.
• Dislocation risks
## Comparison to Predicate Devices
The indications for use, design features and materials of the Mpact Double Mobility System are substantially equivalent to those of the predicate devices. The substantial equivalence of the Mpact Double Mobility System implants is supported by the performance testing, materials information, and data analysis provided within this Premarket Notification.
{5}------------------------------------------------
### Performance Testing
The Mpact Double Mobility System was evaluated by risk analysis to identify any new risks associated with the change. Based on the risk analysis, design verification was conducted to written protocols with pre-defined acceptance criteria. The protocols and pre-defined acceptance criteria were based on the standards, FDA guidance, and comparison to the predicate device system. The Mpact Double Mobility System was compared to the worst case predicate device and it was determined that the Mpact Double Mobility System is not worst case.
Mpact Double Mobility System has similar performance testing as the predicates in terms of:
Acetabular shell:
- Coating adhesion and chemical characterization
- Friction test ASTM G115-04 and ASTM D4518-94 ●
- lon release and corrosion test ASTM F746-04 0
Acetabular shell and liner:
- Excessive wear during motion
- Limited ROM .
- Dislocation Risk
Acetabular liner:
- . Instability risk of the modular connection of the Mpact Double Mobility Liner and femoral head
### Conclusion:
Based on the above information, the Mpact Double Mobility System can be considered as substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.