Shani Darden LED light therapy mask

K214103 · Harpar Grace International · OLP · Mar 22, 2022 · General, Plastic Surgery

Device Facts

Record IDK214103
Device NameShani Darden LED light therapy mask
ApplicantHarpar Grace International
Product CodeOLP · General, Plastic Surgery
Decision DateMar 22, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Shani Darden LED light therapy mask is an over-the-counter device intended to emit energy in the red and blue region of the light spectrum, specifically indicated to treat mild to moderate acne vulgaris of the face. The Shani Darden LED light therapy mask is an over-the-counter device intended to emit energy in the red and near infra-red spectrum and is intended for the use in the treatment of full-face wrinkles.

Device Story

The Shani Darden LED light therapy mask is a wearable, home-use phototherapy device. It consists of a hard-shell mask, a controller, and a mains power adaptor. The mask uses LEDs to emit blue (415nm), red (630nm), and near infra-red (830nm) light. The user operates the device via a controller with a simple GUI and three buttons to select treatment modes (acne or wrinkles) and start/pause a 10-minute timer. The device is worn on the face and held by a Velcro strap. By emitting specific light wavelengths into the skin, the device targets acne-causing bacteria or improves the appearance of wrinkles. It is intended for patient self-use in a home environment. The device does not perform measurements or draw clinical conclusions; it functions as a therapeutic light source.

Clinical Evidence

No clinical data. The submission relies on non-clinical performance testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), home healthcare environment requirements (IEC 60601-1-11), photobiological safety (IEC 62471), and biocompatibility (ISO 10993). A label comprehension study was conducted with 19 subjects (mean age 31.6, 7th-grade reading level) to ensure safe and effective operation by lay users.

Technological Characteristics

Wearable LED mask; outer shell ABS (CHIMEI 757KF), inner shell polycarbonate (CHIMEI PC115P). Wavelengths: Blue 415nm, Red 630nm, NIR 830nm (+/- 10nm). Power: Mains adaptor (100-240V). No wireless/wired data connectivity. Software controls treatment duration via timer. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57, IEC 62471, ISO 10993-1/5/10, EN 62304, ISO 14971.

Indications for Use

Indicated for treatment of mild to moderate acne vulgaris of the face and treatment of full-face wrinkles in the general population.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath. March 22, 2022 Harpar Grace International % Susan D'arcy Owner iSMART Developments Ltd 129 Green Lanes, Sutton Coldfield Birmingham, B735TR United Kingdom Re: K214103 Trade/Device Name: Shani Darden LED light therapy mask Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP, OHS Dated: December 24, 2021 Received: December 29, 2021 Dear Susan D'arcy: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K214103 Device Name Shani Darden LED light therapy mask Indications for Use (Describe) The Shani Darden LED light therapy mask is an over-the-counter device intended to emit energy in the red and blue region of the light spectrum, specifically indicated to treat mild to moderate acne vulgaris of the face. The Shani Darden LED light therapy mask is an over-the-counter device intended to emit energy in the red and Near Infra-red spectrum and is intended for the use in the treatment of full-face wrinkles. | Type of Use (Select one or both, as applicable) | |--------------------------------------------------------------------------------------------------------------------------------------| | <div style="display:flex; align-items:center;"><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</div> | | <div style="display:flex; align-items:center;"><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Section 5: 510(k) Summary This 510(k) Summary is submitted in accordance with 21 CFR Part 807, Section 807.92(c). Submitter's Name: Harpar Grace International (HGI) Submitter's Address: Unit 4 Imperial Court, Magellan Close, Walworth Business Park, Hampshire. SP10 5NT Contact Person: Alana Marie Chalmers Telephone: +44 (0) 2038686242 Date Prepared: December 24th, 2021 Date amended: March 21st, 2022 Device Trade Name: Shani Darden LED light therapy mask ### Device Classification Information: | Regulation<br>Number | Classification<br>Name | Common name | Device<br>Class | Product<br>Code | Classification<br>Panel | Type | |----------------------|----------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------|-----------------|------------------------------|------------------------| | 21 CFR<br>878.4810 | Laser Surgical<br>Instrument for<br>Use in General<br>and Plastic<br>Surgery and In<br>Dermatology | Over-The-Counter<br>Light-Based<br>treatment for<br>Acne and Wrinkle<br>Reduction | Class 2 | OLP<br>OHS | General &<br>Plastic Surgery | Traditional<br>510 (k) | {4}------------------------------------------------ ### 5.1. Device Description The Shani Darden LED light therapy mask consists of the following key components. - Facemask 1. - 2. Controller - 3. Power supply and country specific adaptors - 4. Velcro strap - 5. User manual - 6. Carry Bag The Shani Darden LED light therapy mask is a home use wearable light emitting diode phototherapy device whose purpose is to produce an even, cool, narrow band of light for the treatment of full-face wrinkles and mild to moderate acne vulgaris of the face. The system consists of a hard-shell mask (1) and a controller (2). The outer shell of the mask is manufactured from Acrylonitrile butadiene styrene (ABS) CHIMEI 757KF. The inner shell is a clear Polycarbonate CHIMEI PC115P. The Light emitting diodes are mounted behind the clear Polycarbonate. The LEDs generate the light. The mask is worn on the face and is held in place by an adjustable Velcro strap (5). The LEDs produce blue, red and near infra-red (NIR) light in the visible spectrum (Blue: 415nm +/- 10nm, Red: 630nm +/- 10nm, NIR 830nm +/-10nm.). The device works by emitting the specified wavelengths into the skin in order to improve the appearance of wrinkles or to target bacteria that can cause acne vulgaris. The controller (2) allows the user to select one of three treatment programmes (acne and wrinkles) and switches the LEDs ON/OFF, controlling power to the mask. The controller contains a simple graphical user interface (GUI) and three control buttons. - 1. Standby Button: wakes the device up from Sleep mode - 2. Mode Button: Allows the user to select treatment modes. - 3. Start button. Allows the user to start and pause a treatment The GUI shows the treatment mode selected and a countdown timer that counts downs from 10 minutes in 1-minute increments. {5}------------------------------------------------ The mask receives its power from a mains power adaptor (Input 100v-240v. 50/60hZ. Rated at 0.7A) Output (12V, 1A). The power adaptor is connected to the controller by a standard USB A connector. The Shani Darden LED light therapy mask does NOT contain or does NOT rely on external wired and/or wireless communication interfaces (Wired: USB, ethernet, SD, CD, RGA, etc. or Wireless: Wi-Fi, Bluetooth, RF, inductive, Cloud, etc.). The equipment is not used to make measurements of any sort, or to draw any conclusions regarding the indication to treat. The equipment does not require checks on the light output as the LEDs do not dim with age to any practical extent. ### 5.2. Intended Use The Shani Darden LED light therapy mask is an over-the-counter device intended to emit energy in the red and blue region of the light spectrum, specifically indicated to treat mild to moderate acne vulgaris of the face. The Shani Darden LED light therapy mask is an over-the-counter device intended to emit energy in the red and near infra-red spectrum and is intended for the use in the treatment of full-face wrinkles. #### 5.3. Predicate device K213184 MZ Skin LightMAX Supercharged LED Mask 2.0 {6}------------------------------------------------ #### 5.3.1 Summary of Substantial Equivalence | Description | Shani Darden LED light therapy mask | K213184<br>MZ Skin LightMAX | |---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Manufacturer | Harpar Grace International | MZ SKIN | | Device Trade Name | Shani Darden LED light therapy mask | MZ Skin LightMAX Supercharged LED Mask<br>2.0 | | 510(K) Number | K214103 | K213184 | | Device Product Code | OLP, OHS | OLP, OHS | | Regulation Number | 878.4810<br>Laser surgical instrument for use in general and<br>plastic surgery and in dermatology. | 878.4810<br>Laser surgical instrument for use in general<br>and plastic surgery and in dermatology | | FDA Device<br>Classification | Class II | Class II | | Use | Over the Counter | Over the Counter | | Intended use and<br>Indications | The Shani Darden LED light therapy mask is an<br>over-the-counter device intended to emit energy<br>in the red and blue region of the light spectrum,<br>specifically indicated to treat mild to moderate<br>acne vulgaris of the face.<br>The Shani Darden LED light therapy mask is an<br>over-the-counter device intended to emit energy<br>in the red and Near Infra-red spectrum and is<br>intended for the use in the treatment of full-face<br>wrinkles. | The MZ Skin LightMAX Supercharged LED<br>Mask 2.0 is an over-the-counter device<br>intended to emit energy in the red and blue<br>region of the light spectrum, specifically<br>indicated to treat mild to moderate acne<br>vulgaris of the face.<br>The MZ Skin LightMAX Supercharged LED<br>Mask 2.0 is an over-the-counter device<br>intended to emit energy in the red and Near<br>Infra-red spectrum and is intended for the<br>use in the treatment of full-face wrinkles. | | Description | Shani Darden LED light therapy mask | K213184 MZ Skin LightMAX | | Intended Location of Use | Face | Face | | Energy Type | Light emitting diodes | Light emitting diodes | | Peak Wavelength<br>(FWHM) | Blue: 415nm +/- 10nm,<br>Red: 630nm +/- 10nm,<br>NIR 830nm +/-10nm. | Blue: 415nm +/- 10nm,<br>Red: 630nm +/- 10nm,<br>NIR 830nm +/-10nm. | | Intensity (mW/cm²) | Blue 28 mW/cm²<br>Red 16 mW/cm²<br>Red 18 mW/cm²<br>NIR 11 mW/cm² | Blue 28 mW/cm²<br>Red 16 mW/cm²<br>Red 18 mW/cm²<br>NIR 11 mW/cm² | | Total Intensity (mW/cm²) | Blue/Red 44 mW/cm²<br>Red/NIR 29 mW/cm² | Blue/Red 44 mW/cm²<br>Red/NIR 29 mW/cm² | | Treatment time | 10 Minutes | 10 Minutes | | Dose | Blue 16.8J/cm²<br>Red 9.6J/cm² | Blue 16.8J/cm²<br>Red 9.6J/cm² | | | Red 11J/cm²<br>NIR 7J/cm² | Red 11J/cm²<br>NIR 7J/cm² | | Treatment protocol | Acne: 4 x weekly, 6 weeks | Acne: 4 x weekly, 6 weeks | | | Wrinkles: 5 x weekly, 6 weeks | Wrinkles: 5 x weekly, 6 weeks | | Software Controlled | Device uses a timer and software to control<br>treatment duration | Device uses a timer and software to<br>control treatment duration | | Description | Shani Darden LED light therapy mask | K213184 MZ Skin LightMAX | | Power supply | 100-240V | Lithium-Ion Battery (rechargeable) | | Power input/output | Electrical Input to power adaptor: 100v-240v.<br>50/60Hz. Rated at 0.7A. | Electrical Input from power adaptor: 100-<br>240Vac/47-63Hz. 0.6A | | | Electrical Output from power adaptor: 12V, 1A | Electrical Output; 5V, 9.62Wh or 2200mAH | {7}------------------------------------------------ {8}------------------------------------------------ {9}------------------------------------------------ ### 5.3.2. Comparison of Technological Similarities & Differences The key similarities are. - i. The intended use of the Shani Darden LED light therapy mask is the same as the listed predicate; an over-the-counter device that is intended for the use in the treatment of mild to moderate acne vulgaris and full-face wrinkles. - ii. The devices are phototherapy units utilizing light emitting diodes that emit in the red and blue spectrum for the treatment of mild to moderate acne vulgaris and the red and NIR spectrum for the treatment of full-face wrinkles. - iii. The wavelength spectrum of the devices is similar. - iv. The Shani Darden LED light therapy mask has similar power density and dose to the predicate device. - v. The Shani Darden LED light therapy mask has an identical treatment time and treatment protocol compared to the proposed predicate. - vi. Both devices use software to control the treatment time. The key difference between the Shani Darden LED light therapy mask and predicate device is that the sponsor device uses a power adaptor to control power while the predicate device utilizes Lithium-lon batteries. This difference has been addressed by EMC and electrical safety testing. Where other differences exist, these have been addressed by non-clinical performance testing to applicable standards. ## 5.4. Non- clinical performance testing The Shani Darden LED light therapy mask has been thoroughly evaluated for electrical safety and performance and has been found to conform to the following standards. IEC 60601-1:2005/AMD1:2012/AMD2:2020 Medical electrical equipment Part 1: General requirements for basic safety and essential performance. IEC/EN 60601-1-2: 2015 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests including FCC 47 CFR Part 15, Sub Part B IEC 60601-1-11:2015 Medical electrical equipment -- Part 1-11: General requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. IEC 60601-2-57:2011 Medical Electrical Equipment - Part 2-57: Particular Requirements for The Basic Safety And Essential Performance Of Non-Laser Light {10}------------------------------------------------ Source Equipment Intended For Therapeutic, Diagnostic, Monitoring And Cosmetic/Aesthetic Use IEC 62471:2008. Photobiological safety of lamps and lamp systems. ISO 10993-1: 2018 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management. EN ISO 10993-5:2009 Biological Evaluation of Medical Devices - Part 5: Tests For In Vitro Cytotoxicity EN ISO 10993-10:2010 ISO 10993-10 Third Edition 2010 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization EN 62304: 2006 (ed. 1.0) Medical Device Software - Software Life Cycle Processes. ISO 14971: 2019 Medical Devices - Application of Risk Management to Medical Devices In addition to the above standards the Shani Darden LED light therapy mask labelling was subject to label comprehension testing. With respect to medical devices available without the intervention of a physician, termed 'Over the Counter' (OTC). To determine the effectiveness of labelling pertaining to a medical device, the labelling and device was tested with an appropriate sample of users. A study was conducted and is appended to this submission, demonstrating comprehension of the Shani Darden LED light therapy mask labelling. Nineteen subjects took part in the study, 8:11 M:F, average age 31.6 years (range 16-57). Eight subjects identified English as their second language. In terms of ethnicity 6 subjects identified as Hispanic, 3 Asian, and 1 Indian. The average number of words incorrect in the REALM reading test was 9, giving a mean reading ability of 57 (7th grade). No new use errors, hazards, hazardous situations, or hazard-related use scenarios were discovered during testing. Further improvement of the user interface design as it relates to safety was deemed unnecessary and there were no suggested revisions to the version of the user manual or box packaging tested. The comprehension and use test demonstrated that the Shani Darden LED light therapy mask labelling could be used by lay persons to safely and effectively operate the device to attain its intended use and purpose. {11}------------------------------------------------ ### Clinical Performance Clinical testing was not needed for this 510(k). The non-clinical performance testing described above is sufficient to support that the device can be used safely and effectively. ### Conclusion The Shani Darden LED light therapy mask utilizes the same light-emitting technological characteristics as the predicate device, and the new device does not raise new types of questions regarding safety and efficacy when compared to the predicate device. The device has been tested as described above to show that the device can be used safely and effectively. The Shani Darden LED light therapy device is considered to be substantially equivalent to the predicate device K213184.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...