CFN-CANNULATED FEMUR NAIL, CTN-CANNULATED TIBIA NAIL, CHN-CANNULATED HUMERAL NAIL
Applicant
I.T.S. GmbH
Product Code
HSB · Orthopedic
Decision Date
Mar 7, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The intended use of the I.T.S. IM Nail Systems CFN-CTN-CHN is to stablilize and fix long bone fractures to facilitate healing in an adult patient and is composed of the following categories: The I.T.S. CFN - Cannulated Femur Nail System is indicated for use in long bone femur fracture fixation which include: Open and closed femur shaft fractures; Intertrochanteric, supracondylar and ipsilateral fractures; High subtrochanteric fractures; Combined inter and subrochanteric fractures; Pertrochanteric fractures; Pseudoarthrosis and correction osteotomy; Non-union, mal-union and delayed union fractures; Pathological fractures, impending pathologic fractures, and tumor resections and; Fractures proximal to a total knee arthroplasty; The system is not for spinal use. The I.T.S. CTN - Cannulated Tibia Nail System is indicated for use in long bone tibia fracture fixation which include: Proximal, metaphyseal, epiphyseal and distal shaft fractures; Segmental, simple, compound and comminuted fractures; Transverse, oblique and spiral fractures; Surgically created defects using osteotomies,such as for leg length discrepencies or deformity; Pathologic fractures; Pseudoarthrosis, non-union, mal-union and delayed union of the tibia; Fractures involving osteopenic and osteoporotic bone; Open fractures of the tibia and; Reconstruction of the tibia after tumor resection and/or bone loss. The system is not for spinal use. The I.T.S. CHN - Cannulated Humeral Nail System is indicated for use in long bone humerus fracture fixation which include: Dislocated, unstable 2, 3 and 4 part fractures of the proximal humerus; Valgus-impacted 4 part fractures of the proximal humerus; Proximal hiumeral fractures with diaphyseal extension into the shaft; Pseudoarthrosis, non-unions, mal-unions and malalignments of the proximal humerus and; Pathological and impending pathological fractures. The system is not for spinal use.
Device Story
I.T.S. IM Nail Systems (CFN, CTN, CHN) are intramedullary trauma implants for long bone fracture fixation; components include cannulated nails, end caps, and cortical screws. Nails are inserted antegrade over calibrated guide wires; proximal/distal cross-screw fixation secures the nail. Distal hole/slot location requires x-ray fluoroscopic imaging; proximal insertion uses dedicated guide instrumentation. Used in clinical/OR settings by orthopedic surgeons. System stabilizes fractures to facilitate bone healing; benefits include anatomical alignment and structural support for various fracture types, including osteopenic/osteoporotic bone and tumor-related defects.
Clinical Evidence
No clinical data provided; substantial equivalence determined based on technological characteristics and design similarities to predicate devices.
Technological Characteristics
Materials: Alloyed 6-4 Titanium (ASTM F136) with DOTIZE anodized surface treatment. Design: Cannulated intramedullary nails with proximal/distal cross-screw fixation. Energy: Mechanical fixation. Connectivity: None. Sterilization: Provided non-sterile.
Indications for Use
Indicated for adult patients requiring stabilization and fixation of long bone fractures in the femur, tibia, or humerus, including shaft, metaphyseal, epiphyseal, and proximal fractures, non-unions, mal-unions, pathological fractures, and tumor resections. Contraindicated for spinal use.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
Synthes - Proximal Femoral (PFN) & Cannulated Femoral Nail (CFN) Systems (K970097, K954856)
{0}------------------------------------------------
K132945 (1/4)
I.T.S. GmbH - 510(k) Summary -
## 510(k) Summary of Safety and Effectiveness
MAR 0 7 2014
## SAFE MEDICAL DEVICES ACT OF 1990 510(k) Summary
| NAME OF FIRM: | I.T.S. GmbH<br>Autal 28<br>Lassnitzhöehe A-8301<br>AUSTRIA<br>www.its-implant.com |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) FIRM CONTACT: | Al Lippincott<br>Engineering Consulting Services, Inc.<br>3150 E. 200th St.<br>Prior Lake, MN 55372<br>Tel. No. 952-492-5858<br>e-mail: allippincott@msn.com |
| DATE: | August 31, 2013 |
| TRADE NAME: | I.T.S. IM Nail Systems CFN-CTN-CHN |
| COMMON NAME: | Femoral, Tibial and Humeral - Intramedullary Nail |
| CLASSIFICATION: | Rod, Fixation, Intramedullary and Accessories;<br>Screw, Fixation, Bone;<br>Nail, Fixation, Bone;<br>Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component<br>Intramedullary fixation rod (See 21 CFR, Sec. 888.3020).<br>Smooth or threaded metallic bone fixation fastener (See 21CFR, Sec.<br>888.3040).<br>Single/multiple component metallic bone fixation applieances and<br>accessories (See 21 CRF, Sec. 888.3030). |
| DEVICE PRODUCT CODE: | HSB |
| SUBSEQUENT PRODUCT CODE: | HWC, JDS, KTT |
| SUBSTANTIALLY<br>EQUIVALENT DEVICES | Synthes - Proximal Femoral (PFN) & Cannulated Femoral Nail (CFN)<br>Systems (K970097, K954856)<br>Zimmer - Sirus IM Femur Nail System (K093270)<br>Stryker - T2 RECON Nail System (K032898, K051624, K102992)<br>Biomet - Femoral Locking Nail System (K072161)<br>Synthes - Ti Cannulated Tibial Nail (CTN) System (K962047)<br>Zimmer - Sirus IM Tibia Nail System (K093270, K082770)<br>Stryker - T2 Tibial Nailing System (K021027,K011622,K003018)<br>Biomet - Tibial Locking Nail System (K063570)<br>I.T.S. - CONNEXX Locking Tibia Nail (K080706) |
I.T.S. IM Nail Systems CFN-CTN-CTN-CHN Page 1 of 4 Section XII .
{1}------------------------------------------------
#### I.T.S. GmbH - 510(k) Summary -
I.T.S. - Distal Humeral 3.5mm Cortical Scew (K080184) Synthes - MultiLoc Humeral Nailing System (K120807) Smith & Nephew - TriGen Straight Humeral Nail (K032722) Stryker - T2 Proximal Humeral Nailing System (K042396) Acumed - Polarus Humeral Rod (K920666, K951673, K951740)
#### DEVICE DESCRIPTION:
The I.T.S. IM Nail Systems CFN-CTN-CHN consists of Predicate type intramedullary nail trauma implant components commonly found with large companies with orthopedic markets in the United States. These 'Intramedullary (IM) Nail trauma implant devices' consist of the following categories:
#### 1. CFN - Cannulated Femur Nail System
#### 2. CTN - Cannulated Tibia Nail System
#### 3. CHN - Cannulated Humeral Nail System
A brief and concise description of each system is enclosed as follows:
1. CFN - Cannulated Femur Nail System: The I.T.S. CFN -Cannulated Femur Nail System is a curved/bowed Intramedullary(IM) Nail in a right and left configuration to fit the natural bow of the femur and is inserted proximally/antegrade over 3.5/3.0mm calibrated guide wires. The CFN - IM Nail is composed of various nail diameters in sizes of 9.0, 10.0, 11.0, 12.0 and 13.0mm and various lengths from 240mm to 480mm in 20mm increments. The CFN - IM Nail accepts a 6.5mm Cortical Screw in various lengths for cross-screw fixation in the larger diameter proximal region of the nail and a 4.7mm Triple-lead Cortical Screw in various lengths for crossscrew fixation in the distal portion of the nail. A dynamization slot is located in the distal portion of the nail and an End Cap (in +0, +5, +10, +15, +20, +25 & +30 lengths) is available for proximal closing of the nail. The nail proximal cross-screw insertion uses Insertion Guide instrumentation for location of the hole and screw insertion preparation throught the bone cortex and nail. The nail distal cross-screw insertion uses x-ray fluoroscopic imaging for nail distal hole/slot location.
All CFN - JM Nail, End Cap and Screw components are manufactured from Alloyed 6-4 Titanium material to ASTM F136 and are processed with an anodize DOTIZE surface treatment.
Associated instrumentation such as Insertion Guide accessories, guide pins, drill sleeves, drill guides, drills, screwdriver, depth gauge with removal instruments are available with the system. All CFN - IM Nails, End Caps and Screws are provided Non-Sterile.
#### I.T.S. IM Nail Systems CFN-CTN-CHN
{2}------------------------------------------------
#### I.T.S. GmbH - 510(k) Summary -DEVICE DESCRIPTION CONTINUED:
2. CTN - Cannulated Tibia Nail System: The I.T.S. CTN -Cannulated Tibia Nail System is a universal straight Intramedullary(IM) Naïl with a 50 proximal and distal bend configuration to fit the anatomy of the tibia and is inserted proximally/antegrade over 3.5/3.0mm calibrated guide wires. The CTN - IM Nail is composed of various nail diameters in sizes of 9.0, 10.0. 11.0. and 12.0mm and various lengths from 240mm to 420mm in 1 5mm increments. The CTN - IM Nail accepts only the 4.7mm Triple-lead Cortical Screw in various lengths for cross-screw fixation in both the proximal/distal portion of the nail. A dynamization slot is located in the proximal portion of the nail and an End Cap (in +0, +5, +10, +15, +20, +25, +30 lengths) is available for proximal closing of the nail. The nail proximal cross-screw insertion uses Insertion Guide instrumentation for location of the hole/slot and screw insertion preparation throught the bone cortex and nail. The nail distal cross-screw insertion uses x-ray fluoroscopic imaging for nail distal hole location.
All CTN - IM Nail, End Cap and Screw components are manufactured from Alloved 6-4 Titanium material to ASTM F136 and are processed with an anodize DOTIZE surface treatment.
Associated instrumentation such as Insertion Guide accessories, guide pins, drill sleeves, drill guides, drills, screwdriver, depth gauge and removal instruments are available with the system. All CTN - IM Nails, End Caps and Screws are provided Non-Sterile.
3. CHN - Cannulated Humeral Nail System: The I.T.S. CHN -Cannulated Humeral Nail System is a right and left Intramedullary (IM) straight Nail with a 4 proximal bend configuration to fit the anatomy of the humerus and is inserted proximally over 2.0/2.5mm calibrated guide wires. The CHN - IM Nail is composed of various nail diameters in sizes of 7.0. 8.0. and 9.0mm and various lengths from 140mm to 320mm in 10 and 20mm increments. The CHN - IM Nail accepts only a 3.5mm Double-lead Cortical Screw in various lengths for cross-screw fixation in both the proximal/distal portion of the nail. An End Cap (in +0, +5, +10, +15, +20, +25 & +30 lengths) is available for proximal closing of the nail. The nail proximal cross-screw insertion uses Insertion Guide instrumentation for location of the hole and screw insertion preparation throught the bone cortex and nail. The nail distal cross-screw insertion uses x-ray fluoroscopic imaging for nail distal hole location.
All CHN - IM Nail, End Cap and Screw components are manufactured from Alloyed 6-4 Titanium material to ASTM F136 and are processed with an anodize DOTIZE surface treatment.
Associated instrumentation such as Insertion Guide accessories, guide pin, drill sleeves, drill guides, drills, screwdriver, depth gauge and removal instruments are available with the system. All CHN -IM Nails, End Caps and Screws are provided Non-Sterile.
{3}------------------------------------------------
#### I.T.S. GmbH - 510(k) Summary -
#### INTENDED USE:
The intended use of the I.T.S. IM Nail Systems CFN-CTN-CHN is to stablilize and fix long bone fractures to facilitate healing in an adult patient and is composed of the following categories:
The I.T.S. CFN - Cannulated Femur Nail System is indicated for use in long bone femur fracture fixation which include:
Open and closed femur shaft fractures; Intertrochanteric, supracondylar and ipsilateral fractures; High subtrochanteric fractures; Combined inter and subrochanteric fractures; Pertrochanteric fractures; Pseudoarthrosis and correction osteotomy; Non-union, mal-union and delayed union fractures; Pathological fractures, impending pathologic fractures, and tumor resections and; Fractures proximal to a total knee arthroplasty; The system is not for spinal use.
### The I.T.S. CTN - Cannulated Tibia Nail System is indicated for use in long bone tibia fracture fixation which include:
Proximal, metaphyseal, epiphyseal and distal shaft fractures; Segmental, simple, compound and comminuted fractures; Transverse, oblique and spiral fractures; Surgically created defects using osteotomies,such as for leg length discrepencies or deformity; Pathologic fractures; Pseudoarthrosis, non-union, mal-union and delayed union of the tibia; Fractures involving osteopenic and osteoporotic bone; Open fractures of the tibia and; Reconstruction of the tibia after tumor resection and/or bone loss.
The system is not for spinal use.
The I.T.S. CHN - Cannulated Humeral Nail System is indicated for use in long bone humerus fracture fixation which include: Dislocated, unstable 2, 3 and 4 part fractures of the proximal humerus; Valgus-impacted 4 part fractures of the proximal humerus; Proximal hiumeral fractures with diaphyseal extension into the shaft; Pseudoarthrosis, non-unions, mal-unions and malalignments of the proximal humerus and; Pathological and impending pathological fractures.
The system is not for spinal use.
EQUIVALENCE:
The I.T.S. IM Nail Systems CFN-CTN-CHN are Substantially Equivalent(SE) to the various predicate IM Nail, End Cap and Screw Systems as listed. No nonclinical testing was used in the determination of substantial equivalence.
#### SUMMARY OF TECH-NOLOGICAL CHAR-ACTERISTICS
The I.T.S. IM Nail Systems CEN-CTN-CHN are Similar in Material, Geometry Design/Markings, and Indications to predicate system(s) currently sold in the U.S. market.
#### SUMMARY OF SAFETY AND EFFECTIVENESS:
The I.T.S. IM Nail Systems CFN-CTN-CHN are shown to be safe and effective for use in long bone fracture stablization and fixation in the femur, tibia and humerus.
#### I.T.S. IM Nail Systems CFN-CTN-CHN
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### March 7, 2014
I.T.S. GmbH % Mr. Al Lippincott U.S Agent and Official Correspondent for I.T.S. GmbH Engineering Consulting Services, Inc. 3150 E. 200th Street Prior Lake, Minnesota 55372
Re: K132945
Trade/Device Name: I.T.S. IM Nail Systems CFN-CTN-CHN Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: II Product Codes: HSB, HWC, JDS, KTT Dated: December 16, 2013 Received: December 20, 2013
Dear Mr. Lippincott:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
#### Page 2 - Mr. Al Lippincott
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Vincent JFDAvlin -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
· Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE
#### 510(k) NUMBER: ﺷﻘﻪ
# DEVICE NAME: I.T.S. IM Nail Systems CFN-CTN-CHN
The intended use of the I.T.S. IM Nail Systems is to stablilize and fix long bone fractures to facilitate healing in an adult patient and is composed of the following categories:
The I.T.S. CFN - Cannulated Femur Nail System is indicated for use in long bone femur fracture fixation which include:
Open and closed femur shaft fractures; Intertrochanteric, supracondylar and ipsilateral fractures; High subtrochanteric fractures; Combined inter and subrochanteric fractures; Pertrochanteric fractures; Pseudoarthrosis and correction osteotomy; Non-union, malunion and delayed union fractures; Pathological fractures, impending pathologic fractures, and tumor resections and; Fractures proximal to a total knee arthroplasty; The system is not for spinal use.
The I.T.S. CTN - Cannulated Tibia Nail System is indicated for use in long bone tibia fracture fixation which include:
Proximal, metaphyseal, epiphyseal and distal shaft fractures; Segmental, simple, compound and comminuted fractures; Transverse, oblique and spiral fractures; Surgically created defects using osteotomies,such as for leg length discrepencies or deformity; Pathologic fractures; Pseudoarthrosis, non-union, mal-union and delayed union of the tibia; Fractures involving osteopenic and osteoporotic bone; Open fractures of the tibia and; Reconstruction of the tibia after tumor resection and/or bone loss.
The system is not for spinal use.
The I.T.S. CHN - Cannulated Humeral Nail System is indicated for use in long bone humerus fracture fixation which include:
Dislocated, unstable 2, 3 and 4 part fractures of the proximal humerus; Valgusimpacted 4 part fractures of the proximal humerus; Proximal humeral fractures with diaphyseal extension into the shaft; Pseudoarthrosis, non-unions, mal-unions and malalignments of the proximal humerus and; Pathological and impending pathological fractures.
The system is not for spinal use.
Prescription Use
Over-The-Counter-Use - -
(Per 21 CFR 801 Subpart D)
.
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of Device Evaluation (ODE)
Casey L. Hanley, Ph.D.
1 | Page
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.