The CarboClear® Posterior Cervical Screw System / CarboClear® Hybrid Posterior Cervical Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The CarboClear® Posterior Cervical Screw System / CarboClear® Hybrid Posterior Cervical Screw System is intended to provide immobilization and stabilization of spinal segments with anterior interbody support implanted at the same spinal level/s as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine and the upper thoracic spine (T1 to T3): - Traumatic spinal fractures and/or dislocations - Instability and deformity - Failed previous fusions (e.g., pseudarthrosis) - Degenerative disease, including intractable radiculopathy and/or myelopathy - Neck and/or arm pain of discogenic origin as confirmed by radiographic studies - Degenerative disease of the facets with instability - Patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression The CarboClear® Posterior Cervical Screw System / CarboClear® Hybrid Posterior Cervical Screw System is intended to be used with anterior interbody support implanted at the same spinal level/s with autogenous and /or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
Device Story
System comprises spinal implants (screws, rods) and surgical instruments; used to build spinal constructs for immobilization/stabilization. Screws feature carbon fiber-reinforced PEEK (CFR-PEEK) bodies; some models include titanium alloy tulips/shells and tantalum markers. Rods available in CFR-PEEK or titanium alloy. Used in clinical/surgical settings by surgeons to treat spinal instability or tumors. Implants provide mechanical support to spinal segments; intended to facilitate fusion or provide temporary structural integrity in tumor cases. Benefits include spinal stabilization and restoration of column integrity. Supplied sterile for single use.
Clinical Evidence
Bench testing only. Performance characteristics evaluated via static and dynamic compression bending tests (ASTM F1717), flexion-extension, axial and torsional gripping capacity (ASTM F1798), and wear evaluation. Results demonstrate performance comparable to predicate devices.
Indicated for patients with advanced stage cervical tumors (limited life expectancy); acute/chronic cervical and upper thoracic (T1-T3) instabilities including traumatic fractures/dislocations, deformity, failed fusions, degenerative disease (radiculopathy/myelopathy/discogenic pain/facet instability), and spinal infection. Requires use with anterior interbody support and bone graft.
Regulatory Classification
Identification
Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
Special Controls
*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.
CarboClear® Hybrid Pedicle Screw System (K240846, K243106)
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
March 16, 2026
CarboFix Orthopedics, Ltd.
Yael Rubin
Director of Regulatory Affairs
11 Ha'Hoshlim St.
Herzeliya, 4672411
Israel
Re: K254045
Trade/Device Name: CarboClear® Posterior Cervical Screw System; CarboClear® Hybrid Posterior Cervical Screw System
Regulation Number: 21 CFR 888.3075
Regulation Name: Posterior Cervical Screw System
Regulatory Class: Class II
Product Code: NKG
Dated: December 16, 2025
Received: December 17, 2025
Dear Yael Rubin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254045 - Yael Rubin
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality Management System Regulation (QMSR) (21 CFR Part 820), which includes, but is not limited to, ISO 13485 clause 7.3 (Design controls), ISO 13485 clause 8.3 (Nonconforming product), ISO 13485 clause 8.5.2 (Corrective action), and ISO 13485 clause 8.5.3 (Preventative action). Please note that regardless of whether a change requires premarket review, the QMSR requires device manufacturers to review and approve changes to device design and production (ISO 13485 clause 7.3 and ISO 13485 clause 7.5) and document changes and approvals in the Medical Device File (ISO 13485 clause 4.2.3).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the Quality Management System Regulation (QMSR) (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254045 - Yael Rubin
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
COLIN O'NEILL -S
Colin O'Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K254045 | ? |
| Please provide the device trade name(s). | | ? |
| CarboClear® Posterior Cervical Screw System ;
CarboClear® Hybrid Posterior Cervical Screw System | | |
| Please provide your Indications for Use below. | | ? |
| The CarboClear® Posterior Cervical Screw System / CarboClear® Hybrid Posterior Cervical Screw System
is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time
period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of
insufficient duration to permit achievement of fusion.
The CarboClear® Posterior Cervical Screw System / CarboClear® Hybrid Posterior Cervical Screw System
is intended to provide immobilization and stabilization of spinal segments with anterior interbody support
implanted at the same spinal level/s as an adjunct to fusion for the following acute and chronic instabilities
of the cervical spine and the upper thoracic spine (T1 to T3):
- Traumatic spinal fractures and/or dislocations
- Instability and deformity
- Failed previous fusions (e.g., pseudarthrosis)
- Degenerative disease, including intractable radiculopathy and/or myelopathy
- Neck and/or arm pain of discogenic origin as confirmed by radiographic studies
- Degenerative disease of the facets with instability
- Patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection,
surgical debridement, or decompression
The CarboClear® Posterior Cervical Screw System / CarboClear® Hybrid Posterior Cervical Screw System
is intended to be used with anterior interbody support implanted at the same spinal level/s with autogenous
and /or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
KarboFix Orthopedics Ltd.
CarboClear® Posterior Cervical Screw System
CarboClear® Hybrid Posterior Cervical Screw System
K254045 Page 1 of 4
# 510(k) Summary
CarboFix Orthopedics Ltd.
CarboClear® Posterior Cervical Screw System
CarboClear® Hybrid Posterior Cervical Screw System
## Applicant Name
CarboFix Orthopedics Ltd.
11 Ha’hoshlim St., Herzeliya 4672411, Israel
Tel: +972 9 9511511, Fax: +972 9 9548939
## Contact Person
Yael Rubin
CarboFix Orthopedics Ltd.
11 Ha’hoshlim St., Herzeliya 4672411, Israel
Tel: +972 9 9511511, Fax: +972 9 9548939
## Date Prepared
December 2025
## Trade/Proprietary Name
CarboClear® Posterior Cervical Screw System
CarboClear® Hybrid Posterior Cervical Screw System
## Common Name
Posterior, Cervical Spine Fixation
## Regulation Number and Device Class
Class II; 21 CFR §888.3075
## Product Code, Regulatory Description and Review Panel
NKG, Posterior Cervical Screw System, Orthopedic
{5}
KarboFix Orthopedics Ltd.
CarboClear® Posterior Cervical Screw System
CarboClear® Hybrid Posterior Cervical Screw System
K254045
Page 2 of 4
# Predicate Devices
## Primary Predicate Device:
- CMORE® CT System (icotec; K252327)
## Additional Predicate Devices:
- CarboClear® Posterior Cervical Screw System (CarboFix Orthopedics Ltd.; K233989)
- INFINITY™ OCT System (Medtronic; K163375)
## Reference Device:
- CarboClear® Hybrid Pedicle Screw System (CarboFix Orthopedics Ltd.; K240846, K243106)
# Indications for Use
The CarboClear® Posterior Cervical Screw System / CarboClear® Hybrid Posterior Cervical Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
The CarboClear® Posterior Cervical Screw System / CarboClear® Hybrid Posterior Cervical Screw System is intended to provide immobilization and stabilization of spinal segments with anterior interbody support implanted at the same spinal level/s as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine and the upper thoracic spine (T1 to T3):
- Traumatic spinal fractures and/or dislocations
- Instability and deformity
- Failed previous fusions (e.g., pseudarthrosis)
- Degenerative disease, including intractable radiculopathy and/or myelopathy
- Neck and/or arm pain of discogenic origin as confirmed by radiographic studies
- Degenerative disease of the facets with instability
- Patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression
The CarboClear® Posterior Cervical Screw System / CarboClear® Hybrid Posterior Cervical Screw System is intended to be used with anterior interbody support implanted at the same
{6}
CarboFix Orthopedics Ltd.
CarboClear® Posterior Cervical Screw System
CarboClear® Hybrid Posterior Cervical Screw System
K254045 Page 3 of 4
spinal level/s with autogenous and /or allogenic bone graft comprised of cancellous and/ or corticocancellous bone graft.
## System Description
CarboClear® Posterior Cervical Screw System (“CarboClear® System”) and CarboClear® Hybrid Posterior Cervical Screw System (“CarboClear® Hybrid System”) are composed of implants in various dimensions, used to build a spinal construct; and of a set of instruments, intended to assist in the insertion and placement of the implants.
The screws incorporated into the systems are made of carbon fiber-reinforced polyetheretherketone (CFR-PEEK), with the CarboClear® Hybrid screws incorporating titanium alloy tulip. The screws threaded portion, and spherical head are encased within a thin titanium shell, and they may include a tantalum marker. The CarboClear® posterior cervical screws are used with CFR-PEEK set screw, and the CarboClear® Hybrid posterior cervical screws are used with a titanium alloy set screw. All screws are used with CarboClear® CFR-PEEK rods. CarboClear® titanium alloy rods are also offered.
The implants are supplied sterile, and are intended for single use.
## Purpose of Submission
Modification to the previously cleared CarboClear® Posterior Cervical Screw System (K233989) – modification to indications for use and addition of implants, and addition of CarboClear® Hybrid Posterior Cervical Screw System.
## Performance Data
Performance characteristics included static and dynamic compression bending tests according to ASTM F1717, ASTM F1798 (flexion-extension and axial and torsional gripping capacity), and wear evaluation. The results of the tests are comparable to those of the predicate devices, demonstrating substantially equivalent performance of the subject and predicate devices.
## Substantial Equivalence
The intended use, design, dimensions, materials, technological characteristics, principles of operation, and performance of the CarboClear®- and CarboClear® Hybrid Posterior Cervical Screw Systems are substantially equivalent to those of the predicate devices, as applicable.
{7}
K254045
Page 4 of 4
CarboFix Orthopedics Ltd.
CarboClear® Posterior Cervical Screw System
CarboClear® Hybrid Posterior Cervical Screw System
## Conclusion
Based on the information provided in this Premarket Notification, the CarboClear®- and CarboClear® Hybrid Posterior Cervical Screw Systems are substantially equivalent to their predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.