I.T.S. Clavicle Plates with Angular Stability; I.T.S. Volar Radius Plate with Angular Stability; I.T.S. Humeral Head Plate with Angular Stability; I.T.S. FR.O.H. Calcaneus Repair System; I.T.S. Pilonplate with Angular Stability; I.T.S. Olecranonplate with Angular Stability; I.T.S. Straight Plate with Angular Stability; I.T.S. Screw System; I.T.S. PRS Sacral Rod System; I.T.S. Fibula Plate PROlock with Angular Stability; I.T.S. Distal Humeral Plates with Angular Stability; I.T.S. LRS (L
Applicant
I.T.S. GmbH
Product Code
HRS · Orthopedic
Decision Date
Jun 16, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The I.T.S. Clavicle Plates with Angular Stability is a titanium implant fracture fixation system for repairing fractures located from the middle third to the distal third of the clavicle. Indications for Use include metaphyseal and diaphyseal fracture fixation of acute fractures, malunions, and non-unions of the clavicle. Other indications include corrective osteotomy and open and closed fractures. All fractures of the clavicle in metaphyseal and diaphyseal areas. Hygienisation of pseudo-arthroses with or without spongiosal graft. Corrective osteotomy. Open and closed fractures. The I.T.S. Volar Radius Plates with Angular Stability is a titanium implant fracture fixation system for distal radius fractures of the wrist. Indications for Use include comminuted extra and intra-articular distal radius fractures, failed original fracture fixation, osteotomy and repair of a distal radius malunion, and comminuted volar shearing fractures. Complex intra-articular fractures of the distal radius. Complex extra-articular fractures of the distal radius. Osteotomies of the distal radius. The I.T.S. Humeral Head Plates with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the proximal humerus in the shoulder. Indications for Use include fracture and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone. All stable and unstable humerus fractures with or without shaft involvement. Fractures of the greater or lesser tuberosities. Repair of the greater tuberosity following prior fixation failure or tuberosity "escape". Delayed or nonunion of the proximal humerus. Fixation following osteotomy of proximal humeral malunion. Displaced two, three and four part fracture of the proximal humerus. Displaced anterior and posterior fractures of the proximal humerus and greater tuberosity. Nonunion of two, three and four part fractures of the proximal humerus. Nonunion of anterior and posterior fracture-dislocations of the proximal humerus and greater tuberosity. Dislocated, unstable 2, 3 and 4-segment fractures of the humeral head. Valgus-impacted 4-segment fractures of the humeral head. Non-union of the humeral head. The I.T.S. FR.O.H. Calcaneus Repair System is a titanium implant fracture fixation system for repairing fractures located in the calcaneus heel bone of the foot. Indications for Use include: Intra and extra-articular fracture(s) of the calcaneus, corrective osteotomy, joint depression, non-displaced and tongue type, severely comminuted fractures, multifragmentary fractures, revision procedures, joint fusion, stabilization and fixation of fresh fractures, reconstruction of the calcaneus bones, and open and closed fractures of the calcaneus. The system can be used in both adult and pediatric patients. Complex Fractures of the Calcaneus. All intra-articular fractures with relevant joint distortion and comminution zone in which a semi-operative procedure (screws, drill wires) does not raise expectations of exact repositioning. Fractures of the sustentaculum tali. The I.T.S. Pilon Plates with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the distal tibia in the leg. Indications for Use include fixation of complex intra- and extraarticular fractures, osteotomies, high medial malleolar fractures, and low boot top type rotational distal extra-articular shaft fractures of the distal tibia. Fractures of the tibial pilon of AO classification A3, especially groups C2 and C3. The I.T.S. Olecranon Plates with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the proximal ulna (olecranon) in the elbow. Indications for Use include fixation of complex intra- and extraarticular fractures, osteotomies, nonunions, malunions, Type 1, 2, 3, & 4 simple fractures, and Type 5a, 5b, 5c, & 6 complex fractures (comminuted) of the proximal ulna (olecranon). All fractures of the olecranon. The intended use of the I.T.S. Straight Plates with Angular Stability is to stabilize an osteotomy or fracture of small bones, long bones, the pelvis and the calcaneus in an adult or pediatric patient. Indications for use include commuted fractures, supracondylar fractures, intra-articular and extra-articular condylar fractures, fractures in osteopenic bone, non-unions, and mal-unions, as well, a fracture or osteotomy of the tibia, fibula, femoral condyle, acetabulum, humerus, ulna, middle hand and middle foot bones. Treatment of the calcaneal; hip arthrodesis, and provisional hole fixation. The intended use of the I.T.S. Screw System is for corrective osteotomy or internal fracture fixation of the patella, pelvis, ankle, and long bones in an adult or pediatric patient. For the 4.0mm Cannulated Cancellous Screw indications for use are for radial and ulnar fractures, fractures of the proximal/distal humerus and of the patellas and for tendon fixation, maisonneuve injuries and disruption of the syndesmosis with bimalleolar or supramalleolar fractures and the instability of the talus centering. For the 6.5mm Cannulated Cancellous Screw, indications for use are for fractures of the femoral neck, tibiaplateau, of the sacrum and the articular cavity of the hip joint and the metaphyseal fractures of the distal femur and distal tibia, fixation of the ilio-sacral joint, and fusion of the foot and ankle. For the 7.3mm Cannulated Cancellous Screw, indications for use are for fractures of the calcaneus, femoral neck, tibiaplateau, and of the sacrum and the articular cavity of the hip joint. Fusion of the foot and ankle, fixation of the Ilio-sacral joint, and metaphyseal fractures of the distal femur and distal tibia. The intended use of the I.T.S. Pelvic Reconstruction System (PRS) is to stabilize one or more pelvic bone fractures in the pelvic ring area of the pelvis of an adult or pediatric patient. The indications for use of the I.T.S. PRS Sacral Rod System include fixation of fractures of the posterior pelvis, fixation of fractures of the posterior iliac spine, fixation of fractures of the posterior inferior iliac spine, dorsal stabilization of the posterior pelvic ring for unstable pelvic ring injuries, fixation of sacral fractures, and fracture dislocations of the sacro-iliac joint. The I.T.S. Fibula Plate PROlock with Angular Stability is a titanium implant fracture fixation system for repairing bone fractures located from the middle to the distal third of the fibula. Indications for Use include metaphyseal and diaphyseal fracture fixation of acute fractures, mal-unions, and non-unions of the distal fibula. Other indications include corrective osteotomy and open and closed fractures. Dislocated ankle-fractures group B+C according to Weber (with or without comminuted zones). The I.T.S. Distal Humeral Plates with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the distal humerus in the elbow of a pediatric or adult patient. Indications for Use include intra-articular fractures, supracondylar fractures, osteotomies, and non-unions of the distal humerus. Supra- & diacondylar upper-arm fractures. The I.T.S. LRS Locking Reconstruction System is a titanium implant fracture fixation system for stabilizing fractures in long bones of the distal femur of all pediatric patients (less than or equal to 21 years old) or adult patient. Indications for Use include distal shaft fractures, supracondylar fractures, intra-articular fractures, metaphyseal fractures, osteotomies, nonunions and malunions of the distal femur. The I.T.S. LRS Locking Reconstruction System is a titanium implant fracture fixation system for stabilizing fractures in long bones of the proximal lateral tibia of all pediatric patients (less than or equal to 21 years old) or adult patient. Indications for Use include proximal shaft fractures, supracondylar fractures, intra-articular fractures, metaphyseal fractures, osteotomies, nonunions and malunions of the proximal tibia. The Indications for Use of the I.T.S. GmbH - Distal Ulna Locking (DUL) Plate Systems is to stabilize fractures in the long bone of the distal ulna of an adult patient. Fractures of the ulnar head. Multifragmentary fracture of the ulnar head. Subcapital fractures of the ulnar head. Metaphyseal comminuted fractures of the distal ulna. Combined ulnar head and ulnar shaft fractures. The Indications for Use of the I.T.S. GmbH - Ulna Osteotomy Locking (UOL) Plate Systems is to stabilize osteotomies in the long bone of the mid-ulna of an adult patient. Impaction syndrome of the ulnar wrist. Symptomatic, post-traumatic ulnar malposition in the distal radio-ulnar joint (DRUJ). Degenerative ulnar wrist. Correction of the ulnar positive relative to the unaffected other side up to a maximum of 6mm in one step or 13mm in two steps. The I.T.S. FLS Foot Locking Plate System is indication for use in internal fixation, reconstruction or athrodesis of small bones including the fore, mid and hind foot and ankle. Examples of these procedures may included but are not limited to replantation, lag screw techniques, joint fusions, corrective osteotomies, and treatment of fracture. Metatarsal fractures. TMT I-V Arthrodesis, MTP I-V Arthrodesis, TN, CC Arthrodesis, Corrective Osteotomies. The I.T.S. HCS - Headless Compression Screw System for sizes of 3.5mm or smaller is indicated for use in fixation small bone fractures or for small bone reconstruction including: mono or bicorlical osteotomies in the foot or hand; distal or proximal metatarsal or metacarpal osteotomies; weil osteotomy; fusion of the first metatarsalphalangeal joint and interphalangeal joint: fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc.); Akin type osteotomy; distal radius fractures (articular fragments). ulnar styloid fractures, radial head fractures, capitellum fractures, humeral head fractures, glenoid fractures, intercarpal distal and proximal fusions, malleolar fractures, patellar fractures, osteochondral fractures, talonavicular fusions, tibeo-talar fusions, and cuboid fusions. And for sizes 4.3mmn or larger is indicated for use for fractures, corrective osteotomies, pseudoarthrosis, degenerative transformations of long bones in the hindfoot and large bone intra-articular fractures of the humerus, femur and tibia. The size of the chosen compression screw should be adapted to the specific indication. The I.T.S. HOL - Hallux Osteotomy Locking Plate System is Indicated for use as an intramedullary self-locking plate for distal metatarsal osteotomies and for Hallux Valgus osteotomies up to a corrective angle of 25°. Intramedullary, self locking plate for distal metatarsal osteotomies. Hallux Valgus osteotomies up to a corrective angle of 25°. The I.T.S. Twist-Off Screw System is indicated for use for small bone fixation of bone fractures or for bone reconstruction. Examples include small bone fragments, Weil- Osteotomy, Mono-Cortical fixation, Osteotomies and fractures fixation in the foot and hand. For treatment of fractures, corrective osteotomies, arthrodesis and degenerative transformations of small bones. The I.T.S. Hand Locking Plate System -HLS is indicated for use in fracture fixation of: The phalanges, The metacarpal bones, The carpal bones, For arthrodesis, For corrective osteotomies and For subcapital radial head fractures. Indications for use of the I.T.S. Pelvic Reconstruktion System (PRS Phoenix) include: Fractures involving the Posterior Wall & Posterior Column, Fractures involving the Anterior Column of the Acetabulum, Fractures involving the Quadrilateral Surface, Symphyseal Disruptions & Para-symphseal Fractures, Fractures of the ilium, Fractures of the SIJ, Dorsal neutralization plating for posterior pelvic ring fractures, Osteotomies, arthrodesis and sacroiliac joint dislocations, Revision surgery of pseudoarthroses, non-unions and mal unions. The I.T.S. CTN – Cannulated Tibia Nail System is indicated for use in long bone tibia fracture fixation which include: Proximal, metaphyseal, epiphyseal and distal shaft fractures; Segmental, simple, compound and comminuted fractures; Transverse, oblique and spiral fractures; Surgically created defects using osteotomies, such as for leg length discrepancies or deformity; Pathologic fractures; Pseudoarthrosis, non-union, mal-union and delayed union of the tibia; Fractures involving osteopenic and osteoporotic bone; Open fractures of the tibia and; Reconstruction of the tibia after tumor resection and/or bone loss.
Device Story
Titanium bone fixation system (plates, screws, nails) for fracture stabilization, osteotomy, and arthrodesis; used in clinical settings by surgeons; provides mechanical support to bone fragments during healing; updated to include MR conditional safety labeling; no changes to materials, design, or manufacturing.
Clinical Evidence
No clinical data required; non-clinical MRI safety testing performed per ASTM F2052, F2213, F2182, F2119, and F2503.
Technological Characteristics
Titanium bone fixation implants (plates, screws, nails). MR conditional. No changes to materials, sterilization, or manufacturing processes.
Indications for Use
Indicated for fracture fixation, osteotomy, non-union, mal-union, and arthrodesis of long and small bones (clavicle, radius, humerus, calcaneus, tibia, ulna, fibula, pelvis, hand/foot bones) in adult and pediatric patients.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
I.T.S. Clavicle Plates with Angular Stability (K050852)
I.T.S. Volar Radius Plate with Angular Stability (K033756)
I.T.S. Humeral Head Plate with Angular Stability (K051412)
I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) (K210935)
I.T.S. CTN - Cannulated Tibia Nail System (K132945)
Submission Summary (Full Text)
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 16, 2026
I.T.S. GmbH
% Mindy McCann
VP North America & Principal Consultant
Qserve Group, US, Inc.
350 S Main St.
Suite 309
Doylestown, Pennsylvania 18901
Re: K260958
Trade/Device Name: I.T.S. Clavicle Plates with Angular Stability; I.T.S. Volar Radius Plate with Angular Stability; I.T.S. Humeral Head Plate with Angular Stability; I.T.S. FR.O.H. Calcaneus Repair System; I.T.S. Pilonplate with Angular Stability; I.T.S. Olecranonplate with Angular Stability; I.T.S. Straight Plate with Angular Stability; I.T.S. Screw System; I.T.S. PRS Sacral Rod System; I.T.S. Fibula Plate PROlock with Angular Stability; I.T.S. Distal Humeral Plates with Angular Stability; I.T.S. LRS (Locking Reconstruction System) - DFL (Distal Femur Locking) & PTL (Proximal Lateral Tibia Locking); I.T.S. Ulna Locking Plates - (DUL & UOL) Systems; I.T.S. FLS - Foot Locking Plates System; I.T.S. HCS - Headless Compression Screw System; I.T.S. HOL - Hallux Osteotomy Locking Plate; I.T.S. Twist-Off Screws; I.T.S. Hand Locking Plates System - HLS; I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix); I.T.S. CTN - Cannulated Tibia Nail System
Regulation Number: 21 CFR 888.3030
Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories
Regulatory Class: Class II
Product Code: HRS, HWC, HSB
Dated: March 23, 2026
Received: March 23, 2026
Dear Mindy McCann:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K260958 - Mindy McCann
Page 2
available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
{2}
K260958 - Mindy McCann
Page 3
Also, please note the regulation entitled, 'Misbranding by reference to premarket notification' (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
CHRISTOPHER FERREIRA -S
Christopher Ferreira, M.S.
Assistant Director
DHT6C: Division of Restorative,
Repair, and Trauma Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# Indications for Use
Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.
K260958
?
Please provide the device trade name(s).
?
- I.T.S. Clavicle Plates with Angular Stability;
- I.T.S. Volar Radius Plate with Angular Stability;
- I.T.S. Humeral Head Plate with Angular Stability;
- I.T.S. FR.O.H. Calcaneus Repair System;
- I.T.S. Pilonplate with Angular Stability;
- I.T.S. Olecranonplate with Angular Stability;
- I.T.S. Straight Plate with Angular Stability;
- I.T.S. Screw System;
- I.T.S. PRS Sacral Rod System;
- I.T.S. Fibula Plate PROlock with Angular Stability;
- I.T.S. Distal Humeral Plates with Angular Stability;
- I.T.S. LRS (Locking Reconstruction System) - DFL (Distal Femur Locking) & PTL (Proximal Lateral Tibia Locking);
- I.T.S. Ulna Locking Plates - (DUL & UOL) Systems;
- I.T.S. FLS - Foot Locking Plates System;
- I.T.S. HCS - Headless Compression Screw System;
- I.T.S. HOL - Hallux Osteotomy Locking Plate;
- I.T.S. Twist-Off Screws;
- I.T.S. Hand Locking Plates System - HLS;
- I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix);
- I.T.S. CTN - Cannulated Tibia Nail System
Please provide your Indications for Use below.
?
The I.T.S. Clavicle Plates with Angular Stability is a titanium implant fracture fixation system for repairing fractures located from the middle third to the distal third of the clavicle. Indications for Use include metaphyseal and diaphyseal fracture fixation of acute fractures, malunions, and non-unions of the clavicle. Other indications include corrective osteotomy and open and closed fractures. All fractures of the clavicle in metaphyseal and diaphyseal areas. Hygienisation of pseudo-arthroses with or without spongiosal graft. Corrective osteotomy. Open and closed fractures.
The I.T.S. Volar Radius Plates with Angular Stability is a titanium implant fracture fixation system for distal radius fractures of the wrist. Indications for Use include comminuted extra and intra-articular distal radius fractures, failed original fracture fixation, osteotomy and repair of a distal radius malunion, and comminuted volar shearing fractures. Complex intra-articular fractures of the distal radius. Complex extra-articular fractures of the distal radius. Osteotomies of the distal radius.
The I.T.S. Humeral Head Plates with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the proximal humerus in the shoulder. Indications for Use include fracture and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone. All stable and unstable humerus fractures with or without shaft involvement. Fractures of the greater or lesser tuberosities. Repair of the greater tuberosity following prior fixation failure or tuberosity "escape". Delayed or nonunion of the proximal humerus. Fixation following osteotomy of proximal humeral malunion. Displaced two, three and four part fracture of the proximal humerus. Displaced anterior and posterior fractures of the proximal humerus and greater tuberosity. Nonunion of two, three and four part fractures of the proximal humerus. Nonunion of anterior and posterior fracture-dislocations of the proximal humerus and
K260958 - Page 1 of 4
{4}
greater tuberosity. Dislocated, unstable 2, 3 and 4-segment fractures of the humeral head. Valgus-impacted 4-segment fractures of the humeral head. Non-union of the humeral head.
The I.T.S. FR.O.H. Calcaneus Repair System is a titanium implant fracture fixation system for repairing fractures located in the calcaneus heel bone of the foot. Indications for Use include: Intra and extra-articular fracture(s) of the calcaneus, corrective osteotomy, joint depression, non-displaced and tongue type, severely comminuted fractures, multifragmentary fractures, revision procedures, joint fusion, stabilization and fixation of fresh fractures, reconstruction of the calcaneus bones, and open and closed fractures of the calcaneus. The system can be used in both adult and pediatric patients. Complex Fractures of the Calcaneus. All intra-articular fractures with relevant joint distortion and comminution zone in which a semi-operative procedure (screws, drill wires) does not raise expectations of exact repositioning. Fractures of the sustentaculum tali.
The I.T.S. Pilon Plates with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the distal tibia in the leg. Indications for Use include fixation of complex intra- and extraarticular fractures, osteotomies, high medial malleolar fractures, and low boot top type rotational distal extra-articular shaft fractures of the distal tibia. Fractures of the tibial pilon of AO classification A3, especially groups C2 and C3.
The I.T.S. Olecranon Plates with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the proximal ulna (olecranon) in the elbow. Indications for Use include fixation of complex intra- and extraarticular fractures, osteotomies, nonunions, malunions, Type 1, 2, 3, & 4 simple fractures, and Type 5a, 5b, 5c, & 6 complex fractures (comminuted) of the proximal ulna (olecranon). All fractures of the olecranon.
The intended use of the I.T.S. Straight Plates with Angular Stability is to stabilize an osteotomy or fracture of small bones, long bones, the pelvis and the calcaneus in an adult or pediatric patient. Indications for use include commuted fractures, supracondylar fractures, intra-articular and extra-articular condylar fractures, fractures in osteopenic bone, non-unions, and mal-unions, as well, a fracture or osteotomy of the tibia, fibula, femoral condyle, acetabulum, humerus, ulna, middle hand and middle foot bones. Treatment of the calcaneal; hip arthrodesis, and provisional hole fixation.
The intended use of the I.T.S. Screw System is for corrective osteotomy or internal fracture fixation of the patella, pelvis, ankle, and long bones in an adult or pediatric patient. For the 4.0mm Cannulated Cancellous Screw indications for use are for radial and ulnar fractures, fractures of the proximal/distal humerus and of the patellas and for tendon fixation, maisonneuve injuries and disruption of the syndesmosis with bimalleolar or supramalleolar fractures and the instability of the talus centering. For the 6.5mm Cannulated Cancellous Screw, indications for use are for fractures of the femoral neck, tibiaplateau, of the sacrum and the articular cavity of the hip joint and the metaphyseal fractures of the distal femur and distal tibia, fixation of the ilio-sacral joint, and fusion of the foot and ankle.
For the 7.3mm Cannulated Cancellous Screw, indications for use are for fractures of the calcaneus, femoral neck, tibiaplateau, and of the sacrum and the articular cavity of the hip joint. Fusion of the foot and ankle, fixation of the Ilio-sacral joint, and metaphyseal fractures of the distal femur and distal tibia.
The intended use of the I.T.S. Pelvic Reconstruction System (PRS) is to stabilize one or more pelvic bone fractures in the pelvic ring area of the pelvis of an adult or pediatric patient. The indications for use of the I.T.S. PRS Sacral Rod System include fixation of fractures of the posterior pelvis, fixation of fractures of the posterior iliac spine, fixation of fractures of the posterior inferior iliac spine, dorsal stabilization of the posterior pelvic ring for unstable pelvic ring injuries, fixation of sacral fractures, and fracture dislocations of the sacro-iliac joint.
The I.T.S. Fibula Plate PROlock with Angular Stability is a titanium implant fracture fixation system for
K260958 - Page 2 of 4
{5}
repairing bone fractures located from the middle to the distal third of the fibula. Indications for Use include metaphyseal and diaphyseal fracture fixation of acute fractures, mal-unions, and non-unions of the distal fibula. Other indications include corrective osteotomy and open and closed fractures. Dislocated ankle-fractures group B+C according to Weber (with or without comminuted zones).
The I.T.S. Distal Humeral Plates with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the distal humerus in the elbow of a pediatric or adult patient. Indications for Use include intra-articular fractures, supracondylar fractures, osteotomies, and non-unions of the distal humerus. Supra- & diacondylar upper-arm fractures.
The I.T.S. LRS Locking Reconstruction System is a titanium implant fracture fixation system for stabilizing fractures in long bones of the distal femur of all pediatric patients (less than or equal to 21 years old) or adult patient. Indications for Use include distal shaft fractures, supracondylar fractures, intra-articular fractures, metaphyseal fractures, osteotomies, nonunions and malunions of the distal femur.
The I.T.S. LRS Locking Reconstruction System is a titanium implant fracture fixation system for stabilizing fractures in long bones of the proximal lateral tibia of all pediatric patients (less than or equal to 21 years old) or adult patient. Indications for Use include proximal shaft fractures, supracondylar fractures, intra-articular fractures, metaphyseal fractures, osteotomies, nonunions and malunions of the proximal tibia.
The Indications for Use of the I.T.S. GmbH - Distal Ulna Locking (DUL) Plate Systems is to stabilize fractures in the long bone of the distal ulna of an adult patient. Fractures of the ulnar head. Multifragmentary fracture of the ulnar head. Subcapital fractures of the ulnar head. Metaphyseal comminuted fractures of the distal ulna. Combined ulnar head and ulnar shaft fractures.
The Indications for Use of the I.T.S. GmbH - Ulna Osteotomy Locking (UOL) Plate Systems is to stabilize osteotomies in the long bone of the mid-ulna of an adult patient. Impaction syndrome of the ulnar wrist. Symptomatic, post-traumatic ulnar malposition in the distal radio-ulnar joint (DRUJ). Degenerative ulnar wrist. Correction of the ulnar positive relative to the unaffected other side up to a maximum of 6mm in one step or 13mm in two steps.
The I.T.S. FLS Foot Locking Plate System is indication for use in internal fixation, reconstruction or athrodesis of small bones including the fore, mid and hind foot and ankle. Examples of these procedures may included but are not limited to replantation, lag screw techniques, joint fusions, corrective osteotomies, and treatment of fracture. Metatarsal fractures. TMT I-V Arthrodesis, MTP I-V Arthrodesis, TN, CC Arthrodesis, Corrective Osteotomies.
The I.T.S. HCS - Headless Compression Screw System for sizes of 3.5mm or smaller is indicated for use in fixation small bone fractures or for small bone reconstruction including: mono or bicorlical osteotomies in the foot or hand; distal or proximal metatarsal or metacarpal osteotomies; weil osteotomy; fusion of the first metatarsalphalangeal joint and interphalangeal joint: fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc.); Akin type osteotomy; distal radius fractures (articular fragments). ulnar styloid fractures, radial head fractures, capitellum fractures, humeral head fractures, glenoid fractures, intercarpal distal and proximal fusions, malleolar fractures, patellar fractures, osteochondral fractures, talonavicular fusions, tibeo-talar fusions, and cuboid fusions.
And for sizes 4.3mmn or larger is indicated for use for fractures, corrective osteotomies, pseudoarthrosis, degenerative transformations of long bones in the hindfoot and large bone intra-articular fractures of the humerus, femur and tibia. The size of the chosen compression screw should be adapted to the specific indication.
The I.T.S. HOL - Hallux Osteotomy Locking Plate System is Indicated for use as an intramedullary self-locking plate for distal metatarsal osteotomies and for Hallux Valgus osteotomies up to a corrective angle of 25°. Intramedullary, self locking plate for distal metatarsal osteotomies. Hallux Valgus osteotomies up to a corrective angle of 25°.
K260958 - Page 3 of 4
{6}
The I.T.S. Twist-Off Screw System is indicated for use for small bone fixation of bone fractures or for bone reconstruction. Examples include small bone fragments, Weil- Osteotomy, Mono-Cortical fixation, Osteotomies and fractures fixation in the foot and hand. For treatment of fractures, corrective osteotomies, arthrodesis and degenerative transformations of small bones.
The I.T.S. Hand Locking Plate System -HLS is indicated for use in fracture fixation of: The phalanges, The metacarpal bones, The carpal bones, For arthrodesis, For corrective osteotomies and For subcapital radial head fractures.
Indications for use of the I.T.S. Pelvic Reconstruktion System (PRS Phoenix) include: Fractures involving the Posterior Wall & Posterior Column, Fractures involving the Anterior Column of the Acetabulum, Fractures involving the Quadrilateral Surface, Symphyseal Disruptions & Para-symphseal Fractures, Fractures of the ilium, Fractures of the SIJ, Dorsal neutralization plating for posterior pelvic ring fractures, Osteotomies, arthrodesis and sacroiliac joint dislocations, Revision surgery of pseudoarthroses, non-unions and mal unions.
The I.T.S. CTN – Cannulated Tibia Nail System is indicated for use in long bone tibia fracture fixation which include: Proximal, metaphyseal, epiphyseal and distal shaft fractures; Segmental, simple, compound and comminuted fractures; Transverse, oblique and spiral fractures; Surgically created defects using osteotomies, such as for leg length discrepancies or deformity; Pathologic fractures; Pseudoarthrosis, non-union, mal-union and delayed union of the tibia; Fractures involving osteopenic and osteoporotic bone; Open fractures of the tibia and; Reconstruction of the tibia after tumor resection and/or bone loss.
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
| --- | --- | --- |
| Please select the age group(s) for which the device(s) is to be used. | ☐ Neonates/Newborns (Birth to < 29 days old) ☐ Infants (29 days old to < 2 years old) ☐ Children (2 years old to < 12 years old) ☐ Adolescents (12 years old to < 22 years old) ☑ Adults (22 years old and greater) | ? |
K260958 - Page 4 of 4
{7}
# **K260958**
**510(k) Submission**
**I.T.S. GmbH**
**MR Conditional**
# **510(k) Summary**
**NAME OF FIRM:** I.T.S. GmbH
Autal 28
Lassnitzhoehe, 8301
Austria
www.its-implant.com
**510(k) FIRM CONTACT:** Mindy Mccan
Qserve Group US, Inc.
350 S Main Street, Suite 309
Doylestown, Pennsylvania, 18901, United States
Tel. No. 424-271-8169
e-mail: usagent@qservegroup.com
**TRADE NAME:**
I.T.S. Clavicle Plates with Angular Stability
I.T.S. Volar Radius Plate with Angular Stability
I.T.S. Humeral Head Plate with Angular Stability
I.T.S. FR.O.H. Calcaneus Repair System
I.T.S. Pilonplate with Angular Stability
I.T.S. Olecranonplate with Angular Stability
I.T.S. Straight Plate with Angular Stability
I.T.S. Screw System
I.T.S. PRS Sacral Rod System
I.T.S. Fibula Plate PROlock with Angular Stability
I.T.S. Distal Humeral Plates with Angular Stability
I.T.S. LRS (Locking Reconstruction System) – DFL (Distal Femur Locking) & PTL (Proximal Lateral Tibia Locking)
I.T.S. Ulna Locking Plates – (DUL & UOL) Systems
I.T.S. FLS – Foot Locking Plates System
I.T.S. HCS – Headless Compression Screw System
I.T.S. HOL – Hallux Osteotomy Locking Plate
I.T.S. Twist-Off Screws
I.T.S. Hand Locking Plates System - HLS
I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix)
I.T.S. CTN – Cannulated Tibia Nail System
**DATE:** June 11, 2026
**COMMON NAME:** Plate, Fixation, Bone,
Screw, Fixation, Bone,
Rod, Fixation, Intramedullary and Accessories
**REGULATORY CLASS:** Class II
**DEVICE PRODUCT CODE:** HRS, HWC, HSB
Page 1/6
K260958
{8}
# **510(k) Submission**
**I.T.S. GmbH**
**MR Conditional**
# **SUBSTANTIAL EQUIVALENCE:**
**PRIMARY PREDICATE** I.T.S. Clavicle Plates with Angular Stability (**K050852**)
# **ADDITIONAL PREDICATES**
I.T.S. Volar Radius Plate with Angular Stability (**K033756**)
I.T.S. Humeral Head Plate with Angular Stability (**K051412**)
I.T.S. FR.O.H. Calcaneus Repair System (**K051642**)
I.T.S. Pilonplate with Angular Stability (**K052011**)
I.T.S. Olecranonplate with Angular Stability (**K052368**)
I.T.S. Straight Plate with Angular Stability (**K060156**)
I.T.S. Screw System (**K060156**)
I.T.S. PRS Sacral Rod System (**K063166**)
I.T.S. Fibula Plate PROlock with Angular Stability (**K063672**)
I.T.S. Distal Humeral Plates with Angular Stability (**K080184**)
I.T.S. LRS (Locking Reconstruction System) – DFL (Distal Femur Locking) & PTL (Proximal Lateral Tibia Locking) (**K093868**)
I.T.S. Ulna Locking Plates – (DUL & UOL) Systems (**K130008**)
I.T.S. FLS – Foot Locking Plates System (**K131722**)
I.T.S. HCS – Headless Compression Screw System (**K131722**)
I.T.S. HOL – Hallux Osteotomy Locking Plate (**K131722**)
I.T.S. Twist-Off Screws (**K131722**)
I.T.S. Hand Locking Plates System – HLS (**K142418**)
I.T.S. Pelvic Reconstruction System (PRS RX & PRS Phoenix) (**K210935**)
I.T.S. CTN – Cannulated Tibia Nail System (**K132945**)
# **DEVICE DESCRIPTION:**
The purpose of this Traditional 510(k) is to add the updated MR conditional safety information to the labeling and provide performance data. The intended use, technological characteristics, function and operating principles and anatomical site for implantation of the devices remain unchanged. No modification has been made to the device material, sterilization, packaging, and the manufacturing processes of the currently marketed devices.
# **INDICATION FOR USE:**
The I.T.S. Clavicle Plates with Angular Stability is a titanium implant fracture fixation system for repairing fractures located from the middle third to the distal third of the clavicle. Indications for Use include metaphyseal and diaphyseal fracture fixation of acute fractures, malunions, and non-unions of the clavicle. Other indications include corrective osteotomy and open and closed fractures. All fractures of the clavicle in metaphyseal and diaphyseal areas. Hygienisation of pseudo-arthroses with or without spongiosal graft. Corrective osteotomy. Open and closed fractures.
The I.T.S. Volar Radius Plates with Angular Stability is a titanium implant fracture fixation system for distal radius fractures of the wrist. Indications for Use include comminuted extra and intra-articular distal radius fractures, failed original fracture fixation, osteotomy and repair of a distal radius malunion, and comminuted volar shearing fractures. Complex intra-articular fractures of the distal radius. Complex extra-articular fractures of the distal radius. Osteotomies of the distal radius.
The I.T.S. Humeral Head Plates with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the proximal humerus in the shoulder. Indications for Use include fracture and fracture dislocations, osteotomies, and non-unions of the proximal humerus, particularly in osteopenic bone. All stable and unstable humerus fractures with or without shaft involvement. Fractures of the greater or lesser tuberosities. Repair of the greater tuberosity following prior fixation failure or tuberosity “escape”. Delayed or nonunion of the proximal humerus. Fixation following osteotomy of proximal humeral malunion. Displaced two, three and four part fracture of
Page 2/6
K260958
{9}
# **510(k) Submission**
**I.T.S. GmbH**
**MR Conditional**
the proximal humerus. Displaced anterior and posterior fractures of the proximal humerus and greater tuberosity. Nonunion of two, three and four part fractures of the proximal humerus. Nonunion of anterior and posterior fracture-dislocations of the proximal humerus and greater tuberosity. Dislocated, unstable 2, 3 and 4-segment fractures of the humeral head. Valgus-impacted 4-segment fractures of the humeral head. Non-union of the humeral head.
The I.T.S. F.R.O.H. Calcaneus Repair System is a titanium implant fracture fixation system for repairing fractures located in the calcaneus heel bone of the foot. Indications for Use include: Intra and extra-articular fracture(s) of the calcaneus, corrective osteotomy, joint depression, non-displaced and tongue type, severely comminuted fractures, multifragmentary fractures, revision procedures, joint fusion, stabilization and fixation of fresh fractures, reconstruction of the calcaneus bones, and open and closed fractures of the calcaneus. The system can be used in both adult and pediatric patients. Complex Fractures of the Calcaneus. All intra-articular fractures with relevant joint distortion and comminution zone in which a semi-operative procedure (screws, drill wires) does not raise expectations of exact repositioning. Fractures of the sustentaculum tali.
The I.T.S. Pilon Plates with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the distal tibia in the leg. Indications for Use include fixation of complex intra- and extraarticular fractures, osteotomies, high medial malleolar fractures, and low boot top type rotational distal extra-articular shaft fractures of the distal tibia. Fractures of the tibial pilon of AO classification A3, especially groups C2 and C3.
The I.T.S. Olecranon Plates with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the proximal ulna (olecranon) in the elbow. Indications for Use include fixation of complex intra- and extraarticular fractures, osteotomies, nonunions, malunions, Type 1, 2, 3, & 4 simple fractures, and Type 5a, 5b, 5c, & 6 complex fractures (comminuted) of the proximal ulna (olecranon). All fractures of the olecranon.
The intended use of the I.T.S. Straight Plates with Angular Stability is to stabilize an osteotomy or fracture of small bones, long bones, the pelvis and the calcaneus in an adult or pediatric patient. Indications for use include commuted fractures, supracondylar fractures, intra-articular and extra-articular condylar fractures, fractures in osteopenic bone, non-unions, and mal-unions, as well, a fracture or osteotomy of the tibia, fibula, femoral condyle, acetabulum, humerus, ulna, middle hand and middle foot bones. Treatment of the calcaneal; hip arthrodesis, and provisional hole fixation.
The intended use of the I.T.S. Screw System is for corrective osteotomy or internal fracture fixation of the patella, pelvis, ankle, and long bones in an adult or pediatric patient. For the 4.0mm Cannulated Cancellous Screw indications for use are for radial and ulnar fractures, fractures of the proximal/distal humerus and of the patellas and for tendon fixation, maisonneuve injuries and disruption of the syndesmosis with bimalleolar or supramalleolar fractures and the instability of the talus centering. For the 6.5mm Cannulated Cancellous Screw, indications for use are for fractures of the femoral neck, tiblaplateau, of the sacrum and the articular cavity of the hip joint and the metaphyseal fractures of the distal femur and distal tibia, fixation of the ilio-sacral joint, and fusion of the foot and ankle. For the 7.3mm Cannulated Cancellous Screw, indications for use are for fractures of the calcaneus, femoral neck, tiblaplateau, and of the sacrum and
Page 3/6
K260958
{10}
# 510(k) Submission
## I.T.S. GmbH
## MR Conditional
the articular cavity of the hip joint. Fusion of the foot and ankle, fixation of the Ilio-sacral joint, and metaphyseal fractures of the distal femur and distal tibia.
The intended use of the I.T.S. Pelvic Reconstruction System (PRS) is to stabilize one or more pelvic bone fractures in the pelvic ring area of the pelvis of an adult or pediatric patient. The indications for use of the I.T.S. PRS Sacral Rod System include fixation of fractures of the posterior pelvis, fixation of fractures of the posterior iliac spine, fixation of fractures of the posterior inferior iliac spine, dorsal stabilization of the posterior pelvic ring for unstable pelvic ring injuries, fixation of sacral fractures, and fracture dislocations of the sacro-iliac joint.
The I.T.S. Fibula Plate PROlock with Angular Stability is a titanium implant fracture fixation system for repairing bone fractures located from the middle to the distal third of the fibula. Indications for Use include metaphyseal and diaphyseal fracture fixation of acute fractures, mal-unions, and non-unions of the distal fibula. Other indications include corrective osteotomy and open and closed fractures. Dislocated ankle-fractures group B+C according to Weber (with or without comminuted zones).
The I.T.S. Distal Humeral Plates with Angular Stability is a titanium implant fracture fixation system for stabilizing fractures of the distal humerus in the elbow of a pediatric or adult patient. Indications for Use include intra-articular fractures, supracondylar fractures, osteotomies, and non-unions of the distal humerus. Supra- & diacondylar upper-arm fractures.
The I.T.S. LRS Locking Reconstruction System is a titanium implant fracture fixation system for stabilizing fractures in long bones of the distal femur of all pediatric patients (less than or equal to 21 years old) or adult patient. Indications for Use include distal shaft fractures, supracondylar fractures, intra-articular fractures, metaphyseal fractures, osteotomies, nonunions and malunions of the distal femur.
The I.T.S. LRS Locking Reconstruction System is a titanium implant fracture fixation system for stabilizing fractures in long bones of the proximal lateral tibia of all pediatric patients (less than or equal to 21 years old) or adult patient. Indications for Use include proximal shaft fractures, supracondylar fractures, intra-articular fractures, metaphyseal fractures, osteotomies, nonunions and malunions of the proximal tibia.
The Indications for Use of the I.T.S. GmbH - Distal Ulna Locking (DUL) Plate Systems is to stabilize fractures in the long bone of the distal ulna of an adult patient. Fractures of the ulnar head. Multifragmentary fracture of the ulnar head. Subcapital fractures of the ulnar head. Metaphyseal comminuted fractures of the distal ulna. Combined ulnar head and ulnar shaft fractures. The Indications for Use of the I.T.S. GmbH - Ulna Osteotomy Locking (UOL) Plate Systems is to stabilize osteotomies in the long bone of the mid-ulna of an adult patient. Impaction syndrome of the ulnar wrist. Symptomatic, post-traumatic ulnar malposition in the distal radio-ulnar joint (DRUJ). Degenerative ulnar wrist. Correction of the ulnar positive relative to the unaffected other side up to a maximum of 6mm in one step or 13mm in two steps.
The I.T.S. FLS Foot Locking Plate System is indication for use in internal fixation, reconstruction or athrodesis of small bones including the fore, mid and hind foot and ankle. Examples of these procedures may included but are not limited to replantation, lag screw techniques, joint fusions, corrective osteotomies, and treatment of fracture. Metatarsal fractures. TMT I-V
Page 4/6
K260958
{11}
# 510(k) Submission
## I.T.S. GmbH
## MR Conditional
Arthrodesis, MTP I-V Arthrodesis, TN, CC Arthrodesis, Corrective Osteotomies.
The I.T.S. HCS - Headless Compression Screw System for sizes of 3.5mm or smaller is indicated for use in fixation small bone fractures or for small bone reconstruction including: mono or bicorlical osteotomies in the foot or hand; distal or proximal metatarsal or metacarpal osteotomies; weil osteotomy; fusion of the first metatarsalphalangeal joint and interphalangeal joint: fixation of osteotomies for Hallux Valgus treatment (such as Scarf, Chevron, etc.); Akin type osteotomy; distal radius fractures (articular fragments). ulnar styloid fractures, radial head fractures, capitellum fractures, humeral head fractures, glenoid fractures, intercarpal distal and proximal fusions, malleolar fractures, patellar fractures, osteochondral fractures, talonavicular fusions, tibeo-talar fusions, and cuboid fusions. And for sizes 4.3mmn or larger is indicated for use for fractures, corrective osteotomies, pseudoarthrosis, degenerative transformations of long bones in the hindfoot and large bone intra-articular fractures of the humerus, femur and tibia. The size of the chosen compression screw should be adapted to the specific indication.
The I.T.S. HOL - Hallux Osteotomy Locking Plate System is Indicated for use as an intramedullary self-locking plate for distal metatarsal osteotomies and for Hallux Valgus osteotomies up to a corrective angle of 25°. Intramedullary, self locking plate for distal metatarsal osteotomies. Hallux Valgus osteotomies up to a corrective angle of 25°.
The I.T.S. Twist-Off Screw System is indicated for use for small bone fixation of bone fractures or for bone reconstruction. Examples include small bone fragments, Weil- Osteotomy, Mono-Cortical fixation, Osteotomies and fractures fixation in the foot and hand. For treatment of fractures, corrective osteotomies, arthrodesis and degenerative transformations of small bones.
The I.T.S. Hand Locking Plate System -HLS is indicated for use in fracture fixation of: The phalanges, The metacarpal bones, The carpal bones, For arthrodesis, For corrective osteotomies and For subcapital radial head fractures.
Indications for use of the I.T.S. Pelvic Reconstruktion System (PRS Phoenix) include: Fractures involving the Posterior Wall & Posterior Column, Fractures involving the Anterior Column of the Acetabulum, Fractures involving the Quadrilateral Surface, Symphyseal Disruptions & Para-symphseal Fractures, Fractures of the ilium, Fractures of the SIJ, Dorsal neutralization plating for posterior pelvic ring fractures, Osteotomies, arthrodesis and sacroiliac joint dislocations, Revision surgery of pseudoarthroses, non-unions and mal unions.
The I.T.S. CTN – Cannulated Tibia Nail System is indicated for use in long bone tibia fracture fixation which include: Proximal, metaphyseal, epiphyseal and distal shaft fractures; Segmental, simple, compound and comminuted fractures; Transverse, oblique and spiral fractures; Surgically created defects using osteotomies, such as for leg length discrepancies or deformity; Pathologic fractures; Pseudoarthrosis, non-union, mal-union and delayed union of the tibia; Fractures involving osteopenic and osteoporotic bone; Open fractures of the tibia and; Reconstruction of the tibia after tumor resection and/or bone loss.
Page 5/6
K260958
{12}
# **510(k) Submission**
**I.T.S. GmbH**
**MR Conditional**
**CLINICAL TESTING:** Clinical data was not required for this submission.
**NON-CLINICAL TESTING:** The following non-clinically tests were performed:
- Non-clinical testing was conducted per FDA’s guidance “Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment”, and is provided to support the conditional safety in the MR environment including assessment of magnetically induced displacement force (ASTM F2052) and torque (ASTM F2213), radio frequency (RF) heating (ASTM F2182), image artifacts (ASTM F2119) and standard practice for Marking Medical Devices and Other Items for Safety in the Magnetic Resonance Environment (ASTM F2503).
The non-clinical performance data demonstrate that the subject devices, when exposed to the MR environment under specific MR conditions of use, raise no new questions of safety and efficacy.
**SUMMARY OF TECHNOLOGICAL CHARACTERISTICS:** The intended use, technological characteristics, function and operating principles and anatomical site for implantation of the devices remain unchanged. No modification has been made to the device material, sterilization, packaging, and the manufacturing processes of the currently marketed devices.
**CONCLUSIONS:** Based on the results of the non-clinical MRI safety testing described above, it is concluded that the subject devices are substantially equivalent to the predicate devices. The data demonstrate that the updated labeling does not alter the device performance regarding its intended use. Furthermore, the evaluation confirms that the subject devices remain as safe and effective as the predicates for their cleared indications, raising no new questions of safety or effectiveness.
Page 6/6
K260958
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.