K233609 · Eko Health, Inc. · DQD · Mar 28, 2024 · Cardiovascular
Device Facts
Record ID
K233609
Device Name
CORE 500 Digital Stethoscope
Applicant
Eko Health, Inc.
Product Code
DQD · Cardiovascular
Decision Date
Mar 28, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1875
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The CORE 500 Digital Stethoscope is intended to be used by clinicians or lay users to electronically amplify, filter, and transfer body sounds and three lead electrocardiogram (ECG) waveforms. The CORE 500 Digital Stethoscope also displays ECG waveforms and heart rate on the display and accompanying mobile application (when prescribed or used under the care of a clinician or by lay users). A lay user is not intended to interpret or take clinical action based on the device output without consulting with a qualified healthcare professional.
Device Story
CORE 500 Digital Stethoscope is an electronic stethoscope with integrated electrodes for 3-lead ECG; consists of chestpiece, detachable earpiece, and mobile application (Eko App). Device captures body sounds via diaphragm and ECG via dry electrodes; transmits data via Bluetooth Low Energy (BLE) to mobile app. Features three auscultation modes (Cardiac, Pulmonary, Wide Band) to filter/enhance specific frequency ranges. Computes heart rate in real-time from phonocardiogram (PCG) data. Used in professional healthcare facilities or home settings by clinicians or lay users. Displays ECG waveforms and heart rate on device screen and app. Assists clinicians in physical assessment; lay users must consult professionals for clinical interpretation. Benefits include enhanced auscultation and integrated ECG monitoring.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including audio performance, electrical/mechanical function verification, heart rate measurement, biocompatibility (ISO 10993-1), electrical safety/EMC (IEC 60601-1-11, IEC 60601-2-47, IEC 60601-1-2), software verification/validation, and human factors usability testing (IEC 62366-1).
Technological Characteristics
Electronic stethoscope with 3 dry ECG electrodes. Audio bandwidth 20-2000 Hz; ECG frequency response 0.1-250 Hz. Connectivity: Bluetooth Low Energy (BLE). Power: 3.7V rechargeable lithium-ion battery. Interface: display, mode button, volume button, capacitive touch. Standards: ISO 10993-1 (biocompatibility), IEC 60601-1-11, IEC 60601-2-47, IEC 60601-1-2 (safety/EMC), IEC 62366-1 (usability).
Indications for Use
Indicated for adults and pediatric patients for electronic amplification, filtering, and transfer of body sounds and 3-lead ECG waveforms. Intended for use by clinicians or lay users. Lay users must consult a healthcare professional for interpretation or clinical action.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
March 28, 2024
Eko Health, Inc. Sam Huang Director of Regulatory Affairs 2100 Powell Street Suite 300 Emeryville, California 94608
Re: K233609
Trade/Device Name: CORE 500 Digital Stethoscope Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DQD, DPS Dated: November 8, 2023 Received: November 8, 2023
Dear Sam Huang:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE
by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
K233609
Device Name
CORE 500 Digital Stethoscope
Indications for Use (Describe)
The CORE 500 Digital Stethoscope is intended to be used by clinicians or lay users to electronically amplify, filter, and transfer body sounds and three lead electrocardiogram (ECG) waveforms. The CORE 500 Digital Stethoscope also displays ECG waveforms and heart rate on the display and accompanying mobile application (when prescribed or used under the care of a clinician or by lav users).
A lay user is not intended to interpret or take clinical action based on the device output without consulting with a qualified healthcare professional.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## GENERAL INFORMATION
Applicant: Eko Health, Inc. 2100 Powell Street, Suite 300 Emeryville, CA 94608 Phone: 844-356-3384
Contact Person: Sam Huang, Ph.D. Director of Regulatory Affairs Eko Health, Inc. 2100 Powell Street, Suite 300 Emeryville, CA 94608
Date Prepared: November 9, 2023
## DEVICE INFORMATION
Trade/Proprietary Name: CORE 500 Digital Stethoscope Regulation number: 21 CFR 870.1875 Device Classification Name: Electronic Stethoscope Regulatory Class: Class II Product Code: DQD, DPS
## PREDICATE DEVICE
Primary Predicate: CORE 500 Digital Stethoscope (K230111) Reference Device: Eko CORE (K200776)
## DEVICE DESCRIPTION
CORE 500 Digital Stethoscope (CORE 500) is an electronic stethoscope with integrated electrodes for electrocardiogram (ECG). The device consists of a chestpiece, detachable earpiece (Eko Earpiece) and a mobile application (Eko App) and is intended as a digital auscultation tool on patients requiring physical assessment by the clinicians or lay users. CORE 500 provides the ability to amplify, filter, and transfer body sounds with the chestpiece diaphragm, and three lead ECG through electrodes integrated around the chestpiece. The device can be used in a professional healthcare facility and for home use.
CORE 500 features three auscultation modes for a better auscultation experience by filtering acoustic data and enhancing the primary frequency range of particular body sounds: Cardiac Mode for heart sounds, Pulmonary Mode for lung sounds, and Wide Band Mode for general auscultation. CORE 500 also detects and computes the heart rate in real time based on the phonocardiogram (PCG) data.
{5}------------------------------------------------
# INDICATIONS FOR USE
The CORE 500 Digital Stethoscope is intended to be used by clinicians or lay users to electronically amplify, filter, and transfer body sounds and three lead electrocardiogram (ECG) waveforms. The CORE 500 Digital Stethoscope also displays ECG waveforms and heart rate on the display and accompanying mobile application (when prescribed or used under the care of a clinician or by lay users).
A lay user is not intended to interpret or take clinical action based on the device output without consulting with a qualified healthcare professional.
# SUBSTANTIAL EQUIVALENCE
The CORE 500 Digital Stethoscope (CORE 500) has the same intended use, the same technological characteristics as its predicate device. The differences in user groups do not raise new questions on safety or effectiveness compared to the predicate device. A substantial equivalence comparison table between the subject device and the predicate device is provided below.
| Feature | Subject Device:<br>CORE 500 Digital<br>Stethoscope | Predicate Device:<br>CORE 500 Digital<br>Stethoscope<br>(K230111) | Reference Device:<br>Eko CORE<br>(K200776) | Comparison |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Device<br>Classification<br>Name | Electronic<br>Stethoscope | Electronic<br>Stethoscope | Electronic<br>Stethoscope | Same |
| Regulation<br>Number | 21 CFR 870.1875 | 21 CFR 870.1875 | 21 CFR 870.1875 | Same |
| Classification<br>Product Code | DQD, DPS | DQD, DPS | DQD | Same as the<br>predicate<br>device |
| Indications<br>for Use | The CORE 500<br>Digital Stethoscope<br>is intended to be used<br>by clinicians or lay<br>users to<br>electronically<br>amplify, filter, and<br>transfer body sounds | The CORE 500<br>Digital Stethoscope<br>is intended to be<br>used by clinicians to<br>electronically<br>amplify, filter, and<br>transfer body sounds<br>and three lead | The Eko CORE is<br>an electronic<br>stethoscope that<br>enables<br>amplification,<br>filtering, and<br>transmission of<br>auscultation sound | Same<br>intended use<br>as the<br>predicate<br>device,<br>different user<br>groups.<br>The user |
| Feature | Subject Device:<br>CORE 500 Digital<br>Stethoscope | Predicate Device:<br>CORE 500 Digital<br>Stethoscope<br>(K230111) | Reference Device:<br>Eko CORE<br>(K200776) | Comparison |
| | and three lead<br>electrocardiogram<br>(ECG) waveforms.<br>The CORE 500<br>Digital Stethoscope<br>also displays ECG<br>waveforms and heart<br>rate on the display<br>and accompanying<br>mobile application<br>(when prescribed or<br>used under the care<br>of a clinician or by<br>lay users).<br>A lay user is not<br>intended to interpret<br>or take clinical action<br>based on the device<br>output without<br>consulting with a<br>qualified healthcare<br>professional. | electrocardiogram<br>(ECG) waveforms.<br>The CORE 500<br>Digital Stethoscope<br>also displays ECG<br>waveforms and heart<br>rate on the display<br>and accompanying<br>mobile application<br>(when prescribed or<br>used under the care<br>of a clinician).<br>The data offered by<br>the device is only<br>significant when<br>used in conjunction<br>with clinician<br>evaluation as well as<br>consideration of<br>other relevant patient<br>data. | data (heart, lungs,<br>bowel, arteries, and<br>veins), whereby a<br>clinician at one<br>location on network<br>can listen to the<br>auscultation sounds<br>of a patient on site<br>or at a different<br>location on the<br>network. Eko<br>CORE is intended<br>for use on pediatric<br>and adult patients.<br>The Eko CORE is<br>intended to be used<br>by professional<br>users in a clinical<br>environment or by<br>lay users in a<br>nonclinical<br>environment. The<br>device is not<br>intended for<br>self-diagnosis. | group<br>difference<br>does not raise<br>different<br>questions of<br>safety and<br>effectiveness |
| Patient<br>Population | Adults and pediatric<br>patients | Adults and pediatric<br>patients | Adults and<br>pediatric patients | Same |
| Prescribed | Over-the-Counter<br>Use | Prescription Only | Over-the-Counter<br>Use | The proposed<br>OTC use is<br>the same as<br>the reference<br>device |
| Technological Characteristics Comparison | | | | |
| Type of Data<br>Acquired | Body sounds and<br>ECG | Body sounds and<br>ECG | Body sounds | Same as the<br>predicate<br>device |
| Connectivity | Bluetooth Low | Bluetooth Low | Bluetooth | Same as the |
| Feature | Subject Device:<br>CORE 500 Digital<br>Stethoscope | Predicate Device:<br>CORE 500 Digital<br>Stethoscope<br>(K230111) | Reference Device:<br>Eko CORE<br>(K200776) | Comparison |
| Energy (BLE) | Energy (BLE) | Energy (BLE) | | predicate<br>device |
| Battery Type | Rechargeable<br>Lithium-ion, 3.7V | Rechargeable<br>lithium-ion, 3.7V | Rechargeable 3.7 V<br>Lithium-ion<br>polymer cell | Same as the<br>predicate<br>device |
| Audio<br>Frequency<br>Response<br>Range | Bandwidth of 20 Hz<br>- 2000 Hz | Bandwidth of 20 Hz<br>- 2000 Hz | Bandwidth of 20 Hz - 2000 Hz | Same |
| ECG<br>Performance<br>(ECG<br>Frequency<br>Response<br>Range) | 0.1 - 250 Hz | 0.1 - 250 Hz | N/A | Same as the<br>predicate<br>device |
| Mechanism<br>of Action | User places the<br>device on the body<br>for auscultation. The<br>device<br>simultaneously<br>captures ECG, and<br>Audio data which is<br>transmitted via<br>Bluetooth to the<br>Mobile application. | User places the<br>device on the body<br>for auscultation. The<br>device<br>simultaneously<br>captures ECG, and<br>Audio data which is<br>transmitted via<br>Bluetooth to the<br>Mobile application. | User places the<br>device on the body<br>for auscultation.<br>The device captures<br>Audio data which is<br>transmitted via<br>Bluetooth to the<br>Mobile application. | Same as the<br>predicate<br>device |
| No. of ECG<br>Electrodes | Three (3) dry<br>electrodes | Three (3) dry<br>electrodes | N/A | Same as the<br>predicate<br>device |
| Hardware<br>Interface | Display<br>Mode Button<br>Volume Button<br>Capacitive touch | Display<br>Mode Button<br>Volume Button<br>Capacitive touch | Volume Button | Same as the<br>predicate<br>device |
| Software<br>Interface | Mobile Application<br>(Eko App) | Mobile Application<br>(Eko App) | Mobile Application<br>(Eko App) | Same |
| Feature | Subject Device:<br>CORE 500 Digital<br>Stethoscope | Predicate Device:<br>CORE 500 Digital<br>Stethoscope<br>(K230111) | Reference Device:<br>Eko CORE<br>(K200776) | Comparison |
| Recording<br>and Playback | Yes | Yes | Yes | Same |
| Heart Rate<br>Detection | PCG-based | PCG-based | N/A | Same as the<br>predicate<br>device |
| Visualization<br>of Sound and<br>ECG<br>Waveforms | Yes | Yes | N/A | Same as the<br>predicate<br>device |
| Auscultation<br>Mode | Cardiac Mode<br>Pulmonary Mode<br>Wide Band Mode | Cardiac Mode<br>Pulmonary Mode<br>Wide Band Mode | Cardiac Mode<br>Pulmonary Mode<br>Wide Band Mode | Same as the<br>predicate<br>device |
Table 1 Substantial Equivalence Summary Comparison
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# PERFORMANCE DATA - NONCLINICAL TESTING SUMMARY
The following performance data were provided in support of the substantial equivalence determination.
## Biocompatibility
Biological evaluations were conducted with the CORE 500 Digital Stethoscope according to ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process. The evaluation report concluded that the CORE 500 Digital Stethoscope is biocompatible.
## Electrical Safety and Electromagnetic Compatibility (EMC)
Electrical safety and EMC testing were conducted and demonstrated CORE 500 Digital Stethoscope complies with IEC 60601-1-11, IEC 60601-2-47 standards for safety, IEC 60601-1-2 for electromagnetic compatibility.
## Software Verification and Validation Testing
The software of the subject device is verified and validated according to the FDA guidance for the Content of Premarket Submissions for Device Software Functions.
## Usability Testing
Human factors usability testing was conducted per IEC 62366-1 to evaluate that the intended users of the CORE 500 Digital Stethoscope are able to achieve its intended use with the help of the Instructions for Use.
{9}------------------------------------------------
## Nonclinical Testing
The CORE 500 Digital Stethoscope has gone through rigorous bench testing to demonstrate the product performance, and confirmed that differences between the subject and predicate device do not raise different questions of safety and effectiveness. The testing includes the following:
- Audio performance ●
- Electrical and mechanical function verification, and .
- Heart rate measurement
## CONCLUSIONS
The indications for use, technological characteristics and performance testing support that the proposed device, the CORE 500 Digital Stethoscope is substantially equivalent, and as safe and effective as the predicate device, and raises no new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.