ANT ANGIOGRAPHIC SYRINGES, MODEL(S) 100101,102,103,100201,202,203,204,100301,200101,201,202, 203,200301,101,102,202,301
K072696 · Shenzhen Ant Hi-Tech Industrial Co., Ltd. · DXT · May 2, 2008 · Cardiovascular
Device Facts
Record ID
K072696
Device Name
ANT ANGIOGRAPHIC SYRINGES, MODEL(S) 100101,102,103,100201,202,203,204,100301,200101,201,202, 203,200301,101,102,202,301
Applicant
Shenzhen Ant Hi-Tech Industrial Co., Ltd.
Product Code
DXT · Cardiovascular
Decision Date
May 2, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1650
Device Class
Class 2
Indications for Use
ANT Angiographic Syringes are syringes for the injection of contrast media or saline. This syringe is for single use with US legally marketed angiographic injectors.
Device Story
ANT Angiographic Syringes are single-use, disposable plastic syringes (polypropylene or PET) designed for use with US-marketed angiographic injectors. Available in multiple volumes (10mL to 200mL), these syringes facilitate the delivery of contrast media or saline during angiographic procedures. The device is operated by healthcare professionals in clinical settings. The primary design variation across models involves the nozzle connection to ensure compatibility with specific injector manufacturers (e.g., Medrad). The device functions as a mechanical delivery component; it does not involve electronic processing or software. By providing a sterile, compatible interface for contrast injection, the device enables diagnostic imaging procedures.
Clinical Evidence
Bench testing only. Biocompatibility testing conducted per ISO 10993-1:2003(E) standards, including cytotoxicity, sensitization, irritation/intracutaneous reactivity, systemic toxicity, and haemo-compatibility. All tests met requirements.
Technological Characteristics
Materials: Polypropylene or PET. Dimensions: 10mL to 200mL capacities. Connectivity: Mechanical interface (nozzle) for angiographic injectors. Sterilization: Not specified. Energy source: None (mechanical).
Indications for Use
Indicated for the injection of contrast media or saline using US legally marketed angiographic injectors in patients undergoing angiographic procedures.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Disposable CT/MR Syringes for Nemoto Injectors (K051799)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The Assigned 510(k) Number is:
## 1. Applicant Device Information
Trade/Proprietary Name: ANT Angiographic Syringes Common Name: Angiographic injector and syringe. Classification Name: Injector and syringe, angiographic Device Class: II Product Code: DXT Regulation Number: 870.1650 Intended Use: ANT Angiographic Syringes are syringes for the injection of contrast media or saline. This syringe is for single us with US legally marketed angiographic injectors.
### 2. Submitter Information
### Manufacturer Name:
Shenzhen Ant Hi-Tech Industrial Co., Ltd. Building 11, Lishan Industrial Park, Xinghai Ave, Nanshan District, Shenzhen, China
### Contact Person of the Submission:
Ms. Diana. Hong Mr. Eric. Chen Suite 8D, Zhongxin Zhongshan Mansion, No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai, China 20030 Phone: +86-21-64264467 x 152 Fax: +86-21-64264468 x 809 Email: Diana.hong@mid-link.net Eric.chen@mid-link.net
### 3. Predicate Device
### K051799
{1}------------------------------------------------
Trade/Proprietary Name: Disposable CT/MR Syringes for Nemoto Injectors Common/Usual Name: Syringes for CT and MR Injections Classification Name: Angiographic Injector & Syringe
Submitter Name: Coeur, Inc. Address: 704 Cadet Court Lebanon, TN 37087 Phone: (615) 574-7923 Contact: Debra F. Manning, VP, Q & RA
### 4. Device Description
The applicant device of ANT Angiographic Syringes are plastic, single-use, disposable syringes to be offered in 10 mL, 12 mL, 20mL, 60mL, 60ml, 65ml, 100ml, 115 ml, 125 mL, 130ml, 140ml and 200mL sizes. The syringes will be offered made of polypropylene or PET- both materials, of which, are available in current legally marketed products.
Variants: All variants include 7 series as followings: Series 1: Model 100101, Model 100103, Model 100103, Model 100202, Model 100202, Model 100203, Model 100204, Model 100301, Model 100302, Series 2: Model 200101, Model 200201, Model 200202, Model 200203, Model 200301, Series 3: Model 300101, Model 300102, Model 300202, Model 300301, Model 300302, Series 4: Model 400101, Series 5: Model 500101, Model 500102, Series 6: Model TR0012, Series 7: Model CS0010, Model CS0020,
All 8 series follow same design principle, same material, and same intended use. The only differences are volume and the connection (nozzle) to different angiography injector from different manufacturer. The connection differs to fit with US legally Market Angiographic Injectors from company Medrad. The connector difference is only to fit the injector and does not affect the performance of the syringes.
{2}------------------------------------------------
# 5. Substantially Equivalence Determination
# Comparison Analysis:
The Applicant device has the same classification intended use, sterilizations, performance, biocompatibility, chemical specifications and similar physical and mechanications with the predicate device. The only difference between applicant device and predicate device is some physical specifications variant which is too slight to influence the effectiveness and safety.
# Conclusion:
The applicant device is Substantially Equivalent (SE) to the predicate device in terms of Effectiveness and Safety.
{3}------------------------------------------------
# 6. Effectiveness and Safety Considerations
### Effectiveness:
All the variant models of the applicant device are evaluated regarding the performance.
#### Safety Considerations:
With accordance with the Table 1 Initial Evaluation Tests for Consideration and Table 2 Supplementary Evaluation Tests for Consideration in ISO 10993-1:2003(E), Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing, the necessary tests for Biocompatibility Testing includes: Cytotoxicity, Sensitization, Irritation or Intracutaneous Reactivity, Systemic Toxicity (Acute), Haemo-compatibility.
Per 21 CFR 73.3107, Carbazole violet (Pigment Violet 23) (CAS Reg. No. 6358-30-1, Color Index No. 51319) is exempt from the certification requirement of 721(c) of FD&C Act.
Conclusion: The all conducted Biological Evaluation Tests are in compliance with the standards of ISO 10993, "Biological Evaluation of Medical Devices". The compatibility of all the possible skin-contact component material in the finished product meets the requirement of Biocompatibility
The applicant device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the applicant device is determined as safe and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle with three human profiles incorporated into its design. The eagle is facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Shenzhen Ant Hi-Tech Industrial Co., LTD c/o Ms. Diana Hong Suite 8D, No. 19, Lane 999 Zhongshan No.2 Road (S) Shanghai, 200030. China
Re: K072696
Ant Angiographic Syringes, Model 100101,102,103,100201 Regulation Number: 21 CFR 870.1650 Regulation Name: Injector and Syringe, Angiographic Regulatory Class: Class II (two) Product Code: DXT Dated: Undated Received: April 28, 2008
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
MAY - 2 2008
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Ms. Diana Hong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Duina R. hichner
Bram D. Zuckerman, M.D. Director
Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K072696
Device Name: ANT Angiographic Syringes
Indications for Use:
ANT Angiographic Syringes are syringes for the injection of contrast media or saline. This syringe is for single use with US legally marketed angiographic injectors.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dune D. Achter
(Division Sign-Off) Division or Cardiovascular Devices
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
510(k) Number_K022696
۰۱:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.