LED Light Therapy Mask (K9C-TK, K9C-K, K9P-YK, K9P-K, FM-03BK, FM-09K)
K260946 · Guangdong Newdermo Biotech Co., Ltd. · OHS · Jun 18, 2026 · General, Plastic Surgery
Device Facts
Record ID
K260946
Device Name
LED Light Therapy Mask (K9C-TK, K9C-K, K9P-YK, K9P-K, FM-03BK, FM-09K)
Applicant
Guangdong Newdermo Biotech Co., Ltd.
Product Code
OHS · General, Plastic Surgery
Decision Date
Jun 18, 2026
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The LED Light Therapy Mask (Model:K9C-TK) is an Over-the-Counter (OTC) device intended for treatment of wrinkles and mild to moderate inflammatory acne. 1. Deep Red and Infrared light (660nm + 850nm): Intended for treatment of wrinkles. 2. Red light and Near-Infrared light (630nm + 830nm): Intended for treatment of wrinkles. 3. Amber light(605nm): Intended for treatment of wrinkles. 4. Blue light and Red light (415nm + 630nm): Intended for treatment of mild to moderate inflammatory acne. 5. Blue light(415nm): Intended for treatment of mild to moderate inflammatory acne. The LED Light Therapy Mask (Models:K9C-K, K9P-YK, K9P-K) is an Over-the-Counter (OTC) device intended for treatment of wrinkles and mild to moderate inflammatory acne. 1. Deep Red and Infrared light (660nm + 850nm): Intended for treatment of wrinkles. 2. Red light and Near-Infrared light (630nm + 830nm): Intended for treatment of wrinkles. 3. Blue light and Red light (415nm + 630nm): Intended for treatment of mild to moderate inflammatory acne. 4. Blue light(415nm): Intended for treatment of mild to moderate inflammatory acne. The LED Light Therapy Mask(Model: FM-03BK) is an Over-the-Counter (OTC) device intended for treatment of wrinkles. 1. Red light (630nm): Intended for treatment of wrinkles. 2. Red light and Near-Infrared light (630nm + 830nm): Intended for treatment of wrinkles. The LED Light Therapy Mask (Model: FM-09K) is an Over-the-Counter (OTC) device intended for treatment of wrinkles and mild to moderate inflammatory acne. 1. Red light (630nm): Intended for treatment of wrinkles. 2. Red light and Near-Infrared light (630nm + 830nm): Intended for treatment of wrinkles. 3. Blue light and Red light (415nm + 630nm): Intended for treatment of mild to moderate inflammatory acne. 4. Blue light(415nm): Intended for treatment of mild to moderate inflammatory acne.
Device Story
Wearable LED light therapy mask for home use; treats wrinkles and mild to moderate inflammatory acne. Device consists of mask, controller, USB cable, and straps. Emits specific wavelengths (415nm, 605nm, 630nm, 660nm, 830nm, 850nm) via LED arrays. User selects treatment mode; device operates via manual (10 min) or automatic (15 min) timer. Output is light energy delivered to skin surface. Benefits include wrinkle reduction and acne management. No physician intervention required; patient self-operates.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1 series), photobiological safety (IEC 62471), and biocompatibility (ISO 10993 series).
Technological Characteristics
Wearable LED mask; Li-ion battery powered (3.7V, 2450mAh); wavelengths: 415nm, 605nm, 630nm, 660nm, 830nm, 850nm. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-83, IEC 62471. Biocompatible materials per ISO 10993-5, -10, -23.
Indications for Use
Indicated for OTC treatment of facial/body wrinkles and mild to moderate inflammatory acne in general population.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}
**FDA** **U.S. FOOD & DRUG**
ADMINISTRATION
June 18, 2026
Guangdong Newdermo Biotech Co., Ltd.
Junyi Xing
Certificate Engineer
Contact Address
Re: K260946
Trade/Device Name: LED Light Therapy Mask (K9C-TK, K9C-K, K9P-YK, K9P-K, FM-03BK, FM-09K)
Regulation Number: 21 CFR 878.4810
Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology
Regulatory Class: Class II
Product Code: OHS, OLP
Dated: March 23, 2026
Received: March 23, 2026
Dear Junyi Xing:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K260946 - Junyi Xing
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K260946 - Junyi Xing
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
**TANISHA** Digitally signed by
**L. HITHE -S**
**TANISHA L. HITHE -S**
Date: 2026.06.18
16:18:07 -04'00'
Tanisha Hithe
Assistant Director
DHT4A: Division of General Surgery Devices
OHT4: Office of Surgical and
Infection Control Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
# Indications for Use
510(k) Number (if known)
K260946
Device Name
LED Light Therapy Mask (K9C-TK, K9C-K, K9P-YK, K9P-K, FM-03BK, FM-09K)
Indications for Use (Describe)
The LED Light Therapy Mask adopts the latest LED light therapy technology, combining 6 kinds of Model: K9C-TK, K9C-K, K9P-YK, K9P-K, FM-03BK, FM-09K.
The LED Light Therapy Mask (Model:K9C-TK) is an Over-the-Counter (OTC) device intended for treatment of wrinkles and mild to moderate inflammatory acne.
1. Deep Red and Infrared light (660nm + 850nm): Intended for treatment of wrinkles.
2. Red light and Near-Infrared light (630nm + 830nm): Intended for treatment of wrinkles.
3. Amber light(605nm): Intended for treatment of wrinkles.
4. Blue light and Red light (415nm + 630nm): Intended for treatment of mild to moderate inflammatory acne.
5. Blue light(415nm): Intended for treatment of mild to moderate inflammatory acne.
The LED Light Therapy Mask (Models:K9C-K, K9P-YK, K9P-K) is an Over-the-Counter (OTC) device intended for treatment of wrinkles and mild to moderate inflammatory acne.
1. Deep Red and Infrared light (660nm + 850nm): Intended for treatment of wrinkles.
2. Red light and Near-Infrared light (630nm + 830nm): Intended for treatment of wrinkles.
3. Blue light and Red light (415nm + 630nm): Intended for treatment of mild to moderate inflammatory acne.
4. Blue light(415nm): Intended for treatment of mild to moderate inflammatory acne.
The LED Light Therapy Mask(Model: FM-03BK) is an Over-the-Counter (OTC) device intended for treatment of wrinkles.
1. Red light (630nm): Intended for treatment of wrinkles.
2. Red light and Near-Infrared light (630nm + 830nm): Intended for treatment of wrinkles.
The LED Light Therapy Mask (Model: FM-09K) is an Over-the-Counter (OTC) device intended for treatment of wrinkles and mild to moderate inflammatory acne.
1. Red light (630nm): Intended for treatment of wrinkles.
2. Red light and Near-Infrared light (630nm + 830nm): Intended for treatment of wrinkles.
3. Blue light and Red light (415nm + 630nm): Intended for treatment of mild to moderate inflammatory acne.
4. Blue light(415nm): Intended for treatment of mild to moderate inflammatory acne.
Type of Use (Select one or both, as applicable)
☐ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740 EF
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**\*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.\***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*'An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.'*
---FORM FDA 3881 (8/23)
Page 2 of 2
{5}
# 510(K) Summary
K260946
Prepared in accordance with the content and format regulatory requirements of
21 CFR Part 807.92
1. Submitter:
510(k) owner's name: Guangdong Newdermo Biotech Co., Ltd.
Establishment Registration Number: 3013178978
Address: Building 70(C28), No.9 Huateng Road, Jinshan Village, Shiqi Town, Panyu District, Guangzhou, Guangdong, 511450, P.R. CHINA
Tel: +86 15012419132
Contact person (including title): Junyi Xing (Manager)
Email: sales13@newdermo.com
Preparing date: June 10, 2026
2. Device name and classification:
Device Name: LED Light Therapy Mask
Model: K9C-TK, K9C-K, K9P-YK, K9P-K, FM-03BK, FM-09K
Classification Name: Light Based Over The Counter Wrinkle Reduction Over-The-Counter Powered Light Based Laser For Acne
Product code: OHS, OLP
Regulation Number: 21 CFR 878.4810
Regulatory Class: Class II
Review Panel: General & Plastic Surgery
3. Premarket Notification Class III Certification and Summary
Not applicable, the subject device is Class II.
Guangdong Newdermo Biotech Co.,Ltd
1 / 17
{6}
#### **4. Predicate Device(s):**
| Predicate # | Sponsor | Device name | 510(k) Number | Regulation Number | Product Code |
| --- | --- | --- | --- | --- | --- |
| **Predicate device1** | Harpar Grace International | Shani Darden LED light therapy mask | K214103 | 21 CFR 878.4810 | OLP, OHS |
| **Reference Device1** | Therabody, Inc. | TheraFace Mask | K230293 | 21 CFR 878.4810 | OHS, OLP |
| **Reference Device2** | Shenzhen Qianyu Technology Co., Ltd. | Wrinkle Treatment Device (JM2) | K240360 | 21 CFR 878.4810 | OHS |
| **Reference Device3** | Sunglor Technology Co., Ltd. | LED Light Therapy Device (SG-FSM, SG-NK, SG-CT) | K253073 | 21 CFR 878.4810 | OHS, ILY |
| **Predicate device2** | Shenzhen Kaiyan Medical Equipment Co., Ltd. | TRUDERMAL Pro (ZLD-390) | K251012 | 21 CFR 878.4810 | OHS, OLP, ILY |
| **Reference Device4** | Ismart Developments, Ltd. | d o LITE LED Device | K242382 | 21 CFR 878.4810 | OHS |
#### **5. Reason for Submission**
New device, there were no prior submissions for the device.
#### **6. Pre-Submission, IDE**
Not applicable, there is no prior submission.
#### **7. Device Description:**
LED Light Therapy Mask is a home use wearable LED photo therapy device. LED Light Therapy Mask intended for treatment of wrinkles and mild to moderate inflammatory acne. LED Light Therapy Mask is consisting of mask, controller, USB cable and straps. There are 6 models of LED Light Therapy Mask, including K9C-TK, K9P-YK, K9C-K, K9P-K, FM-03BK, FM-09K.
There are 6 kinds of light which include Red light (wavelength 630nm), Deep Red light ( wavelength 660nm), Blue light (wavelength 415nm), Amber Light (wavelength 605nm), Near-Infrared light (wavelength 830nm), Infrared light (wavelength 850nm).
For the FM-03BK model, the available modes are Red Light only (630 nm) and Red + Near-Infrared Light (630nm+830nm).
#### **8. Indications for Use:**
The LED Light Therapy Mask adopts the latest LED light therapy technology, combining 6 kinds of Model: K9C-TK, K9C-K, K9P-YK, K9P-K, FM-03BK, FM-09K.
The LED Light Therapy Mask (Model: K9C-TK) is an Over-the-Counter (OTC) device intended for treatment of wrinkles and mild to moderate inflammatory acne.
Guangdong Newdermo Biotech Co.,Ltd
2 / 17
{7}
1. Deep Red and Infrared light (660nm + 850nm): Intended for treatment of wrinkles.
2. Red light and Near-Infrared light (630nm + 830nm): Intended for treatment of wrinkles.
3. Amber light (605nm): Intended for treatment of wrinkles.
4. Blue light and Red light (415nm + 630nm): Intended for treatment of mild to moderate inflammatory acne.
5. Blue light(415nm): Intended for treatment of mild to moderate inflammatory acne.
The LED Light Therapy Mask (Models: K9C-K, K9P-YK, K9P-K) is an Over-the-Counter (OTC) device intended for treatment of wrinkles and mild to moderate inflammatory acne.
1. Deep Red and Infrared light (660nm + 850nm): Intended for treatment of wrinkles.
2. Red light and Near-Infrared light (630nm + 830nm): Intended for treatment of wrinkles.
3. Blue light and Red light (415nm + 630nm): Intended for treatment of mild to moderate inflammatory acne.
4. Blue light(415nm): Intended for treatment of mild to moderate inflammatory acne.
The LED Light Therapy Mask (Model: FM-03BK) is an Over-the-Counter (OTC) device intended for treatment of wrinkles.
1. Red light (630nm): Intended for treatment of wrinkles.
2. Red light and Near-Infrared light (630nm + 830nm): Intended for treatment of wrinkles.
The LED Light Therapy Mask (Model: FM-09K) is an Over-the-Counter (OTC) device intended for treatment of wrinkles and mild to moderate inflammatory acne.
1. Red light (630nm): Intended for treatment of wrinkles.
2. Red light and Near-Infrared light (630nm + 830nm): Intended for treatment of wrinkles.
3. Blue light and Red light (415nm + 630nm): Intended for treatment of mild to moderate inflammatory acne.
4. Blue light(415nm): Intended for treatment of mild to moderate inflammatory acne.
Guangdong Newdermo Biotech Co.,Ltd
3 / 17
{8}
## 9. Predicate Device Comparison
**Table1 Comparison to Predicate/Reference Device for K9C-TK, K9C-K, K9P-YK, K9P-K, for Face**
| Item | Subject Device | Predicate Device1 | Reference Device1 | Reference Device2 | Reference Device3 | Comparison Result |
| --- | --- | --- | --- | --- | --- | --- |
| Trade name | LED Light Therapy Mask | Shani Darden LED light therapy mask | TheraFace Mask | Wrinkle Treatment Device (JM2) | LED Light Therapy Device (SG-FSM, SG-NK, SG-CT) | / |
| 510 (k) number | K260946 | K214103 | K230293 | K240360 | K253073 | / |
| Manufacturer | Guangdong Newdermo Biotech Co., Ltd. | Harpar Grace International | Therabody, Inc. | Shenzhen Qianyu Technology Co., Ltd. | Sunglor Technology Co., Ltd. | / |
| Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Regulation Name | Light Based Over The Counter Wrinkle Reduction, Over-The-Counter Powered Light Based Laser For Acne | Light Based Over The Counter Wrinkle Reduction, Over-The-Counter Powered Light Based Laser For Acne | Light Based Over The Counter Wrinkle Reduction, Over-The-Counter Powered Light Based Laser For Acne | Light Based Over The Counter Wrinkle Reduction | Light Based Over The Counter Wrinkle Reduction, Lamp, Infrared, Therapeutic Heating | Similar Note1 |
| Product code | OHS, OLP | OHS, OLP | OHS, OLP | OHS | OHS, ILY | Similar Note1 |
| Classification | II | II | II | II | II | Same |
Guangdong Newdermo Biotech Co.,Ltd
4 / 17
{9}
| Indications for use/ Intended use | See Indications for Use Section | The Shani Darden LED light therapy mask is an over-the-counter device intended to emit energy in the red and blue region of the light spectrum, specifically indicated to treat mild to moderate acne vulgaris of the face. The Shani Darden LED light therapy mask is an over-the-counter device intended to emit energy in the red and Near Infra-Red spectrum and is intended for the use in the treatment of full-face wrinkles. | Red Light is intended to treat full face wrinkles. Blue Light is intended to treat mild to moderate inflammatory acne. Red + Infrared Light is intended to treat full face wrinkles. | The wrinkle treatment device (model: JM2) is an Over-The-Counter (OTC) light based device intended for the use in treating full-face wrinkles. | The Led Light Therapy Device (Model:SG-FSM) is an Over-the-Counter medical device intended for treatment of wrinkles in the face and decolletage area. Red light: Treatment of wrinkles. Red+Infrared Light: Treatment of wrinkles. Amber light: Treatment of wrinkles. Amber+Infrared Light: Treatment of wrinkles. The Led Light Therapy Device (Model: SG-NK) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area. Red light: Treatment of wrinkles. Red+Infrared Light: Treatment of wrinkles. Amber light: Treatment of wrinkles. Amber+Infrared Light: Treatment of wrinkles. The Led Light Therapy Device (Model:SG-CT) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area. Red light: Treatment of wrinkles. Red+Infrared light: Treatment of wrinkles. Infrared light: Provide topical heating for the purpose of elevating tissue temperature, and to temporarily increase local blood circulation. | Similar Note2 |
| --- | --- | --- | --- | --- | --- | --- |
| Location for use | Face and Body | Face | Face | Face | Face, Body | Similar Note3 |
Guangdong Newdermo Biotech Co.,Ltd
5 / 17
{10}
| OTC or prescription | OTC | OTC | OTC | OTC | OTC | Same |
| --- | --- | --- | --- | --- | --- | --- |
| Power supply | Input:DC 5.0V , 1A; Li-ion Polymer Battery: DC 3.7V, 2450mAh, 9.065Wh | Electrical Input to power adaptor: 100V-240V, 50/60Hz. Rated at 0.7A. Electrical Output from power adaptor: 12V, 1A | 5-15V DC 2.5A max powered by 2 Li-Ion Batteries 3.7V 1500mAh) is charged via Universal USB charger cord or fast charger adaptor | Not display | Input: DC 5V, 1ABattery: Li-ion rechargeable battery 3.7V, 3000mAh | Similar Note4 |
| Power Source | Lithium-Ion Battery (rechargeable) | 100V-240V | Lithium-Ion Battery (rechargeable) | Lithium-Ion Battery (rechargeable) | Lithium-Ion Battery (rechargeable) | Similar Note4 |
| Light source | LED | LED | LED | LED | LED | same |
| Wavelengths | Red light: 630nm Deep Red light: 660nm; Blue light: 415nm; Amber Light: 605nm (Only K9C-TK); Near-Infrared light: 830nm; Infrared light: 850nm. | Blue: 415nm ± 10nm, Red: 630nm ± 10nm, NIR 830nm ± 10nm. | Red: 633 ± 10nm Blue: 415 ± 10nm Red + IR: 633 ± 10nm/830 ± 10nm | 660 ± 20nm 850 ± 20nm | SG-FSM: NIR: 850nm (face)and 830nm (upper chest)±10nm; Red: 630nm±10nm; Amber:605nm±10nm; SG-NK: NIR: 830nm±10nm; Red: 630nm±10nm; Amber:605nm±10nm; SG-CT: NIR: 830nm±10nm; Red: 630nm±10nm; | Similar Note5 |
Guangdong Newdermo Biotech Co.,Ltd
6 / 17
{11}
| Power Density | Deep light+Infrared light: 40mw/cm^{2} Blue light: 19mw/cm^{2} Amber light: 15mw/cm^{2} (Only K9C-TK) Red light+Near-Infrared light: 29mw/cm^{2} Red light+Blue light: 44mw/cm^{2} | Blue 28mW/cm^{2} Red 16mW/cm^{2} Red And Blue Total: 44mW/cm^{2} Red 18mW/cm^{2} NIR 11mW/cm^{2} Red And NIR Total: 29mW/cm^{2} | Red: 73 ± 5mW/cm^{2} Blue: 64 ± 5mW/cm^{2} Red + IR: 73 ± 5mW/cm^{2}/55 ± 5mW/cm^{2} | 40/70/100mW/cm^{2} | Max Irradiance:850/830nm:3 ± 20%mW/cm^{2}; 630nm:22 ± 20%mW/cm^{2}; 605nm:14 ± 20%mW/cm^{2}; | Similar Note6 |
| --- | --- | --- | --- | --- | --- | --- |
Guangdong Newdermo Biotech Co.,Ltd
7 / 17
{12}
| Treatment time | Manual Mode: 10minutes Automatic Mode: 15minutes | 10 Minutes | LED: 3 minutes each light mode for a total of 9 minutes per treatment, recommended to use 2 to 5 times per week. Vibration: accompanies LED treatments or can be used without LED's active. 3 vibration patterns, 5 minutes each, for a total of 15 minutes. During blue light treatment mode, vibration is not active around the eyes. Vibration is included for a more relaxing experience. | Not displayed | 10/15/20/25/30minutes | Similar Note6 |
| --- | --- | --- | --- | --- | --- | --- |
| Safety and EMC | IEC 60601-1; IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 IEC 62471 | IEC 60601-1; IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-57 IEC 62471 | IEC 60601-1; IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-57 IEC 62471 | EC 60601-1; IEC 60601-1-2 IEC 60601-1-11 IEC 60825-1 | IEC 60601-1; IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 IEC 62471 | Similar Note7 |
#### Note1
The product code/regulation name of the predicate/reference devices includes the product code/regulation name of the subject device. This slight difference will not raise safety and effective issue.
Guangdong Newdermo Biotech Co.,Ltd
8 / 17
{13}
Note2
The indications for use/intended use of the predicate/reference devices includes the indications for use/intended use of the subject device. This slight difference will not raise safety and effective issue.
Note3
The subject device location for use is face(model: K9C-TK\ K9P-YK\ K9C-K\ K9P-K) and body(model: FM-03BK\ FM-09K). The predicate device and the reference devices are include the location for use. This slight difference will not raise safety and effective issue.
Note4
The power supply of the devices are very similar but not identical, the IEC60601-1test demonstrated the safety of the power source. The subject device shows more information about the battery, with other content remaining the same. The slight difference will not raise safety and effective issue.
Note5
The predictive device and the reference device have fully covered the wavelengths of the subject device. This slight difference will not raise safety and effective issue.(see Note6 for details)
Note6
1. The Red light + Near-Infrared light mode and the Red light + Blue light mode of the subject device as same as the predicate device1;
2. The Blue light mode of the subject device's total power density(19×60×10÷1000=11.4J/cm²) as same as the reference device1(64×60×3÷1000=11.52J/cm²), so this slight deference will not raise safety and effective issue;
3. The Deep Red light + Infrared light mode of the subject devices same as the reference device2;
4. The Amber light mode(wavelength 605nm, for face)'s total power density(15×60×10÷1000= 9J/cm² ) is between the reference device4(min:14×60×10÷1000= 8.4J/cm² ; max:14×60×30÷1000= 25.2J/cm² ). So, it will not raise safety and effective issue.
Note7
The subject device complies with more laws, making it safer compared to the predicate device.
The definition OF HOME LIGHT THERAPY EQUIPMENT in IEC 60601-2-83 standard is broader than the corresponding definition of light-source equipment in IEC 60601-2-57 and includes both eye-mediated and skin-mediated photo biological effects, which can be visual and non-visual as explained in the introduction. The definition is restricted to equipment intended to be used.
Guangdong Newdermo Biotech Co.,Ltd
9 / 17
{14}
**Table2 Comparison to Predicate/Reference Device for FM-03BK, FM-09K, for Body**
| Item | Subject Device | Predicate Device2 | Reference Device3 | Reference Device4 | Comparison Result |
| --- | --- | --- | --- | --- | --- |
| Trade name | LED light therapy mask | TRUDERMAL Pro (ZLD-390) | LED Light Therapy Device (SG-FSM, SG-NK, SG-CT) | décoLITE LED Device | / |
| 510 (k) number | K260946 | K251012 | K253073 | K242382 | / |
| Manufacturer | Guangdong Newdermo Biotech Co., Ltd. | Shenzhen Kaiyan Medical Equipment Co., Ltd. | Sunglor Technology Co., Ltd. | Ismart Developments, Ltd. | / |
| Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Regulation Name | Light Based Over The Counter Wrinkle Reduction, Over-The-Counter Powered Light Based Laser For Acne | Light Based Over The Counter Wrinkle Reduction, Over-The-Counter Powered Light Based Laser For Acne Lamp, Infrared, Therapeutic Heating | Light Based Over The Counter Wrinkle Reduction, Lamp, Infrared, Therapeutic Heating | Light Based Over The Counter Wrinkle Reduction, | Similar Note8 |
| Product code | OHS, OLP | OHS, OLP, ILY | OHS, ILY | OHS | Similar Note8 |
| Classification | II | II | II | II | Same |
Guangdong Newdermo Biotech Co.,Ltd
10 / 17
{15}
| Indications for use/ Intended use | See Indications for Use Section | TRUDERMAL Pro (Model: ZLD-390) is an over-the-counter (OTC) device that delivering targeted photo therapy using three wavelengths (blue:415nm, red:630nm, Infrared (IR):850nm): Red light: Reduces full-face wrinkles. Blue light: Treats mild to moderate inflammatory acne. IR light: Provides topical heating for the purpose of elevating tissue temperature, temporarily relieving minor muscle and joint pain, arthritis and muscle spasms; relieving stiffness; promoting relaxation of muscle tissues and temporarily increasing local blood circulation. Mixed lights: Red+IR light: Reduces full-face wrinkles. Blue+Red light:Treats mild to moderate inflammatory acne. | The Led Light Therapy Device (Model:SG-FSM) is an Over-the-Counter medical device intended for treatment of wrinkles in the face and decolletage area. Red light: Treatment of wrinkles. Red+Infrared Light: Treatment of wrinkles. Amber light: Treatment of wrinkles. Amber+Infrared Light: Treatment of wrinkles. The Led Light Therapy Device (Model: SG-NK) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area. Red light: Treatment of wrinkles. Red+Infrared Light: Treatment of wrinkles. Amber light: Treatment of wrinkles. Amber+Infrared Light: Treatment of wrinkles. The Led Light Therapy Device (Model:SG-CT) is an Over-the-Counter medical device intended for treatment of wrinkles in the decolletage area. Red light: Treatment of wrinkles. Red+Infrared light: Treatment of wrinkles. Infrared light: Provide topical heating for the purpose of elevating tissue temperature, and to temporarily increase local blood circulation. | The décoLITE LED device is an over-the-counter device that is intended for the use in the treatment of wrinkles in the decolletage area. | Similar Note9 |
| --- | --- | --- | --- | --- | --- |
| Location for use | Face and Body | Face, Body | Face, Body | Body | Similar Note10 |
| OTC or prescription | OTC | OTC | OTC | OTC | Same |
Guangdong Newdermo Biotech Co.,Ltd
11 / 17
{16}
| Power supply | Input:DC 5.0V, 1A; Li-ion Polymer Battery: DC 3.7V, 2450mAh, 9.065Wh | Adaptor Input: 100-240Vac, 50-60Hz, 0.8A Output: 12V ~ 3A | Input: DC 5V, 1ABattery: Li-ion rechargeable battery 3.7V, 3000mAh | not shown | Similar Note4 |
| --- | --- | --- | --- | --- | --- |
| Power Source | Lithium-Ion Battery (rechargeable) | 100-240Vac | Lithium-Ion Battery (rechargeable) | Lithium-Ion Battery (rechargeable) | Similar Note11 |
| Light source | LED | LED | LED | LED | same |
| Wavelengths | Red light: 630nm Deep Red light:660nm Blue light: 415nm, Near-Infrared light:830nm, Infrared light (wavelength 850nm). | Blue: 415±10nm Red: 630±10nm IR: 850±10nm Mixed light: Blue+Red:415nm+630nm Red+IR:630nm+850nm | SG-FSM: NIR: 850nm (face)and 830nm (upper chest)±10nm; Red: 630nm±10nm; Amber:605nm±10nm; SG-NK: NIR: 830nm±10nm; Red: 630nm±10nm; Amber:605nm±10nm; SG-CT: NIR: 830nm±10nm; Red: 630nm±10nm; | Red: 630nm±10nm NIR: 830nm± 10nm | Similar Note12 |
Guangdong Newdermo Biotech Co.,Ltd
12 / 17
{17}
| Power Density | 1. Deep Red light + Infrared light(wavelengths: 660nm + 850nm):40mw/cm^{2}(Only K9C-TK \ K9P-YK \ K9C-K \ K9P-K) 2.Blue light(415nm): 19mw/cm^{2}(Only K9C-TK \ K9P-YK \ K9C-K \ K9P-K) or 15mw/cm^{2}(Only FM-09K) 3.Amber Light (wavelength 605nm): 15mw/cm^{2}(Only K9C-TK) 4.Red light + Near-Infrared light(wavelength 630nm + 830nm): 29mw/cm^{2} 5.Red light + Blue light(wavelength 630nm + 415nm): 44mw/cm^{2}(Only K9C-TK \ K9P-YK \ K9C-K \ K9P-K) or 36mw/cm^{2}(Only FM-09K) 6.Red light(wavelength 630nm): 21mw/cm^{2}(Only FM-03BK \ FM-09K) | Blue:6-10mW/cm^{2} Red:10-15mW/cm^{2} IR:6-10mW/cm^{2} Mixed light: Blue+Red: 16-20mW/cm^{2} Red+IR:16-20mW/cm^{2} | Max Irradiance:850/830nm:3 ± 20%mW/cm^{2}; 630nm:22 ± 20%mW/cm^{2}; 605nm:14 ± 20%mW/cm^{2}; | 30mW/cm^{2} total | Similar Note13 |
| --- | --- | --- | --- | --- | --- |
| Treatment time | Manual Mode: 10minutes Automatic Mode: 15minutes | 20 min per treatment Acne and Wrinkles: 3-4x weekly; 4 weeks. Pain relief: 3x weekly | 10/15/20/25/30minutes | 600 seconds (10 minutes) | Similar Note13 |
| Safety and EMC | IEC 60601-1; IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 IEC 62471 | IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-57 IEC 62471 | IEC 60601-1; IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 IEC 62471 | IEC 60601-1; IEC 60601-1-2 IEC 60601-1-11 IEC 60601-2-83 | Similar Note14 |
Guangdong Newdermo Biotech Co.,Ltd
13 / 17
{18}
Note8
The product code/regulation name/regulation number of the predicate/reference devices includes the product code/regulation name of the subject device. This slight difference will not raise safety and effective issue.
Note9
The indications for use/intended use of the predicate/reference devices includes the indications for use/intended use of the subject device. t. This slight difference will not raise safety and effective issue.
Note10
The subject device location for use is face(model: K9C-TK\ K9P-YK\ K9C-K\ K9P-K) and body(model: FM-03BK\ FM-09K). The predicate device and the reference devices are include the location for use. This slight difference will not raise safety and effective issue.
Note11
The power supply of the devices are very similar but not identical, the IEC60601-1test demonstrated the safety of the power source. The subject device shows more information about the battery, with other content remaining the same. The slight difference will not raise safety and effective issue.
Note12
The predictive device and the reference device have fully covered the wavelengths of the subject device. This slight difference will not raise safety and effective issue.(see Note13 for details)
Note13
1.1 The Blue light mode of the subject device's total power density(15 × 60 × 10 ÷1000=9J/cm²) is between the predicate device2(min: 6 × 60 × 20 ÷1000= 7.2 J/cm²;max:10 × 60 × 20 ÷1000= 12 J/cm²), so this slight deference will not raise safety and effective issue;
1.2 The Red light + Blue light mode of the subject device's total power density (36 × 60 × 10 ÷1000= 21.6 J/cm²) is between the predicate device2(min: 16 × 60 × 20 ÷1000= 19.2 J/cm²;max:20 × 60 × 20 ÷1000= 24 J/cm²); so this slight deference will not raise safety and effective issue;
2. The Red light mode of the subject device as same as the reference device3.
3. The Red light + Near-Infrared light mode of the subject device as same as the reference device4.
Note14
The subject device complies with more laws, making it safer compared to the predicate device.
Guangdong Newdermo Biotech Co.,Ltd
14 / 17
{19}
The definition Of HOME LIGHT THERAPY EQUIPMENT in IEC 60601-2-83 standard is broader than the corresponding definition of light-source equipment in IEC 60601-2-57 and includes both eye-mediated and skin-mediated photo biological effects, which can be visual and non-visual as explained in the introduction. The definition is restricted to equipment intended to be used.
Guangdong Newdermo Biotech Co.,Ltd
15 / 17
{20}
# **10. Performance Data:**
# **Non-clinical data:**
Non-clinical tests have been conducted to verify that the Beauty LED Mask meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate device. The testing results demonstrate that the subject device complies with the following standards:
- ANSI AAMI ES 60601-1:2005/AMD1:2012/AMD2:2020, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014+A1:2020, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic disturbances – Requirements and tests
- IEC/TR 60601-4-2:2024 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems
- IEC 60601-1-11:2015/AMD1:2020, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- IEC 60601-2-83:2019 Medical electrical equipment - Part 2-83: Particular requirements for the basicsafety and essential performance of home light therapy equipment
- IEC 62133-2:2017/AMD1:2021 Secondary cells and batteries containing alkaline or other non-acid electrolytes - Safety requirements for portable sealed secondary cells and for batteries made from them, for use in portable applications -Part 2: Lithium systems
- IEC 62471:2016 Photobiological safety of lamps and lamp systems
# **Biocompatibility Test**
The device has been tested for biocompatibility, it complies with the following standards.
- ISO 10993-5:2009, Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity
- ISO 10993-10:2021, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization
- ISO 10993-23:2021, Biological Evaluation of Medical Devices - Part 23: Tests for irritation
# **Software Verification and Validation Testing**
Software verification and validation testing were conducted and documentation was
Guangdong Newdermo Biotech Co.,Ltd
16 / 17
{21}
provided as recommended by FDA’s 2023 Guidance for Industry and FDA Staff, “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.”
**Clinical data:** Not applicable.
### **Summary**
Based on the non-clinical and clinical performance as documented in the system development, the subject devices were found to have a safety and effectiveness profile that is similar to the predicate device.
### **11. Conclusion**
Based on comparing to predicate devices and reference devices, the proposed device of K9C-TK, K9C-K, K9P-YK, K9P-K, FM-03BK and FM-09K are determined to be Substantially Equivalent (SE) to the predicate devices, in respect of safety and effectiveness.
Guangdong Newdermo Biotech Co.,Ltd
17 / 17
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.