Sterile High-pressure Angiographic Syringes for Single-use
K151960 · Shenzhen Baoan Medical Supplies Co., Ltd. · DXT · Jan 7, 2016 · Cardiovascular
Device Facts
Record ID
K151960
Device Name
Sterile High-pressure Angiographic Syringes for Single-use
Applicant
Shenzhen Baoan Medical Supplies Co., Ltd.
Product Code
DXT · Cardiovascular
Decision Date
Jan 7, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1650
Device Class
Class 2
Indications for Use
Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with an US legally marketed angiographic injectors.
Device Story
Sterile, single-use plastic syringes designed for high-pressure delivery of contrast media or saline; used in conjunction with US-marketed angiographic injectors. Device includes syringe barrel, connecting tubes (straight, Y, or T configurations), and J-shape draw tubes. Operated by clinicians in clinical settings; device attaches to injector systems to facilitate fluid delivery into patient vasculature. Output is controlled fluid injection; assists in diagnostic imaging procedures. Benefits include reliable, sterile, single-use fluid delivery for angiography.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-4, 5, 10, 11), sterility (ISO 11737-2, ISO 10993-7), seal integrity (ASTM F88, F1929), internal pressure resistance (ASTM F1140), and mechanical performance (ISO 7886-1/2, ISO 594-1/2).
Technological Characteristics
Plastic, single-use, high-pressure angiographic syringes. Features Luer taper fittings (ISO 594-1/2). Sterilized via Ethylene Oxide (EO). Compatible with various power-driven angiographic injectors. Materials meet ISO 10993 biocompatibility requirements. Includes connecting tubes and J-shape draw tubes.
Indications for Use
Indicated for patients requiring injection of contrast media or saline during angiographic procedures using legally marketed angiographic injectors.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 7, 2016
Shenzhen Baoan Medical Supplies Co., Ltd % Diana Hong General Manager Mid-link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China
Re: K151960
Trade/Device Name: Sterile High-pressure Angiographic Syringes For Single-use Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector And Syringe Regulatory Class: Class II Product Code: DXT Dated: November 26, 2015 Received: December 1, 2015
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151960
### Device Name
Sterile High-pressure Angiographic Syringes for Single-use
Indications for Use (Describe)
Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with an US legally marketed angiographic injectors.
Type of Use (Select one or both, as applicable)
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Exhibit #2 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: K151960
- 1. Date of Preparation: 11/26/2015
- 2. Sponsor Identification
Shenzhen BaoAn Medical Supplies Co., Ltd Industrial Zone, Shenkeng, Liuyue Community, Henggang Street, Longgang District, Shenzhen, Guangdong, 518173, China
Establishment Registration Number: Not yet registered
Contact Person: Mr. Mingan Mu Position: Deputy General Manager Tel: +86-755-28638515 Fax: +86-755-28638033 Email: boon@szboon.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Mr. Lee Fu (Alternative Contact Person)
## Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{4}------------------------------------------------
#### 4. Identification of Proposed Device
Trade Name: Sterile High-pressure Angiographic Syringes for Single-use Common Name: Disposable angiographic syringe Models: 100101, 100103, 100104, 100113, 100114, 100121, 100124, 200101, 200102, 200103, 200104, 300101, 300103 and 300105
Regulatory Information
Classification Name: Angiographic injector and syringe Classification: II Product Code: DXT Regulation Number: CFR 870.1650 Review Panel: Cardiovascular
Indication for Use:
Sterile High-pressure Angiographic Syringes for Single-use are intended for the injection of contrast media or saline; they shall be used with an US legally marketed angiographic injectors.
Device Description
The proposed devices are available in packs, which may include different configurations of syringes and accessories. The syringes are plastic, disposable syringes, which are available in various models and configurations. They are intended to be used with an U.S. legally marketed angiography injector, compatibilities are shown in Table 1.
| Model (Syringe) | Volume (ml) | Type | Compatible Angiographic Injector |
|----------------------------------------------------------------------------------------|-------------|-------------|-----------------------------------------------------------------------------|
| The Sterile High-pressure Angiographic Syringes for Single-use models listed below are | | | |
| compatible with the following Medrad Injector models | | | |
| 100101 | 200ml | Single Shot | MCT & MCT plus CT, K924116<br>Visitron CT, K991557<br>En Vision CT, K993782 |
| 100103 | 200ml | Single Shot | Stellant CT, K023183 |
| 100113 | 200/200ml | Dual Shots | Stellant CT, K023183 |
| 200101 | 65/65ml | Dual Shots | Solaris, K033247 |
| 200102 | 65/115ml | Dual Shots | Solaris, K033247 |
| 300101 | 150ml | Single Shot | Mark V, K903390 |
| 300105 | 130ml | Single Shot | Mark III & Mark IV, K903390 |
| The Sterile High-pressure Angiographic Syringes for Single-use models listed below are | | | |
| compatible with the following LIEBEL – FLARSHEIM Injector models | | | |
| 100104 | 200ml | Single Shot | CT 9000 & CT9000 ADV, K912944 |
| 100114 | 200/200ml | Dual Shots | CT 9000 & CT9000 ADV, K912944 |
Table 1 Compatibility between Syringes and Injectors
{5}------------------------------------------------
| 200103 | 60/60ml | Dual Shots | Optistar LE, Elite, K073592 |
|----------------------------------------------------------------------------------------|-----------|-------------|-----------------------------|
| 300103 | 150ml | Single Shot | ILLUMENA, K963071 |
| The Sterile High-pressure Angiographic Syringes for Single-use models listed below are | | | |
| compatible with the following E-Z-EM Injector models | | | |
| 100111 | 200ml | Single Shot | Empower CT, K073178 |
| 100121 | 200/200ml | Dual Shots | Empower CT, K073178 |
| The Sterile High-pressure Angiographic Syringes for Single-use models listed below are | | | |
| compatible with the following Nemoto Injector models | | | |
| 100124 | 60/100ml | Dual Shots | CT, K052633 |
| 200104 | 60/60ml | Dual Shots | Sonic Shot, K091734 |
The connecting tubes, which is used to connect the syringe and the catheter. The tubes are also available in three configurations, which are straight tube (used with single shot syringe), type Y or type T tube (used with dual shot syringe) and high pressure tube, used with syringes which can withstanding maximum pressure of 1200 psi.
J shape tube, which is used to draw contrast media/saline into the syringe barrel.
- ડ. Identification of Predicate Device
510(k) Number: K072696 Product Name: ANT Angiographic Syringes Manufacturer: Shenzhen Ant Hi-Tech Industrial Co., Ltd.
- 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- A ISO 10993-7:2008 Biological evaluation of medical devices- Part 7: Ethylene oxide sterilization residuals:
- > ASTM F 88/F88M-09 Standard test method for seal strength of flexible barrier materials;
- > ASTM F1140/F1140M-13 Standard test methods for internal pressurization failure resistance of unrestrained packages:
- > ASTM F1929-12 Standard Test Method for Detecting Seal Leaks in Porous Medical Package by Dye Penetration;
- > USP37-NF32 <85> Bacterial Endotoxins Limit;
- > ISO 11737-2:2009 Sterilization of medical devices- Microbiological methods- Part 2: Test of sterility performed in the definition, validation and maintenance of a sterilization process;
- > ISO 7886-1:1993 Sterile hypodermic syringes for single use - Part 1: Svringes for manual use;
{6}------------------------------------------------
- > ISO 7886-2:1996 Sterile hypodermic syringes for single use - Part 2: Syringes for use with power-driven Syringe Pumps;
- > ISO594:1986 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 1: General Requirements;
- > ISO594:1998 Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment - Part 2: Lock Fitting;
- > ISO 10993-4:2002 AMD 2006 Biological evaluation of medical devices -- Part 4: Selection of tests for interactions with blood:
- > ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity;
- > ISO 10993-10:2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization;
- > ISO 10993-11:2006 Biological evaluation of medical devices -- Part 11: Tests for systemic toxicity;
- > ASTM F 756-08, Standard practice for assessment of hemolytic properties of material
- > USP37-NF32 <151> Pyrogen test
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
- Substantially Equivalent (SE) Comparison 8.
| Item | Proposed Device | Predicate Device |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | DXT | DXT |
| Regulation Number | CFR 870.1650 | CFR 870.1650 |
| Intended Use | Sterile High-pressure Angiographic<br>Syringes for Single-use are intended for<br>the injection of contrast media or saline;<br>they shall be used with an US legally<br>marketed angiographic injectors. | ANT Angiographic Syringes are<br>syringes for the injection of contrast<br>media or saline. This syringe is for<br>single use with US legally marketed<br>angiographic injectors. |
| Mode of operation | Power-driven operation, single use | Power-driven operation, single use |
| Configuration | Angiographic Syringe<br>Connecting tube<br>J shape tube/Spike | Angiographic Syringe<br>Connecting tube<br>J shape tube/Spike |
| Sterility | EO Sterilized | EO Sterilized |
| Single Use | Yes | Yes |
| Performance | Comply with:<br>ISO 7886-1 | Comply with:<br>ISO 7886-1 |
## Table 2 Comparison of Technology Characteristics
{7}------------------------------------------------
| | ISO 594-1 | ISO 594-1 |
|------------------|-------------------------------------|-------------------------------------|
| | ISO 594-2 | ISO 594-2 |
| Biocompatibility | Conforms to the requirements of ISO | Conforms to the requirements of ISO |
| | 10993 series Standards | 10993 series Standards |
{8}------------------------------------------------
#### Substantially Equivalent (SE) Conclusion 9.
Based on the comparison, the intended use, mode of operation, and sterility of proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
In addition, the results of performance tests performed on the proposed device can also demonstrate the proposed devices are complied with FDA recognized standards, which the predicate device was also complied with. The results of biocompatibility studies performed on the proposed device demonstrate that the patient materials used in proposed device are biocompatible.
Based on the comparison above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.