K253073 · Sunglor Technology Co., Ltd. · OHS · Nov 19, 2025 · General, Plastic Surgery
Device Facts
Record ID
K253073
Device Name
LED Light Therapy Device (SG-FSM, SG-NK, SG-CT)
Applicant
Sunglor Technology Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Nov 19, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Led Light Therapy Device (Model:SG-FSM) is an Over-the-Counter medical device intended for treatment of wrinkles in the face and decolletage area. Red light: Treatment of wrinkles. Red+Infrared Light: Treatment of wrinkles. Amber light: Treatment of wrinkles. Amber+Infrared Light: Treatment of wrinkles. The Led Light Therapy Device (Model:SG-NK) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area. Red light: Treatment of wrinkles. Red+Infrared Light: Treatment of wrinkles. Amber light: Treatment of wrinkles. Amber+Infrared Light: Treatment of wrinkles. The Led Light Therapy Device (Model:SG-CT) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area. Red light: Treatment of wrinkles. Red+Infrared light: Treatment of wrinkles. Infrared light: Provide topical heating for the purpose of elevating tissue temperature, and to temporarily increase local blood circulation.
Device Story
Over-the-counter LED light therapy device; treats wrinkles in face/decolletage; provides topical heating/increased local blood circulation (SG-CT model). Input: electrical energy from internal rechargeable Li-ion battery. Transformation: LEDs emit specific wavelengths (Red 630nm, Amber 605nm, Infrared 830/850nm) applied directly to skin via flexible silicone panels. Output: light energy/heat. Used in home environment by patient. User selects treatment mode via controller; device provides consistent light administration. Benefits: wrinkle reduction; temporary increase in local blood circulation.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993-5, -10, -23), electrical safety (IEC 60601-1, -1-11, -2-83), electromagnetic compatibility (IEC 60601-1-2), and photobiological safety (IEC 62471).
Indicated for OTC treatment of wrinkles in the face and/or decolletage area using red, amber, and/or infrared light. Model SG-CT also indicated for topical heating to elevate tissue temperature and temporarily increase local blood circulation.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 19, 2025
Sunglor Technology Co.,ltd.
Jane Li
Regulatory supervisor
Room 306, 3nd Floor, building 4 , Fuhai science and Tech.
Industrial Pk, Fuyong community, Fuyong street, Baoan Dist.
Shenzhen, 518000
China
Re: K253073
Trade/Device Name: LED Light Therapy Device (SG-FSM, SG-NK, SG-CT)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, ILY
Dated: September 23, 2025
Received: September 23, 2025
Dear Jane Li:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K253073 - Jane Li
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K253073 - Jane Li
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tanish a Hithe
Digitally signed
by Tanisha Hithe
Date: 2025.11.19
11:23:44 -05'00"
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K253073 | |
| Device Name LED Light Therapy Device (SG-FSM, SG-NK, SG-CT) | |
| Indications for Use (Describe) The Led Light Therapy Device (Model:SG-FSM) is an Over-the-Counter medical device intended for treatment of wrinkles in the face and decolletage area. Red light: Treatment of wrinkles. Red+Infrared Light: Treatment of wrinkles. Amber light: Treatment of wrinkles. Amber+Infrared Light: Treatment of wrinkles. | |
| The Led Light Therapy Device (Model:SG-NK) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area. Red light: Treatment of wrinkles. Red+Infrared Light: Treatment of wrinkles. Amber light: Treatment of wrinkles. Amber+Infrared Light: Treatment of wrinkles. | |
| The Led Light Therapy Device (Model:SG-CT) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area. Red light: Treatment of wrinkles. Red+Infrared light: Treatment of wrinkles. Infrared light: Provide topical heating for the purpose of elevating tissue temperature, and to temporarily increase local blood circulation. | |
| Type of Use (Select one or both, as applicable) ☐ Prescription Use (Part 21 CFR 801 Subpart D) ☑ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
Sponsor: SUNGLOR TECHNOLOGY CO.,LTD
Subject Device: LED Light Therapy Device
# 013_510k Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K253073
Date of Preparation: 2025-11-19
# 1. Information of Submitter and correspondent
## Submitter's information
Submitter Name: SUNGLOR TECHNOLOGY CO.,LTD.
Submitter Address: Room 306, 3nd Floor, building 4, Fuhai science and Technology, Industrial Park, Fuyong community, Fuyong street, Baoan District, Shenzhen, Guangdong, China.
Contact Person (including Title): Mr. Allen.Wu
Tel: +86 13312905381
E-mail: Allen@sunglor-led.com
## Submission correspondent's information:
SUNGLOR TECHNOLOGY CO.,LTD.
Address: Room 306, 3nd Floor, building 4, Fuhai science and Technology, Industrial Park, Fuyong community, Fuyong street, Baoan District, Shenzhen, Guangdong, China.
Contact Person: Jane Li
E-mail:357627482@qq.com
# 2. Subject Device Identification
Type of 510(k) submission: Traditional
Device Trade Name: LED Light Therapy Device
Common Name: Light Based Over The Counter Wrinkle Reduction Lamp, Infrared, Therapeutic Heating
Regulation Name:
Laser surgical instrument for use in general and plastic surgery and in dermatolog Infrared lamp
Regulation Medical Specialty: General & Plastic Surgery, Physical Medicine
Review Panel: General & Plastic Surgery, Physical Medicine
Product Code: OHS, ILY
Regulation Number: 878.4810, 890.5500
{5}
Sponsor: SUNGLOR TECHNOLOGY CO.,LTD
Subject Device: LED Light Therapy Device
Regulation Class: II
# 3. Predicate Device Information
| No. | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 |
| --- | --- | --- | --- |
| Sponsor | Dongguan Boyuan Intelligent Technology Co.,Ltd | ISMART Developments Ltd | Guangdong Newdermo Biotech Co.,Ltd |
| Device Name | LED Light Therapy Device (KFB290, KFB291) | décoLITE LED Device (TN2037) | LED light therapy mask (FM-01, FM-02, FM-03) |
| 510(k) Number | K241857 | K242382 | K223544 |
| Product Code | OHS, OLP, ILY | OHS | OHS, OLP, ILY |
| Regulation Class | Class II | Class II | Class II |
# 4. Device Description
The Led Light Therapy Device is an over-the-counter light emitting diode (LED) device that emits energy for use in dermatology for the treatment of wrinkles.
LED Light Therapy Device(Model:SG-FSM) is consisting of Flexible silicone panel (LED mask and upper chest), controller, silicone eye protector, USB charging cable and straps.
LED Light Therapy Device(Model:SG-NK, SG-CT) is consisting of Flexible silicone panel (upper chest), controller, USB charging cable and straps.
The LED Light Therapy Device(Model:SG-FSM) have 3 types of light, which include Red light (wavelength 630nm), Infrared light (wavelength 850nm for LED mask, wavelength 830nm for upper chest) and Amber light (wavelength 605nm).
The LED Light Therapy Device(model:SG-NK) have 3 types of light, which include Red light (wavelength 630nm), Infrared light (wavelength 830nm) and Amber light (wavelength 605nm).
The LED Light Therapy Device (Model:SG-CT) have 2 types of light, which include Red light (wavelength 630nm) and Infrared light (wavelength 830nm).
The LED Light Therapy Device is applied directly to the skin to ensure consistent administration of light during each treatment. The device does not contain any user serviceable components.
{6}
Sponsor: SUNGLOR TECHNOLOGY CO.,LTD
Subject Device: LED Light Therapy Device
The device operates on an internal lithium rechargeable battery. The rechargeable battery can be charged from the external charging adapter through the provided USB Charging Cable.
The USB charging cable is only used for charging, and the device cannot be used while charging.
# 5. Intended Use/Indications for Use
(1) The Led Light Therapy Device (Model:SG-FSM) is an Over-the-Counter medical device intended for treatment of wrinkles in the face and decolletage area.
Red light: Treatment of wrinkles.
Red+Infrared Light: Treatment of wrinkles.
Amber light: Treatment of wrinkles.
Amber+Infrared Light: Treatment of wrinkles
(2) The Led Light Therapy Device (Model:SG-NK) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area.
Red light: Treatment of wrinkles.
Red+Infrared Light: Treatment of wrinkles.
Amber light: Treatment of wrinkles.
Amber+Infrared Light: Treatment of wrinkles.
(3) The Led Light Therapy Device (Model:SG-CT) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area.
Red light: Treatment of wrinkles.
Red+Infrared light: Treatment of wrinkles.
Infrared light: Provide topical heating for the purpose of elevating tissue temperature, and to temporarily increase local blood circulation.
# 6. Substantial Equivalence(SE) Comparison
Compared with the predicate devices, the subject device is very similar in design principle, intended use, indications for use, functions and the applicable standards. The differences between the subject device and predicate devices do not raise new questions of safety or effectiveness.
| Elements of Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 |
| --- | --- | --- | --- | --- |
| Sponsor | SUNGLOR TECHNOLOGY CO.,LTD | Dongguan Boyuan Intelligent Technology Co.,Ltd | ISMART Developments Ltd | Guangdong Newdermo Biotech Co.,Ltd |
{7}
Sponsor: SUNGLOR TECHNOLOGY CO.,LTD
Subject Device: LED Light Therapy Device
| Elements of Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 |
| --- | --- | --- | --- | --- |
| Device Name and Model | Led Light Therapy Device (Model: SG-FSM, SG-NK, SG-CT) | LED Light Therapy Device (KFB290, KFB291) | décoLITE LED Device (TN2037) | LED light therapy mask (FM-01, FM-02, FM-03) |
| 510 (K) Number | K253073 | K241857 | K242382 | K223544 |
| Regulation number | 878.4810, 890.5500 | 878.4810, 890.5500 | 878.4810 | 878.4810, 890.5500 |
| Classification Panel | General & Plastic Surgery; Physical Medicine; | General & Plastic Surgery; Physical Medicine; | General & Plastic Surgery; | General & Plastic Surgery; Physical Medicine; |
| Classification | Class II | Class II | Class II | Class II |
| Product code | OHS, ILY | OHS, OLP, ILY | OHS | OHS, OLP, ILY |
| Indication for Use | The Led Light Therapy Device (Model:SG-FSM) is an Over-the-Counter medical device intended for treatment of wrinkles in the face and decolletage area.
Red light: Treatment of wrinkles.
Red+Infrared Light: Treatment of wrinkles.
Amber light: Treatment of wrinkles.
Amber+Infrared Light: Treatment of wrinkles. | -Red light: Treatment of full-face wrinkles.
-Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; relieving stiffness; promoting the relaxation of muscle tissue; and to temporarily increase local blood circulation.
-Red+Infrared Light: Treatment of full-face wrinkles.
-Amber light: Treatment of full-face wrinkles.
-Blue light: Treatment | The décoLITE LED device is an over-the-counter device that is intended for the use in the treatment of wrinkles in the decolletage area. | Red light: Treatment of full-face wrinkles.
Blue light: Treatment of mild to moderate inflammatory acne.
Infrared light: Provide topical heating for the purpose of elevating tissue temperature; arthritis and muscle spasm; Relieving stiffness; promoting the relaxation of Muscle tissue; and to temporarily |
4 / 10
{8}
Sponsor: SUNGLOR TECHNOLOGY CO.,LTD
Subject Device: LED Light Therapy Device
| Elements of Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 |
| --- | --- | --- | --- | --- |
| | The Led Light Therapy Device (Model:SG-NK) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area.
Red light: Treatment of wrinkles.
Red+Infrared Light: Treatment of wrinkles.
Amber light: Treatment of wrinkles.
Amber+Infrared Light: Treatment of wrinkles.
The Led Light Therapy Device (Model:SG-CT) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area.
Red light: Treatment of wrinkles.
Red+Infrared light: Treatment of wrinkles.
Infrared light: Provide topical heating for the | of mild to moderate inflammatory acne.
-Mixed light(Red+Blue +Infrared): Treatment of mild to moderate inflammatory acne. | | increase local blood circulation.
Mixed light: Treatment of Mild to moderate inflammatory acne. |
5 / 10
{9}
Sponsor: SUNGLOR TECHNOLOGY CO.,LTD
Subject Device: LED Light Therapy Device
| Elements of Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 |
| --- | --- | --- | --- | --- |
| | purpose of elevating tissue temperature, and to temporarily increase local blood circulation. | | | |
| Location for use | SG-FSM: Face and Decolletage (upper chest)
SG-NK and SG-CT: Decolletage (upper chest) | Face | Decolletage (upper chest) | Face and body |
| Intended Environment | Home use | Home use | Home use | Home use |
| Type of use | OTC | OTC | OTC | OTC |
| Dimensions | LED mask for SG-FSM: 304mm*196mm*4 mm
Upper chest for SG-FSM and SG-NK: 345mm*224mm*4 mm
Upper chest for SG-CT: 337mm*245mm*4 mm
Controller: 119.5mm*43.5mm*30.2mm; | KFB290: Approximately 302 mm x 197 mm x 22 mm
KFB291: Approximately 412 mm x 195 mm x 22 mm | Controller: 23mm max (dia) x 104mm (l) | FM-01: 207x 277x 43mm,
FM-02: 198x 383x 33.5mm,
FM-03: 237.5x 108x 8.1mm |
| Power source | Input: DC 5V, 1A
Battery: Li-ion rechargeable battery, | Input: 100 -240 V, 50/60 Hz
Output: 5V, 1A | 100-240Vac/47-63Hz. 0.6A
Electrical Output: 5V | Input: 100-240V,50/60Hz, 0.25 A; Output: DC |
6 / 10
{10}
Sponsor: SUNGLOR TECHNOLOGY CO.,LTD
Subject Device: LED Light Therapy Device
| Elements of Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 |
| --- | --- | --- | --- | --- |
| | 3.7V, 3000mAh | Lithium ion battery: 1300mAh | | 5 V, 500mA |
| Type of Energy | LEDs | LEDs | LEDs | LEDs |
| Wavelength | SG-FSM: NIR: 850nm (face) and 830nm (upper chest)±10nm; Red: 630nm±10nm; Amber: 605nm±10nm; SG-NK: NIR: 830nm±10nm; Red: 630nm±10nm; Amber: 605nm±10nm; SG-CT: NIR: 830nm±10nm; Red: 630nm±10nm; | 635nm ± 5nm visible red light; 850nm±5nm Invisible red light; 465±5nm blue light; 605±5nm amber light; | Red: 630nm ± 10nm NIR: 830nm ± 10nm | Red: 620nm Blue: 460nm Infrared: 850nm Mixed: 620nm and 850nm and 460nm |
| Irradiance | Max Irradiance: 850/830nm: 3±20%mW/cm²; 630nm: 22±20%mW/cm²; 605nm: 14±20%mW/cm²; | Red: 25mW/cm2; IR: 6mW/cm2; Red+IR: 30mW/cm2; Blue: 18mW/cm2 ; Amber: 20mW/cm2; Mixed light: 9mW/cm2; | 30mW/cm2 total | Red light:: 2.0~3.0mW/cm² Blue light: 2.0~4.0mW/cm² Infrared light:: 2.0~4.0 mW/cm² Mixed light: 9.0~12.0 mW/cm² |
7 / 10
{11}
Sponsor: SUNGLOR TECHNOLOGY CO.,LTD
Subject Device: LED Light Therapy Device
| Elements of Comparison | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 |
| --- | --- | --- | --- | --- |
| Treatment time | 10/15/20/25/30 minutes | For red, blue and red+infrared: 10, 20, 30 minutes; For infrared, amber light and mixed light: 10, 20 minutes. | 600 seconds (10 minutes) | Manual Mode: 15 minutes each time, Automatic Mode: 10 minutes each time. 3-4 treatment a week, reduce to 1-2treatment a week once the results shown. |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 |
| Safety | IEC 60601-1
IEC 60601-1-11
IEC 62471
IEC 60601-2-83
IEC 62133-2 | IEC 60601-1
IEC 60601-1-11
IEC 62471
IEC 60601-2-83 | IEC 60601-1
IEC 60601-1-11
IEC 62471
IEC 60601-2-57
IEC 60601-2-83
IEC 62133-2 | IEC 60601-1
IEC 60601-1-11
IEC 60601-2-57 |
| Biocompatibility compliance | ISO 10993-5
ISO 10993-10
ISO 10993-23 | ISO 10993-5
ISO 10993-10
ISO 10993-23 | Unknown | All body-contacting materials are complied with ISO10993-5 and ISO 10993-10 |
## 7. Testing Summary
## 7.1 Non-clinical testing
All necessary performance testing was conducted on the Led Light Therapy Device to support a determination of substantial equivalence to the predicate device.
## 1) Biocompatibility Testing
The biocompatibility evaluation for the body-contacting components of the LED Light Therapy Device was conducted in accordance with the “Use of International Standard ISO 10993-1, “Biological evaluation of medical devices - Part 1 Evaluation and testing within
{12}
Sponsor: SUNGLOR TECHNOLOGY CO.,LTD
Subject Device: LED Light Therapy Device
a risk management process” Document Issued on September 8, 2023”, as recommended by FDA. The following testing was performed to, and passed, including:
- ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity.
- ISO 10993-10:2021 Biological evaluation of medical devices - Part 10: Tests for skin sensitization.
- ISO 10993-23:2021 Biological evaluation of medical devices - Part 23: Tests for irritation
## 2) Electrical Safety
Non-clinical tests were performed on the subject device for validate the design and to assure conformance with the following voluntary design standards in connection with medical device electrical safety and electromagnetic compatibility:
- IEC 60601-1: 2005 +AMD1:2012+AMD2:2020 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance.
- IEC 60601-1-11:2015+A1:2020 Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 60601-1-2:2014+A1:2020 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
- IEC TS 60601-4-2:2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems.
- IEC 60601-2-83:2019+A1:2022 Medical electrical equipment Part 2-83: Particular requirements for the basic safety and essential performance of home light therapy equipment.
- IEC 62133-2:2017/AMD1:2021 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems.
## 3) Eye Safety
- IEC 62471:2006 Photobiological safety of lamps and lamp systems
{13}
Sponsor: SUNGLOR TECHNOLOGY CO.,LTD
Subject Device: LED Light Therapy Device
## 4) Software verification and validation
Software verification and validation testing were conducted and documentation was provided as recommended by FDA’s Guidance for Industry and FDA Staff, “Content of Premarket Submissions for Device Software Functions - Guidance for Industry and Food and Drug Administration Staff.” The software for this device was considered as a Basic Documentation Level, since a failure or latent flaw of the device software function(s) would not present a hazardous situation with a probable risk of death or serious injury to either a patient, user of the device, or others in the environment of use, prior to the implementation of risk control measures.
## 7.2 Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
## 8. Final Conclusion
Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device is as safe, as effective, and performs as well as than the legally marketed predicated devices (K241857, K242382 and K223544).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.