CMORE® CT System; CMORE® CT System Navigated Instruments
K252327 · Icotec AG · NKG · Nov 12, 2025 · Orthopedic
Device Facts
Record ID
K252327
Device Name
CMORE® CT System; CMORE® CT System Navigated Instruments
Applicant
Icotec AG
Product Code
NKG · Orthopedic
Decision Date
Nov 12, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3075
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CMORE® CT System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. In patients requiring adjuvant radiotherapy following spinal stabilization, metal markers at the screw tips can be used as fiducial markers in radiotherapy treatments (e.g., SBRT, proton therapy). The CMORE® CT System is intended to provide immobilization and stabilization of spinal segments with anterior interbody support implanted at the same spinal level/s as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1 to C7) and the upper thoracic spine (T1 to T3): • Traumatic spinal fractures and/or traumatic dislocations • Instability or deformity • Failed previous fusions (e.g. pseudarthrosis) • Degenerative disease, including intractable radiculopathy and/or myelopathy • Neck and/or arm pain of discogenic origin as confirmed by radiographic studies • Degenerative disease of the facets with instability • Patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression. The CMORE® CT System is intended to be used with anterior interbody support implanted at the same spinal level/s with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. The CMORE® CT System may be connected to the VADER® Pedicle System using connectors. CMORE® CT System Navigated Instruments are intended to be used during the preparation and placement of the CMORE® CT System during spinal surgery to assist the surgeon in precisely locating anatomical structures in open procedures. These instruments are designed for use with the Medtronic StealthStation® and Brainlab Navigation Systems which are indicated for any medical condition in which the use of stereotactic surgery may by appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT- or MR- based model, fluoroscopy images, or digitized landmarks of anatomy.
Device Story
System provides posterior cervical/upper thoracic spinal fixation; comprises Carbon/PEEK screws, rods, nuts, and connectors. Navigated instruments (awls, drills, probes, taps, drivers, adapters) assist surgeons in open procedures by locating anatomical structures. Instruments integrate with Medtronic StealthStation® or Brainlab Navigation Systems using instrument-mounted tracking arrays for CT/MR/fluoroscopy-based anatomical referencing. Metal markers at screw tips serve as fiducials for radiotherapy (SBRT, proton therapy). Used by surgeons in OR to stabilize spinal segments; provides structural support for tumor patients or adjunct to fusion for instabilities. Benefits include spinal stabilization, compatibility with radiotherapy, and precise surgical navigation.
Clinical Evidence
Bench testing only. Mechanical performance assessed via static/dynamic compression bending (ASTM F1717), torsion (ASTM F1717), and axial/torque gripping (ASTM F1798). Navigated instruments tested per ASTM F2554. Summative usability evaluation conducted in wetlab with cadaver specimens; confirmed radiographic visibility of tantalum/titanium markers via CT/MRI.
Technological Characteristics
Materials: Carbon/PEEK (implants), Tantalum/Titanium (markers). Standards: ASTM F1717, F1798, F2554. Components: Screws, rods, nuts, connectors, reusable navigated instruments. Navigation: Instrument-mounted tracking arrays for integration with external navigation systems (StealthStation/Brainlab).
Indications for Use
Indicated for patients with advanced stage cervical spine tumors (limited life expectancy); acute/chronic cervical (C1-C7) and upper thoracic (T1-T3) instabilities including traumatic fractures/dislocations, deformity, failed fusions, degenerative disease (radiculopathy/myelopathy/discogenic pain/facet instability), and spinal infection.
Regulatory Classification
Identification
Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
Special Controls
*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant.
(3) Device components must be demonstrated to be biocompatible.
(4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(5) Labeling must include the following:
(i) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(ii) Intended use and indications for use including levels of fixation;
(iii) Device specific warnings, precautions, and contraindications that include the following statements:
(A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies (
*e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.”
(iv) Identification of magnetic resonance (MR) compatibility status;
(v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and;
(vi) Detailed instructions of each surgical step, including device removal.
Predicate Devices
EUROPA™ Posterior Cervical Fusion System (K242516)
CarboClear® Posterior Cervical Screw System (K233989)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
November 12, 2025
icotec ag
% Michael Coladonato
Associate Director, Spine Regulatory Affairs
MCRA, LLC
803 7th Street, NW
Washington, District of Columbia 20001
Re: K252327
Trade/Device Name: CMORE® CT System; CMORE® CT System Navigated Instruments
Regulation Number: 21 CFR 888.3075
Regulation Name: Posterior Cervical Screw System
Regulatory Class: Class II
Product Code: NKG, OLO
Dated: October 8, 2025
Received: October 8, 2025
Dear Justin Eggleton:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}
K252327 - Justin Eggleton
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252327 - Justin Eggleton
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Colin O'Neill -S
Colin O'Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions. | K252327 | ? |
| Please provide the device trade name(s). | | ? |
| CMORE® CT System;
CMORE® CT System Navigated Instruments | | |
| Please provide your Indications for Use below. | | ? |
| CMORE® CT System
The CMORE® CT System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In patients requiring adjuvant radiotherapy following spinal stabilization, metal markers at the screw tips can be used as fiducial markers in radiotherapy treatments (e.g., SBRT, proton therapy).
The CMORE® CT System is intended to provide immobilization and stabilization of spinal segments with anterior interbody support implanted at the same spinal level/s as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1 to C7) and the upper thoracic spine (T1 to T3):
• Traumatic spinal fractures and/or traumatic dislocations
• Instability or deformity
• Failed previous fusions (e.g. pseudarthrosis)
• Degenerative disease, including intractable radiculopathy and/or myelopathy
• Neck and/or arm pain of discogenic origin as confirmed by radiographic studies
• Degenerative disease of the facets with instability
• Patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression.
The CMORE® CT System is intended to be used with anterior interbody support implanted at the same spinal level/s with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
The CMORE® CT System may be connected to the VADER® Pedicle System using connectors.
CMORE® CT System Navigated Instruments
CMORE® CT System Navigated Instruments are intended to be used during the preparation and placement of the CMORE® CT System during spinal surgery to assist the surgeon in precisely locating anatomical structures in open procedures. These instruments are designed for use with the Medtronic StealthStation® and Brainlab Navigation Systems which are indicated for any medical condition in which the use of stereotactic surgery may by appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT- or MR- based model, fluoroscopy images, or digitized landmarks of anatomy. | | |
| Please select the types of uses (select one or both, as applicable). | ☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
K252327
Page 1 of 4
# 510(k) Summary
Device Trade Name: CMORE® CT System, CMORE® CT System Navigated Instruments
Manufacturer: icotec ag
Industriestrasse 12
9450 Altstaetten
Switzerland
www.icotec-medical.com
Contact: Michael Coladonato
Associate Director, Spine Regulatory Affairs
MCRA, LLC
803 7th Street, NW
Washington, DC 20001
Phone: 202.552.5804
mcoladonato@mcra.com
Prepared by: MCRA, LLC
803 7th Street, NW
Washington, DC 20001
Office: 202.552.5800
Date Prepared: 11/10/2025
Classifications: 21 CFR §888.3075, Posterior cervical screw system
21 CFR §882.4560, Stereotaxic instrument
Class: II
Product Codes: NKG, OLO
Primary Predicate: EUROPA™ Posterior Cervical Fusion System (K242516)
Additional Predicate: CarboClear® Posterior Cervical Screw System (K233989)
ExcelsiusHub™ (K231850)
Reference Device: SYMPHONY™ OCT System (K181949)
Indications For Use:
**CMORE® CT System**
The CMORE® CT System is intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving
{5}
K252327
Page 2 of 4
the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.
In patients requiring adjuvant radiotherapy following spinal stabilization, metal markers at the screw tips can be used as fiducial markers in radiotherapy treatments (e.g., SBRT, proton therapy).
The CMORE® CT System is intended to provide immobilization and stabilization of spinal segments with anterior interbody support implanted at the same spinal level/s as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1 to C7) and the upper thoracic spine (T1 to T3):
- Traumatic spinal fractures and/or traumatic dislocations
- Instability or deformity
- Failed previous fusions (e.g. pseudarthrosis)
- Degenerative disease, including intractable radiculopathy and/or myelopathy
- Neck and/or arm pain of discogenic origin as confirmed by radiographic studies
- Degenerative disease of the facets with instability
- Patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression.
The CMORE® CT System is intended to be used with anterior interbody support implanted at the same spinal level/s with autogenous and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft.
The CMORE® CT System may be connected to the VADER® Pedicle System using connectors.
## CMORE® CT System Navigated Instruments
CMORE® CT System Navigated Instruments are intended to be used during the preparation and placement of the CMORE® CT System during spinal surgery to assist the surgeon in precisely locating anatomical structures in open procedures. These instruments are designed for use with the Medtronic StealthStation® and Brainlab Navigation Systems which are indicated for any medical condition in which the use of stereotactic surgery may by appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT- or MR- based model, fluoroscopy images, or digitized landmarks of anatomy.
## Device Description:
## CMORE® CT System
The CMORE® CT System is a posterior cervical fixation system intended to provide structural stability and mechanical support to the cervical and upper thoracic spine. The CMORE® CT System is manufactured from Carbon/PEEK using a proprietary manufacturing process and is comprised of screws, rods, nut screws, and connectors as well as corresponding instruments.
## CMORE® CT System Navigated Instruments
The CMORE® CT System Navigated Instruments are reusable instruments used for the preparation and placement of FDA-cleared spinal bone screws during spinal surgery to assist the surgeon in precisely locating anatomical structures. These instruments are designed to be
{6}
K252327
Page 3 of 4
used with the Medtronic StealthStation® and Brainlab Navigation Systems. The instruments include awls, drills, probes, taps, drivers, and adapters.
## Predicate Device:
### CMORE® CT System
icotec ag submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA’s regulation of medical devices, the CMORE® CT System is substantially equivalent in indications, design principles, and performance to the following predicate devices, which have been determined by FDA to be substantially equivalent to pre-amendment devices:
Primary Predicate: EUROPA™ Posterior Cervical Fusion System (K242516)
Additional Predicate: CarboClear® Posterior Cervical Screw System (K233989)
Reference Device: SYMPHONY™ OCT System (K181949)
### CMORE® CT System Navigated Instruments
icotec ag submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA’s regulation of medical devices, the CMORE® CT System Navigated Instruments are substantially equivalent in indications, design principles, and performance to the following predicate devices, which have been determined by FDA to be substantially equivalent to pre-amendment devices:
Additional Predicate: ExcelsiusHub™ (K231850)
## Performance Testing Summary:
### CMORE® CT System
The mechanical performance profile of the CMORE® CT System was assessed through static and dynamic testing in accordance with the following test methods:
- Static and dynamic compression bending per ASTM F1717
- Static torsion strength testing per ASTM F1717
- Static axial and torque gripping capacity per ASTM F1798
- Static flexion/extension bending strength per ASTM F1798
The summative usability evaluation was conducted on the CMORE® CT System markers as part of the final assessment of cervical posterior fixation implants. The evaluation took place in a wetlab environment using cadaver specimens and various imaging modalities. Orthopedic surgeons and radiation oncologists participated by providing feedback on implant visibility and imaging compatibility.
Results demonstrated that tantalum and titanium markers provided radiographic visibility with CT and MRI imaging modalities. The usability testing found no safety-related issues or deviations during the evaluation.
Overall, the implant markers were confirmed to be safe and effective for intra- and postoperative imaging, supporting their use in the device as intended.
{7}
K252327
Page 4 of 4
CMORE® CT System Navigated Instruments
The CMORE® CT System Navigated Instruments have been tested per ASTM F2554.
## Substantial Equivalence:
The CMORE® CT System is composed of Carbon/PEEK, while the primary predicate device (K242516) is made of a Molybdenum-Rhenium composite. An additional predicate device (K233989) made of CFR-PEEK has been cleared for cervical and upper thoracic spine applications. Despite the differences in material composition, the technological characteristics of the CMORE® CT System do not raise new safety or effectiveness concerns.
The CMORE® CT System Navigated instruments and the corresponding additional predicate device (K231850) both use instrument mounted tracking arrays for navigation and allow for integration with CT or fluoroscopy-based imaging for anatomical referencing. The subject device and the predicate device have different navigation system compatibility, but this does not raise questions of safety and effectiveness.
The subject devices were demonstrated to be substantially equivalent to the predicate devices cited in the section above with respect to intended use, indications, design, and performance.
## Conclusion:
The subject devices and the predicate devices have the same intended use, have similar technological characteristics, and are made of similar materials. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above. The CMORE® CT System and the CMORE® CT System Navigated Instruments are as safe, as effective, and performs as well as, or better, than the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.