K073592 · Mallinckrodt Inc., Liebel-Flarsheim Business · DXT · May 7, 2008 · Cardiovascular
Device Facts
Record ID
K073592
Device Name
OPTISTAR ELITE INJECTION SYSTEM
Applicant
Mallinckrodt Inc., Liebel-Flarsheim Business
Product Code
DXT · Cardiovascular
Decision Date
May 7, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1650
Device Class
Class 2
Indications for Use
The Optistar Elite Injection System is a contrast delivery system and is designed to inject MR contrast media and flushing solutions into a patient's vascular system to obtain diagnostic images when used with Magnetic Resonance Imaging equipment.
Device Story
Optistar Elite is a contrast delivery system for MRI; injects contrast media and saline into patient vascular system. System comprises power head, console, power control, and power supply. Power head uses two piezoelectric motors driving lead screw rams to expel fluid from syringes. Console provides touch-screen interface for programming injection protocols, viewing status, and controlling system. Remote hand switch allows start/stop. System supports multi-phasic injections, patency checks, and timing bolus features. Used in clinical imaging environments; operated by healthcare professionals. Output is controlled delivery of contrast/saline to enhance diagnostic images. Benefits include improved workflow, user-friendly interface, and compatibility with high-field MRI scanners (up to 3.0 Tesla).
Clinical Evidence
Bench testing only. Verification included unit, integration, and system-level software testing. Validation performed under simulated clinical conditions. Performance in 3.0 Tesla MR environments verified for volume, flow, and pressure accuracy; scanner manufacturers confirmed no interference with MR system operation.
Technological Characteristics
Piezoelectric motor-driven ram system; plastic and non-ferrous materials. Components: Power head, console (color LCD touch screen), power control, power supply. Connectivity: Standalone system. Software-controlled. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4. Sterilization: Injector is non-sterile; syringes/disposables provided sterile.
Indications for Use
Indicated for patients undergoing diagnostic imaging procedures requiring the injection of MR contrast media and flushing solutions into the vascular system using Magnetic Resonance Imaging equipment.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
K073592
**MAY - 7 2008**
COVIDIE
# 510(k) Summary
## 1. Company Identification
Mallinckrodt Inc., Liebel-Flarsheim Business 2111 East Galbraith Road Cincinnati, OH 45237
Establishment Registration: 1518293
## 2. Contact Person
Dale Moore Quality Manager Phone: (513) 948-5771 Fax: (513) 948-5708 Email: Dale.Moore@Covidien.com
## 3. 510(k) Preparation Date
12/19/07
## 4. Device Name
Optistar Elite Injection System Trade Name: Common Name: Power Injector Classification Name Iniector and Syringe, Angiographic (21 CFR 870.1650, product code DXT)
## 5. Device Classification
Class II
#### 6. Indications for Use
The Optistar Elite Injection System is a contrast delivery system and is designed to inject MR contrast media and flushing solutions into a patient's vascular system to obtain diagnostic images when used with Magnetic Resonance Imaging equipment.
### 7. Description of Device
The Optistar Elite is a delivery system designed to inject image enhancing MR contrast media and flushing solutions into the vascular system for the purpose of obtaining enhanced diagnostic images. The Optistar Elite consists of 4 main components and syringes, just like the predicate device:
Power Head- Contains two piezoelectric motors that drive two lead screw rams, the syringe holding mechanisms, the main microprocessor, two sets of forward and reverse keys, one set of accelerator keys, check air key, start key, stop kev, syringe size sensors and indication lamps, and pivot adjustment knob. The Optistar Elite uses a drive system that is similar to the existing Mallinckrodt MR injector in the marketplace. The piezoelectric motors drive the rams that are designed to push against the syringe pushrods (or plunger in 125 ml syringes) to expel the fluid from the barrel of the syringe.
LIEBEL FLARSHEIM BUSINESS 2111 EAST GALBRAITH ROAD CINCINNATI, OH 45237
Page 52 of 78
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Covidien. The logo consists of a black square with a white cross in the middle, followed by the word "COVIDIEN" in black, sans-serif font. The logo is simple and modern, and the text is easy to read.
Console - The console is the main user interface for the injector system. It allows the user to input injection parameters via a touch screen as well as view the results of an injection. The console incorporates a remote hand switch for starting or stopping an injection. It communicates with the Power Head to program and initiate injection protocols, displays the injection status, and displays a timer. The console contains the system on/off control.
Power Control - The power control is the interface between the console and the power head. The power control processes the information input by the user at the console and relays this information to the power head. The power control also supplies DC voltage to the power control receives its power from the power supply. The power control is the servo drive for the motors in the power head.
Power Supply - The power supply converts AC line voltage to power the Console and Power Control. It also passes communications from the console to power control.
Syringes - The Optistar Elite Injection System accommodates the Mallinokrodt Optimark pre-filled syringes (10 ml, 15 ml, 20 ml, 30 ml), the Mallinckrodt pre-filled saline syringes (50ml and 125 ml) and the empty disposable 60 ml Mallinckrodt syringes.
## 8. Substantial Equivalence
The predicate injector to the Optistar Elite Injection System is the Optistar MR Injection System, 510(k) number K984088
The Optistar Elite Injector System maintains the same intended use as the predicate device. The Optistar Elite is intended for injecting MR contrast and flushing solutions for the purpose of enhancing diagnostic imaging of humans. The stated Indications for Use differ from those of the predicate device, in that the predicate device included the magnetic field strength limitation of 1.5 Tesla in the Indications statement itself. The proposed device has been validated for use in a magnetic field strength up to 3.0 Testing it with a variety of scanners with field strengths up to 3.0 Tesla. This limitation has been removed from the Indications statement, and will be addressed in the Warnings and Precautions.
The Optistar Elite Injector System consists of four main components and syringes in the same mamer as the predicate device: a Power Head, a Console, a Power Control and Syringes. Both the Optistar Elite Injector System and the predicate device consist of the same fundamental technology. They are piezoelectric mechanical devices which are software controlled.
The Optistar Elite Injector System is similar to the predicate device in that it contains a second identical drive system as a feature to conveniently inject a flushing solution (i.e. saline) in conjunction with the MR contrast injection.
The primary difference between the Optistar Elite and the predicate device is the addition of several features that make the Elite more user friendly. The addition of a patency check function and timing bolus are the main examples of functionality that was present in the predicate device, but is much more convenient in the Elite. The graphical user interface on the Elite is easier to program and has a more modern look.
Below is a table that compares the predicate device to the proposed Optistar Elite Injection System.
LIEBEL FLARSHEIM BUSINESS 2111 EAST GALBRAITH ROAD Cincinnati, OH 45237
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "COVIDIEN" in bold, black letters. To the left of the word is a black square with a white cross in the middle. The cross is made up of two rectangles, one horizontal and one vertical.
| | Optistar Elite Injection System | Optistar MR Injection System<br>Injection System<br>Predicate Device<br>(K984088) |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | | |
| Intended Use | Intended for injecting MR contrast agents<br>and flushing solutions for the purpose of<br>enhancing diagnostic imaging of humans. | Intended for injecting MR contrast agents<br>and flushing solutions for the purpose of<br>enhancing diagnostic imaging of humans. |
| MR Environment | Up to 3 Tesla. | Up to 1.5 Tesla. |
| Multi-phasic Injections | Yes - multiple phases per protocol | Yes - multiple phases per protocol |
| Protocol Storage | 40 protocols | 40 protocols |
| Injection History | Stores last 48 injection results | Stores last 12 injection results |
| Syringe Sizes | Prefilled Contrast 10ml, 15ml, 20ml, 30ml<br>syringes<br>Prefilled Saline 50ml & 125 ml syringes<br>60 ml disposable syringes | Prefilled Contrast 10ml, 15ml, 20ml,<br>30ml syringes<br>25ml & 60 ml disposable syringes |
| Single or Dual Syringe<br>Modality | Single syringe and Dual syringe capability | Single syringe and Dual syringe<br>capability |
| Syringe Drive System | Piezoelectric | Piezoelectric |
| Syringe Heater | No | No |
| Flow Rate | A side: MR contrast media<br>10, 15, 20, 30ml syringes: 0.1-8.0 ml/sec<br>60 ml syringes: 0.1-10 ml/sec<br>B side: Saline<br>60 ml syringes: 0.1-8.0 ml/sec<br>50 & 125 ml syringes: 0.1-8.0 ml/sec | A side: MR contrast media<br>10, 15, 20, 25, 30ml syringes: 0.1-8.0<br>ml/sec<br>60 ml syringes: 0.1-10 ml/sec<br>B side: Saline<br>60 ml syringes: 0.1-7.0 ml/sec |
| Max Pressure Limit | 200 psi | 200 psi |
| Pressure Limit Control | User-settable<br>10 psi increments | User-settable<br>10 psi increments |
| Flushing System | Manual or via injector | Manual or via injector |
| Ability to check<br>patency | Yes - Manual or using Patency Check | Yes - Manual |
| Ability to run a Test<br>Injection | Yes - Using Timing Bolus feature | Yes - Running 2 separate injections |
| Remote Start | Yes | Yes |
| Scan Delay | 0 - 60 seconds | 0-60 Seconds |
| Phase Delays | Yes | No |
| Inject Delay | Yes | No |
| Console Controls | Touch screen Color LCD | Touch screen Color LCD |
| Power head Controls | Keypad | Buttons |
| Post Injection Readout | Yes | Yes |
| Enable Sequence | Yes | Yes |
| Check for Air Button | Yes | Yes |
| Volume Remaining<br>Display | Displayed on Console | Displayed on Console |
| Mode for keeping fluid<br>path open | Yes - Drip Mode | Yes -- Drip Mode |
| Ram Retraction | Manual or automatic | Manual |
| Materials | Plastic and non-ferrous | Plastic and non-ferrous |
| Sterility | Injectors are not sterile products. Syringes<br>and Disposables are provided sterile | Injectors are not sterile products. Syringes<br>and Disposables are provided sterile. |
이 대한민국 대학교
LIEBEL FLARSHEIM BUSINESS 2111 EAST GALBRAITH ROAD CINCINNATI, OH 45237
: 上一篇:
·
: 上
{3}------------------------------------------------
## 9. Performance Testing
Substantial equivalence between the subject device (Optistar Elite) and the predicate device (Optistar MR) can be demonstrated mostly through verification steps. The following steps were taken:
- Evaluation of Conformity to Standards The Optistar MR and the Optistar Elite both conform 1. to recognized consensus standards, specifically: IEC 60601-1 IEC 60601-1-2 IEC 60601-1-4
- Software Verification and Validation as can be seen by reviewing the Safety and 2. Effectiveness comparison chart, the software differences between the Optistar MR and the Optistar Elite do not impact the safety and effectiveness of the contrast media injector. They have been implemented as conveniences to the user. Software implemented for the Optistar Elite has been verified and validated.
- Verification testing was accomplished through unit, integration, exploratory, and o system level testing. Production equivalent software passed all acceptance criteria in the verification protocol.
- Validation testing was performed by customers on production equivalent units in o simulated clinical conditions. Based on the results, it can be concluded that the Optistar Elite software meets its intended use and meets defined customer needs when thoroughly assessed against the product's Customer Requirements.
- 3. Non-Software Verification
- The product requirements were verified by reviewing the part (or assembly) drawing, . bill of material, labeling and manuals, or other form of data. The Optistar Elite is mechanically and electrically equivalent to Optistar LE, and therefore documentation for Optistar LE was used to provide verification for the Elite.
- When the Optistar MR was introduced, it was validated in a 1.5 Tesla environment, . which represented the standard MR scanner found in imaging facilities at that time. Since then, MR scanners having a higher field strength have been introduced to the market. The Optistar LE injector has been tested with a variety of MR scanners from various manufacturers. These MR systems had magnetic field strengths up to 3.0 Tesla. In this environment, the injector performed as specified, delivering accurate volume, flow and pressures. The power consumption of the injector did not change when operating in this environment. The scanner manufacturers verified that the presence of the Optistar LE injector near the MR scanner did not interfere with the operation of the MR systems.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized caduceus, which is a traditional symbol of medicine, with three intertwined snakes and a staff.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY - 7 2008
Mr. Dale Moore Quality Manager Mallinckrodt Inc., Liebel-Flarsheim Business 2111 East Galbraith Road Cincinnati, OH 45237
Re: K073592
Trade/Device Name: Optistar Elite Injector System Regulation Number: 21 CFR 870.1650 Regulation Name: Angiographic Injector and Syringe Regulatory Class: Class II (two) Product Code: DXT Dated: April 11, 2008 Received: April 14, 2008
Dear Mr. Moore:
We have reviewed your Section 510(k) premarket notification of intent to market the device w oneve and have and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, atores or visions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device to such additional controls. Existing major regulations affecting your device can may or subject to back academic egulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Mr. Dale Moore
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
D. Vachner
A Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
KOZ 3592
510(k) Number (if known): KXXXXX
Device Name: Optistar Elite Injection System
Indications For Use:
The Optistar Elite Injection System is a contrast delivery system and is designed to inject MR contrast media and flushing solutions into a patient's vascular system to obtain diagnostic images when used with Magnetic Resonance Imaging equipment.
Prescription Use _ X (Part 21 CFR 801 Subpart D)
·
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
vmer R., v. Ames
vision Sian-Of Division of Cardiovascular Devices
510(k) Number_k073592
Page 15 of 78
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.