The Eko CORE is an electronic stethoscope that enables amplification, filtering, and transmission of auscultation sound data (heart, lungs, bowel, arteries, and veins), whereby a clinician at one location on network can listen to the auscultation sounds of a patient on site or at a different location on the network. Eko CORE is intended for use on pediatric and adult patients. The Eko CORE is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis.
Device Story
Electronic stethoscope attachment for analog stethoscopes; captures heart, lung, bowel, arterial, and venous sounds. Device converts analog audio to digital data; transmits via Bluetooth to mobile application. App provides visualization, storage, playback, and sharing. Used by clinicians for remote monitoring/telemedicine or by lay users to record/share sounds with physicians. Enables transition between traditional and digital auscultation. Benefits include remote specialist consultation and improved patient monitoring. Device includes volume control, power switch, and LED indicator.
Clinical Evidence
Bench testing only. Includes software verification and validation, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and wireless coexistence testing. Human factors testing conducted for OTC use with passing results.
Technological Characteristics
Electronic attachment for analog stethoscopes; digital/analog auscultation modes. Bluetooth connectivity to mobile app. Features volume adjustment, power switch, LED indicator. Standards met: IEC 60601-1 (General Safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (Home Healthcare).
Indications for Use
Indicated for pediatric and adult patients for amplification, filtering, and transmission of auscultation sounds (heart, lungs, bowel, arteries, veins). Intended for use by professional users in clinical settings or lay users in nonclinical settings. Not for self-diagnosis.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Eko Electronic Stethoscope System (EESS) (K151319)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 6, 2020
Eko Devices, Inc. Arezou Azar Sr. Dir. Regulatory Affairs 1212 Broadway, Suite 100 Oakland, California 94612
Re: K200776
Trade/Device Name: Eko CORE Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DOD Dated: March 23, 2020 Received: March 25, 2020
Dear Arezou Azar:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200776
Device Name Eko CORE
#### Indications for Use (Describe)
The Eko CORE is an electronic stethoscope that enables amplification, filtering, and transmission of auscultation sound data (heart, lungs, bowel, arteries, and veins), whereby a clinician at one location on network can listen to the auscultation sounds of a patient on site or at a different location on the network. Eko CORE is intended for use on pediatric and adult patients. The Eko CORE is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CER 801 Subpart D) </span> </div> | <div> <span> <span style="font-size: 16px;">☑</span> Over-The-Counter Use (21 CER 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
#### Eko Devices, Inc. Eko CORE
Submitter Name Eko Devices, Inc. and Address: 1212 Broadway, Suite 100 Oakland, CA 94612
Contact Person: Arezou Azar, PhD. Sr.Dir of Regulatory Affairs Email: arezou.azar@ekohealth.com Phone: 650-804-0285
Device Trade Name(s): Eko CORE Other Names Used: Eko Electronic Stethoscope System CORE by Eko Common or Usual Name: Electronic Stethoscope Classification Name: Electronic Stethoscope Regulatory Class: Class II Product Code: DQD, 870.1875 Predicate Devices: Eko Electronic Stethoscope System (EESS)(K151319), Eko Devices, Inc. Tyto Stethoscope (K181612) (OTC), Tyto Care Ltd.
#### Device Description
The Eko CORE (OTC) is a digital stethoscope device designed for use by health care professionals as well as lay users in clinical or non-clinical environments. It enables physicians transition between traditional and digital auscultation. It can electronically amplify, filter and transfer sounds to the accompanying mobile application for storage and sharing or can transmit the data for telemedicine use. It also enables regular users to listen to their body sounds (lungs, heart, arteries, veins, Gastrointestinal tract, etc.) and record and share it with their physicians.
lt consists of two primary components: 1) The CORE: an electronic attachment for an analog stethoscope and 2) Eko App, and application that pairs with the CORE attachment.
- . CORE is used to record audio in the stethoscope, convert it to digital data points, and transmit data to a mobile device via Bluetooth". It includes a volume adjustment button, an analog-to-digital power switch, and an LED light indicator.
- . The app captures audio data from the CORE and provides data visualization, secure data storage, audio playback, and sharing features.
These features enable a healthcare professional to monitor patients, seek out second opinions from a specialist or use the device for telemedicine use.
{4}------------------------------------------------
### Intended Use / Indications for Use
The Eko CORE is an electronic stethoscope that enables amplification, filtering, and transmission of auscultation sound data (heart, lungs, bowel, arteries, and veins), whereby a clinician at one location on network can listen to the auscultation sounds of a patient on site or at a different location on the network. Eko CORE is intended for use on pediatric and adult patients. The Eko CORE is intended to be used by professional users in a clinical environment or by lay users in a nonclinical environment. The device is not intended for self-diagnosis.
#### Summary of Technological Characteristics
The Eko CORE is a modified version of the previously-cleared predicate device and has very similar technological characteristics. Minor changes have been made, but the devices have the same overall operational and technological characteristics. Both devices include the ability to attach to a standard analog stethoscope and feature both digital and analog auscultation modes. Additionally, both devices connect via Bluetooth to the Eko App for visualization, recording and transfer of data. Both devices included on/off power buttons and volume adjustment controls. Both devices have the same frequency range and have the same maximum sound level.
A table comparing the key features of the subject and predicate devices is provided below.
#### Performance Data
The device has undergone extensive bench verification, as well as software verification and validation. Electrical safety, EMC and wireless coexistence testing were also successfully conducted.
The OTC use of the device was assessed in human factors testing with passing results.
In all instances, the device functioned as intended. Based on the testing results, the device has a safety and effectiveness profile that is similar to the predicate device.
#### Conclusions
The Eko CORE is as safe and effective as the predicate devices. The Eko CORE has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device. The minor differences in indications do not introduce a new intended use and do not raise different questions of safety and effectiveness when used as labeled. In addition, the minor technological differences between the Eko CORE
{5}------------------------------------------------
and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Eko CORE is as safe and effective as the predicate device. Thus, the Eko CORE is substantially equivalent.
| Parameters | Eko Electronic<br>Stethoscope System<br>(OTC) | Eko Electronic<br>Stethoscope System<br>(Rx) - K151319 | Tyto Stethoscope<br>(OTC)- K181612 |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The Eko CORE is an<br>electronic stethoscope<br>that enables<br>amplification, filtering,<br>and transmission of<br>auscultation sound data<br>(heart, lungs, bowel,<br>arteries, and veins),<br>whereby a clinician at<br>one location on<br>network can listen to<br>the auscultation sounds<br>of a patient on site or at<br>a different location on<br>the network. Eko CORE<br>is intended for use on<br>pediatric and adult<br>patients. The Eko CORE<br>is intended to be used<br>by professional users in<br>a clinical environment<br>or by lay users in a<br>nonclinical<br>environment. The<br>device is not intended<br>for self-diagnosis. | The Eko Electronic<br>Stethoscope System is<br>intended to be used as a<br>part of a physical<br>assessment of a patient<br>by healthcare<br>professionals for<br>diagnostic decision<br>support in clinical<br>settings. Eko is intended<br>for use on pediatric and<br>adult patients. It can<br>electronically amplify,<br>filter and transfer<br>sounds to the<br>accompanying mobile<br>application for storage<br>and sharing. It can used<br>to record heart sounds<br>and cardiac murmurs,<br>bruits, respiratory<br>sounds and abdominal<br>sounds during physical<br>examination in normal<br>patients or those with<br>suspected diseases of<br>the cardiac, vascular,<br>pulmonary or<br>abdominal organ<br>systems. | The Tyto Stethoscope is<br>an electronic<br>stethoscope that<br>enables transmission of<br>auscultation sound data,<br>whereby a<br>clinician at one location<br>on an IP network can<br>listen to the auscultation<br>sounds of a patient on<br>site or at a different<br>location<br>on the IP network with<br>the signal carried on an<br>IP connection between<br>the two locations. The<br>Tyto Stethoscope is<br>intended<br>to be used by<br>professional users in a<br>clinical environment or<br>by lay users in a<br>nonclinical<br>environment. The<br>device is for<br>medical diagnostics<br>purposes only. The<br>device is not intended<br>for self-diagnosis. |
| Standards Met | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 |
| General Safety | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 |
| EMC Safety | IEC 60601-1-11 | | IEC 60601-1-11 |
| Device<br>Classification | DQD: Electronic<br>stethoscope | DQD: Electronic<br>stethoscope | DQD: Electronic<br>stethoscope |
| Prescribed | OTC | Rx | OTC |
| Parameters | Eko Electronic<br>Stethoscope System<br>(OTC) | Eko Electronic<br>Stethoscope System<br>(Rx) - K151319 | Tyto Stethoscope<br>(OTC)- K181612 |
| Classification | Class II | Class II | Class II |
| Stethoscope<br>Type | Attachment to an<br>analog stethoscope<br>(Core) | Attachment to an analog<br>stethoscope (Core) | Stand-alone electronic<br>stethoscope |
| User Interface | On/Off Power button<br>Volume adjustment<br>LED status indicator<br>Mobile App | On/Off Power button<br>Volume adjustment<br>LED status indicator<br>Mobile App | On/Off Power button<br>Volume adjustment<br>LED screen<br>Mobile App |
| Analog/ Digital<br>Interoperability | Yes | Yes | No |
| Connectivity | Bluetooth® | Bluetooth® | Wi-Fi |
| Sound<br>Amplification | Yes | Yes | No |
| Record and<br>Playback<br>Sounds | Yes | Yes | Yes |
| Data Transfer<br>to Compatible<br>Computing<br>platform | Yes | Yes | Yes |
### Table1. Comparison with Predicate Devices
{6}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.