CarboClear® Hybrid Pedicle Screw System; CarboClear® Hybrid Navigated Instruments; CarboClear® Hybrid Fenestrated Pedicle Screw System and High V+® Bone Cement; CarboClear® X Fenestrated Pedicle Screw System and High V+® Bone Cement
CarboClear® Hybrid Pedicle Screw System; CarboClear® Hybrid Navigated Instruments; CarboClear® Hybrid Fenestrated Pedicle Screw System and High V+® Bone Cement; CarboClear® X Fenestrated Pedicle Screw System and High V+® Bone Cement
Applicant
CarboFix Orthopedics , Ltd.
Product Code
NKB · Orthopedic
Decision Date
Apr 25, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3070
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CarboClear® Hybrid Pedicle Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The CarboClear® Hybrid Pedicle Screw System is intended to provide rigid immobilization and stabilization of lumbar and/or sacral segments as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level from L2 to S1, with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature and have at least six months of non-operative treatment. The CarboClear® Hybrid Pedicle Screw System is intended to be used with an intervertebral body fusion device implanted at the same spinal level with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
Device Story
System comprises pedicle screws, rods, and navigated instruments for spinal surgery. Screws feature carbon fiber-reinforced PEEK (CFR-PEEK) bodies with titanium alloy tulips and shells; fenestrated versions allow PMMA bone cement delivery for vertebral augmentation in oncological cases. Instruments are manually operated and compatible with Medtronic StealthStation for image-guided navigation. Surgeons use the system to stabilize spinal segments; implants provide mechanical support. Navigation assists in precise screw placement relative to patient anatomy (CT/MR/fluoroscopy). Benefits include spinal stabilization and integrity restoration in tumor patients or fusion support in DDD patients. System is used in clinical settings by surgeons.
Clinical Evidence
Bench testing only. Performance evaluated via static and dynamic compression bending tests (ASTM F1717), static flexion-extension, axial gripping capacity, and torsional gripping capacity (ASTM F1798). Results demonstrate performance comparable to predicate devices.
Technological Characteristics
Implants: CFR-PEEK screw body, titanium alloy tulip/shell, tantalum markers. Bone cement: Radiopaque PMMA (polymethylmethacrylate) with barium sulfate/hydroxyapatite. Instruments: Manually operated, compatible with Medtronic StealthStation. Sterilization: Supplied sterile. Connectivity: None (mechanical/manual).
Indications for Use
Indicated for patients with advanced stage thoracic/lumbar tumors requiring spinal column integrity restoration without fusion; or skeletally mature patients with degenerative disc disease (DDD) at one level (L2-S1) with up to Grade I spondylolisthesis requiring rigid immobilization/stabilization as an adjunct to fusion. Fenestrated screws with High V+ Bone Cement indicated for tumor patients where spinal integrity is not severely compromised.
Regulatory Classification
Identification
(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
Special Controls
*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.
{0}------------------------------------------------
April 25, 2024
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
CarboFix Orthopedics Ltd. Hila Wachsler-Avrahami Regulatory Affairs 11 Ha'Hoshlim Street Herzeliya, 4672411 Israel
### Re: K240846
Trade/Device Name: CarboClear® Hybrid Pedicle Screw System; CarboClear® Hybrid Navieated Instruments; CarboClear® Hybrid Fenestrated Pedicle Screw System and High V+® Bone Cement: CarboClear® X Fenestrated Pedicle Screw System and High V+® Bone Cement Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral Pedicle Screw System Regulatory Class: Class II Product Code: NKB, PML, OLO Dated: March 27, 2024 Received: March 27, 2024
Dear Hila Wachsler-Avrahami:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
### K240846 - Hila Wachsler-Avrahami
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (OS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuingeducation/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumereducation-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Eileen
Digitally signed
by Eileen Cadel -
S
Cadel -S Date: 2024.04.25
13:25:22 -04'00'
for
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K240846
Device Name
CarboClear® Hybrid Pedicle Screw System;
CarboClear® Hybrid Navigated Instruments;
CarboClear® Hybrid Fenestrated Pedicle Screw System and High V+® Bone Cement;
CarboClear® X Fenestrated Pedicle Screw System and High V+® Bone Cement
Indications for Use (Describe)
CarboClear® Hybrid Pedicle Screw System (Oncological)
The CarboClear® Hybrid Pedicle Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
CarboClear® Hybrid Pedicle Screw System (DDD)
The CarboClear® Hybrid Pedicle Screw System is intended to provide rigid immobilization and stabilization of lumbar and/or sacral segments as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level from L2 to S1, with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature and have at least six months of non-operative treatment. The CarboClear® Hybrid Pedicle Screw System is intended to be used with an intervertebral body fusion device implanted at the same spinal level with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
# CarboClear® Hybrid Navigated Instruments
CarboClear® Hybrid Navigated Instruments are intended to be used during the preparation and placement of CarboClear® Hybrid Pedicle Screws during spinal surgery to assist the surgeon in precisely locating anatomical structures. CarboClear® Hybrid Navigated Instruments are designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
# CarboClear® Hybrid Fenestrated Pedicle Screw System
When used in conjunction with High V+® Bone Cement, the CarboClear® Hybrid Fenestrated Pedicle Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The CarboClear® Hybrid Fenestrated Pedicle Screw System augmented with High V+® Bone Cement is for use at spinal levels where the structural integrity of the spine is not severely compromised.
When used without cement, the CarboClear® Hybrid Fenestrated Pedicle Screws are intended: (a) to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
(b) to provide rigid immobilization and stabilization of lumbar and/or sacral segments as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level from L2 to S1, with up to
{3}------------------------------------------------
Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature and have at least six months of non-operative treatment. The CarboClear® Hybrid Pedicle Screw System is intended to be used with an intervertebral body fusion device implanted at the same spinal level with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
# High V+® Bone Cement
When used in conjunction with the CarboClear® Hybrid Fenestrated Pedicle Screw System, High V +® Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. High V+® Bone Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
# CarboClear® X Fenestrated Pedicle Screw System
When used in conjunction with High V+® Bone Cement, the CarboClear® X Fenestrated Pedicle Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The CarboClear® X Fenestrated Pedicle Screw System augmented with High V+® Bone Cement is for use at spinal levels where the structural integrity of the spine is not severely compromised.
When used without cement, the CarboClear® X Fenestrated Pedicle Screw System is intended: (a) to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
(b) to provide rigid immobilization and stabilization of lumbar and/or sacral segments as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level from L2 to S1, with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature and have at least six months of non-operative treatment. The CarboClear® X Pedicle Screw System is intended to be used with an intervertebral body fusion device implanted at the same spinal level with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
## High V+® Bone Cement
When used in conjunction with the CarboClear® X Fenestrated Pedicle Screw System, High V+® Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. High V+® Bone Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{4}------------------------------------------------
# 510(k) Summary
## CarboFix Orthopedics Ltd.
# CarboClear® Hybrid Pedicle Screw System, CarboClear® Hybrid Navigated Instruments, CarboClear® Hybrid Fenestrated Pedicle Screw System and High V+® Bone Cement, CarboClear® X Fenestrated Pedicle Screw System and High V+® Bone Cement
#### APPLICANT NAME
CarboFix Orthopedics Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel Tel: +972 9 9511511, Fax: +972 9 9548939
### CONTACT PERSON
Hila Wachsler-Avrahami CarboFix Orthopedics Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel Tel: +972 9 9511511, Fax: +972 9 9548939
### DATE PREPARED
March 27, 2024
### TRADE/PROPRIETARY NAME
- CarboClear® Hybrid Pedicle Screw System 1.
- 2. CarboClear® Hybrid Navigated Instruments
- CarboClear® Hybrid Fenestrated Pedicle Screw System and High V+ Bone Cement 3.
- 4. CarboClear® X Fenestrated Pedicle Screw System and High V+ Bone Cement
### COMMON NAME
- 1. Thoracolumbosacral Pedicle Screw System
- 2. Stereotaxic Instrument
- Polymethylmethacrylate (PMMA) Bone Cement 3.
#### REGULATION NUMBER AND DEVICE CLASS
- 1. Class II; 21 CFR §888.3070
- 2. Class II; 21 CFR §882.4560
- Class II; 21 CFR §888.3027 3.
{5}------------------------------------------------
## PRODUCT CODE, REGULATORY DESCRIPTION AND REVIEW PANEL
- 1. NKB: Thoracolumbosacral Pedicle Screw System: Orthopedic
- 2. OLO; Orthopedic Stereotaxic Instrument; Orthopedic
- PML; Bone Cement, Posterior Screw Augmentation 3.
# PREDICATE DEVICES
# Primary Predicate Device:
CarboClear® X Pedicle Screw System (CarboFix Orthopedics Ltd.; K210716, K231280, K233944).
# Additional Predicate Devices:
- 1. CarboClear® X Navigated Instruments (CarboFix Orthopedics Ltd.; K210716, K231280, K233944)
- 2. CarboClear® X Fenestrated Pedicle Screw System, High V+® Bone Cement (CarboFix Orthopedics Ltd.; K231280)
- 3. VADER®one Pedicle System MIS and LightMore® Pedicle System 6.0 (Icotec ag; K190545, K193423)
- 4. VADER® Pedicle System, G21 Cement (Icotec ag; K200596, K222789)
- 5. CD HORIZON® Spinal System (Medtronic Sofamor Danek; K082236 and more)
- 6. CD HORIZON™ Fenestrated Screw Set (Medtronic Sofamor Danek; K170347)
- 7. Medtronic Navigated Manual Reusable Instruments (Medtronic Sofamor Danek; K153442 and more)
## INDICATIONS FOR USE
# CarboClear® Hybrid Pedicle Screw System
# CarboClear® Hybrid Pedicle Screw System (Oncology)
The CarboClear® Hybrid Pedicle Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
## CarboClear® Hybrid Pedicle Screw System (DDD)
The CarboClear® Hybrid Pedicle Screw System is intended to provide rigid immobilization and stabilization of lumbar and/or sacral segments as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level from L2 to S1, with up to Grade I
{6}------------------------------------------------
spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature and have at least six months of non-operative treatment.
The CarboClear® Hybrid Pedicle Screw System is intended to be used with an intervertebral body fusion device implanted at the same spinal level with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
#### CarboClear® Hybrid Navigated Instruments
CarboClear® Hybrid Navigated Instruments are intended to be used during the preparation and placement of CarboClear® Hybrid Pedicle Screws during spinal surgery to assist the surgeon in precisely locating anatomical structures. CarboClear® Hybrid Navigated Instruments are designed for use with the Medtronic StealthStation® System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as vertebra, can be identified relative to a CT or MR based model, fluoroscopy images, or digitized landmarks of the anatomy.
#### CarboClear® Hybrid Fenestrated Pedicle Screws System
When used in conjunction with High V+® Bone Cement, the CarboClear® Hybrid Fenestrated Pedicle Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The CarboClear® Hybrid Fenestrated Pedicle Screw System augmented with High V+® Bone Cement is for use at spinal levels where the structural integrity of the spine is not severely compromised.
When used without cement, the CarboClear® Hybrid Fenestrated Pedicle Screws are intended:
- (a) to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
- (b) to provide rigid immobilization and stabilization of lumbar and/or sacral segments as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level from L2 to S1, with up to Grade I spondylolisthesis. DDD is defined as back pain of
{7}------------------------------------------------
discogenic origin with degeneration of the disc confirmed by history and radiographic studies. Patients should be skeletally mature and have at least six months of non-operative treatment.
The CarboClear® Hybrid Pedicle Screw System is intended to be used with an intervertebral body fusion device implanted at the same spinal level with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
#### High V+® Bone Cement
When used in conjunction with the CarboClear® Hybrid Fenestrated Pedicle Screw System, High V+® Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. High V+® Bone Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
#### CarboClear® X Pedicle Screw System
#### CarboClear® X Fenestrated Pedicle Screw System
When used in conjunction with High V+® Bone Cement, the CarboClear® X Fenestrated Pedicle Screw System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. The CarboClear® X Fenestrated Pedicle Screw System augmented with High V+® Bone Cement is for use at spinal levels where the structural integrity of the spine is not severely compromised.
When used without cement, the CarboClear® X Fenestrated Pedicle Screws are intended:
- (a) to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
- (b) to provide rigid immobilization and stabilization of lumbar and/or sacral segments as an adjunct to fusion in patients with degenerative disc disease (DDD) at one level from L2 to S1, with up to Grade I spondylolisthesis. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and
{8}------------------------------------------------
radiographic studies. Patients should be skeletally mature and have at least six months of non-operative treatment.
The CarboClear® X Pedicle Screw System is intended to be used with an intervertebral body fusion device implanted at the same spinal level with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
#### High V+® Bone Cement
When used in conjunction with the CarboClear® X Fenestrated Pedicle Screw System, High V+® Bone Cement is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracic and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion. High V+® Bone Cement is limited to use at spinal levels where the structural integrity of the spine is not severely compromised.
#### SYSTEM DESCRIPTION
The CarboClear® Hybrid Pedicle Screw System is composed of implants in various dimensions, used to build a spinal construct; and of a set of instruments, intended to assist in the insertion and placement of the implants.
The CarboClear® Hybrid Pedicle Screws, including CarboClear® Hybrid Fenestrated Screws, are made of carbon fiber-reinforced polyetheretherketone (CFR-PEEK), with titanium alloy tulip. Their threaded portion is encased within a thin titanium shell. They are used with a compatible titanium alloy locking element, and with CarboClear X CFR-PEEK rods and trans-connectors. The implants may include tantalum markers. CarboClear X titanium alloy rod is also offered.
The implants are supplied sterile, and are intended for single use.
The CarboClear® Hybrid implants may be implanted in an open approach or using a minimally invasive surgery (MIS) approach.
The CarboClear® Hybrid Navigated Instruments are manually operated instruments, intended to be used with Medtronic's StealthStation® Navigation System, to assist surgeons in precisely locating anatomical structures for preparation and placement of CarboClear® Hybrid pedicle screws, respectively, during spinal surgery.
{9}------------------------------------------------
The CarboClear® Hybrid Fenestrated Pedicle Screws and CarboClear® X Fenestrated Pedicle Screws are cannulated polyaxial pedicle screws in various dimensions, with lateral fenestrations near screws' distal tip, which allow controlled delivery of polymethylmethacrylate (PMMA) bone cement (High V+® Bone Cement) into the vertebral body in oncological patients.
These implants may also serve as traditional cannulated pedicle screws when used without bone cement.
The High V+ Bone Cement is a self-curing, high viscosity, radiopaque PMMA based bone cement. It is provided sterile in two components: 20 grams of powder and 8.6 grams of liquid. The powder component consists of polymethylmethacrylate, with barium sulfate and hydroxyapatite as radiopacifier, and benzoyl peroxide as an initiator. The liquid component comprises methylmethacrylate monomer, with N.N-dimethyl-p-toluidine as a promoter, and hydroquinone as a stabilizer. The powder and liquid components are mixed into homogenous paste, to initiate the polymerization reaction.
### PURPOSE OF SUBMISSION
Modification to the previously cleared CarboClear® X Pedicle Screw System (K210716, K231280, K233944).
# COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH PREDICATE DEVICES
- . The CarboClear® Hybrid Pedicle Screw System differs from the primary predicate in the screw tulip and locking element material.
- . The CarboClear® Hybrid pedicle screws are used with rods identical to the rods of the primary predicate.
- The subject CarboClear® X fenestrated screws and CarboClear® Hybrid fenestrated ● screws have the same indications as CarboFix fenestrated screws previously cleared in K190526 and K231280, when augmented with cement (High V+® Bone Cement). The subject fenestrated screws may also be used without cement, for the non-cement augmented indications stated in the Indications for Use section.
#### PERFORMANCE DATA
Performance characteristics included static and dynamic compression bending tests according to ASTM F1717, as well as static flexion-extension test, axial gripping capacity test, and torsional gripping capacity test according to ASTM F1798. The results of the tests
{10}------------------------------------------------
are comparable to those of the predicate devices, demonstrating substantially equivalent performance of the subject and predicate devices.
### SUBSTANTIAL EQUIVALENCE
The intended use, design, dimensions, materials, technological characteristics, principles of operation, and performance of the CarboClear® Hybrid Pedicle Screw System (including the CarboClear® Hybrid Navigated Instruments and the CarboClear® Hybrid Fenestrated Pedicle Screw System (with High V+® Bone Cement)), and the CarboClear X Fenestrated Pedicle Screw System, are substantially equivalent to those of the predicate devices.
#### CONCLUSION
Based on the information provided in this 510(k) Premarket Notification, the CarboClear® Hybrid Pedicle Screw System, including the CarboClear® Hybrid Navigated Instruments and the CarboClear Hybrid Fenestrated Pedicle Screw System (with High V+® Bone Cement); and the CarboClear X Fenestrated Pedicle Screw System, are substantially equivalent to their predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.