LED Light Therapy Masks (LED Face Mask Pro FAC08NA, LED Chest Mask Pro FAC09NA)
K261052 · Guangdong Newdermo Biotech Co., Ltd. · OHS · Jun 29, 2026 · General, Plastic Surgery
Device Facts
Record ID
K261052
Device Name
LED Light Therapy Masks (LED Face Mask Pro FAC08NA, LED Chest Mask Pro FAC09NA)
Applicant
Guangdong Newdermo Biotech Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Jun 29, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LED Light Therapy Mask (Model: LED Face Mask Pro FAC08NA) is an Over-the-Counter (OTC) device. 1. Red light (630 nm): Intended for treatment of wrinkles. 2. Amber light (605 nm) : Intended for treatment of wrinkles. 3. Red+Near-Infrared (630nm and 830 nm): Intended for treatment of wrinkles. 4. Blue light (415 nm): Intended for treatment of mild to moderate inflammatory acne. 5. Blue and Red (415 nm and 630 nm): Intended for treatment of mild to moderate inflammatory acne. 6. Near-Infrared (830 nm): The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. The LED Light Therapy Mask (Model: LED Chest Mask Pro FAC09NA) is an Over-the-Counter (OTC) device. 1. Red light (630 nm): Intended for treatment of wrinkles. 2. Amber light (605 nm) : Intended for treatment of wrinkles. 3. Red+Near-Infrared (630nm and 830 nm): Intended for treatment of wrinkles. 4. Near-Infrared (830 nm): The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
Device Story
Home-use LED light therapy masks (Face/Chest models) utilize LED energy at specific wavelengths (415nm, 605nm, 630nm, 830nm) to treat wrinkles, acne, and provide therapeutic heating. Device consists of mask, straps (face model), remote control, and charging cable; powered by rechargeable Li-ion battery. User operates device at home; selects light mode via remote. Output is non-ionizing light energy; therapeutic effect achieved through photobiomodulation and topical heating. Benefits include wrinkle reduction, acne management, and temporary pain relief. Healthcare providers do not operate; output used by patient for self-care.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1 series), electromagnetic compatibility (IEC 60601-1-2), photobiological safety (IEC 62471), and biocompatibility (ISO 10993-5, -10, -23).
Indicated for OTC use by adults for treatment of facial/chest wrinkles, mild to moderate inflammatory acne, and topical heating for temporary relief of minor muscle/joint pain, arthritis, muscle spasms, stiffness, and increased local blood circulation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
LED Light Therapy Device (SG-FSM, SG-NK, SG-CT) (K253073)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 29, 2026
Guangdong Newdermo Biotech Co., Ltd.
Junyi Xing
Primary Contact
Bldg. 70(C28), # 9 Huateng Rd., Jinshan Village, Shiqi Town, Panyu District
Guangzhou, Guangdong 511450
China
Re: K261052
Trade/Device Name: LED Light Therapy Masks (LED Face Mask Pro (FAC08NA), LED Chest Mask Pro (FAC09NA))
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, OLP, ILY
Dated: March 31, 2026
Received: March 31, 2026
Dear Junyi Xing:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of
{1}
K261052 - Junyi Xing
Page 2
Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the
{2}
K261052 - Junyi Xing
Page 3
Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
TANISHA L. HITHE -S
Digitally signed by
TANISHA L. HITHE -S
Date: 2026.06.29
11:08:01 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and Infection Control Devices
Office of Product Evaluation and Quality Center for
Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K261052
Device Name
LED Light Therapy Mask
Indications for Use (Describe)
The LED Light Therapy Mask (Model: LED Face Mask Pro FAC08NA) is an Over-the-Counter (OTC) device.
1. Red light (630 nm): Intended for treatment of wrinkles.
2. Amber light (605 nm) : Intended for treatment of wrinkles.
3. Red+Near-Infrared (630nm and 830 nm): Intended for treatment of wrinkles.
4. Blue light (415 nm): Intended for treatment of mild to moderate inflammatory acne.
5. Blue and Red (415 nm and 630 nm): Intended for treatment of mild to moderate inflammatory acne.
6. Near-Infrared (830 nm): The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
The LED Light Therapy Mask (Model: LED Chest Mask Pro FAC09NA) is an Over-the-Counter (OTC) device.
1. Red light (630 nm): Intended for treatment of wrinkles.
2. Amber light (605 nm) : Intended for treatment of wrinkles.
3. Red+Near-Infrared (630nm and 830 nm): Intended for treatment of wrinkles.
4. Near-Infrared (830 nm): The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740 EF
{4}
Traditional 510(k) Submission of LED Light Therapy Mask---
# 510(K) Summary K261052
**Prepared in accordance with the content and format regulatory requirements of 21 CFR Part 807.92**
## 1. Submitter:
510(k) owner's name: Guangdong Newdermo Biotech Co., Ltd.
Establishment Registration Number: 3013178978
Address: Building 70(C28), No.9 Huateng Road, Jinshan Village, Shiqi Town, Panyu District, Guangzhou, Guangdong, 511450, P.R. CHINA
Tel: +86 15012419132
Contact person (including title): Junyi Xing (Manager)
Email: sales13@newdermo.com
Preparing date: February 24, 2026
## 2. Device name and classification:
Device Name: LED Light Therapy Mask
Model: LED Face Mask Pro FAC08NA, LED Chest Mask Pro FAC09NA
Classification Name: Light Based Over The Counter Wrinkle Reduction;
Over-The-Counter Powered Light Based Laser For Acne
Lamp, Infrared, Therapeutic Heating
Product code: OHS, OLP, ILY
Regulation Number: 21 CFR 878.4810
Regulatory Class: Class II
Review Panel: General & Plastic Surgery
## 3. Premarket Notification Class III Certification and Summary
Not applicable, the subject device is Class II.
## 4. Predicate Device(s):
### 1) Predicate device
Sponsor: Sunglor Technology Co., Ltd.
Device name: LED Light Therapy Device (SG-FSM, SG-NK, SG-CT)
510(k) Number: K253073
Regulation Number: 21 CFR 878.4810
Classification Name: Light Based Over The Counter Wrinkle Reduction;
Lamp, Infrared, Therapeutic Heating
Product Code: OHS, ILY
### 2) Reference Device
Guangdong Newdermo Biotech Co., Ltd.
2 / 18
{5}
Traditional 510(k) Submission of LED Light Therapy Mask---
Sponsor: Shenzhen Kaiyan Medical Equipment Co., Ltd.
Device name: TRUDERMAL Pro (ZLD-390)
510(k) Number: K251012
Regulation Number: 21 CFR 878.4810
Classification Name: Light Based Over The Counter Wrinkle Reduction;
Over-The-Counter Powered Light Based Laser For
Acne Lamp, Infrared, Therapeutic Heating
Product Code: OHS, OLP, ILY
# **3) Reference Device2**
Sponsor: Harpar Grace International
Device name: Shani Darden LED light therapy mask
510(k) Number: K214103
Regulation Number: 21 CFR 878.4810
Classification Name: Light Based Over The Counter Wrinkle Reduction;
Over-The-Counter Powered Light Based Laser For
Product Code: Acne OHS, OLP
# **4) Reference Device3**
Sponsor: Ismart Developments, Ltd.
Device name: docoLITE LED Device
510(k) Number: K242382
Regulation Number: 21 CFR 878.4810
Classification Name: Light Based over the counter Wrinkle Reduction
Product Code: OHS
# **5. Reason for Submission**
New device, there were no prior submissions for the device.
# **6. Pre-Submission, IDE**
Not applicable, there is no prior submission.
# **7. Device Description:**
LED Light Therapy Mask is a home use device. LED Light Therapy Mask uses Light Emitting Diode (LED) energy to treatment of wrinkles, treatment of mild to moderate inflammatory acne, provide topical heating for the purpose of elevating tissue temperature, for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm, relieving stiffness, promoting the relaxation of muscle tissue, and to temporarily increase local blood circulation.
There are 4 kinds of light which include Red light (wavelength 630nm), Blue light (wavelength 415nm), Near-Infrared light (wavelength 830nm), Amber light (wavelength 605nm).
There are 2 models of LED Light Therapy Mask, including LED Face Mask Pro FAC08NA, LED Chest Mask Pro FAC09NA.
The product consists of LED Light Therapy Mask, straps (Only LED Face Mask Pro FAC08NA),
Guangdong Newdermo Biotech Co., Ltd.
3 / 18
{6}
Traditional 510(k) Submission of LED Light Therapy Mask
remote control, charging cable.
LED Face Mask Pro FAC08NA is designed for facial application, whereas LED Chest Mask Pro FAC09NA is intended for use on the upper chest.
### **8. Intended Use:**
The LED Light Therapy Mask (Model: LED Face Mask Pro FAC08NA) is an Over-the-Counter (OTC) device.
1. Red light (630 nm): Intended for treatment of wrinkles.
2. Amber light (605 nm) : Intended for treatment of wrinkles.
3. Red+Near-Infrared (630nm and 830 nm): Intended for treatment of wrinkles.
4. Blue light (415 nm): Intended for treatment of mild to moderate inflammatory acne.
5. Blue and Red (415 nm and 630 nm): Intended for treatment of mild to moderate inflammatory acne.
6. Near-Infrared (830 nm): The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
The LED Light Therapy Mask (Model: LED Chest Mask Pro FAC09NA) is an Over-the-Counter (OTC) device.
1. Red light (630 nm): Intended for treatment of wrinkles.
2. Amber light (605 nm) : Intended for treatment of wrinkles.
3. Red+Near-Infrared (630nm and 830 nm): Intended for treatment of wrinkles.
4. Near-Infrared (830 nm): The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
Guangdong Newdermo Biotech Co., Ltd.
4 / 18
{7}
Traditional 510(k) Submission of LED Light Therapy Mask---
## 9. Predicate Device Comparison
**Table 1 Comparison to Predicate/Reference Device for LED Face Mask Pro FAC08NA, for Face**
Guangdong Newdermo Biotech Co., Ltd.
5 / 18
{8}
Traditional 510(k) Submission of LED Light Therapy Mask
| Item | Subject Device | Predicate Device | Reference Device1 | Reference Device2 | Comparison Result |
| --- | --- | --- | --- | --- | --- |
| Trade name | LED Light Therapy Mask | LED Light Therapy Device (SG-FSM, SG-NK, SG-CT) | TRUDERMAL Pro (ZLD-390) | Shani Darden LED light therapy mask | / |
| 510 (k) number | K261052 | K253073 | K251012 | K214103 | / |
| Manufacturer | Guangdong Newdermo Biotech Co., Ltd | Sunglor Technology Co., Ltd. | Shenzhen Kaiyan Medical Equipment Co., Ltd. | Harpar Grace International | / |
| Regulation number | 21 CFR 878.4810; | 21 CFR 878.4810 | 21 CFR 878.4810; | 21 CFR 878.4810 | Same |
| Regulation Name | Light Based Over The Counter Wrinkle Reduction; Over-The-Counter Powered Light Based Laser For Acne; Lamp, Infrared, Therapeutic Heating | Light Based Over The Counter Wrinkle Reduction; Lamp, Infrared, Therapeutic Heating | Light Based Over The Counter Wrinkle Reduction; Over-The-Counter Powered Light Based Laser For Acne; Lamp, Infrared, Therapeutic Heating | Light Based Over The Counter Wrinkle Reduction; Over-The-Counter Powered Light Based Laser For Acne | Similar Note1 |
| Product code | OHS, OLP, ILY | OHS,ILY | OHS, OLP, ILY | OHS, OLP | Similar Note1 |
| Classification | II | II | II | II | Same |
Guangdong Newdermo Biotech Co., Ltd.
6 / 18
{9}
Traditional 510(k) Submission of LED Light Therapy Mask
| Indications for use/ Intended use | The LED Light Therapy Mask (Model: LED Face Mask Pro FAC08NA) is an Over-the-Counter (OTC) device. 1. Red light(630 nm): Intended for treatment of wrinkles. 2. Amber light (605 nm) : Intended for treatment of wrinkles. 3. Red+Near-Infrared(630nm and 830 nm): Intended for treatment of wrinkles. 4. Blue light(415 nm): Intended for treatment of mild to moderate inflammatory acne. 5. Blue and Red (415 nm and 630 nm): Intended for treatment of mild to moderate inflammatory acne. 6. Near-Infrared(830 nm): The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local | The Led Light Therapy Device (Model:SG-FSM) is an Over-the-Counter medical device intended for treatment of wrinkles in the face and decolletage area. Red light: Treatment of wrinkles. Red+Infrared Light: Treatment of wrinkles. Amber light: Treatment of wrinkles. Amber+Infrared Light: Treatment of wrinkles. The Led Light Therapy Device (Model: SG-NK) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area. Red light: Treatment of wrinkles. Red+Infrared Light: Treatment of wrinkles. Amber light: Treatment of wrinkles. Amber+Infrared Light: Treatment of wrinkles. The Led Light Therapy Device (Model:SG-CT) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area. Red light: Treatment of wrinkles. Red+Infrared light: Treatment of wrinkles. Infrared light: Provide topical heating for the purpose of elevating tissue temperature, and to temporarily increase local blood circulation. | TRUDERMAL Pro (Model: ZLD-390) is an over-the-counter(OTC) device that delivering targeted photo therapy using three wavelengths(blue:415nm, red:630nm, Infrared(IR):850nm): Red light: Reduces full-face wrinkles. Blue light: Treats mild to moderate inflammatory acne. IR light: Provides topical heating for the purpose of elevating tissue temperature, temporarily relieving minor muscle and joint pain, arthritis and muscle spasms; relieving stiffness; promoting relaxation of muscle tissues and temporarily increasing local blood circulation. Mixed lights: Red+IR light: Reduces full-face wrinkles. Blue+Red light: Treats mild to moderate inflammatory acne. | The Shani Darden LED light therapy mask is an over-the-counter device intended to emit energy in the red and blue region of the light spectrum, specifically indicated to treat mild to moderate acne vulgaris of the face. The Shani Darden LED light therapy mask is an over-the-counter device intended to emit energy in the red and Near Infra-Red spectrum and is intended for the use in the treatment of full-face wrinkles. | Similar Note2 |
| --- | --- | --- | --- | --- | --- |
Guangdong Newdermo Biotech Co., Ltd.
7 / 18
{10}
Traditional 510(k) Submission of LED Light Therapy Mask
| | blood circulation. | | | | |
| --- | --- | --- | --- | --- | --- |
| Location for use | Face | Face and Decolletage (upper chest) Chest | Face and Body | Face | Similar Note3 |
| OTC or prescription | OTC | OTC | OTC | OTC | Same |
| Power supply | Input: 5V–1A; Li-ion Polymer Battery: DC 3.7V, 2000mAh, 7.4Wh | Input: DC 5V, 1ABattery: Li-ion rechargeable battery 3.7V, 3000mAh | Adaptor Input: 100-240Vac, 50-60Hz, 0.8A Output: 12V – 3A | Electrical Input to power adaptor: 100V-240V, 50/60Hz. Rated at 0.7A. Electrical Output from power adaptor: 12V, 1A | Similar Note4 |
| Power Source | Lithium-Ion Battery (rechargeable) | Lithium-Ion Battery (rechargeable) | 100-240Vac | 100V-240V | Similar Note4 |
| Light source | LED | LED | LED | LED | same |
| Wavelengths | 605nm; 630nm; 660nm; 830nm; 415nm | SG-FSM: NIR: 850nm (face)and 830nm (upper chest) ± 10nm; Red: 630nm ± 10nm; Amber:605nm ± 10nm; SG-NK: NIR: 830nm ± 10nm; Red: 630nm ± 10nm; Amber:605nm ± 10nm; SG-CT: NIR: 830nm ± 10nm; Red: 630nm ± 10nm; | Blue: 415 ± 10nm Red: 630 ± 10nm IR: 850 ± 10nm Mixed light: Blue+Red:415nm+630nm Red+IR:630nm+850nm | Blue: 415nm ± 10nm, Red: 630nm ± 10nm, NIR 830nm ± 10nm. | Similar Note5 |
Guangdong Newdermo Biotech Co., Ltd.
8 / 18
{11}
Traditional 510(k) Submission of LED Light Therapy Mask
| Power Density | M1-Red light (630nm): 21mW/cm² ± 20% M2-Blue light (415nm): 19mW/cm² ± 20% M3-Amber light (605nm): 15mW/cm² ± 20% M4-Near-Infrared light (830nm): 4mW/cm² ± 20% M5-Red light (630nm) +Blue light (415nm):40mW/cm² ± 20% M6-Red light (630nm) + Near-Infrared light (830nm): 29mW/cm² ± 20% | Max Irradiance:850/830nm:3 ± 20%mW/cm²; 630nm:22 ± 20%mW/cm²; 605nm:14 ± 20%mW/cm²; | Blue:6-10mW/cm² Red:10-15mW/cm² IR:6-10mW/cm² Mixed light: Blue+Red: 16-20mW/cm² Red+IR:16-20mW/cm² | Blue 28mW/cm² Red 16mW/cm² Red And Blue Total: 44mW/cm² Red 18mW/cm² NIR 11mW/cm² Red And NIR Total: 29mW/cm² | Similar Note6 |
| --- | --- | --- | --- | --- | --- |
| Treatment time | 10 Minutes | 10/15/20/25/30minutes | 20 min per treatment Acne and Wrinkles: 3-4x weekly; 4 weeks. Pain relief: 3x weekly | 10 Minutes | Similar Note6 |
| Safety and EMC | IEC 60601-1; IEC 60601-1-2 IEC 60601-4-2 IEC 60601-1-11 IEC 60601-2-83 IEC 662471 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 IEC 62471 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-57 IEC 62471 | IEC 60601-1; IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-57 IEC 62471 | Similar |
| Biocompatibility feature | ISO 10993-5 ISO 10993-10 ISO 10993-23 | ISO 10993-1 ISO 10993-5 ISO 10993-10 | ISO 10993-5 ISO 10993-10 | ISO 10993-1 ISO 10993-5 ISO 10993-10 | Similar |
Guangdong Newdermo Biotech Co., Ltd.
9 / 18
{12}
Traditional 510(k) Submission of LED Light Therapy Mask
Comparison in Detail(s):
Note1
The product code/regulation name of the predicate/reference devices includes the product code/regulation name of the subject device. This slight difference will not raise safety and effective issue.
Note2
The indications for use/intended use of the predicate/reference devices includes the indications for use/intended use of the subject device. This slight difference will not raise safety and effective issue.
Note3
The subject device location for use is face (model: LED Face Mask Pro FAC08NA) and upper Chest (model:LED Chest Mask Pro FAC09NA ). The predicate device and the reference devices are include the location for use. This slight difference will not raise safety and effective issue.
Note4
The power supply and the power source of four devices are very similar but not identical, the IEC60601-1test demonstrated the safety of the power adapter. The slight difference will not raise safety and effective issue.
Note5
The predicate device and the reference device have fully covered the wavelengths of the subject device. This slight difference will not raise safety and effective issue.(see Note6 for details).
Note6
1. The M1-Red light(630nm, 21mW/cm² ± 20%, 10mins) of the subject device is close to the Red light (630nm, 22mW/cm² ± 20%, 10/15/20/25/30minutes)of the Predicate device. So, it will not raise safety and effective issue.
2. The M2-Blue light(415nm, 19mW/cm² ± 20%, 10mins) of the subject device's total power density(19×60×10÷1000= 11.4J/cm² ) is between the Reference device1(min:6×60×20÷1000= 7.2J/cm² ; max:10×60×20÷1000= 12J/cm² ). So, it will not raise safety and effective issue.
3. The M3-Amber light(605nm, 15mW/cm² ± 20%, 10mins) of the subject device is close to the Amber light (630nm, 14mW/cm² ± 20%, 10/15/20/25/30minutes)of the Predicate device. So, it will not raise safety and effective issue.
4. The M4-Near-Infrared light(830nm, 4mW/cm² ± 20%, 10mins) of the subject device is close to the Max Irradiance (830/850nm, 3mW/cm² ± 20%, 10/15/20/25/30minutes)of the Predicate device. So, it will not raise safety and effective issue.
5. The M5-Red light and Blue light(630nm+415nm, 40mW/cm² ± 20%, 10mins) is close to the Red light and Blue light(630nm+415nm, 44mW/cm² , 10mins) of
Guangdong Newdermo Biotech Co., Ltd.
10 / 18
{13}
Traditional 510(k) Submission of LED Light Therapy Mask---
the Reference device2. So, it will not raise safety and effective issue.
6. The M6-Red light and Near-Infrared light(630+830nm, 29mW/cm² ± 20%, 10mins) as same as the Red light and Near-Infrared light(630nm+830nm,29mW/cm², 10mins) of the Reference device2. So, it will not raise safety and effective issue.
All the differences don’t affect the safety and effectiveness which is concluded after all the required testing, so no safety and effectiveness issues relating to the system come into conclusion.
Guangdong Newdermo Biotech Co., Ltd.
11 / 18
{14}
Traditional 510(k) Submission of LED Light Therapy Mask
**Table 2 Comparison to Predicate/Reference Device for LED Chest Mask Pro FAC09NA.**
Guangdong Newdermo Biotech Co., Ltd.
12 / 18
{15}
Traditional 510(k) Submission of LED Light Therapy Mask
| Item | Subject Device | Predicate Device | Reference Device3 | Comparison Result |
| --- | --- | --- | --- | --- |
| Trade name | LED Light Therapy Mask | LED Light Therapy Device (SG-FSM, SG-NK, SG-CT) | décoLITE LED Device | / |
| 510 (k) number | K261052 | K253073 | K242382 | / |
| Manufacturer | Guangdong Newdermo Biotech Co.,Ltd | Sunglor Technology Co., Ltd. | Ismart Developments, Ltd. | / |
| Regulation number | 21 CFR 878.4810; | 21 CFR 878.4810 | 21 CFR 878.4810, | Same |
| Regulation Name | Light Based Over The Counter Wrinkle Reduction; Over-The-Counter Powered Light Based Laser For Acne; Lamp, Infrared, Therapeutic Heating | Light Based Over The Counter Wrinkle Reduction ; Lamp, Infrared, Therapeutic Heating | Light Based Over The Counter Wrinkle Reduction | Similar Note8 |
| Product code | OHS, OLP, ILY | OHS,ILY | OHS | Similar Note8 |
| Classification | II | II | II | Same |
Guangdong Newdermo Biotech Co., Ltd.
13 / 18
{16}
Traditional 510(k) Submission of LED Light Therapy Mask
| Indications for use/ Intended use | The LED Light Therapy Mask (Model: LED Chest Mask Pro FAC09NA) is an Over-the-Counter (OTC) device. 1. Red light(630 nm): Intended for treatment of wrinkles. 2. Amber light (605 nm) : Intended for treatment of wrinkles. 3. Red+Near-Infrared(630nm and 830 nm): Intended for treatment of wrinkles. 4. Near-Infrared(830 nm): The device is intended to provide topical heating for the purpose of elevating tissue temperature; for the temporary relief of minor muscle and joint pain, arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation. | The Led Light Therapy Device (Model:SG-FSM) is an Over-the-Counter medical device intended for treatment of wrinkles in the face and decolletage area. Red light: Treatment of wrinkles. Red+Infrared Light: Treatment of wrinkles. Amber light: Treatment of wrinkles. Amber+Infrared Light: Treatment of wrinkles. The Led Light Therapy Device (Model: SG-NK) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area. Red light: Treatment of wrinkles. Red+Infrared Light: Treatment of wrinkles. Amber light: Treatment of wrinkles. Amber+Infrared Light: Treatment of wrinkles. The Led Light Therapy Device (Model:SG-CT) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area. Red light: Treatment of wrinkles. Red+Infrared light: Treatment of wrinkles. Infrared light: Provide topical heating for the purpose of elevating tissue temperature, and to temporarily increase local blood circulation. | The décoLITE LED device is an over-the-counter device that is intended for the use in the treatment of wrinkles in the decolletage area. | Similar Note9 |
| --- | --- | --- | --- | --- |
| Location for use | Decolletage (upper chest) | Face and Decolletage (upper chest) Chest | Decolletage (upper chest) | Similar Note10 |
| OTC or prescription | OTC | OTC | OTC | Same |
| Power supply | Input: 5V~1A; Li-ion Polymer Battery: DC 3.7V, 2000mAh, 7.4Wh | Input: DC 5V, 1ABattery: Li-ion rechargeable battery 3.7V, 3000mAh | not shown | Similar Note11 |
| Power Source | Lithium-Ion Battery (rechargeable) | Lithium-Ion Battery (rechargeable) | Lithium-Ion Battery (rechargeable) | Similar Note11 |
Guangdong Newdermo Biotech Co., Ltd.
14 / 18
{17}
Traditional 510(k) Submission of LED Light Therapy Mask
| Light source | LED | LED | LED | same |
| --- | --- | --- | --- | --- |
| Wavelengths | 605nm; 630nm; 660nm; 830nm; | SG-FSM: NIR: 850nm (face)and 830nm (upper chest) ± 10nm; Red: 630nm ± 10nm; Amber:605nm ± 10nm; SG-NK: NIR: 830nm ± 10nm; Red: 630nm ± 10nm; Amber:605nm ± 10nm; SG-CT: NIR: 830nm ± 10nm; Red: 630nm ± 10nm; | Red: 630nm ± 10nm NIR: 830nm ± 10nm | Similar Note12 |
| Power Density | M1-Red light (630nm): 21mW/cm² ± 20% M2-Amber light (605nm): 15mW/cm² ± 20% M3-Near-Infrared light (830nm): 4mW/cm² ± 20% M4-Red light (630nm) + Near-Infrared light (830nm): 29mW/cm² ± 20% | Max Irradiance:850/830nm:3 ± 20%mW/cm²; 630nm:22 ± 20%mW/cm²; 605nm:14 ± 20%mW/cm²; | 30mW/cm² total | Similar Note13 |
| Treatment time | The session begins and lasts for 10 minutes for modes M1, M2,M3,M4. | 10/15/20/25/30minutes | 600 seconds (10 minutes) | Similar Note13 |
| Safety and EMC | IEC 60601-1; IEC 60601-1-2 IEC 60601-4-2 IEC 60601-1-11 IEC 60601-2-83 IEC 662471 | IEC 60601-1 IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 IEC 62471 | IEC 60601-1; IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 | Similar |
| Biocompatibility feature | ISO 10993-5 ISO 10993-10 ISO 10993-23 | ISO 10993-1 ISO 10993-5 ISO 10993-10 | ISO 10993-1 ISO 10993-5 ISO 10993-10 ISO 10993-11 ISO 10993-23 | Similar |
Guangdong Newdermo Biotech Co., Ltd.
15 / 18
{18}
Traditional 510(k) Submission of LED Light Therapy Mask
Comparison in Detail(s):
Note8
The product code/regulation name of the predicate/reference devices includes the product code/regulation name of the subject device. This slight difference will not raise safety and effective issue.
Note9
The indications for use/intended use of the predicate/reference devices includes the indications for use/intended use of the subject device. This slight difference will not raise safety and effective issue.
Note10
The subject device location for use is face(model: LED Face Mask Pro FAC08NA) and upper Chest (model:LED Chest Mask Pro FAC09NA ). The predicate device and the reference devices are include the location for use. This slight difference will not raise safety and effective issue.
Note11
The power supply and the power source of four devices are very similar but not identical, the IEC60601-1test demonstrated the safety of the power adapter. The slight difference will not raise safety and effective issue.
Note12
The predictive device and the reference device have fully covered the wavelengths of the subject device. This slight difference will not raise safety and effective issue.(see Note13 for details).
Note13
1. The M1-Red light(630nm, 21mW/cm² ± 20%, 10mins) of the subject device is close to the Red light (630nm, 22mW/cm² ± 20%, 10/15/20/25/30minutes)of the Predicate device. So, it will not raise safety and effective issue.
2. The M2-Amber light(605nm, 15mW/cm² ± 20%, 10mins) of the subject device is close to the Amber light (630nm, 14mW/cm² ± 20%, 10/15/20/25/30minutes)of the Predicate device. So, it will not raise safety and effective issue.
3. The M3-Near-Infrared light(830nm, 4mW/cm² ± 20%, 10mins) of the subject device is close to the Max Irradiance (830/850nm, 3mW/cm² ± 20%, 10/15/20/25/30minutes)of the Predicate device. So, it will not raise safety and effective issue.
4. The M4-Red light and Near-Infrared light(630+830nm, 29mW/cm² ±20%, 10mins) is close to the Red light and Near-Infrared light(630nm+830nm, 30mW/cm² , 10mins) of the Reference device3. So, it will not raise safety and effective issue.
All the differences don't affect the safety and effectiveness which is concluded after all the required testing, so no safety and effectiveness issues relating to the system come into conclusion.
10. Performance Data:
Non-clinical data:
Non-clinical tests have been conducted to verify that the Beauty LED Mask meets all Guangdong Newdermo Biotech Co., Ltd.
16 / 18
{19}
Traditional 510(k) Submission of LED Light Therapy Mask
design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate device. The testing results demonstrate that the subject device complies with the following standards:
- ANSI AAMI ES 60601-1:2005/AMD1:2012/AMD2:2020, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014+A1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances – Requirements and tests
- IEC/TR 60601-4-2:2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- IEC 60601-1-11:2015/AMD1:2020, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-83:2019 Medical electrical equipment - Part 2-83: Particular requirements for the basicsafety and essential performance of home light therapy equipment
- IEC 62133-2:2017/AMD1:2021 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells and for batteries made from them, for use in portable applications -Part 2: Lithium systems
- IEC 62471:2016 Photobiological safety of lamps and lamp systems
# Biocompatibility Test
The device has been tested for biocompatibility; it complies with the following standards.
- ISO 10993-5:2009, Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity
- ISO 10993-10:2021, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization
- ISO 10993-23:2021, Biological Evaluation of Medical Devices - Part 23: Tests for irritation
# Software Verification and Validation Testing
Software verification and validation testing were conducted, and Basic documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices."
**Clinical data:** Not applicable.
Guangdong Newdermo Biotech Co., Ltd.
17 / 18
{20}
Traditional 510(k) Submission of LED Light Therapy Mask---
## Summary
Based on the non-clinical and clinical performance as documented in the system development, the subject devices were found to have a safety and effectiveness profile that is similar to the predicate device.
## 11. Conclusion
The non-clinical data supports the safety of the device and the hardware and software verification, and validation demonstrates that LED Light Therapy Mask should perform as intended in the specified use conditions. The device has been tested as described above to show that the device can be used safely and effectively. The subject device LED Light Therapy Mask is as safe, as effective, and performs as well as than the legally marketed predicated devices (K253073, K251012, K214103 and K242382.)
Guangdong Newdermo Biotech Co., Ltd.
18 / 18
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.