The CORE 300 Digital Stethoscope is intended to electronically amplify, filter, and transmit body sounds, and to calculate heart rate. The device is intended for use by healthcare professionals or lay users on pediatric and adult patients. For lay users, the device output is not intended to be interpreted or used to take clinical action without consultation with a healthcare professional.
Device Story
Electronic stethoscope; amplifies, filters, and transmits body sounds; calculates heart rate via phonocardiogram (PCG) data. Operates in traditional (passive) or digital (active) modes; mode selected via chestpiece-binaural twist mechanism. Digital mode offers Cardiac, Pulmonary, and All-Purpose audio filter modes. Features integrated chestpiece display for heart rate and device info; connects via Bluetooth Low Energy (BLE) to mobile application. Used by clinicians or lay users in clinical or non-clinical settings. Output aids auscultation; heart rate displayed on device and app. Benefits include enhanced sound visualization and real-time heart rate monitoring.
Clinical Evidence
Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601-1, 60601-1-11, 60601-1-2), software verification/validation, cybersecurity, and bench performance (audio, functional, heart rate measurement).
Technological Characteristics
Electronic stethoscope; 3.7V rechargeable Lithium-ion polymer battery; 20 Hz - 2000 Hz frequency response. Connectivity via Bluetooth Low Energy (BLE). Integrated display and mode/volume buttons. Materials biocompatible per ISO 10993-1. Software-based audio filtering and PCG-based heart rate calculation.
Indications for Use
Indicated for pediatric and adult patients; intended for use by healthcare professionals or lay users to amplify, filter, and transmit body sounds and calculate heart rate.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
July 23, 2026
Eko Health, Inc.
Min Cao
Project Manager, Regulatory Affairs
2100 Powell St.
Suite 300
Emeryville, California 94608
Re: K261694
Trade/Device Name: CORE 300 Digital Stethoscope
Regulation Number: 21 CFR 870.1875
Regulation Name: Stethoscope
Regulatory Class: Class II
Product Code: DQD
Dated: May 22, 2026
Received: May 22, 2026
Dear Min Cao:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K261694 - Min Cao
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K261694 - Min Cao
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
STEPHEN C. BROWNING -S
CDR Stephen Browning
Assistant Director
Division of Cardiac Electrophysiology,
Diagnostics, and Monitoring Devices
Office of Cardiovascular Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
# **Indications for Use**
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2023
See PRA Statement below.
510(k) Number (if known)
K261694
Device Name
CORE 300 Digital Stethoscope
Indications for Use (Describe)
The CORE 300 Digital Stethoscope is intended to electronically amplify, filter, and transmit body sounds, and to calculate heart rate. The device is intended for use by healthcare professionals or lay users on pediatric and adult patients.
For lay users, the device output is not intended to be interpreted or used to take clinical action without consultation with a healthcare professional.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
FORM FDA 3881 (6/20)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
# GENERAL INFORMATION
Applicant:
Eko Health, Inc.
2100 Powell Street, Suite 300
Emeryville, CA 94608
Phone: 844-356-3384
Contact Person:
Min Cao
Project Manager, Regulatory Affairs
Eko Health, Inc.
2100 Powell Street, Suite 300
Emeryville, CA 94608
Date Prepared: May 20, 2026
# DEVICE INFORMATION
Trade/Proprietary Name: CORE 300 Digital Stethoscope
Regulation number: 21 CFR 870.1875
Device Classification Name: Electronic Stethoscope
Regulatory Class: Class II
Product Code: DQD
# PREDICATE DEVICE
Predicate Device: Eko CORE (K200776)
Reference Device: CORE 500 Digital Stethoscope (K233609)
# DEVICE DESCRIPTION
The CORE 300 Digital Stethoscope (CORE 300) is an electronic stethoscope intended to amplify, filter, and transmit body sounds. The device has a form factor consistent with a traditional stethoscope and includes a chestpiece with an integrated display and a semi-permanently attached binaural earpiece. The device operates in two modes: traditional mode, when powered off, and digital mode, when powered on. In traditional mode, the device functions as a traditional stethoscope without requiring battery power. The operating mode is selected via a twist mechanism at the connection between the chestpiece and the binaural; rotating to the ON position activates digital mode, and rotating to the OFF position returns the device to traditional mode.
In digital mode, the CORE 300 Digital Stethoscope provides three audio filter modes for
K261694
Page 1 of 7
{5}
auscultation that process acoustic data to emphasize frequency ranges associated with specific body sounds: Cardiac Mode for heart sounds, Pulmonary Mode for lung sounds, and All-Purpose Mode for general auscultation. The device also calculates heart rate in real time using phonocardiogram (PCG) data. Heart rate is displayed on the chestpiece display and, when connected, in the associated software applications.
## INDICATIONS FOR USE
The CORE 300 Digital Stethoscope is intended to electronically amplify, filter, and transmit body sounds, and to calculate heart rate. The device is intended for use by healthcare professionals or lay users on pediatric and adult patients.
For lay users, the device output is not intended to be interpreted or used to take clinical action without consultation with a healthcare professional.
## SUBSTANTIAL EQUIVALENCE
The CORE 300 Digital Stethoscope has the same general intended use and similar technological characteristics as the predicate device, Eko CORE (K200776). Both devices are electronic stethoscopes intended to electronically amplify, filter, and transmit body sounds for use on pediatric and adult patients by healthcare professionals or lay users.
CORE 300 additionally calculates heart rate using phonocardiogram (PCG) data and includes an integrated display for presentation of device-related information, including heart rate, as applicable. These differences do not alter the fundamental intended use or principle of operation of the device as an electronic stethoscope because the heart-rate calculation is derived from the same body sound data acquired during auscultation, and the display represents an additional location for presenting device information. These differences do not introduce a new signal modality. CORE 500 Digital Stethoscope (K233609) is cited as a reference device to support the PCG-based heart-rate calculation methodology, display-related comparison, and related performance considerations.
The differences in technological characteristics between CORE 300 and the predicate device do not raise different questions of safety and effectiveness. Performance testing supports that CORE 300 is substantially equivalent to the predicate device.
A substantial equivalence comparison table between the subject device, predicate device, and reference device is provided below.
K261694
Page 2 of 7
{6}
Table 1 Substantial Equivalence Summary Comparison
| Feature | Subject Device: CORE 300 Digital Stethoscope | Predicate Device: Eko CORE (K200776) | Reference Device: CORE 500 Digital Stethoscope (K233609) | Comparison Conclusion |
| --- | --- | --- | --- | --- |
| Device Classification Name | Electronic Stethoscope | Electronic Stethoscope | Electronic Stethoscope | Same as the predicate device |
| Regulation Number | 21 CFR 870.1875 | 21 CFR 870.1875 | 21 CFR 870.1875 | Same as the predicate device |
| Classification Product Code | DQD | DQD | DQD, DPS | Same as the predicate device (The reference device includes an additional DPS product code due to ECG functionality, which is not included in CORE 300) |
| Indications for Use | The CORE 300 Digital Stethoscope is intended to electronically amplify, filter, and transmit body sounds, and to calculate heart rate. The device is intended for use by healthcare professionals or lay users on pediatric and adult patients. For lay users, the device output is not intended to be interpreted or used | The Eko CORE is an electronic stethoscope that enables amplification, filtering, and transmission of auscultation sound data (heart, lungs, bowel, arteries, and veins), whereby a clinician at one location on network can listen to the auscultation sounds of a patient on site or at a different location on the network. Eko CORE is intended for | The CORE 500 Digital Stethoscope is intended to be used by clinicians or lay users to electronically amplify, filter, and transfer body sounds and three lead electrocardiogram (ECG) waveforms. The CORE 500 Digital Stethoscope also displays ECG waveforms and heart rate on the display and accompanying mobile application (when prescribed or | Same general intended use and user groups as the predicate device |
K261694
Page 3 of 7
{7}
| Feature | Subject Device: CORE 300 Digital Stethoscope | Predicate Device: Eko CORE (K200776) | Reference Device: CORE 500 Digital Stethoscope (K233609) | Comparison Conclusion |
| --- | --- | --- | --- | --- |
| | to take clinical action without consultation with a healthcare professional. | use on pediatric and adult patients. The Eko CORE is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis. | used under the care of a clinician or by lay users). A lay user is not intended to interpret or take clinical action based on the device output without consulting with a qualified healthcare professional. | |
| Patient Population | Adults and pediatric patients | Adults and pediatric patients | Adults and pediatric patients | Same as the predicate device |
| Prescription/ OTC Status | Over-the-Counter Use | Over-the-Counter Use | Over-the-Counter Use | Same as the predicate device |
| **Technological Characteristics Comparison** | | | | |
| Type of Data Acquired | Body sounds | Body sounds | Body sounds and ECG | Same as the predicate device |
| Connectivity | Bluetooth Low Energy (BLE) | Bluetooth Low Energy (BLE) | Bluetooth Low Energy (BLE) | Same as the predicate device |
| Battery Type | Rechargeable 3.7 V Lithium-ion polymer cell | Rechargeable 3.7 V Lithium-ion polymer cell | Rechargeable Lithium-ion, 3.7V | Same as the predicate device |
| Audio Frequency Response Range | Bandwidth of 20 Hz - 2000 Hz | Bandwidth of 20 Hz - 2000 Hz | Bandwidth of 20 Hz - 2000 Hz | Same as the predicate device |
| Mechanism of Action | User places the device on the body for auscultation. The device captures Audio data which is | User places the device on the body for auscultation. The device captures Audio data which is | User places the device on the body for auscultation. The device simultaneously | Same as the predicate device |
K261694
Page 4 of 7
{8}
| Feature | Subject Device: CORE 300 Digital Stethoscope | Predicate Device: Eko CORE (K200776) | Reference Device: CORE 500 Digital Stethoscope (K233609) | Comparison Conclusion |
| --- | --- | --- | --- | --- |
| | transmitted via Bluetooth to the Mobile application | transmitted via Bluetooth to the Mobile application. | captures ECG, and Audio data which is transmitted via Bluetooth to the Mobile application. | |
| Hardware Interface | Display Mode Button Volume Button | Mode Button Volume Button | Display Mode Button Volume Button Capacitive Touch | Similar to the predicate device. CORE 300 includes a display, which is a technological difference from the predicate device and is consistent with the reference device. This difference does not alter the device's fundamental principle of operation or raise different questions of safety and effectiveness |
| Software Interface | Mobile Application (Eko App) | Mobile Application (Eko App) | Mobile Application (Eko App) | Same as the predicate device |
| Recording and Playback | Yes | Yes | Yes | Same as the predicate device |
| Heart Rate Detection | PCG-based | N/A | PCG-based | Additional feature compared to the predicate device; PCG-based |
K261694
Page 5 of 7
{9}
| Feature | Subject Device: CORE 300 Digital Stethoscope | Predicate Device: Eko CORE (K200776) | Reference Device: CORE 500 Digital Stethoscope (K233609) | Comparison Conclusion |
| --- | --- | --- | --- | --- |
| | | | | heart-rate calculation is consistent with the reference device and supported by performance testing |
| Audio Filter | Cardiac Mode Pulmonary Mode All-Purpose Mode | Cardiac Mode Pulmonary Mode Wide Band Mode | Cardiac Mode Pulmonary Mode Wide Band Mode | Same audio filter mode categories as the predicate device. CORE 300 uses the user-facing term “All-Purpose Mode” for the Wide Band Mode. This terminology difference does not change the underlying audio filtering functionality or raise different questions of safety and effectiveness. |
| Operating Mode | Traditional Mode Digital Mode | Traditional Mode Digital Mode | Digital Mode | Same as the predicate device |
K261694
Page 6 of 7
{10}
# PERFORMANCE DATA - NONCLINICAL TESTING SUMMARY
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility
A biological evaluation was conducted for the patient-contacting materials of the CORE 300 Digital Stethoscope in accordance with the applicable FDA-recognized biocompatibility framework and ISO 10993-1. The evaluation concluded that the device is biocompatible for its intended patient-contacting use.
## Electrical Safety and Electromagnetic Compatibility
Electrical safety and electromagnetic compatibility testing were conducted to demonstrate that CORE 300 complies with applicable IEC 60601-1, IEC 60601-1-11, and IEC 60601-1-2 requirements.
## Software Verification and Validation
Software verification and validation activities were conducted in accordance with FDA's guidance, Content of Premarket Submissions for Device Software Functions. The testing verified that the device software functions perform as intended and support the device's intended use.
## Cybersecurity
Cybersecurity documentation and risk management activities were provided in accordance with FDA's current cybersecurity guidance for connected medical devices. The assessment supports that the cybersecurity risks associated with the device and associated mobile application are appropriately controlled for the intended use environment.
## Bench Performance Testing
Bench performance testing was conducted to demonstrate that the CORE 300 Digital Stethoscope performs as intended and that differences from the predicate device do not raise different questions of safety and effectiveness. Testing included:
- Audio performance
- Electrical and mechanical functional verification
- Heart rate measurement performance
# CONCLUSIONS
Based on the intended use, technological characteristics, risk analysis, and performance testing, the CORE 300 Digital Stethoscope is substantially equivalent to the predicate device, Eko CORE (K200776). The differences between CORE 300 and the predicate device do not raise different questions of safety and effectiveness. The nonclinical performance data support that CORE 300 is as safe and effective as, and performs as well as, the legally marketed predicate device.
K261694
Page 7 of 7
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.