The Vistron CT Injection System is a syringe-based fluid delivery system indicated for delivery of contrast media during computed tomography applications. It is intended to be used for the specific purpose of injecting intravenous contrast medium into the human vascular system for diagnostic studies in computed tomography.
Device Story
Programmable, syringe-based fluid delivery system for CT contrast media; delivers contrast into human vascular system. Components: Injector Head (user interface, mechanical motor/piston, sensors), System Power Unit (power, Imaging System Interface), Remote Monitor (Intelligent Hand Unit for remote control/monitoring), and manual startswitch. Operator (clinician) programs injection parameters (volume, flow rate, pressure limit, scan delay) via Injector Head or Remote Monitor. System monitors injection values against programmed limits; safety stop mechanism triggers if deviations occur. Imaging System Interface (ISI) enables bidirectional communication with CT scanner to initiate/stop injections or trigger scans. Used in clinical settings; provides automated, controlled contrast delivery to enhance diagnostic imaging. Benefits include precise, reproducible contrast administration and integrated scanner synchronization.
Clinical Evidence
Bench testing only. Design verification and validation testing performed to confirm compliance with design specifications and safety standards (IEC 601-1, IEC 601-1-2, IEC 1000 series, UL 2601, CSA C22.2). No clinical data presented.
Technological Characteristics
Syringe-based fluid delivery; 125ml/200ml syringe capacity. Components: Injector Head, System Power Unit, Remote Monitor. Connectivity: DB25 connector for CT scanner interface. Power: 110/220 VAC. Safety: Position encoder, pressure/stall monitoring, software backup system. Standards: IEC 601-1, IEC 601-1-2, IEC 1000-4-6/8/11, IEC 1000-3-2/3, IEC 529, UL 2601, CSA C22.2 No. 601.1-M90. Software: Imaging System Interface module for scanner synchronization.
Indications for Use
Indicated for patients undergoing computed tomography (CT) diagnostic studies requiring intravenous contrast media delivery.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
JUL 2 1 1999
MEDRAD VISTRON CT INJECTION SY
Mav 3, 1999
| Official Contact: | Jan Dobscha, Senior Regulatory Affairs Coordinator<br>Medrad, Inc.<br>One Medrad Drive<br>Indianola, PA 15051<br>(412) 767-2400 ext. 3280 |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Classification: | Injector with Syringe, Angiographic |
| Common/Usual Name: | Powered Injector with Syringe |
| Proprietary Name: | Medrad Vistron CT Injection System |
| Predicate Device: | Medrad Vistron CT Injection System (K982814) |
Medrad, Inc.
Technical Description
One Medrad Drive
Indianola, PA 15051-0780
(412) 767-2400
The Medrad Vistron CT Injection System is a programmable, syringe-based fluid delivery system for delivery of contrast media during computed tomography procedures. The Vistron CT Injector is a modified version of the Medrad Vistron CT Injector. It has the same indications for use and the same intended use as the predicate device. No changes to the fundamental scientific technology were made nor are any new risks introduced by this modification.
The Vistron CT Injector is comprised of three main components - Injector Head, System Power Unit (SPU), and Remote Monitor (a.k.a. Intelligent Hand Unit). The Medrad Vistron CT Injector System, like the predicate Vistron CT Injector System, uses a 125ml and 200ml Front Load Sterile Disposable Svringe. The only difference is that the modified Injector accommodates the use of an alternate 200ml contrast pre-filled syringe, and it now has an Imaging System Interface.
- SYSTEM POWER UNIT (SPU) The System Power Unit houses the injector's main power . supply. It receives 110/220 VAC input and distributes power to the Injector Head via the Injector Head cable and to the Remote Monitor. The SPU has essentially been unchaged from the predicate device. The only change is that it now also contains circuitry for the imaging system interface. (See description below)
- INJECTOR HEAD The Injector Head is the main user interface. It is the main point of user data entry and display feedback data. It contains the mechanical subsystem (motor and plunger piston) that delivers fluid through an attached syringe. It also consists of the syringe interface, position encoder, syringe position/size sensors, user display/controlinterface, electronic controls, and safety circuits. The syringe heat maintainer and the hand held startswitch connect externally to the head.
The Iniector Head processes keyboard inputs from the hard keys on the control panel. Remote Monitor, or hand held startswitch. It displays injection parameters and operating information on the control panel's LEDs, and processes handswitch input. It also provides a standby switch which places the system in a standby mode: that is, it disables power to functional system components but keeps power supplied to all system microprocessors. The only change is that it now contains software for the imaging system interface. (See description below)
Image /page/0/Picture/13 description: The image shows a black and white graphic of the letter Q. The letter is in a bold, sans-serif font and appears to be slightly distorted or textured. Below the letter, there is some indistinct text that is difficult to read due to the image quality. The overall impression is that the image is a close-up of a logo or a design element featuring the letter Q.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "medrad" in a stylized, bold, black font. The letters are slightly slanted to the right, giving the word a dynamic appearance. A registered trademark symbol is present to the upper right of the last letter, indicating that the brand name is protected. The overall design is clean and modern, making the logo easily recognizable.
- REMOTE MONITOR (a.k.a. Intelligent Hand Unit) The purpose of the Remote Monitor is to allow the operator to interact with the Injector Head without being near the Head. It permits the operator to view two phases of the injector's programmed parameters and to control the iniector's Start/Hold and Disarm functions. The Remote Monitor provides LED readouts to display Scan Delay, Injected Volume, Injection, Phase, Flow Rate, and Volume information. It connects to the System Power Unit by means of an integral cord set. It can be placed up to 100 feet from the System Power Unit. The Remote Monitor has essentially been unchanged from the predicate device. The only change is that it now contains software for the imaging system interface. (See description below)
- MANUAL STARTSWITCH A hand-held push button handswitch, connected to the injector . Head via a cable, allows the operator to initiate, suspend, or resume a programmed injection.
- lmaging System Interface (ISI) (Also known as Autolink) A small printed circuit board (pcb) . that resides in the System Power Unit. The pcb acts as an interconnect card for the Injector Head and Remote Monitor, and provides an isolated interface connection between the Injector and CT Scanner. There is one input and one output to the operator. The input is composed of two signal lines and the output has three signal lines. All signal lines are fully isolated in order to provide protection to the Injector and CT Scanner. Physical connection to the CT Scanner is via a DB25 connector. The ISI pcb is controlled by the Rernote Controller. An Injector System Interface software module was added to the Remote Controller software to effect this function. The Injector System Interface module receives inputs from CT Scanner which are used to initiate or stop an injection. It also controls outputs to the CT Scanner which are used to activate Throughout its processing the Imaging System Interface software verifies the a scan. functionality of all input and output signals.
Functional Features: Except for the change to the Syringe Sensing and the addition of the Imaging System Interface, the functional specifications have not changed. The following is a summary of the Injector's specifications.
Programmed Volume - The volume of contrast medium delivered is settable for each phase of the injection profile. The range is 10 mi to 125 ml or 10 ml to 200 ml depending on the size of syringe used.
Programmed Flow Rate - The flow rate is the ratio of fluid volume delivered to the injection duration (milsec) and is settable, in 0.1 ml/sec increments, for each phase of the injection profile. The range is 0.1 ml/sec to 9.9 ml/sec.
Pressure Limiting - The system has a settable pressure limit. The range is 50 psi to 300 psi with increments of 50 psi. In the event of a pressure limit being reached, the injection will proceed at a reduced flow rate until the programmed voiume has been delivered, unless a stall condition occurs. A stall condition will automatically terminate the injection. The user is alerted via audible beeps and an indicator light on the Injector Head control panel.
Program Parameters - The user can program the injector with fluid volume. flow rate. injection duration, pressure limit value, program phase/level, and scan delay parameters.
Injection Profile - The user can program a sequence of one or more program phases, up to a maximum of 4, in which each phase delivers fluid according to the program parameters defined for that phase.
Image /page/1/Picture/10 description: The image shows a logo with a stylized letter 'Q' in a square. Below the square, the words 'Quality for Life' are printed in a simple, sans-serif font. The 'Q' is large and bold, dominating the logo's design.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "medrad" in a stylized, bold, black font. The letters are slightly slanted to the right, giving the word a dynamic appearance. A small registered trademark symbol is located to the upper right of the letter "d".
Hold - "Hold" allows the user to temporarily interrupt an injection. When the user enables the "Hold" function, the system remains armed, but the injection stops until the user cancels the hold function, whereupon the injection resumes from the point at which it was interrupted. The injector will remain in the "Hold" condition for a maximum of ten minutes, after which the unit disarms, an audible beep is emitted and a message is displayed on the Injector Head and Remote Monitor.
Scan Delay - "Scan Delay" is a function that the user can program into the injection profile to indicate to the user that they should begin the scan. The "Scan Delay" interval begins simultaneously with the first phase of the injection profile. The interval duration can be set fiorn seconds to 99 seconds. A visible "Scan Delay" countdown clock, which shows the time remaining in 1 second increments, is located on the Iniector Head control panel and Remote Monitor. When the countdown clock reaches 0 an audible beep emitted from the Injector Head and Remote Monitor alerts the user.
Autofill - "Autofill" allows the user io load a specified amount of fluid into a syringe by actuating a control on the Injector Head. Once the "Autofili" control is activated the head will load the specified amount of fluid into the syringe without further operator interaction. The user may select volumes of 25, 50, 75, 100. 125, 150, 175, or 200ml depending on the syringe size used. The fill rate is fixed at 2ml/sec. "The Autofili feature is non-functionalwhen an alternate 200ml syringe is installed.
Single/Multi Arm · · The user can select one of two arming modes. "Single Arm" allows the execution of a single injection profile after pressing the start switch After the programmed injection has been completed, the unit disarms. "Multipie Arm" allows the programmed injection to be repeated without re-arning the injector. After completing its programmed injection, the injector remains in the armed state. The user can press the Start switch again to repeat the same programmed injection. Repeated activation of the Start switch will deliver the programmed injection until there is insufficient contrast media in the syringe to complete a full injection. At this point, the injector will inform the user, by flashing the Volume Remaining display, that there is insufficient volume for another injection. The operator is complete the injection with iess than the programmed volume available in the syringe.
Syringe Sensing - "Syringe Sensing" is a svstem on the injector head that determines which of the two compatible Medrad syringe sizes ( 125ml, 200ml cr an alternate 200ml contrast prefilled syringe) is installed on the injector nead, and also senses whetner or not the syringe has been installed properly. After determining the size of the syringe that has been installed, the injector adjusts its pressure, volume and flow characteristics accordingly. If the syringe has not been properly installed, the injector issues an error message on the Injector Head control panel alerting the user that the svringe is not correctly installed.
Syringe Heater - The Syringe heater maintains the temperature of preheated contrast medium within the syringe while the syringe is attached to the Injector. The heater operature range is 87.8° to 105.8°F.
Safety Stop - All critical system functions are continuously monitored during an injection to insure that the actual injection values do not deviate from the programmed values beyond a specified limit. A secondary microprocessorin the Injector Head constantly monitors the electrical signals of the injector system. It monitors for over-volume, over flow rate, and pressure/stall; a hardware comparitor circuit monitors for over-pressure. When the backup monitoring system detects one of these conditions, it will open a relay to cut power to the piston motor and the injector system is placed in a maintenance state.
Image /page/2/Picture/8 description: The image shows a logo with a large, stylized letter 'Q' in black, positioned above the words 'Quality for Life' in a smaller font size. The 'Q' is bold and slightly tilted, giving it a dynamic appearance. The text below is aligned to the left, suggesting a brand or organization focused on quality of life.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "medrad" in a stylized, bold, black font. The letters are slightly slanted to the right, giving the word a dynamic appearance. A small registered trademark symbol is located to the upper right of the letter "d".
Retract Control - A control provided to return the piston to its fully retracted station. A single touch of a button is required to activate and complete this operation. This function is only operational when the syringe is detached from the injector system. The Syringe Sensing system detects whether or not a syringe is in place.
Imaging System Interface - An interface between the Injector and CT Scanner. Permits the operator to initiate a scan sequence from the Injector, or to trigger an injection from the CT Scanner. This is a new feature, described in detail above.
The Vistron CT Injection System is a syringe-based fluid delivery system Intended Use: indicated for delivery of cor:trast media during computed tomography applications. It is intended to be used for the specific purpose of injecting intravenous contrast medium into the human vascular system for diagnostic studies in computed tomography.
| Feature | Medrad Vistron CT Injector<br>with ISI & PPCM<br>(Modified Device) | Medrad Vistron CT Injector<br>(K982814)<br>Predicate |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Information Display | Large Colored LEDs | Large Colored LEDs |
| Programming Keys | Non-Dedicated Keys - Software<br>Determined | Dedicated Function Keys |
| Multi-Phase | 1 - 4 Phases per Injection | 1 - 4 Phases per Injection |
| Arming Modes | Single/Multi Arm | Single/Multi Arm |
| Protocol Storage Capability | Stores last protocol used | Stores last protocol used |
| Hold Capability | 0 - 600 seconds | 0 - 600 seconds |
| Scan Delay | 0-99 seconds | 0 - 99 seconds |
| Safety Stop Mechanism | Electrical Stop with a Software<br>Backup System | Electrical Stop with a Software<br>Backup System |
| Syringe System | 200ml, 125ml, Alternate 200ml | 200ml, 125ml |
| Programmed Volume | 1 to 125ml or 1 to 200 ml<br>Depending on Syringe Size | 1 to 125 ml or 1 to 200 ml<br>Depending on Syringe Size |
| Volume Remaining<br>Readout | LED | LED |
| Fill Rate | 2 ml/sec to 9.9 ml/sec | 2 ml/sec to 9.9 ml/sec |
| Flow Rate | Settable .1 ml/sec to 9.9 ml/sec<br>(Settable for each injection during<br>programming and during injection.) | Settable .1 ml/sec to 9.9 ml/sec<br>(Settable for each injection during<br>programming and during injection.) |
| Pressure Limit | Settable from 50 to 300 PSI | Settable from 50 to 300 PSI |
| Autofill | Fill rate fixed at 2 ml/sec. (Non-<br>functional when PPCM is used.) | Fill rate fixed at 2 ml/sec |
| Retract Control | Yes | Yes |
| Imaging System Interface<br>(AutoLink) | Yes | No |
| Remote Start Switch | Yes | Yes |
## Technological Characteristics
Image /page/3/Picture/6 description: The image shows a logo with a large letter Q in a black square. Below the Q, the words "Quality for life" are written in a smaller font. The logo appears to be a symbol or emblem representing quality of life.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "medrad" in a stylized, bold, sans-serif font. The letters are black, and there is a registered trademark symbol to the right of the letter "d". The font appears to be slightly italicized or slanted to the right.
## Supporting Data:
Medrad has established, as part of its Quality System, design controls in compliance with FDA's Quality System Requirements. These design controls are applied to all Medrad product development processes and product design changes. These design controls were applied to the development of the Vistron CT Injector and meet the requirements of the FDA's QSRs.
As part of the design control a risk analysis was performed, and design verification and validation tesling was conducted to support the conclusion drawn by the risk analysis.
## Conclusion:
Test results concluded that the design specifications for the Vistron CT Iniection System were met. The Vistron CT Injection System meets the applicable requirements of the following standards: IEC 601-1, IEC 601-1-2, IEC 1000-4-6, IEC 1000-4-8, IEC 1000-4-11, IEC 1000-3-2, IEC 1000-3-3, IEC 529, UL 2601, and CSA C22.2 No. 601.1-M90. Therefore, it has been determined that the Vistron CT Injection System is substantially equivalent to the predicate device for its intended use when used as prescribed in the User Operation Manual.
Image /page/4/Picture/6 description: The image shows a logo with a stylized letter 'Q' at the top, rendered in a bold, blocky font. Below the 'Q', the words 'Quality' and 'for life' are stacked vertically, written in a smaller, simpler font. The overall design is clean and suggests a focus on quality.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's name around the perimeter. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 2 1 1999
Ms. Jan Dobscha Sr. Regulatory Affairs Coordinator Medrad, Inc. One Medrad Drive Indianola, PA 15051
Re: K991557 Medrad Vistron CT Injection System Trade Name: Requlatory Class: II Product Code: DXT Dated: June 18, 1999 Received: June 21, 1999
Dear Ms. Dobscha:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices
{6}------------------------------------------------
Page 2 - Ms. Jan E. Dobscha
under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4586. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "MEDRAD" in a stylized, blocky font. The letters are all capitalized and appear to be bold. A circled "R" symbol, indicating a registered trademark, is located to the upper right of the word.
:
14 17:11
. . . . . . .
## INDICATION FOR USE
510(k) Number: _ _____________________________________________________________________________________________________________________________________________________________
| Device Name: Medrad Vistron CT Injection System | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
.
Indications for Use/Intended Use:
.
The Vistron CT Injection System is a syringe-based fluid delivery system indicated for delivery of contrast media during computed tomography applications. It is intended to be used for the specific purpose of injecting intravenous contrast medium into the human vascular system for diagnostic studies in computed tomography.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ohrstyn M. Soe Callahan
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number 199155
Prescription Use
(Per 21 CFR 801.109)
✓
OR
Over-The-CounterUse __________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.